JoVE Logo

サインイン

このコンテンツを視聴するには、JoVE 購読が必要です。 サインイン又は無料トライアルを申し込む。

この記事について

  • Overview
  • プロトコル
  • 開示事項
  • 資料
  • 参考文献

Overview

In this video, we demonstrate SPME-probe-based extraction of metabolites from a pre-harvested donor kidney. The extracted metabolites can be used to perform metabolomic and lipidomic analysis in order to assess the organ quality for transplantation.

プロトコル

All procedures involving human participants have been performed in compliance with the institutional, national and international guidelines for human welfare and have been reviewed by the local institutional review board.

1. Preparation of Probes

  1. Prepare titanium-nickel alloy probes (40 mm length; 0.2 mm diameter) coated with 7 mm mix-mode sorbent. The number of probes depends on the time points targeted during the entire procedure and the number of replicates (3 is recommended per time point).
    NOTE: The length and type of extraction phase may be adjusted based on the mode of study, the polarity of metabolites, and the sample matrix.
  2. Prepare a cleaning mixture composed of 2:1 chloroform:methanol (v/v). Pipette 1.0 mL of the solution to each 2.0 mL glass vial and place one probe, previously pierced through the cap, in each vial.
    NOTE: Before use, clean all probes to remove contaminating particles.
  3. Put the vials on the agitator and set the agitation speed at 1,200 rpm. After 45 min, stop the device and rinse the coatings with LC-MS grade water.
  4. As coatings must undergo a preconditioning step to activate them, prepare a preconditioning mixture composed of 1:1 methanol:water (v/v). Pipette 1.0 mL of the solution to each 2.0 mL glass vial and place one probe, previously pierced through the cap, in each vial.
  5. Put the vials on the vortex agitator and set the agitation speed at 1,200 rpm.
  6. After 60 min, stop the agitator and rinse the coatings with LC-MS grade water.
  7. Sterilize probes according to the standard surgical equipment sterilization protocol.

2. Extraction

  1. Open the sterile package right before sampling to ensure a sterile environment.
  2. Insert two probes directly into the kidney cortex for 10 min at each time point. The entire length of the coating must be covered by the tissue matrix; no specific angle is required, but ca. 90 deg is usually used.
    NOTE: The entire medical procedure follows standard protocols in a given institution. No modification concerning SPME sampling is considered. The procedure involves the six following sampling time points: a) before kidney resection, in vivo from donor; b)–e) after 1 h, 3 h, 5 h, 7 h of kidney perfusion, ex vivo in organ chamber; f) after reperfusion, in vivo from recipient.
  3. Retract the probes by pulling them out from the tissue and then rinse coatings with LC-MS grade water to clear any remaining blood from the coating surface. Rinse away from the surgical site, and immediately after removal of the probes.

3. Transport and Storage

  1. Place probes in separate vials and close them.
  2. Place vials in a Styrofoam box filled with dry ice or in liquid nitrogen for transport.
  3. Store samples in a freezer (-80 °C), or immediately commence the desorption step.

4. Desorption

  1. Prepare desorption solutions composed of 80:20 acetonitrile:water (v/v) for metabolomic analysis, and 1:1 isopropanol:methanol (v/v) for lipidomic analysis.
  2. Pipette 100 μL of the solution to inserts placed into 2.0 mL labeled vials and place one probe, previously pierced through the cap, in each vial.
  3. Put the vials on the vortex agitator and set the agitation speed at 1,200 rpm for 120 min.
  4. Remove probes from the vials. Obtained extracts are now ready for instrumental analysis.

