The signal-to-noise ratio of data is one of the most important considerations in performing X-ray diffraction measurements from microcrystals. The VMXm beamline provides a low-noise environment and microbeam for such experiments. Here, we describe sample preparation methods for mounting and cooling microcrystals for VMXm and other microfocus macromolecular crystallography beamlines.
The mounting of microcrystals (<10 µm) for single crystal cryo-crystallography presents a non-trivial challenge. Improvements in data quality have been seen for microcrystals with the development of beamline optics, beam stability and variable beam size focusing from submicron to microns, such as at the VMXm beamline at Diamond Light Source1. Further improvements in data quality will be gained through improvements in sample environment and sample preparation. Microcrystals inherently generate weaker diffraction, therefore improving the signal-to-noise is key to collecting quality X-ray diffraction data and will predominantly come from reductions in background noise. Major sources of X-ray background noise in a diffraction experiment are from their interaction with the air path before and after the sample, excess crystallization solution surrounding the sample, the presence of crystalline ice and scatter from any other beamline instrumentation or X-ray windows. The VMXm beamline comprises instrumentation and a sample preparation protocol to reduce all these sources of noise.
Firstly, an in-vacuum sample environment at VMXm removes the air path between X-ray source and sample. Next, sample preparation protocols for macromolecular crystallography at VMXm utilize a number of processes and tools adapted from cryoTEM. These include copper grids with holey carbon support films, automated blotting and plunge cooling robotics making use of liquid ethane. These tools enable the preparation of hundreds of microcrystals on a single cryoTEM grid with minimal surrounding liquid on a low-noise support. They also minimize the formation of crystalline ice from any remaining liquid surrounding the crystals.
We present the process for preparing and assessing the quality of soluble protein microcrystals using visible light and scanning electron microscopy before mounting the samples on the VMXm beamline for X-ray diffraction experiments. We will also provide examples of good quality samples as well as those which require further optimization and strategies to do so.
A major barrier for the determination of high-resolution structures of biological molecules by macromolecular crystallography (MX) remains the production of well diffracting crystals at an amenable size. There are many strategies for achieving this goal from recombinant protein gene construct design through to large sparse matrix searches for chemical cocktails that may generate initial crystals2. For the latter, it is often the case that the crystallographer will need to optimize any initial hits to obtain crystals with sufficient diffraction quality and size for structure determination studies3. Despite these options, some target molecules may never generate large (>10 µm), well diffracting crystals and as a result the crystallographer must persevere with their microcrystals and the challenges that such samples present. These include appropriately mounting and cryo-protecting the crystals, managing inherently weaker diffraction and increased radiation sensitivity. Microcrystals are formed from fewer unit cells and molecules than larger crystals and as such, the diffraction is not amplified to the same extent compared to larger crystals, resulting in inherently weaker diffraction intensities. It is important that the background signal does not mask these reflections, particularly at higher resolution where weak reflection intensities can be lost4. In addition, microcrystals are more sensitive to radiation damage and despite recording diffraction at liquid nitrogen temperatures5, it may not be possible to collect complete data from a single crystal, making it necessary to collect data from a very large number of crystals to produce a single complete dataset6.
The increasing availability of X-ray free electron lasers (XFELs) and the evolution of serial crystallography methods (SFX)7 have provided routes to collecting data from smaller microcrystals. However, these are bespoke sample delivery methods, which require a significant amount of hardware and software expertise, where experiments are limited to room-temperature and typically sample consumption is high (hundreds of microliters) and still may require further optimisation8. As such, projects where only a limited quantity of microcrystals can be made are not appropriate for SFX.
Meanwhile, synchrotron beamline technology over recent decades has progressed to produce smaller, more stable beams9 with a brilliance that has permitted data collection from ever smaller crystals10,11. Microfocus beamlines such as FMX at NSLS-II and I24 at Diamond Light Source have been able to determine novel structures from crystals with maximum dimensions of ~3 µm12 and demonstrate the ability to collect usable data from even smaller crystals measuring ~1 µm13. The beamline must be precisely configured, with excellent, high resolution on-axis-viewing optics, a minimal sphere of confusion for sample rotation and a precisely aligned rotation axis that is coincident with the X-ray beam. It is important to closely match the X-ray beam profile to the crystal volume and ensure the crystal is precisely aligned in the X-ray beam - a challenge for crystals <5 µm14. Meeting these experimental conditions at the beamline is essential to recording the best quality data from microcrystals.
