JoVE Logo

Oturum Aç

Bu içeriği görüntülemek için JoVE aboneliği gereklidir. Oturum açın veya ücretsiz deneme sürümünü başlatın.

Bu Makalede

  • Overview
  • Protokol
  • Sonuçlar
  • Açıklamalar
  • Malzemeler
  • Referanslar

Overview

This video demonstrates a microelectrode array-based assay for studying neuronal network activity in mouse spinal cord sections. First, the electrophysiological activity from the superficial dorsal horn (SDH) of the slice is recorded. Then, a potassium channel inhibitor is introduced to prolong depolarization, resulting in synchronous rhythmic activity across the neuronal network.

Protokol

All procedures involving sample collection have been performed in accordance with the institute's IRB guidelines.

1. In vitro electrophysiology

  1. Preparation of solutions for spinal cord slice preparation and recording
    1. Artificial cerebrospinal fluid
      ​NOTE: Artificial cerebrospinal fluid (aCSF) is used in an interface incubation chamber, where slices are stored until recording commences and during experiments as both perfusate and diluent for drugs. See Table 1 for the detailed composition.
  2. Prepare aCSF containing (in mM) 118 NaCl, 25 NaHCO3, 10 glucose, 2.5 KCl, 1 NaH2PO4, 1 MgCl2, and 2.5 CaCl2 by adding the required quantities of the above, excluding CaCl2, to 2 L of distilled water.
  3. Bubble the above solution with carbogen (95% O2, 5% CO2) for 5 min and add CaCl2.
    NOTE: This step prevents CaCl2 precipitation, i.e., the solution should not turn cloudy. For drug application during experiments, dilute the drug stock solutions in aCSF to desired final concentrations.

2. Sucrose-substituted artificial cerebrospinal fluid

NOTE: Sucrose-substituted aCSF is used during dissection and spinal cord slicing. As indicated by the name, sucrose is substituted for NaCl to reduce neuronal excitation during these procedures while maintaining osmolarity. See Table 1 for the detailed composition.

  1. Prepare sucrose-substituted aCSF containing (in mM) 250 sucrose, 25 NaHCO3, 10 glucose, 2.5 KCl, 1 NaH2PO4, 1 MgCl2, and 2.5 CaCl2 by adding the required quantities of all of the above, excluding CaCl2, to 300 mL of distilled water.
  2. Bubble the solution with carbogen for 5 min and then add CaCl2.
  3. Store the solution in a -80 °C freezer for approximately 40 min or until the solution forms a slurry. Avoid freezing solid and use while in slurry consistency.

3. Microelectrode array preparation

NOTE: The contact surface of the MEA requires a pretreatment to make it hydrophilic.

  1. Before the experiment, fill the MEA well with either fetal bovine serum (FBS) or horse serum (HS) for 30 min.
  2. Remove the FBS or HS and thoroughly rinse MEA with approximately five washes of distilled water until the distilled water is no longer foamy. Fill the well with aCSF, ready for use.

4. Acute spinal cord slice preparation

  1. Deeply anesthetize the mouse with 100 mg/kg ketamine (i.p.) and then decapitate it using large surgical scissors.
  2. Remove the skin over the abdominal region by making a small cut in the skin at the level of the hips. Pull the skin on either side of the cut rostrally until all the skin is removed, i.e., from the top of the rib cage to the top of the pelvis (both ventrally and dorsally).
  3. Place the body on ice and use a ventral approach to expose the vertebral column by removing all the viscera and cutting through the ribs lateral to the sternum.
  4. Remove the ventral rib cage, both scapulae (cut off at approximately T2), and the lower limbs and pelvis (cut off at approximately the top of the sacrum).
  5. Transfer the vertebral column and rib preparation to a dissecting bath containing ice-cold sucrose aCSF. Pin all four corners of the preparation (ventral surface upwards) by placing pins through the lower back muscles and the attached upper ribs.
  6. Remove all muscle and connective tissue overlying the ventral surface of the vertebrae with rongeurs and identify the vertebral region over the lumbosacral enlargement, which lies approximately beneath the T12 to L2 vertebral bodies.
  7. Remove a vertebral body that is caudal to the lumbosacral enlargement region to provide access to the spinal cord as it sits in the vertebral canal.
  8. Using curved spring scissors, cut through the vertebral pedicles bilaterally while lifting and pulling the vertebral body rostrally to separate the ventral and dorsal aspects of the vertebrae and expose the spinal cord.
  9. Once the vertebral bodies are removed to reveal the lumbosacral enlargement, carefully clear the remaining roots that anchor the spinal cord with spring scissors until the cord floats free.
  10. Isolate the spinal cord with rostral and caudal cuts well above and below the lumbosacral enlargement, allowing the target region of the cord to 'float free.'
    NOTE: The preferred slice orientation will determine how the cord is subsequently mounted for sectioning (Figure 1).
  11. For transverse slices, lift the lumbosacral segment by an attached root and place it on a pre-cut polystyrene (Styrofoam) block (1 cm x 1 cm x 1 cm) with a shallow channel cut in the center. Use cyanoacrylate adhesive (see the Table of Materials) to attach the block and cord to the sectioning platform and place it in the cutting bath containing ice-cold sucrose aCSF (slurry).
    NOTE: The shallow channel helps secure and orient the spinal cord, with the dorsal side exposed and the thoracic end of the cord at the bottom of the block.
  12. For sagittal slices, lay a thin line of cyanoacrylate adhesive on the sectioning platform, lift the lumbosacral enlargement by an attached root, and place the cord along the line of glue, ensuring one lateral surface is in the adhesive and the other faces upwards. Place it in the cutting bath containing ice-cold sucrose aCSF (slurry).
  13. For horizontal slices, put a thin line of cyanoacrylate adhesive on the sectioning platform. Lift the lumbosacral enlargement by an attached root, and place the lumbosacral enlargement along the line of adhesive, ensuring the ventral surface is in the adhesive and the dorsal surface faces upwards. Use attached roots to position the cord. Place it in the cutting bath containing ice-cold sucrose aCSF (slurry).

