The video describes a single-cell western blotting technique that identifies the composition of individual cells captured on a single-cell western blotting chip in a heterogeneous suspension. The technique detects the fluorescence intensity of the fluorophore tagged to a protein marker specific to the target cell.

研究方案

1. Determining Cell Composition in the Cell Suspension Using Single-cell Western Blot (scWB) Analysis

  1. Rehydrate the scWB chip before use in 15 mL of 1x suspension buffer in a Petri dish for at least 10 min at RT.
  2. Add 1 mL of the diluted single-cell suspension (10,000-100,000 cells/mL) to the rehydrated chip. Set the surface onto the bottom of a 10 cm Petri dish. Ensure that the surface is set flat.
  3. Cover the Petri dish with a lid to prevent drying and let the cells settle for 5-15 min by a gradient.
  4. Place the chip under a bright-field microscope and inspect the wells at 10x magnification.
    NOTE: Approximately 15-20% of microwells should be occupied by a single cell, and fewer than 2% of wells should contain 2 or more cells.
  5. Tilt the 10 cm Petri dish containing the chip by 45 degrees.
  6. Wash the chip with the 1x suspension buffer to remove uncaptured cells from the surface of the chip by gentle pipetting from the top to the bottom of the chip. Repeat 3 times.
  7. Prepare the single-cell Western instrument.
  8. Set the lysis time, electrophoresis time, and UV capture time. To analyze recruited macrophages recovered from the basement membrane-derived gel plug, use the following settings: lysis time: 0 s; electrophoresis time: 160 s; UV capture time 240 s.
  9. Carefully load the chip into the electrophoresis cell of single cell Western instrument, gel-side facing. Use caution not to damage the gel side of the chip.
  10. Pour the lysis/running buffer into the chamber and completely cover the entire single-cell Western chip. Start the cell lysis.
  11. Run the scWB instrument using the setting listed in 1.8.
  12. Once the run is complete, transfer the chip to a 10 cm Petri dish and wash the chip twice with 1x washing buffer for 10 min at RT.
  13. Prepare dilutions of primary antibody (anti-vinculin: 1:10; anti-CD45: 1:15, anti-CD11b: 1:20, anti-F4/80: 1:10) in a total volume of 80 µL of antibody diluent (mix 8 µL of antibody 1 plus 8 µL of antibody 2 plus 64 µL of diluent).
  14. Add 80 µL of the primary antibody solution to the antibody probing chamber and lower the chip gel-side down so that the antibody solution wicks across the chip.
  15. Incubate with the primary antibody solution for 2 h at RT.
  16. Wash the chip three times for 10 min in 1x washing buffer on the shaker.
  17. Wash the chip once for 10 min in the water on the shaker to desalt the gel.
  18. Prepare a 1:20 dilution of secondary antibody in a total volume of 80 µL (mix 2 µL of secondary antibody plus 78 µL of diluent).
  19. Incubate the chip with a secondary antibody for 1 h at RT protected from light.
  20. Wash the chip three times for 10 min with 1x washing buffer on the shaker.
  21. Spin the chip using a slide spinner to remove any remaining washing buffer.
  22. Scan the chip on a dual-laser microarray scanner at 5 µm resolution at the spectral channel of the fluorophore coupled to the secondary antibodies.
  23. Analyze data using scWB-specific software.

2. Stripping the scWB Chip

  1. Store the scanned chip in a wash buffer until ready to strip.
  2. Place the water bath in a fume hood.
  3. Place a 50 mL tube rack in the water bath with the water just 1 cm above the rack and set the water temperature at 60 °C.
  4. Prepare the stripping buffer. For 1 L of stripping buffer solution, dissolve 9.85 g of Tris-HCl (pH 6.8) and 20 g of SDS in 900 mL of distilled water and adjust pH. Then fill with distilled water to 1L. Add 0.8% (v/v) of β-mercaptoethanol (β-ME; 14.3 M) immediately before each use.
  5. After the initial scan, place the chip in a 10 cm Petri dish before stripping.
  6. For each chip, take 40 mL of stripping buffer and add 320 µl of β-ME.
    CAUTION: β-ME is toxic and hazardous to humans and the environment. The following procedure should be done in a fume hood.
  7. Place the chip in the canister and seal the canister with parafilm to prevent water from leaking into the canister
  8. Place the canister inside the tube rack in the pre-warmed water bath.
  9. Incubate for 90 min.
  10. Carefully remove the chip from the canister and place it in a fresh Petri dish.
  11. Briefly wash the chip once with 1x wash buffer. Then add 15 mL of 1x washing buffer to the Petri dish.
  12. Wash the chip for 15 min on a shaker. Repeat the wash step four times.
    NOTE: Chip is ready for the next primary antibody (see steps 1.12 - 1.21)

披露声明

No conflicts of interest declared.

材料

NameCompanyCatalog NumberComments
10 cm Petri dishgriner bio-one664 160
10x suspension bufferproteinsimpleR101
15 cm Petri dishFalcon353025
2-MercaptoethanolMP BIOMEDICALS2194705
5% washing bufferproteinsimpleR252
Antibody diluentproteinsimple042-203
Bench Top CentrifugeBeckman CoulterMicrofuge 22R Centrifuge
Bovine Serum AlbuminSigma-AldrichA7906-100G
Calcein AMThermoFisherC30991 mL
CD11bNovus BiologicalsNB110-89474
CD45 (D4H7K) rabbit mAbCell Signal Technologies727987S
CD68/SR-D1 (FA-11)Novus BiologicalsNBP2-33337SS
Cellometer VisionNexcelom
DispaseBD354235100 mL
Dissecting scissor
Donkey anti-Rabbit IgG secondary antibody [NL557]Novus BiologicalsNL004NL 557 conjugate
Donkey anti-Rat IgG (H+L) secondary antibody [DyLight 650]Novus BiologicalsNBP1-75655DyLight 650 conjugate
F4/80 (CI-A3-1)Novus BiologicalsNB600-404SS
Heat Blockeffendorf22331Thermomixer
Lysis/Running bufferproteinsimpleR200
Matrigel Matrix (Growth Factor Reduced)BD35423010 ml
Microarray scannerPerkinElmerScanArray Gx
Microcentrifuge tubeFisher Scientific05-408-129
MicroscopeThermoFisherEvos fl
MiloproteinsimpleSingle cell western
NeedleBD30511126 G
Parafilm
Probing chamberproteinsimple035-020
Recombinant mouse CCL2/JE/ MCP-1 proteinR&D479-JE-050
scWEST chipproteinsimpleC300
ShakerBioexpressGene mate orbital shaker
Single cell western chip canisterproteinsimple035-118
Slide spinnerLabnetC1303-T
Sodium Dodecyl Sulfate (SDS)Fisher ScientificBP 166-500
SyringeBD3096021 mL
Tris-HydrochlorideFisher ScientificBP 153-500
Tweezerproteinsimple035-020
Water bathThermoFisher

This article has been published

Video Coming Soon

JoVE Logo

政策

使用条款

隐私

科研

教育

关于 JoVE

版权所属 © 2025 MyJoVE 公司版权所有,本公司不涉及任何医疗业务和医疗服务。