The remaining and possibly most important aspect of data collection from microcrystals is the presentation of the crystal to the X-ray beam. Microcrystals have often been mounted on micromesh sample mounts, manufactured from polyimide, a low X-ray scattering material with apertures as small as 10 µm15,16. The polyimide mesh is mounted on a standard pin that is set into a magnetic SPINE base, making it compatible with most MX beamlines17. The mesh mount is used to fish crystals from the crystallization drop often following the same procedure as mounting a 100 µm crystal using a standard loop style mount. While the crystals may be distributed across the mesh, a key disadvantage is that a relatively large volume of liquid can be carried by the mesh and the pin while harvesting (Figure 1C,D). This volume of liquid, that can be many times larger than the crystals themselves, will contribute to background noise when illuminated with X-rays. This background scatter can be even stronger if the liquid forms crystalline ice during flash cooling, diminishing the signal-to-noise ratio of already weak intensities within the resolutions of ice diffraction. Therefore, it is key that excess liquid is removed from the sample, to ensure that all possible signals can be recorded. This challenge is even greater in the case of membrane protein crystals formed within the lipid cubic phase (LCP), where the LCP generates strong background scatter and is also difficult to remove from around the crystals 18.
The new Versatile Macromolecular Crystallography microfocus (VMXm) beamline at Diamond Light Source provides the conditions with which to collect data from crystals potentially measuring less than a micron in size. The beamline has been designed to deliver a beam profile measuring 0.3 µm x 0.5 µm (VxH)1, a goniometer with a sphere of confusion no greater than 60 nm and an in vacuo sample environment. These design features of the VMXm endstation minimize the generation of background X-ray noise by the beamline apparatus during data collection with the largest remaining source of background generated by the sample14.
Specific sample preparation methods designed for the VMXm beamline provide an opportunity to reduce this background and further improve the signal-to-noise of diffraction data, maximizing the quality of the data that can be recorded from microcrystals measuring <10 µm. Many of the requirements outlined here for low-background diffraction from microcrystals are also common to cryogenic transmission electron microscopy (cryoTEM)19 and microcrystal electron diffraction (microED)20. As a result, many of the tools that have already been developed for the preparation of cryoTEM samples are suitable, with some adaptations, for the preparation of microcrystals. In the preparation of samples for single particle cryoTEM, the particles under investigation are embedded in very thin layers (typically <100 nm) of vitreous ice such that electrons are able to transmit through the sample. The thin uniform layer is achieved by blotting away excess liquid and vitrification of the sample is achieved by rapid cooling of the sample (~104 K s-1)21 through plunging into liquid ethane held at ~93 K22. In contrast, liquid nitrogen, as used routinely for MX sample preparation, is a less efficient cryogen than ethane and has a greater propensity for crystalline ice formation within the sample21. The formation of crystalline ice, which can degrade diffraction and generate background noise, is normally mitigated through the use of cryo-protectant compounds23. Low molecular weight polymers such as poly-ethylene glycol (PEG) 400 and methyl-2,4-pentanediol (MPD), sugars, oils or saturated salts can be added to an aliquot of crystallization solution in low concentrations24- there is not a 'one size fits all' solution to selecting the most appropriate cryoprotectant and this often requires optimization25. The crystal also undergoes multiple manipulations during the harvesting and cryo-protecting process which may result in damage to the crystal, the opportunity to utilize liquid ethane allows the omission of this step and helps protect the integrity of the crystal.
While liquid ethane is an effective cryogen for microcrystals (<10 µm) due to the thinness of the sample, there are alternative methods for preventing crystalline ice formation, particularly in larger crystals, including reducing the water content of the crystal by use of a tightly controlled humid environment26, or through the wicking of excess liquid away from both the loop and surface of the crystal27, however, these again require greater manipulation of the sample. The use of automated blotting and plunge freezing with liquid ethane, as in cryoTEM, together remove excess crystallization solution and provide a means to flash cool microcrystals in a controlled manner while attempting to minimize manipulation.
