NMR spectrometers consist of a strong magnet, a radiofrequency transmitter, and a detector attached to a computer console for recording spectra of samples containing NMR-active nuclei. In first-generation NMR instruments called continuous-wave spectrometers, the resonance frequencies of the nuclei are determined by frequency-sweep or field-sweep methods. The magnetic field strength is fixed and the rf signal is swept in the former, while the radiofrequency signal is fixed and the magnetic field strength is gradually increased in the latter.
In modern pulsed-Fourier-Transform-NMR instruments (FT-NMR), the magnetic field is held constant while a sample is irradiated with a series of short pulses. A pulse is a short burst of energy distributed over a range of radio frequencies that excites the nuclei into higher energy states. As each nucleus relaxes in the interval or delay between pulses and returns to its original spin state, the energy is released as an electrical impulse called free induction decay or FID. The FID is then converted into a frequency-versus-amplitude signal via a mathematical technique called Fourier transform and recorded as an NMR spectrum.
Del capítulo 7:
Now Playing
Principles of Nuclear Magnetic Resonance
1.1K Vistas
Principles of Nuclear Magnetic Resonance
2.2K Vistas
Principles of Nuclear Magnetic Resonance
1.7K Vistas
Principles of Nuclear Magnetic Resonance
1.0K Vistas
Principles of Nuclear Magnetic Resonance
866 Vistas
Principles of Nuclear Magnetic Resonance
940 Vistas
Principles of Nuclear Magnetic Resonance
1.2K Vistas
Principles of Nuclear Magnetic Resonance
624 Vistas
Principles of Nuclear Magnetic Resonance
622 Vistas
Principles of Nuclear Magnetic Resonance
254 Vistas
Principles of Nuclear Magnetic Resonance
752 Vistas
Principles of Nuclear Magnetic Resonance
664 Vistas
Principles of Nuclear Magnetic Resonance
832 Vistas
ACERCA DE JoVE
Copyright © 2025 MyJoVE Corporation. Todos los derechos reservados