JoVE Logo

Zaloguj się

Aby wyświetlić tę treść, wymagana jest subskrypcja JoVE. Zaloguj się lub rozpocznij bezpłatny okres próbny.

W tym Artykule

  • Podsumowanie
  • Streszczenie
  • Wprowadzenie
  • Protokół
  • Wyniki
  • Dyskusje
  • Ujawnienia
  • Podziękowania
  • Materiały
  • Odniesienia
  • Przedruki i uprawnienia

Podsumowanie

Accuracy is a major demand in dental medicine. To verify accuracy, reference scanners are needed. This article presents a new reference scanner with an adjusted scanning method to acquire a broad variety of dental morphologies with high trueness and precision.

Streszczenie

Reference scanners are used in dental medicine to verify a lot of procedures. The main interest is to verify impression methods as they serve as a base for dental restorations. The current limitation of many reference scanners is the lack of accuracy scanning large objects like full dental arches, or the limited possibility to assess detailed tooth surfaces. A new reference scanner, based on focus variation scanning technique, was evaluated with regards to highest local and general accuracy. A specific scanning protocol was tested to scan original tooth surface from dental impressions. Also, different model materials were verified. The results showed a high scanning accuracy of the reference scanner with a mean deviation of 5.3 ± 1.1 µm for trueness and 1.6 ± 0.6 µm for precision in case of full arch scans. Current dental impression methods showed much higher deviations (trueness: 20.4 ± 2.2 µm, precision: 12.5 ± 2.5 µm) than the internal scanning accuracy of the reference scanner. Smaller objects like single tooth surface can be scanned with an even higher accuracy, enabling the system to assess erosive and abrasive tooth surface loss. The reference scanner can be used to measure differences for a lot of dental research fields. The different magnification levels combined with a high local and general accuracy can be used to assess changes of single teeth or restorations up to full arch changes.

Wprowadzenie

Accuracy is a major interest in many fields in dental medicine. Replacing dental hard tissue needs an exact fitting prosthesis to ensure proper function and prevent further destroying the remaining tooth structure1,2. Fixed partial dentures and total prosthesis are especially critical for exact fitting at supporting structures like prepared teeth or implants3. This is why a highly accurate reproduction is needed, especially in the field of dental impressions and dental laboratory workflow. However, other fields of dental treatment also benefit from a true and precise metrical outcome, to verify treatment success and evaluate new treatment strategies, e.g. soft and hard tissue augmentation, erosion and abrasion monitoring, periodontal treatments, and orthodontic treatments4,5. In many of these fields, current validation procedures are linear distance measurements with calipers or microscopes6,7. These methods are limited to only few measuring points and limited information of three-dimensional (3D) changes of the testing area. Newer measuring methods include the optical or radiographic capturing of the entire surface of the test object8,9. Here, the entire surface or volume is measured and displayed as a 3D object on the computer screen. Linear measurements are possible, as well as superimpositions of models from different scan times. With this superimposition, an evaluation of the surface changes at every scan point is possible. This enables the possibility of monitoring a specific area or displaying deformations in all three coordinate axes. Also, volumetric changes can be measured10. The limiting point with these new methods is the accuracy of the scanner, used to capture the test object. None of the changes within the accuracy of the reference scanner can be divided into changes of the test object or scan errors. Scan accuracy is often a value given by the manufacturer derived from scanning small, calibrated objects11. This minimal scan error is different when scanning large objects like a dental arch. Accuracy consists of trueness and precision. Trueness is the deviation of the scanned object from its real geometry. Precision is the deviation between repeated scans (ISO 5725-1). In this study, a new optical reference scanner, based on the focus variation scanning technique, was introduced to scan specimens from single tooth up to full arch models with highest accuracy. This reference scanner was used as a base for several studies, comparing dental impression accuracy from conventional and digital techniques12–14 and for actual projects concerning dental occlusion and abrasion of dental materials. The aim of this study was to provide basic information of the accuracy of the reference scanner and some possibilities to use this device in the field of dental research.

Protokół

1. Specimen Preparation/Base

  1. Apply a flat base around the specimen. Place the specimen on the scanning table. Orientate the occlusal surface to the horizontal plane (Figure 1).