Here, we present a protocol that can be utilized not only by both users of the VMXm beamline and at other microfocus beamlines to collect high signal-to-noise diffraction data but may also be useful to those preparing soluble protein crystal and detergent based membrane protein crystal samples for microED experiments. While all of the facilities to prepare and assess samples are available at VMXm, many structural biology laboratories are increasingly equipped for cryoTEM sample preparation. As a result, we envisage that some users may wish to use their own facilities to prepare their samples for beamtime at VMXm.
1. Equipment setup
NOTE: The methods described here use a plunge freezing instrument with a single blotting arm. Some instruments are equipped with two blotting arms and we advise the user to check manufacturers' instructions to adjust the instrument so only one blotting arm is in use.
2. Determining initial blotting parameters
3. Harvesting crystals
4. Sample density assessment by light microscopy
CAUTION: This method is destructive. If there is limited availability of sample, such as only one or two crystallization droplets, it is recommended that this step is skipped. After plunge freezing samples (step 3.7) the distribution of crystals across the grid can be assessed.
5. Sample assessment by scanning electron microscopy
NOTE: Microcrystal preparation on cryoTEM grids is best assessed with a scanning electron microscopy (SEM) under cryogenic conditions, this can be achieved with an SEM fitted with a cryo-stage. At VMXm, a JEOL JSM-IT100 SEM (tungsten source) with a Quorum PP3006 airlock and cryostage are utilized. To ensure minimal radiation damage when viewing samples on VMXm28,29, the following settings are used: 5 kV accelerating voltage; a spot size of 40 (arbitrary units on the JEOL JSM-IT100); 10 mm working distance. Images are recorded using the secondary electron detector, for sample alignment and focusing a dwell time of 0.8 µs is used while high-resolution images of single crystals are captured using a dwell time of 16 µs. It is important before loading a sample into the SEM that the microscope is aligned as per manufacturer instructions. It is advised that only a single grid is loaded into the SEM while any remaining grids prepared with the same parameters are reserved for X-ray diffraction experiments.
6. Preparing grid for diffraction experiments at VMXm
The aim of this protocol is to achieve vitrified microcrystals with a minimal volume of liquid surrounding the crystal that enable X-ray diffraction experiments with minimal background scatter to improve the diffraction signal. Example SEM images of microcrystals prepared on cryoTEM grids for minimal background scatter are shown in Figure 1A,B and Figure 2. Grids with evenly distributed single crystals will provide the most efficient use of the grid, with good signal-to-noise. However, this is not usually possible across the entirety of the grid and some regions may display some amount of clumping (Figure 2A,B). Despite this clumping, these examples still display a useful number of single isolated crystals that will provide low-background diffraction (Figure 5). The level of blotting that still maintains low-background scatter can vary. Strong blotting such that the holes in the carbon support film are clearly visible but the crystals remain hydrated is the aim (Figure 2C,D). However, good quality grids can still display some liquid within the holes of the support film although the position of the holes should still be identifiable (Figure 2A,B). Importantly, all of these examples display single, isolated crystals with a vitrified halo of liquid surrounding the crystal to maintain hydration between blotting and plunge freezing.
Many samples may require further optimization (Figure 3), which can include variation of the blotting time or the concentration of the microcrystals. Grids overloaded with crystals will demonstrate reduced blotting efficiency and can result in multiple lattices being recorded in single diffraction images (Figure 3A,B). More viscous crystallization conditions, such as that for trypsin which contains 8% PEG 4000 and 15% ethylene glycol (Figure 3C), can result in the need for longer blotting times (>10 s). Conversely, crystallization conditions with very low viscosity that blot very quickly, can suffer from distribution problems due to gravity causing settling before blotting occurs, resulting in all the microcrystals settling along one side of the grid (Figure 3D).
Optimum preparation of samples allows the full capabilities of VMXm to be exploited to collect high quality X-ray diffraction data at the highest possible resolution with high signal-to-noise (Figure 5). These samples benefit from being compatible with the in-vacuo sample environment, resulting in very low or zero average background counts in diffraction images. The use of liquid ethane, without cryo-protectant, results in an absence of ice-rings (Figure 5), although where crystals lay close to the copper grid bars, the X-ray beam can glance the bars resulting in copper powder diffraction rings at ~2.1 Ã… and ~1.8 Ã….
