JoVE Logo

Sign In

β-Lactamase-Based Conductimetric Biosensor Assay for Protein-Protein Interactions


Transcript


To begin the β-lactamase-based conductimetric biosensor assay, take a transducer chip with working, counter, and reference electrodes. The working electrode has a conductive polymer coating, facilitating target antigen immobilization. The reference and counter electrodes lack this polymer coating.

Add a solution containing the target antigen onto the chip. Incubate. The antigen gets immobilized on the working electrode surface via non-covalent interactions. Wash to remove unbound antigens.

Add a blocking solution. Proteins in the solution attach to unbound sites on the working electrode and block them.

Add a bifunctional chimeric protein to the chip to detect the immobilized antigen. The chimeric protein comprises a β-lactamase enzyme — a bacterial protein responsible for antibiotic resistance. The enzyme is fused to a nanobody — the N-terminal variable region of single-domain antibodies.

Upon incubation, the nanobody of the protein binds to the immobilized target antigen. Wash to remove unbound chimeric proteins.

Plug the chip into a computer-controlled digital multimeter. Add a detection solution containing the antibiotic benzylpenicillin — a substrate for β-lactamase.

Benzylpenicillin binds to the β-lactamase of the antigen-bound chimeric protein and gets hydrolyzed — causing the release of protons. The protons induce a change in the electrical conductance of the polymer.

Plot the real-time difference in conductance between the reference electrode and the working electrode with the immobilized target antigen, confirming its interaction with the chimeric protein.

USAGE STATISTICS
JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2024 MyJoVE Corporation. All rights reserved