The eVOLVER framework enables high-throughput continuous microbial culture with high resolution and dynamic control over experimental parameters. This protocol demonstrates how to apply the system to conduct a complex fitness experiment, guiding users on programming automated control over many individual cultures, measuring, collecting, and interacting with experimental data in real-time.
Continuous culture methods enable cells to be grown under quantitatively controlled environmental conditions, and are thus broadly useful for measuring fitness phenotypes and improving our understanding of how genotypes are shaped by selection. Extensive recent efforts to develop and apply niche continuous culture devices have revealed the benefits of conducting new forms of cell culture control. This includes defining custom selection pressures and increasing throughput for studies ranging from long-term experimental evolution to genome-wide library selections and synthetic gene circuit characterization. The eVOLVER platform was recently developed to meet this growing demand: a continuous culture platform with a high degree of scalability, flexibility, and automation. eVOLVER provides a single standardizing platform that can be (re)-configured and scaled with minimal effort to perform many different types of high-throughput or multi-dimensional growth selection experiments. Here, a protocol is presented to provide users of the eVOLVER framework a description for configuring the system to conduct a custom, large-scale continuous growth experiment. Specifically, the protocol guides users on how to program the system to multiplex two selection pressures — temperature and osmolarity — across many eVOLVER vials in order to quantify fitness landscapes of Saccharomyces cerevisiae mutants at fine resolution. We show how the device can be configured both programmatically, through its open-source web-based software, and physically, by arranging fluidic and hardware layouts. The process of physically setting up the device, programming the culture routine, monitoring and interacting with the experiment in real-time over the internet, sampling vials for subsequent offline analysis, and post experiment data analysis are detailed. This should serve as a starting point for researchers across diverse disciplines to apply eVOLVER in the design of their own complex and high-throughput cell growth experiments to study and manipulate biological systems.
Continuous cell culture techniques, first developed nearly 70 years ago1,2, are enjoying a recent revival3,4. This is due to a confluence of factors. First, the development of high throughput -omics techniques, which have made it possible to read out and generate large numbers of genotypes5,6, has created a concomitant demand for experimental techniques that facilitate well-controlled cell growth and phenotyping. To this end, continuous culture represents a powerful experimental approach to capitalize on emergent genomic advances. By facilitating growth selections/experiments on cellular populations in precisely controlled (and dynamic) environmental conditions, continuous culture provides a means to rigorously map genotypes to phenotypes7,8, quantitatively characterize engineered strains and organisms9, and track adaptive genetic changes in laboratory evolution studies10,11,12.
Second, the recent emergence of accessible prototyping techniques, such as additive manufacturing and open-source hardware and software elements, has enabled a wider set of users to design and build their own cost-effective forms of continuous culture systems directly in the laboratory. All of this has led to an exciting array of do-it-yourself (DIY) devices that perform continuous culture functionalities, such as the chemostat13, turbidostat14, or morbidostat15. Unfortunately, though successful in addressing specific (niche) problems for which they were designed, these ad hoc solutions generally lack the ability to scale in throughput and/or experimental design complexity.
The eVOLVER system was designed with the goal of creating a single platform that can accommodate the growing experimental needs of continuous culture and match the speed and scale of emergent genomic techniques16 (Figure 1A). eVOLVER’s design implements common tenets underlying highly scalable technologies from other disciplines17, including standardized footprints, modular components, and open-source design principles. Thus, solutions for new niche applications can be designed without major modifications to the system. Comprised of highly modular and open-source wetware, hardware, electronics, and web-based software, eVOLVER is the first automated continuous culture system that can be cost-effectively and readily re-configured to carry out virtually any type of high-throughput growth experiment. Through modular and programmable Smart Sleeves which house all the sensors and actuators needed to control individual cultures, eVOLVER uniquely enables scaling of both throughput and individual control of culture conditions. Moreover, as a web-based platform, eVOLVER exchanges data and information with remote computers in real-time, permitting simultaneous monitoring of hundreds of individual cultures and automated culture perturbations through arbitrarily defined control algorithms.