Figure 1: Comparison of microcrystal mounting on micromesh mounts and cyoTEM grids. Scanning electron micrographs of plunge frozen microcrystals of virus polyhedral measuring 2.5 µm across (A, B) 5 kV accelerating voltage, a spot size of 39 (arbitrary units) and dwell time of 16 µs. The grid is free from excess liquid (A), a narrow halo of liquid is observed surrounding the crystals (B). The same samples mounted on a micromesh (20 µm apertures) before flash cooling in liquid nitrogen and observed using the on-axis-viewing system at beamline I24 face-on (C) and side on (D). The optical distortions (C) when viewing face-on give an indication of the extent of the liquid thickness across the micromesh, this is clearer when viewing the micromesh side-on (D). The red target in C and D represents the X-ray beam position and size. Please click here to view a larger version of this figure.
Figure 2: Examples of good quality samples. Polyhedra crystals (A) observed across a single ~100 µm grid square. While some of the crystals are slightly clumped and show some connectivity of the surrounding liquid, there are a number of isolated single crystals with a small halo of liquid surrounding the crystal. Slightly larger insulin crystals (B) measuring ~5 x 5 µm also show some clumping but again there are well isolated individual crystals. Needle crystals can have a very narrow dimension and require a microbeam, such as these needle shaped lysozyme crystals (C). A narrow band of liquid can be seen surrounding these crystals (light grey halo). Larger microcrystals up to tens of microns also can be well mounted on cryoTEM grids, such as these ~7 µm proteinase K crystals (D). In both (C) and (D) and to a lesser extent in (A), the holes of the carbon support film are clearly visible, demonstrating the presence of very little/no liquid. In example B, while no empty holes are observed, the position of the holes can still be identified, indicating that the liquid is only filling the holes in the support film. In all of these examples, a halo of liquid can be seen around the edge of the grid square (a rounded inner square), where the holes have a lighter grey appearance. This is a common feature of well-prepared grids. Please click here to view a larger version of this figure.
Figure 3: Examples of samples requiring further optimisation. Grids overloaded with microcrystals (A, B) can increase the background scatter as the samples block the holes in the support film reducing the blotting efficiency, and with a higher surface tension between the crystals more liquid remains. As well as a degradation in the signal-to-noise resulting in a loss of information, it is likely that multiple lattices will be recorded. Using a blotting time that is not long enough, or a highly viscous crystallization solution can result in an overly wet sample (C), also producing low signal-to-noise. Crystallization conditions with no precipitants, such as those for bovine insulin (D), have a very low viscosity. While this results in a very short blotting time, it can also result in the movement of crystals across the grid before and during blotting due to the effects of gravity. This usually results in a largely empty grid with a high concentration of crystals along one edge (D). It can be advantageous to add a viscous agent such as ethylene glycol to the crystal slurry shortly before applying them to the grid to reduce crystal flow and improve the distribution of the crystals. This can also lengthen the blotting time, making it easier to observe blotting. Please click here to view a larger version of this figure.
Figure 4: Tools for sample loading at VMXm. A bespoke set of tools has been designed to load the VMXm sample holders (A). The sample loader (B) has space for a single VMXm sample holder, grid box and tool storage while working. It is also designed to allow working just below the surface of the liquid nitrogen and allow the sample cartridge to fit beneath (B). The airlock dewar (C), which fits the airlock nitrogen gas box on the VMXm endstation, is used to take the loaded sample cartridge from the lab to the beamline experimental hutch. To view samples in the offline SEM, a bespoke shuttle (D) comprising of the VMXm sample holder without base has been made to enable assessment in the sample holder that is used on the beamline. Please click here to view a larger version of this figure.
Figure 5: Example diffraction from microcrystals at VMXm. A single diffraction image recorded using a Pilatus 3 6M detector at VMXm of a ~3 µm virus polyhedral crystal mounted on a cryoTEM grid using the plunge freezing method. Diffraction is observed to beyond 1.7 Å and a reflection (blue square) at 1.74 Å is inset, demonstrating the low background scatter, with a high number of pixels with zero counts. Ethylene glycol to a final concentration of 50% v/v was added to the crystal slurry before applying it to the grid. The low background nature of the VMXm sample position is demonstrated by the constant background across the image as shown by intensities plotted beneath the blue dashed line, even near the beam center. The plotted intensities show the background remains below 3 counts. Two faint rings are observed at 2.15 Å and 1.86 Å which are generated by powder diffraction of the copper of the cryoTEM grid. There are no detectable ice-rings, demonstrating the effectiveness of blotting followed by cooling with liquid ethane. Please click here to view a larger version of this figure.