5. Microelectrode array recordings

NOTE: The following steps detail how to use record data from MEA-based experiments on spinal cord slices. Several MEA designs can be used depending on the experiment. Design details for MEAs used in these experiments are shown in Table 2 and Figure 2. Detailed design information has been published by Egert et al. and Thiebaud et al. for planar and 3-dimensional (3D) MEAs, respectively. Both MEA types are composed of 60 titanium nitride electrodes, with a silicon nitride insulating layer and titanium nitride tracks and contact pads.

  1. Experimental setup
    1. Turn on the computer and interface board, and start the recording software.
    2. Load the pre-assembled recording template (Figure 3A). Name the files for the day in the recorder tab.
    3. Continuously bubble aCSF with carbogen (5% CO2, 95% O2) for the duration of the experiment.
    4. Turn the perfusion system on, which is controlled by a peristaltic pump. Place the inlet line into aCSF and the inlet end in a waste beaker. Prime the perfusion lines with aCSF.
    5. Prepare 4-AP and any other drug solutions by diluting stocks in 50 mL of aCSF to the required final concentration (e.g., 200 µM for 4-AP).
  2. 4-aminopyridine (4-AP) activity
    1. Following incubation, transfer a single slice from the incubator using a large-tip Pasteur pipette filled with aCSF.
    2. Place the slice in the MEA well and add additional aCSF.
    3. Position the slice over the 60-electrode recording array using a fine short hair paintbrush. Avoid contacting the electrodes with the paintbrush or dragging the tissue across the electrodes, especially if using 3D arrays.
      NOTE: Depending on the MEA layout, this can be done with or without the assistance of a microscope for accurate positioning.
    4. After positioning the slice, place a weighted net over the tissue to hold it in place and promote good contact with MEA electrodes.
      NOTE: The slice may need repositioning following net placement.
    5. Place the MEA in the recording headstage (Figure 2A, B).
    6. Check the position of the tissue over the electrodes using an inverted microscope (2x magnification) to confirm that as many electrodes as possible are under the superficial dorsal horn (SDH). Ensure that at least 2-6 electrodes do not contact the slice as these electrodes are important for subtracting noise and recording artefacts during analysis (Figure 2E).
    7. Turn on the camera, connect it to the device, and take a reference image of the slice relative to the MEA for use during analysis.
    8. Press Start DAQ (data acquisition) in the recording software and confirm that all electrodes are receiving a clear signal.
      NOTE: If the signal is noisy, unclip the headstage, and clean both the MEA contact pads and gold spring contacts with 70% ethanol (use a laboratory wipe to ensure that the pads and contacts are dry after cleaning). If the signal is still noisy, turn off the malfunctioning electrodes in the recording software or note down for exclusion later during analysis.
    9. Attach the perfusion inlet and outlet lines to the MEA-well (previously filled with aCSF) and turn the perfusion system on. Check the flow rate, ideally 4-6 bath volumes per minute, and ensure that the outflow is sufficient to prevent overflow of the superfusate.
    10. Allow the tissue to equilibrate for 5 min and then record 5 min of raw, unfiltered baseline data.
    11. Move the perfusion inlet line from aCSF to a 4-AP solution and wait for 12 min for the 4-AP-induced rhythmic activity to reach steady state (2 min for drugs to reach the bath and 10 min for the activity to peak and then plateau).
    12. Record 5 min of 4-AP-induced activity. Be prepared for subsequent recordings to test the drugs or to check the stability of 4-AP.

Table 1: Artificial Cerebrospinal Fluid compositions.