In previous work16, eVOLVER’s robust performance was demonstrated in long-term experiments over hundreds of hours of operation, and its ability to cultivate various organisms, from E. coli and S. cerevisiae to undomesticated microbes. A series of distinct growth selection experiments were performed, in which programmatically defined multi-dimensional selection gradients were applied across an array of individual culture conditions and the resulting cellular fitness landscapes were quantified. Here, the objective is to provide eVOLVER users with a description of how to use the system to design and these types of experiments. As an illustrative example, methods quantifying the fitness landscape of S. cerevisiae mutants across a two-dimensional environmental gradient composed of temperature and osmotic stress are presented. The protocol guides users through configuring the eVOLVER framework for this experiment both programmatically, in using the software to set customized turbidity and temperature control routines for each of 16 parallel continuous cultures, and physically, through the fluidics layout to appropriately route medias of different salt concentrations. This protocol should serve as a general rubric for configuring eVOLVER to execute a wide variety of automated continuous culture experiments for diverse studies and disciplines.
1. Preparing Media, Vials, and Inoculum
NOTE: This protocol assumes users have already calibrated the eVOLVER system and are using the Smart Sleeve configuration described in prior work16. Smart Sleeves are easily redesigned or modified, but setup details of volume and control parameters may differ for alternative configurations.
2. Setting Up an Experiment
3. Configuring eVOLVER Software and Programming Algorithmic Culture Routines
4. Initiating Experiment and Monitoring in Real-time
5. Sampling Yeast Cells from eVOLVER Vials
6. Experiment Break Down and Clean Up
7. Quantifying Fitness Using Flow Cytometry Analysis of Fractional Populations
The protocol described here was used to construct fitness maps for genetic variants of S. cerevisiae across a two-dimensional stress gradient of temperature and salinity. Specifically, for each variant strain, we used eVOLVER to conduct pairwise competition experiments against a fluorescently-labeled reference strain (S. cerevisiae strain BY4741 with an integrated constitutive mNeonGreen reporter) in 16 different combinations of these stresses (Figure 1). For variants, two knockout strains were selected each predicted to be sensitive to one of the stress condition axes: ΔPRX1 which has been reported to have fitness defects in high temperatures20 and ΔPBS2 with fitness defects at high salt concentrations21. As a control variant, an unlabeled wild-type BY4741 strain was used, expected to perform similarly to the reference strain. In total, these three 72-hour competition experiments were conducted simultaneously in 48 vials across three eVOLVER (16-vial) devices all run from a single computer.
The eVOLVER interactive dashboard is used to navigate the large amount of data that is generated during each experiment (Figure 2A). Representative OD traces across all 16 vials of a single eVOLVER device are shown in Figure 2B. Each trace displays a characteristic sawtooth pattern from the turbidostat control algorithm, which uses feedback on OD to trigger dilutions and maintain cultures in a narrow density range (0.2-0.3 in this case). OD and temperature data are continuously measured, streamed to the cloud-enabled database, and updated in real-time in the eVOLVER interactive dashboard. From continuously streamed data, higher-order metrics (e.g. growth rate, cumulative generations) may be calculated and plotted in the dashboard (Figure 2C) and even used as feedback parameters in different selection schemes (e.g., morbidostat).
To generate fitness maps, we measure the rate at which the reference strain outcompetes the variant strain by incorporating both flow cytometry measurements and eVOLVER growth data. Specifically, population fractions are calculated as the log-ratio of variant strain over reference strain using flow cytometry data from each timepoint sample. For each culture and time point, the elapsed number of generations is interpolated from eVOLVER growth rate data in each dilution period. Plotting the log-ratios against generations, qualitative differences between conditions begin to emerge (Figure 3A). To calculate fitness, a slope is fit through the linear portion of each plot18,19, yielding a quantitative fitness value (with both sign and rate) that describes how the variant strain competes against the reference strain (Figure 3B). In this experiment, negative fitness values indicate that the variant strain is outcompeted by the reference strain. Constructing heatmaps from all the fitness calculations permits visualization of the two-dimensional fitness landscape, revealing subtle quantitative differences in performance between strains and conditions (Figure 3C).
From the fitness heatmaps, we see that each variant strain exhibits fitness defects that manifest in different ways. ΔPRX1 is primarily sensitive to high temperatures, but salt stress appears to have an additive effect, increasing the fitness defect. Conversely, ΔPBS2 shows fitness defects primarily in response to high salt concentrations, with minimal differences along the temperature axis. These results highlight the utility of high-resolution selection experiments particularly for deciphering interactions of multiple environmental stressors, which could have additive, synergistic, or epistatic effects.
