A scanning electron microscope (SEM) is an instrument that uses an electron beam to nondestructively image and characterize conductive materials in a vacuum. As an analogy, an electron beam is to the SEM as light is to the optical microscope. The difference is that the electron microscope yields images of much higher resolution and magnification. The best optical microscopes typically have a resolution down to 200 nm, whereas SEMs usually claim a resolution of 0.5 nm. This is due to the fact that optical microscopes are limited by the diffraction of waves, a function of the wavelength, which is around 500 nm for visible light. Conversely, the SEM uses an energized electron beam, which as a wavelength of 1 nm. This characteristic makes them very dependable tools for the study of nano and microstructures. Electron microscopes also enable the study of biological samples with feature sizes too small for optical microscopy.

This demonstration provides an introduction to sample preparation and initial image acquisition of biological samples using a scanning electron microscope. In this case, a collagen-hydroxyapatite (HA) cellular scaffold will be studied. The vacuum environment of the SEM and the induced charging by the electron beam on non-conductive samples (such as organic matter) creates challenges that will be addressed in the preparation. The advantages and disadvantages of different imaging methods as they relate to resolution, depth of focus and sample type will also be discussed. The purpose of this demonstration is to give the participant more information on SEM to determine if this microscopy module is the best fit for a type of biological sample.

Procedura

1. Sample Preparation

  1. Wear gloves and take precautions to avoid contamination when handling the sample.
  2. Make sure that the sample on the slide is dried and there is no contamination on the sample. This is because SEM measures surface characterization, and these defects can severely hinder the signal.
  3. If the sample is loaded on a standard glass slide, decrease the size of the sample by scoring the slide with a diamond tipped glass cutter in a straight line and gently push on the scored line

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Risultati

The SEM images in Figures 3 and 4 show that the imaged structure is highly three dimensional with microscale features. Image quality is affected by the focus and the thickness of the sputter coating.

Figure 4
Figure 3: The following images demonstrate how the sample focus can affect image quality.  In the image on the right, the

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Riferimenti
  1. Oatley, C. W., W. C. Nixon, and R. F. W. Pease. "Scanning electron microscopy." Advances in Electronics and Electron Physics 21 (1966): 181-247.
  2. Goldstein, Joseph, et al. Scanning electron microscopy and X-ray microanalysis: a text for biologists, materials scientists, and geologists. Springer Science & Business Media, 2012.
  3. Carol Heckman, et al. Preparation of cultural cells for scanning electron microscope. Nature Protocols Network, 2007, doi:10.1038/nprot.2007.504
Tags
SEM ImagingBiological SamplesScanning Electron MicroscopyNano ScaleOptical MicroscopesResolutionDepth Of FieldElectron BeamCondenser LensesDetectorHigh Energy Electron BeamFilament CathodeObjective LensRaster ScanningTopographyElemental CompositionCrystallinitySecondary ElectronsBackscattered Electrons

Vai a...

0:07

Overview

1:04

Principles of SEM

3:33

Preparing and Loading the Sample

4:53

Imaging the Sample with SEM

5:59

Results

6:22

Applications

8:38

Summary

Video da questa raccolta:

article

Now Playing

SEM Imaging of Biological Samples

Biomedical Engineering

22.8K Visualizzazioni

article

Imaging Biological Samples with Optical and Confocal Microscopy

Biomedical Engineering

34.8K Visualizzazioni

article

Biodistribution of Nano-drug Carriers: Applications of SEM

Biomedical Engineering

9.1K Visualizzazioni

article

High-frequency Ultrasound Imaging of the Abdominal Aorta

Biomedical Engineering

14.1K Visualizzazioni

article

Quantitative Strain Mapping of an Abdominal Aortic Aneurysm

Biomedical Engineering

4.6K Visualizzazioni

article

Photoacoustic Tomography to Image Blood and Lipids in the Infrarenal Aorta

Biomedical Engineering

5.6K Visualizzazioni

article

Cardiac Magnetic Resonance Imaging

Biomedical Engineering

14.4K Visualizzazioni

article

Computational Fluid Dynamics Simulations of Blood Flow in a Cerebral Aneurysm

Biomedical Engineering

11.5K Visualizzazioni

article

Near-infrared Fluorescence Imaging of Abdominal Aortic Aneurysms

Biomedical Engineering

8.2K Visualizzazioni

article

Noninvasive Blood Pressure Measurement Techniques

Biomedical Engineering

11.6K Visualizzazioni

article

Acquisition and Analysis of an ECG (electrocardiography) Signal

Biomedical Engineering

100.5K Visualizzazioni

article

Tensile Strength of Resorbable Biomaterials

Biomedical Engineering

7.4K Visualizzazioni

article

Micro-CT Imaging of a Mouse Spinal Cord

Biomedical Engineering

7.9K Visualizzazioni

article

Visualization of Knee Joint Degeneration after Non-invasive ACL Injury in Rats

Biomedical Engineering

8.1K Visualizzazioni

article

Combined SPECT and CT Imaging to Visualize Cardiac Functionality

Biomedical Engineering

10.9K Visualizzazioni

JoVE Logo

Riservatezza

Condizioni di utilizzo

Politiche

Ricerca

Didattica

CHI SIAMO

Copyright © 2025 MyJoVE Corporation. Tutti i diritti riservati