Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor

DOI :

10.3791/61670-v

7:51 min

September 29th, 2020

September 29th, 2020

7,495 Views

1Division of Respiratory Care, UT Health Science Center at San Antonio, 2Research and Development, OEM Engineering, Patient Monitoring, Medtronic, Jerusalem, 3Minimally Invasive Therapies Group, Scientific Communications, Medtronic, Minneapolis

The goal of this study was to evaluate the accuracy of capnography sampling lines used in conjunction with a portable bedside capnography monitor. Sampling lines from 7 manufacturers were evaluated for tensile strength, rise time, and ETCO2 accuracy as a function of respiratory rate or supplemental oxygen flow rate.

Tags

Capnography

-- Views

Related Videos

article

Hickman Catheter Use for Long-Term Vascular Access in a Preclinical Swine Model

article

Optimization of Processing Technology for Tiebangchui with Zanba Based on CRITIC Combined with Box-Behnken Response Surface Method

article

A 3D Digital Model for the Diagnosis and Treatment of Pulmonary Nodules

article

Establishment and Confirmation of a Postnatal Right Ventricular Volume Overload Mouse Model

article

A Stable Phantom Material for Optical and Acoustic Imaging

article

Author Spotlight: Context Dependent Blood Collection Methods in Rats

article

Author Spotlight: Unveiling Prognostic Indicators in Heart Failure - The Role of Phase Angle and Bioelectrical Impedance Analysis

article

Analgesic Effect of Tuina on Rat Models with Compression of the Dorsal Root Ganglion Pain

article

Author Spotlight: Integrating Tai Chi with Mindfulness Training to Achieve an Effective Mind-Body Exercise

article

Author Spotlight: Investigating the Effectiveness of Tuina Therapy for Frozen Shoulder

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2024 MyJoVE Corporation. All rights reserved