Finally, it is important to note that in some cases, particularly for severe stress conditions, fitness calculations can lead to exaggerated or skewed results. For example, ΔPBS2 appears to show a steep positive slope in the 39 °C/1 M NaCl condition relative to the reference strain (Figure 3A). This is attributed to poor growth of both strains, which is reflected in the wild type data and limits the utility of this particular metric in this condition. An insufficient number of generations results in 1) poor separation of timepoints which interferes with slope fitting, and 2) sensitivity to stochastic effects originating from lag phase. While we did not elect to do so in this study, the real-time growth data collected by eVOLVER during the experiment could have been used to inform the decision to extend the experiment and collect additional timepoints for this condition with little effort. In follow-up experiments, starting ratios could also be modulated to extend the length of the linear region before saturation occurs.
Figure 1: Overview of the eVOLVER framework and experimental design. (A) eVOLVER is an automated, continuous culture platform allowing multiparameter control of culture conditions. The platform consists of Smart Sleeves, which house the sensors and actuators for controlling culture conditions, a peristaltic pump array for flowing media and waste in and out of the cultures, a touchscreen for interacting manually with the device, and a computer used for interacting with the cloud infrastructure where data from the platform is streamed. (B) eVOLVER can be configured in multiple ways: physically, through cell and media inputs, and programmatically by adjusting the algorithms controlling Smart Sleeve culture vial modules. This can be utilized to program multidimensional selection/environmental gradients at high-resolution, with outputs consisting of physically collected cells at defined time points and real-time streaming of culture data. (C) Configuring eVOLVER to conduct a growth fitness experiment across a two-dimensional selection gradient, composed of temperature and osmotic stress. eVOLVER culture vials were seeded with equal proportions of a reference and variant yeast strain. A turbidostat routine was programmed to maintain all the cultures in exponential phase in a defined OD window (OD 0.2-0.3). Temperature and media conditions were varied across the Smart Sleeve array to form two independent stress gradients. Base media was composed of YPD (2% glucose) + 100 µg/mL carbenicillin + 25 µg/mL chloramphenicol as a precaution against bacterial contamination. Media compositions were varied by adding supplemental NaCl to 0 M, 0.6 M, 0.8 M, or 1.0 M. Vial temperatures were programmed to one of four values: 30 °C, 35 °C, 37 °C, or 39 °C. (D) Examples of raw output for culture OD and population fractions from three tested conditions. The competition experiment was maintained for 72 h, sampling at 24 h and subsequently every 12 h until the conclusion of the experiment. Please click here to view a larger version of this figure.
Figure 2: eVOLVER real-time data analysis dashboard. (A) eVOLVER data is processed and streamed in real time via cloud-based software to an interactive dashboard. Through the dashboard, users can define and initiate their own culture routines on a connected eVOLVER, monitor currently running experiments across all eVOLVER units, and vary experimental conditions manually on an individual vial or set of vials if needed. (B) OD traces for each vial across the matrix of conditions tested for the entire experiment duration. Traces can be viewed in real-time to ensure experiment is running as desired and used post-experiment for further analysis. (C) Growth rates and cell generations can be calculated from OD traces on the fly to track experiment progress. Please click here to view a larger version of this figure.
Figure 3: Construction of fitness surfaces. (A) Cell population fractions natural log (variant strain/reference strain) plotted against cell generations. Color indicates the temperature of selection while dashes distinguish salt concentrations. (B) To quantify relative fitness of the variant to the reference strain, a fitness metric is derived from the rate at which the cellular population fraction changes over generations. This metric is calculated from the linear region of the cellular fraction plots using a minimum of three points. (C) The relative fitness metrics are displayed in a heatmap to construct a fitness surface over the environmental stress gradients. The top right corner condition (39 °C/1.0 M NaCl) for the wild type and ΔPBS2 are not included in this analysis as the cultures went through insufficient generations (see Representative Results). Please click here to view a larger version of this figure.
Growth selection is an indispensable tool in biology, broadly used to generate and characterize phenotypic differences between cellular populations. While batch cultures do permit growth selection in a limited way, continuous culture techniques dramatically expand the degree of control and predictability of these experiments, by exerting precise regulation over the form and dynamics of selection to generate repeatable, quantitative results22. Continuous culture has been employed to rigorously control selection for high-diversity libraries20,23,24,25, and to implement sophisticated adaptive regimes in experimental and directed evolution11,12,26,27. Continuous culture also enables precise characterization of cells across an array of quantitatively controlled conditions to better understand complex genetic systems and optimize engineered bioproduction strains9,14,28.