2. Software Analysis

  1. Start the program and then the Laboratory Measurement Module (Figure 2)
  2. Position the specimen at the center of the scanning table.
  3. Choose the correct magnification lens. For large objects, like full arch scans, use the 5X objective.
  4. Move the scan optic by using the 3D mouse until the surface of the specimen is displayed on the live view window (Figure 3).
  5. Use sensor control to set Exposure and Contrast to achieve optimal scanning parameters. For metal surfaces, use an exposure between 400 and 800 µsec and a contrast between 0.3 and 0.8 (Figure 4).
  6. Check image quality by checking the button "Show Image Quality in Live Preview".
  7. Set the correct parameters in the section "Measurement Control". The measurement type is "3D Dataset"; ImageField Type is "General ImageField" (Figure 5).
  8. Click on "New ImageField".
  9. Define the scan volume. The software needs the bounding coordinates of the measure volume.
    1. Move the specimen to the highest and lowest scan level and click on "Add Position" at both points. The "Z Range" value shows the actual height of the scan volume (Figure 6).
    2. Move the specimen to the X-Y boundings of the scan volume. Click on "Add Position" to define X-axis and Y-axis lengths. The dimension of the scan volume is shown at the "Information" tab and should exceed the dimensions of the specimen by 1 cm in the X-axis and Y-axis.
  10. Check the "Points" number. The software is capable of scanning 100 million surface points in one "ImageField" scan. The actual number of points exceeds this limit. "Decimate" the measure point's number by a "Lateral Downsampling" of the point size.
    1. Click on "Advanced Settings" (Figure 7).
    2. Move the "Lateral Downsampling" slider to the right until the "Points" number is reduced below 100 M (Figure 7). Downsampling reduces the lateral resolution of the surface points, resulting in larger pixel size of the scanned object. To ensure optimal scan results, the vertical resolution is increased.
  11. Click on "Start Measurement". This will start the "Preview Mode". The software performs a prescan with the selected X- and Y-axis dimensions.
  12. After the prescan is completed, select the region of interest. This helps to reduce file size and scanning time (Figure 8).
  13. Select all parts of the prescan except the specimen and two measure fields around the specimen, containing the flat base (Figure 9).
  14. Click on "Start" to begin the scan.
  15. Control the scan in the "ImageViewer" using the mouse and press the left mouse button (Figure 10).
  16. Close the window and click on "View Pseudo Color Only".
  17. Click on "Settings" and "Pseudo Coloring", and choose "Repeatability".
  18. Set the "Max." value to 0.2 and click on "Apply Range".
  19. Control the repeatability, it should be equal for areas with the same incline and material. In particular, the base around the specimen should display a homogeneous repeatability (Figure 11).
  20. Click on "Database" and save the scan to the appropriate folder.
  21. Export the scan to different file formats, if needed. Click on "File/Export/3D Data As/…." The protocol can be paused at that point and continued later.

3. Difference Analysis

  1. Use the "3D-Editor" to cut away the base. This area will not be used for the difference analysis.
  2. To compare and analyze two scans, start the "Difference Measurement" software (Figure 12).
  3. Choose the second model to compare.
  4. Click on "Automatic Rough Alignment" to perform a first match of the models.
  5. Click on "Manual Adjustment" and align the models with rotating and moving until they are in the same orientation (Figure 13).
  6. Click on "Automatic Alignment" and "Apply" to start the best fit algorithm for the optimal model matching.
  7. Click on "Differences" to view the difference map of the two matched models.
  8. Select a proper color range to display the deviations (Figure 14). Save the visual deviation pattern as a screenshot for analysis.

Click on "Statistics" to display the statistical values of the differences. Choose 1 µm class size and save histogram data to a text file for statistical comparison (Figure 15).

Wyniki

Figure 16A shows the scan of a gypsum specimen. To verify optimal scan parameters, control the repeatability for the material. A section of the specimen is scanned with different contrast and brightness settings and the repeatability is checked after each scan to find the optimal scan settings. Metal surfaces show the best results with a contrast lower 1.0 and gypsum or resin materials with a higher contrast. Figure 16B shows the repeatability without optimal scanning parameters after th...

Dyskusje

Accuracy is a basic demand in dental medicine. The reference scanner is capable of scanning small and large objects with high trueness and precision. With the optimal scanning method, even individual morphological detailed dental surfaces can be scanned with a high resolution and repeatability. With the different magnification levels of the scanner, macro and micro morphological structures can be acquired. It is possible to scan a variety of model materials.

The evaluation of the reference sca...

Ujawnienia

The authors declare that they have no competing financial interests or other conflicts of interest.

Podziękowania

The authors thank the dental technician Nicola Lanfranconi for producing the master reference model and the Alicona Company for their ongoing support with improving the scan software.