This protocol demonstrates how tools from cryoTEM sample preparation can be used for the preparation of microcrystals for X-ray diffraction experiments at microfocus beamlines. Standard beamline instrumentation is centered around a pin mounted sample and while efforts have been made to provide a sample support on these mounts for microcrystals, they are often challenging to load with sample whilst ensuring that the highest signal-to-noise is achieved (Figure 1C,D). Many of these samples may also require optimization of the cryo-protection conditions to ensure the sample is vitreous. The plunge freezing method provides a repeatable manner to remove excess liquid and flash cool the sample in an efficient cryogen (Figure 1A,B). While the grid can then be mounted on a standard beamline with a tweezer-based pin mount, VMXm sample holders have been specifically designed to accept grids and hold them below the glass transition temperature in a vacuum environment via conductive cooling. The sample environment of VMXm enables low background data collection, where the sample is the remaining source of background, and provides a microbeam that can be used to match crystals with dimensions less than 10 µm. This sample preparation method can also be used to prepare nanocrystals for electron diffraction where there is also a requirement for very little excess liquid and a vitreous sample due to the weak penetration of electrons. While cryoTEM grids are fragile, those experienced in the harvesting of crystals in loops will quickly adapt to the handling of the grids. With a small amount of experience, few grids will be lost during the blotting, freezing and loading stages of the protocol. The optimization steps are, however, critical to this success and careful preparation will reduce the chances of losing crystals or reducing crystal integrity.
CryoTEM grids provide a relatively large single mount that can contain many hundreds of crystals, thus improving throughput where it may only be possible to record a small wedge of diffraction data. A single grid may also provide enough crystals to determine the protein structure, particularly in crystals of high symmetry. Where only one or two single crystallization drops have generated microcrystals, trial blotting of the crystallization condition alone can help ensure that when the microcrystals are blotted, the times used are as close as possible to those needed to generate initial good quality samples. The carbon-film supports are invisible to X-rays and are available with differing hole spacing, which can be used to suit a particular morphology. We most commonly utilize support films with 2 µm holes at a 2 µm spacing, but smaller holes with greater spacing may be more appropriate for crystals smaller than 2 µm. Other support films such as those with 1 µm holes with a 4 µm spacing as well as support films with different shaped holes are available, all of which will affect the blotting time. A grid square mesh size of 200 (200 squares per inch) also provides enough space (~100 µm) between the copper grid bars so that the X-ray beam does not strongly interact with the copper while providing enough structural support for the carbon-film loaded with crystals. The use of liquid ethane negates the need for cryoprotectants, which in turn reduces the requirement on sample volume that would have been used in the optimization of cryoprotectant conditions.
The main parameters to be optimized during the process are the blotting times and the sample dilution. Blotting times should be long enough to observe the 'popping' effect across the entirety of the grid before plunge freezing. Over blotting could result in the dehydration of the crystals, however, control of the humidity within the sample chamber is used to minimize this effect. While it is suggested that a relative humidity of 90% used, some samples may benefit in optimization of the humidity. The humidity may affect the blotting efficiency of the blotting paper which can slowly become saturated with water. Additionally, humidity control within the sample chamber could be used to improve diffraction quality of the crystals30. It is recommended that small changes (<5%) in humidity are made before checking diffraction integrity to ensure the diffraction quality is not degraded.
Optimization of non-precious samples can be conducted using a light microscope in place of an SEM. Although destructive, it is useful for assessing the density of crystals across the grid and to enable decision making on whether a sample should be diluted or concentrated to better disperse crystals across the grid. This step is most useful when there are a large number of crystals available and particularly highly concentrated samples. Clumping together of crystals should be avoided (Figure 3), as while it is not a significant problem if two crystals are illuminated at the same time during data collection6, there will likely be a larger volume of liquid surrounding the clump, thus reducing the signal-to-noise (Figure 5). While it is possible to observe large excesses of liquid globally across the grid using a light microscope, assessment of the volume of liquid surrounding the microcrystals and the presence of crystalline ice can only be made using an electron microscope fitted with a cryogenic vacuum transfer system and stage. Sometimes, after application of the crystals to the grid and before blotting occurs, crystals in low viscosity solutions may settle along one edge of the grid. We have found that adding up to a 50% final concentration of ethylene glycol can slow the movement of crystals through the droplet, ensuring a better distribution of microcrystals across the grid as well as providing greater control over blotting by increasing the blotting time (Figure 3D).