However, there is no universal protocol for continuous culture, as subtle changes to the selective conditions can lead to dramatic changes in biological outcomes4,29,30. Experimenters must be able to choose between selection regimes and adapt experimental protocols and equipment accordingly. In addition to offering a choice between control parameters, such systems would ideally be sophisticated enough to independently manage several parameters simultaneously in highly-parallel experiments that are needed to decipher interacting inputs in complex biological systems (e.g. epistasis). eVOLVER addresses this challenge by enabling users to arbitrarily program feedback control between culture conditions and fluidic functions in order to specify highly specialized environmental niches.
To overcome limitations in the current setup and expand or change control parameters, the Smart Sleeve could easily be redesigned to add new sensors or actuators. Additionally, reducing vial volume would decrease media expenditures, which can be significant in continuous culture. While the current design permits measurement and control of temperature, culture agitation, light induction, turbidity, and fluidics, other parameters must be measured externally by sampling from the vials. Current work includes incorporating the ability to monitor enzymatic activity via luciferase and regulate dissolved oxygen and pH directly in eVOLVER cultures. Additionally, while not demonstrated in this work, eVOLVER can interface with novel millifluidic multiplexing devices16 that draw on principles of large-scale integration (originating from electronics and adopted by microfluidics) in order to inexpensively enable more complex fluidic handling (e.g. multiplexed fluidic inputs, vial-to-vial transfers). These wetware modules may be designed and manufactured completely in the lab, allowing users to route fluids by programmatically actuating different combinations of valves in automated fluidic routines. This allows users to overcome the rigid fluidic designs traditionally used in continuous culture, but also to scale fluidic capabilities to high-throughput with a smaller number of costly control elements (e.g. peristaltic pumps). Lastly, we are hoping to incorporate an autosampling platform which will utilize these millifluidics and DIY components, overcoming the limitation of manual interaction during longer and larger experiments where sampling cultures would be cumbersome.
In addition to physical modifications to the platform, the web-based software opens new degrees of freedom by allowing users to write, edit, and share custom eVOLVER scripts, generating fully automated, feedback-enabled culture programs (e.g., turbidostat). Users may programmatically sweep across parameter ranges in subtle variations on the same selection scheme or connect control algorithms in novel combinations to specify any number of sophisticated selection schemes. Furthermore, the ability to easily monitor cultures in real-time transforms the way in which experiments are conducted. With real-time monitoring, users may 1) check for consistency between runs, a critical feature for bioproduction applications and high-throughput experiments, and 2) intervene during experiments if necessary, to troubleshoot challenging strains that exhibit poor growth or biofilm formation, or diagnose user errors (e.g., contamination). Finally, with multiple data streams being collected and interpreted in real-time for each individual culture, eVOLVER generates a high density of data, which may facilitate machine learning approaches for novel downstream analysis.
Beyond demonstrated uses for fitness characterization, library selection, and laboratory evolution, we view a number of related fields as ripe for implementation in eVOLVER with integrated fluidics. eVOLVER experiments with microbiome samples could assay community stability in controlled environments31,32, explore microbiota composition using culturomics techniques33, or dynamically mix species to interrogate ecological dynamics of colonization or invasion34,35. Numerous methods for continuous directed evolution of biomolecules could easily be implemented on the device as well26,36,37, greatly increasing accessibility and throughput of these systems. The ability to optimize growing conditions such as media composition, temperature, and strains in a dynamic, high throughput nature can aid in optimization efforts for industrial biomanufacturing applications9. We further envision vertically integrating eVOLVER with other analysis techniques such as microscopy and flow cytometry in a closed loop fashion, providing a fully automated system for growth and analysis of cellular cultures at both single cell and population levels. Moreover, with some hardware modifications to the Smart Sleeve such as sealing the vessel and controlling gas content, eVOLVER could potentially be adapted to support the growth of a wider range of cell types, such as suspension mammalian cells. It is also feasible to place the entire framework into an anaerobic chamber for anaerobic cell culture. Looking forward, we aim to build our software framework into a centralized cloud infrastructure and believe this would allow users to easily configure, analyze, and share their data remotely without needing to physically be present in the lab. Functioning as a data curator, the cloud infrastructure would also lend itself to large-scale meta analyses across experiments. We anticipate that eVOLVER and these future advances will greatly expand the scope of possible growth selection experiments by facilitating automation and innovation in continuous culture.