ChemicalaCSF (mM)aCSF (g/100 mL)Sucrose-substituted aCSF (mM)Sucrose-substituted aCSF (g/100 mL)High-potassium aCSF (mM)High-potassium aCSF (g/100 mL)
Sodium chloride (NaCl)1180.690--1180.690
Sodium hydrogen carbonate (NaHCO3)250.210250.210250.210
Glucose100.180100.180100.180
Potasium chloride (KCl)2.50.0192.50.0194.50.034
Sodium dihydrogen phosphate (NaH2PO4)10.01210.01210.012
Magnesium cloride (MgCl2)10.0110.0110.01
Calcium chloride (CaCl2)2.50.0282.50.0282.50.028
Sucrose--2508.558--

Table 2: Microelectrode array layouts.

Microelectrode Array Layouts
Microelectrode Array Model60MEA 200/30iR-Ti60-3DMEA 100/12/40iR-Ti60-3DMEA 200/12/50iR-Ti60MEA 500/30iR-Ti
Planar or 3-Dimensional (3D)Planar3D3DPlanar
Electrode Grid8 x 88 x 88 x 86 x 10
Electrode Spacing200 µm100 µm200 µm500 µm
Electrode Diameter30 µm12 µm12 µm30 µm
Electrode Height (3D)N/A40 µm50 µmN/A
ExperimentsTransverse sliceTransverse sliceSagittal + HorizontalSagittal + Horizontal

Sonuçlar

figure-results-58
Figure 1: Spinal cord slice orientations, mounting and cutting methods. (A) Transverse slices require a Styrofoam cutting block with a supporting groove cut into it. The spinal cord is rested against the block in the support groove, the dorsal side of the cord facing away from the block. The block and cord are glued onto a cutting stage with cyanoacrylate adhesive. (B

Açıklamalar

No conflicts of interest declared.

Malzemeler

NameCompanyCatalog NumberComments
4-aminopyridineSigma-Aldrich275875-5G
100% ethanolThermo FisherAJA214-2.5LPL
CaCl2 1MBanksia Scientific0430/1L
Carbonox (Carbogen - 95% O2, 5% CO2)Coregas219122
Curved long handle spring scissorsFine Science Tools15015-11
Custom made air interface incubation chamber
Foetal bovine serumThermo Fisher10091130
Forceps Dumont #5Fine Science Tools11251-30
GlucoseThermo FisherAJA783-500G
Horse serumThermo Fisher16050130
Inverted microscopeZeissAxiovert10
KClThermo FisherAJA383-500G
KetamineCevaKETALAB04
Large surgical scissorsFine Science Tools14007-14
Loctite 454 Instant AdhesiveBolts and Industrial SuppliesL4543G
MATLABMathWorksR2018b
MEAs, 3-DimensionalMultichannel Systems60-3DMEA100/12/40iR-Ti, 60-3DMEA200/12/50iR-Ti60 titanium nitride (TiN) electrodes with 1 internal reference electrode, organised in an 8x8 square grid. Electrodes are 12 µm in diameter, 40 µm (100/12/40) or 50 µm (200/12/50) high and equidistantly spaced 100 µm (100/12/40) or 200 µm (200/12/50) apart.
MEA headstageMultichannel SystemsMEA2100-HS60
MEA interface boardMultichannel SystemsMCS-IFB 3.0 Multiboot
MEA netMultichannel SystemsALA HSG-MEA-5BD
MEA perfusion systemMultichannel SystemsPPS2
MEAs, PlanarMultichannel Systems60MEA200/30iR-Ti, 60MEA500/30iR-Ti60 titanium nitride (TiN) electrodes with 1 internal reference electrode, organised in either a 8x8 square grid (200/30) or a 6x10 rectangular grid (500/30). Electrodes are 30 µm in diameter and equidistantly spaced 200 µm (200/30) or 500 µm (500/30) apart.
MgCl2Thermo FisherAJA296-500G
Microscope cameraMoticMoticam X Wi-Fi
Multi Channel Analyser softwareMultichannel SystemsV 2.17.4
Multi Channel Experimenter softwareMultichannel SystemsV 2.17.4
NaClThermo FisherAJA465-500G
NaHCO3Thermo FisherAJA475-500G
NaH2PO4Thermo FisherACR207805000
RongeursFine Science Tools16021-14
Small spring scissorsFine Science Tools91500-09
Small surgical scissorsFine Science Tools14060-09
SucroseThermo FisherAJA530-500G
Supergluecyanoacrylate adhesive
TetrodotoxinAbcamAB120055
Vibration isolation tableNewportVH3048W-OPT
Vibrating microtomeLeicaVT1200 S

Referanslar

This article has been published

Video Coming Soon

JoVE Logo

Gizlilik

Kullanım Şartları

İlkeler

Araştırma

Eğitim

JoVE Hakkında

Telif Hakkı © 2020 MyJove Corporation. Tüm hakları saklıdır