Atomic Absorption Spectroscopy (AAS) atomizes samples through flame atomization or electrothermal atomization. Flame atomization typically involves a nebulizer and spray chamber assembly to combine the sample with a fuel–oxidant mixture, creating a fine aerosol mist that enters a burner. Typically, the fuel and oxidant are combined in an approximately stoichiometric ratio. However, for atoms that are easily oxidized, a fuel-rich mixture may be more advantageous. Only about 5% of the aerosol droplets reach the flame, where they undergo desolvation in the primary combustion zone, leaving bare particles atomized in the inner flame. The gaseous atoms, ions, and molecular species quickly pass through the interzonal region for analysis before exiting the flame. Flame atomizers have low atomization efficiency due to large aerosol droplets not reaching the flame and significant sample dilution from combustion gases. However, the flame atomization efficiencies can be enhanced by continuously aspirating the sample, optimizing fuel-to-oxidant ratios, adjusting nebulizer flow rates, and setting the burner height.

Flame atomization is not suitable for samples with low analyte concentrations or limited volumes due to the small number of samples successfully atomized and detected. In contrast, electrothermal atomization, also known as graphite furnace atomization, uses a graphite tube to capture and concentrate the analytes and works well for small, discrete samples. In this technique, the sample is dried and charred before being atomized at high temperatures.

Elements such as As, Se, Sb, Bi, Ge, Sn, Te, and Pb can be atomized under milder conditions by chemically converting them into volatile hydrides before carrying them to the flame. Additionally, mercury determination can utilize a unique cold-vapor method due to its natural volatility.

来自章节 14:

article

Now Playing

14.6 : Atomic Absorption Spectroscopy: Atomization Methods

Atomic Spectroscopy

173 Views

article

14.1 : Atomic Spectroscopy: Absorption, Emission, and Fluorescence

Atomic Spectroscopy

376 Views

article

14.2 : Atomic Spectroscopy: Effects of Temperature

Atomic Spectroscopy

139 Views

article

14.3 : Atomic Absorption Spectroscopy: Overview

Atomic Spectroscopy

282 Views

article

14.4 : Atomic Absorption Spectroscopy: Instrumentation

Atomic Spectroscopy

218 Views

article

14.5 : Atomic Absorption Spectroscopy: Radiation and Light Sources

Atomic Spectroscopy

172 Views

article

14.7 : Atomic Absorption Spectroscopy: Interference

Atomic Spectroscopy

254 Views

article

14.8 : Atomic Absorption Spectroscopy: Lab

Atomic Spectroscopy

144 Views

article

14.9 : Atomic Emission Spectroscopy: Overview

Atomic Spectroscopy

307 Views

article

14.10 : Atomic Emission Spectroscopy: Instrumentation

Atomic Spectroscopy

145 Views

article

14.11 : Atomic Emission Spectroscopy: Interference

Atomic Spectroscopy

72 Views

article

14.12 : Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle

Atomic Spectroscopy

292 Views

article

14.13 : Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation

Atomic Spectroscopy

98 Views

article

14.14 : Atomic Emission Spectroscopy: Lab

Atomic Spectroscopy

74 Views

article

14.15 : Atomic Fluorescence Spectroscopy

Atomic Spectroscopy

112 Views

See More

JoVE Logo

政策

使用条款

隐私

科研

教育

关于 JoVE

版权所属 © 2025 MyJoVE 公司版权所有,本公司不涉及任何医疗业务和医疗服务。