We thank B. Stafford for his assistance in design of the system, and H. Khalil, A. Soltanianzadeh, A. Sun, S. Pipe, and A. Cavale for help with construction of the system. We acknowledge the Electronics Design Facility (EDF), Engineering Product Innovation Center (EPIC), and the Software & Application Innovation Lab (SAIL) at the Hariri Institute for Computing at Boston University for their services. This work was supported by an NSF CAREER Award (MCB-1350949 to A.S.K.), and DARPA grants HR0011-15-C-0091 and HR0011-18-2-0014 (to A.S.K.). A.S.K. also acknowledges funding from the NIH Director’s New Innovator Award (1DP2AI131083-01), DARPA Young Faculty Award (D16AP00142), and NSF Expeditions in Computing (CCF-1522074).
Name | Company | Catalog Number | Comments |
5 Gallon Plastic Hedpack with cap | Midwest Brewing and Winemaking Supplies | 45-56Y8-E2FR | For waste collection |
a-D(+)-Glucose | Chem-Impex | 00805 | For YPD Medium |
Attune NxT Autosampler | Thermo Fisher | Allows Flow Cytometer to run samples from 96 well plate | |
Attune NxT Flow Cytometer | Thermo Fisher | Used to determine population fractions via single cell fluoresence | |
Bacto Peptone | Fisher Scientific | DF0118-07-0 | For YPD Medium |
Carbenicillin | Fisher Scientific | BP2648250 | For YPD Medium |
Chemical-Resistant Barbed Tube Fitting Tee Connector, for 1/8" Tube ID, 250°F Maximum Temperature | McMaster- Carr | 5121K731 | For media input branching |
Chloramphenicol | Fisher Scientific | BP904-100 | For YPD Medium |
CLOROX GERMICIDAL Bleach 8.25 | Fisher Scientific | 50371500 | For Sterilization of fluidic lines |
Custom eVOLVER vial lid | FynchBio | Lid has ports for sampling and fluidic input/output | |
Cycloheximide | Fisher Scientific | ICN10018301 | For flow cytometry sampling plates |
Ethanol, Anhydrous (Histological) | Fisher Scientific | A405P-4 | For sterilization of fluidic lines |
eVOLVER Unit | FynchBio | ||
Fisherbrand Extended-Length Tips (Lift Off Rack; 1 to 200 ul) | Fisher Scientific | 02-681-420 | For vial sampling |
Fisherbrand Octagon Spinbar Magnetic Stirring Bars | Fisher Scientific | 14-513-57 | Diameter: 4.5 mm, Length, 12 mm |
Fisherbrand Reusable Glass Media Bottles with Cap | Fisher Scientific | FB8002000 | Must be fitted with tubing |
High-Temperature Silicone Rubber Tubing Semi-Clear White, Durometer 70A, 1/8" ID, 1/4" OD | McMaster- Carr | 51135K73 | For media bottles |
Mac Mini | Apple | For running the experiment/collecting data | |
Phosphate Buffered Saline (PBS) | Fisher Scientific | BP243820 | For flow cytometry sampling plates |
Pipettes | Eppendorf | ||
Plastic Quick-Turn Tube Coupling, Plugs, for 1/16" Barbed Tube ID, Polypropylene | McMaster- Carr | 51525K141 | For media bottles |
Plastic Quick-Turn Tube Coupling, Plugs, for 5/32" Barbed Tube ID, Polypropylene | McMaster- Carr | 51525K144 | For media bottles |
Plastic Quick-Turn Tube Coupling, Sockets, for 1/16" Barbed Tube ID, Polypropylene | McMaster- Carr | 51525K291 | For media bottles |
Plastic Quick-Turn Tube Coupling, Sockets, for 5/32" Barbed Tube ID, Polypropylene | McMaster- Carr | 51525K294 | For media bottles |
SCREW CAPS, OPEN TOP, WITH PTFE FACED SILICONE SEPTA, LAB-PAC, SEPTUM. Screw thread size: 24-400, GREEN | Chemglass | CG-4910-04 | Culture vials |
Sodium Chloride (NaCl) | Fsher Scientific | S271-3 | For YPD Medium |
SpectraMax M5 Multi-Mode Microplate Reader | Molecular Devices | For measuring OD600 of overnight cell cultures | |
Vial Only, Sample, 40mL, Clear, 28x95mm, GPI 24-400 | Chemglass | CG-4902-08 | Culture vials |
Yeast Extract | Fisher Scientific | BP1422-500 | For YPD Medium |