JoVE Logo

サインイン

6.6 : Mass Spectrum: Interpretation

An unknown compound can be established by identifying the molecular ion peak in the mass spectrum. The molecular ion peak is often weak or absent due to the predominance of fragmentation in high-energy electron beams. In such cases, a low-energy electron beam can be used to scan the spectrum to enhance the intensity of the molecular ion peak. Additionally, chemical ionization, field ionization, and desorption ionization spectra are used to obtain a relatively intense molecular ion peak.

To determine the empirical formula from the molecular ion mass, a high-resolution mass spectrometer like a double-focusing mass spectrometer, a time-of-flight mass spectrometer, an orbitrap mass spectrometer, or a Fourier transform ion cyclotron resonance mass spectrometer is required.

Comparing the intensities of isotope peaks (M+1 and M+2), which arise from the natural abundances of higher-mass isotopes, with the molecular ion peak (M+) gives a fair idea of the empirical formula of an unknown compound.

The number of carbon atoms present in the unknown compound can be deduced by comparing the M+1 to M+ peak intensity ratio to the expected M+1 contribution per carbon atom.

The nitrogen rule, stating that a molecule having an even molecular weight contains zero or even number of nitrogen atoms, predicts the number of nitrogen atoms present in the unknown compound.

Further, the degree of unsaturation predicted from the formula

Figure1

indicates the presence of multiple bonds or rings in the unknown compound. Lastly, the study of fragmentation patterns for compounds based on a series of guidelines helps identify signature fragments at characteristic m/z values, suggesting the presence of certain structural features.

タグ

Mass SpectrumMolecular Ion PeakFragmentationLow energy Electron BeamChemical IonizationHigh resolution Mass SpectrometerEmpirical FormulaIsotope PeaksCarbon AtomsNitrogen RuleDegree Of UnsaturationFragmentation PatternsM z Values

章から 6:

article

Now Playing

6.6 : Mass Spectrum: Interpretation

Principles of Mass Spectrometry

1.1K 閲覧数

article

6.1 : Mass Spectrometry: Overview

Principles of Mass Spectrometry

4.1K 閲覧数

article

6.2 : Mass Spectrometry: Isotope Effect

Principles of Mass Spectrometry

1.9K 閲覧数

article

6.3 : Mass Spectrometry: Molecular Fragmentation Overview

Principles of Mass Spectrometry

2.8K 閲覧数

article

6.4 : Mass Spectrometers

Principles of Mass Spectrometry

5.1K 閲覧数

article

6.5 : Mass Spectrum

Principles of Mass Spectrometry

1.7K 閲覧数

article

6.7 : Mass Analyzers: Overview

Principles of Mass Spectrometry

559 閲覧数

article

6.8 : Mass Analyzers: Common Types

Principles of Mass Spectrometry

542 閲覧数

article

6.9 : High-Resolution Mass Spectrometry (HRMS)

Principles of Mass Spectrometry

1.2K 閲覧数

article

6.10 : Mass Spectrometry: Complex Analysis

Principles of Mass Spectrometry

687 閲覧数

article

6.11 : Tandem Mass Spectrometry

Principles of Mass Spectrometry

853 閲覧数

article

6.12 : Gas Chromatography–Mass Spectrometry (GC–MS)

Principles of Mass Spectrometry

3.8K 閲覧数

article

6.13 : Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview

Principles of Mass Spectrometry

606 閲覧数

article

6.14 : Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Principles of Mass Spectrometry

365 閲覧数

JoVE Logo

個人情報保護方針

利用規約

一般データ保護規則

研究

教育

JoVEについて

Copyright © 2023 MyJoVE Corporation. All rights reserved