Materiały

NameCompanyCatalog NumberComments
Reference modelindividual non-precious metal model, derived from a patient impression
Araldit repairHuntsmen Advanced Material, Basel, Switzerlandused for making the base of the reference model
CamBaseDentona, Dortmund, GermanyType IV dental ston for pouring conventional impressions
IdentiumKettenbach, Eschenburg, GermanyVinylsiloxanether impression material for conventional impression
inEOS model holderSirona Dental Systems, Bensheim, Germanyused for fixing stone models at the reference scanner
AccutransColtene Whaledent, Altstätten, Switzerlandused for making the base of thestone models
President puttyColtene Whaledent, Altstätten, Switzerlandmix with accutrans for betterstability of the base
Alicona Infinite FocusAlicona Imaging, Graz, AustriaReference scanner 

Odniesienia

  1. Wettstein, F., Sailer, I., Roos, M., Hämmerle, C. H. Clinical study of the internal gaps of zirconia and metal frameworks for fixed partial dentures. Eur. J. Oral Sci. 116 (3), 272-279 (2008).
  2. Persson, A. S., Oden, A., Andersson, M., Sandborgh-Englund, G. Digitization of simulated clinical dental impressions: virtual three-dimensional analysis of exactness. Dent. Mater. 25 (7), 929-936 (2009).
  3. Del'Acqua, M. A., Arioli-Filho, J. N., Compagnoni, M. A., Mollo, F. de A Accuracy of impression and pouring techniques for an implant-supported prosthesis. Int. J. Oral Maxillofac. Implants. 23 (2), 226-236 (2008).
  4. Schneider, D., Grunder, U., Ender, A., Hämmerle, C. H. F., Jung, R. E. Volume gain and stability of peri-implant tissue following bone and soft tissue augmentation: 1-year results from a prospective cohort study. Clinical Oral Implants Research. 22 (1), 28-37 (2011).
  5. Windisch, S. I., Jung, R. E., Sailer, I., Studer, S. P., Ender, A., Hämmerle, C. H. F. A new optical method to evaluate three-dimensional volume changes of alveolar contours: a methodological in vitro study. Clinical Oral Implants Research. 18 (5), 545-551 (2007).
  6. Caputi, S., Varvara, G. Dimensional accuracy of resultant casts made by a monophase, one-step and two-step, and a novel two-step putty/light-body impression technique: an in vitro. 99 (4), 274-281 (2008).
  7. Hoyos, A., Soderholm, K. J. Influence of tray rigidity and impression technique on accuracy of polyvinyl siloxane impressions. Int. J. Prosthodont. 24 (1), 49-54 (2011).
  8. Luthardt, R. G., Kuhmstedt, P., Walter, M. H. A new method for the computer-aided evaluation of three-dimensional changes in gypsum materials. Dent. Mater. 19 (1), 19-24 (2003).
  9. Mehl, A., Ender, A., Mörmann, W., Attin, T. Accuracy testing of a new intraoral 3D camera. International Journal of Computerized Dentistry. 12 (1), 11-28 (2009).
  10. Fickl, S., et al. Dimensional changes of the ridge contour after socket preservation and buccal overbuilding: an animal study. J. Clin. Periodontol. 36 (5), 442-448 (2009).
  11. Vlaar, S. T., vander Zel, J. M. Accuracy of dental digitizers. Int. Dent. J. 56 (5), 301-309 (2006).
  12. Ender, A., Mehl, A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. 109 (2), 121-128 (2013).
  13. Ender, A., Mehl, A. Influence of scanning strategies on the accuracy of digital intraoral scanning systems. Int. J. Comput. Dent. 16 (1), 11-21 (2013).
  14. Ender, A., Mehl, A. Full arch scans: conventional versus digital impressions--an in-vitro study. Int. J. Comput. Dent. 14 (1), 11-21 (2011).
  15. Meer, W. J., Andriessen, F. S., Wismeijer, D., Ren, Y. Application of intra-oral dental scanners in the digital workflow of implantology. PLoS One. 7 (8), (2012).

Przedruki i uprawnienia

Zapytaj o uprawnienia na użycie tekstu lub obrazów z tego artykułu JoVE

Zapytaj o uprawnienia

Przeglądaj więcej artyków

Dental MedicineDental ImpressionsReference ScannerFocus Variation ScanningTruenessPrecisionDental RestorationsTooth SurfaceScanning AccuracyErosionAbrasion

This article has been published

Video Coming Soon

JoVE Logo

Prywatność

Warunki Korzystania

Zasady

Badania

Edukacja

O JoVE

Copyright © 2025 MyJoVE Corporation. Wszelkie prawa zastrzeżone