Some crystallization solutions containing viscous precipitating agents such as high molecular weight PEGs can prove challenging to blot, requiring increasingly long blotting times (>10 s). In such cases, it can be helpful to reduce the volume of liquid deposited on the back of the grid as well as the volume of the crystal containing solution to the support-film side of the grid. Strategies such as using 2 layers of blotting paper or glass fibers may also aid blotting in these difficult cases31.
While this pipeline is suited to soluble protein crystals, those that form in very viscous mediums such as membrane proteins in LCP present a different challenge for which this protocol is not suited. However, strategies are emerging for preparing LCP crystals on cryoTEM grids for microED which include reducing the viscosity of the samples by inducing a phase change to the LCP. This permits the samples to be applied to grids in a similar manner to that described in this article. Finally, the sample can be milled with a focused ion beam to remove excess non-crystal material32,33,34.
Overall, this pipeline will generally take 1-2 h (including equipment setup time) to follow from the sample arriving at VMXm to providing optimized grids with well dispersed, vitrified samples, depending upon sample availability, the concentration of crystals and the viscosity of the crystallization solution. These methods have already been successfully employed for the preparation of microcrystals for X-ray diffraction experiments exploring radiation damage on microcrystals where a minimal volume of liquid surrounding the sample was essential28,35. It should be noted that the protocol can be applied to all soluble microcrystal samples, not just to well diffracting samples that have already been optimized. A crystallization experiment that produces microcrystalline material would traditionally be a target for optimization with the aim of obtaining larger crystals, however, this sample preparation method and the capabilities of VMXm may allow adequate data to be collected from such samples without further optimization. Alternatively, if such microcrystalline samples diffract poorly, the data collected from VMXm using this sample preparation method could still act as a useful guide for further optimization of crystallization conditions. The tools for preparing grids, including glow discharging and plunge freezing are now widely available in research institutes equipped for cryoTEM experiments and will be available to many users enabling them to prepare samples in advance of beamtime at VMXm.
The authors would like to thank Jeremy Keown, Jon Grimes, Geoff Sutton and Dave Stuart, STRUBI University of Oxford and Rachel Bolton, University of Southampton for kindly providing microcrystal samples for the development and demonstration of sample preparation methods for the VMXm beamline in addition to enabling commissioning of the beamline. The authors would also like to thank iNEXT-Discovery (project number 871037) for the opportunity and support in publishing this manuscript.
Name | Company | Catalog Number | Comments |
Automated Cryo-EM plunge freezing instrument | Leica or ThermoFisher | Various | |
Benchtop light microscope with light source | Various | Various | |
Blade/Scalpel | Fisher Scientific | Various | |
CryoTEM Copper 200 mesh grids with carbon support film with 2 µm holes | Quantifoil | N1-C16nCu20-50 | |
CryoTEM grid storage boxes | Agar Scientific | AGG3727 | |
ddH2O | n/a | n/a | |
Ethane gas supply | n/a | n/a | |
Ethylene Glycol | Acros Organics | 146750010 | |
Glass microscope slides | FisherBrand | 12383118 | |
Glass petri dish | FisherBrand | 455732 | |
Glow discharging device | Pelco | 91000S | |
Laboratory wrapping film (Parafilm) | Bemis | HS234526B | |
Large and small, fine forceps | Agar Scientific | Various | |
Liquid nitrogen supply | n/a | n/a | |
Pipette tips | Various | Various | |
Pipetting devices | Various | Various | |
Sealing tape for crystallisation plates. | Molecular Dimensions | MD6-01 | |
Small/medium liquid nitrogen dewars | Spearlab | Various | |
Sprung circlip clipping tool | Subangstrom | SCT08 | |
Whatmann No.1 pre-cut filter paper | Leica | 16706440 |