5. LC-MS analysis

  1. Place vials containing extracts in the autosampler of the LC-MS system.
    NOTE: Liquid chromatography (RP, HILIC) coupled with high-resolution mass spectrometry and an orbitrap mass analyzer were used for this study. For metabolomic analysis parameters, go to step 5.2. For lipidomic analysis parameters, go to step 5.6.
  2. Use a pentafluorophenyl (PFP) column (2.1 mm x 100 mm, 3 μm) for reversed-phase separation.
    1. Set the flow rate to 300 µL/min, and autosampler and column temperatures to 4 °C and 25 °C, respectively.
    2. Prepare mobile phases according to the following proportions: mobile phase A: water:formic acid (99.9:0.1, v/v), and mobile phase B: acetonitrile:formic acid (99.9:0.1, v/v). Set the mobile phase flow according to the following parameters: starting mobile phase flow (0–3 min): 100% A followed by a linear gradient to 10% A (3–25 min), ending with the isocratic flow of 10% A until 34 min, followed by 6 min of column re-equilibrium time.
  3. For HILIC separation, use a HILIC column (2.0 mm x 100 mm, 3 μm, 200A). Set the flow rate to 400 μL/min.
    1. Prepare mobile phases according to the following proportions: mobile phase A: acetonitrile:ammonium acetate buffer (9:1, v/v, effective salt concentration 20 mM), mobile phase B: acetonitrile:ammonium acetate buffer (1:1, v/v, effective salt concentration 20 mM).
    2. Set the mobile phase flow according to the following parameters: start mobile phase flow (0–3 min) at 100% A, hold for 3.0 min and then ramp to 100% B within 5 min. Hold with 100% B until 12 min, followed by 8 min of column re-equilibrium time.
  4. Set the scan range to 85–1000 m/z. Set HESI ion source parameters in positive ionization mode to: spray voltage 1,500 V, capillary temperature 300 °C, sheath gas 40 a.u., aux gas flow rate 15 a.u., probe heater temperature 300 °C, S-Lens RF level 55%.
  5. Run QC samples composed of 10 µL of each analyzed sample (regularly, every 10–12 samples) to monitor instrument performance.
  6. For reversed-phase separation, use a C18 column (2.1 mm x 75 mm, 3.5 μm).
    1. Set the flow rate to 200 µL/min, and autosampler and column temperatures to 4 °C and 55 °C, respectively. Prepare mobile phases according to the following proportions: mobile phase A: H2O:MeOH (60: 40, v/v), 10 mM ammonium acetate and 1 mM acetic acid; mobile phase B: IPA:MeOH (90:10, v/v), 10 mM ammonium acetate and 1 mM acetic acid.
    2. Set the mobile phase flow according to the following parameters: 0–1 min (20% B), 1–1.5 min (20–50% B), 1.5–7.5 min (50–70% B), 7.5–13 min (70–95% B), 13–17 min (95% B), 17–17.1 min (95–20% B), 17.1–23 min (20%).
  7. For HILIC separation, use a HILIC column (100 x 2.1 mm, 3 μm).
    1. Set the flow rate to 400 µL/min, and autosampler and column temperatures to 4 °C and 40 °C, respectively.
    2. Prepare mobile phases according to the following proportions: mobile phase A:ACN; mobile phase B: 5 mM ammonium acetate in water. Set the mobile phase flow according to the following parameters: 0–2 min (96% B), 2–15 min (96–80% B), 15–15.1 min (80–96% B), 15.1–21 min (96% B).
  8. Set HESI ion source parameters in positive ionization mode to spray voltage 3,500 V, capillary temperature 275 °C, sheath gas 20 a.u., aux gas flow rate 10 a.u., probe heater temperature 300 °C, S-Lens RF level 55%.
  9. Run QC samples composed of 10 µL of each analyzed sample (every 10–12 samples) to monitor instrument performance.
  10. Perform data acquisition in software compatible with the instrument.

6. Data Analysis

  1. Perform data processing, putative identification, and statistical analysis using software dedicated to untargeted metabolomics and lipidomics analysis.
    NOTE: Principal components analysis (PCA) and box-whisker plots can be obtained to visualize data structure.

開示事項

No conflicts of interest declared.

資料

NameCompanyCatalog NumberComments
Acetic acidMerck5330010050Mobile phase additive
AcetonitrileAlchem696-34967-4X2.5LHPLC solvent
Ammonium acetateMerck5330040050Mobile phase additive
BENCHMIXER XL MULTI-TUBE VORTEXERBenchmark ScientificBV1010Vortex mixer
CapsPerlan Technologies5183-2076Blue scrw tp, pre-slit PTFE/Si spta, 100PK
ChloroformMerck1024441000
Discovery HS F5 Supelguard Cartridge, 3 μm, L × I.D. 2 cm × 2.1 mmMerck567570-UHPLC guard column
Discovery HS F5, 2.1 mm x 100 mm, 3 μmMerck567502-UHPLC column
Formic acidAlchem497-94318-50MLMobile phase additive
Glass vialsPerlan Technologies5182-0714
HILIC Luna 3 μm, 200A, 100 x 2.0 mmShim-PolPHX-00D-4449-B0HPLC column
HILIC SecurityGuard Cartridge, 3 μm, 4 x 2.0 mmShim-PolPHX-AJ0-8328HPLC guard column
IsopropanolAlchem231-AL03262500HPLC solvent
MethanolAlchem696-34966-4X2.5LHPLC solvent
Nano-pure waterMerck1037281002HPLC solvent
Q Exactive Focus hybrid quadrupole-Orbitrap MSThermo ScientificQ Exactive FocusMass Spectrometer
SeQuant ZIC-cHILIC 3µm,100Å 100 x 2.1 mmMerck1506570001HPLC column
SeQuant ZIC-HILIC Guard Kit 20 x 2.1 mmMerck1504360001HPLC guard column
SPME LC fiber probes, mixed modeSupelcoprototype fibers
UltiMate 3000 HPLC systemsThermo ScientificUltiMate 3000HPLC system
Vial inserts (deactivated)Perlan Technologies5181-8872
XSelect CSH C18 3.5μm 2.1x75mmWaters186005644HPLC column
XSelect CSH C18 VanGuard Cartridge 3.5μm, 2.1x5mmWaters186007811HPLC guard column

参考文献

This article has been published

Video Coming Soon

Source: Stryjak, I. et al. Using a Chemical Biopsy for Graft Quality Assessment. J. Vis. Exp. (2020)

JoVE Logo

個人情報保護方針

利用規約

一般データ保護規則

研究

教育

JoVEについて

Copyright © 2023 MyJoVE Corporation. All rights reserved