High-frequency Ultrasound Imaging of the Abdominal Aorta

Panoramica

Source: Amelia R. Adelsperger, Evan H. Phillips, and Craig J. Goergen, Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana

High-frequency ultrasound systems are used to acquire high resolution images. Here, the use of a state-of-the-art system will be demonstrated to image the morphology and hemodynamics of small pulsatile arteries and veins found in mice and rats. Ultrasound is a relatively inexpensive, portable, and versatile method for the noninvasive assessment of vessels in humans as well as large and small animals. These are several key advantages that ultraound offers compared to other techniques, such as computed tomography (CT), magnetic resonance imaging (MRI), and near-infrared fluorescence tomography (NIRF). CT requires ionizing radiation and MRI can be prohibitively expensive and even impractical in some scenarios. NIRF, on the other hand, is limited by the penetration depth of light required to excite the fluorescent contrast agents.

Ultrasound has limitations in terms of imaging depth; however, this may be overcome by sacrificing resolution and using a lower frequency transducer. Abdominal gas and excess body weight can severely diminish image quality. In the first case, the propagation of sound waves is limited, while in the latter case, they are attenuated by overlying tissues, such as fat and connective tissue. As a result, no contrast or faint contrast may be observed. Finally, ultrasound is a highly user-dependent technique, requiring the sonographer to be familiar with anatomy and to be able to work around issues, such as the appearance of imaging artifacts or acoustic interference.

Procedura

1. Image setup

  1. Turn the ultrasound system on using the switch on the back. Turn the monitor on.
  2. Plug in the physiological monitoring unit and turn on heart rate and temperature monitoring. Turn on the gel warmer and ensure the light is on.
  3. Check the isoflurane level in the anesthetic vaporizer and refill if necessary.
  4. Turn on your O2 tank or filtered air source and adjust the air flow on the vaporizer to approximately 1 L/min.
  5. Attach the mouse or rat stage, a

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Risultati

This procedure allowed the anatomical and functional imaging of the abdominal aorta. Acquiring real-time images in short-axis and long-axis by B-mode, M-mode, and Doppler ultrasound takes at least thirty minutes and therefore requires careful monitoring of the anesthetized animal. Some data are readily analyzed on-the-fly, such as two-dimensional B-mode scans (Fig. 1). These data can provide aortic diameter or cross-sectional area measurements. Other data, such as three-dimensional B-mode (Fig. 2), M-mode (Fig. 3), Color

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Tags
High frequency Ultrasound ImagingAbdominal AortaNoninvasive Imaging TechnologyClinical ImagingUltrasound TechnologyLive ImagesAnatomical StructuresOrgansAdvantages Of UltrasoundLimitations Of UltrasoundHigh frequency Ultrasound SystemImaging Blood VesselsUltrasound Imaging ApplicationsAcoustic WavesTransducerEchoesAcoustic ImpedanceTissue DensitySpeed Of Sound Wave

Vai a...

0:07

Overview

1:05

Principles of Ultrasound Imaging

3:07

Ultrasound Imaging Set-up

5:42

Ultrasound Image Acquisition

10:25

Results

12:29

Applications

13:58

Summary

Video da questa raccolta:

article

Now Playing

High-frequency Ultrasound Imaging of the Abdominal Aorta

Biomedical Engineering

14.1K Visualizzazioni

article

Imaging Biological Samples with Optical and Confocal Microscopy

Biomedical Engineering

34.8K Visualizzazioni

article

SEM Imaging of Biological Samples

Biomedical Engineering

22.8K Visualizzazioni

article

Biodistribution of Nano-drug Carriers: Applications of SEM

Biomedical Engineering

9.1K Visualizzazioni

article

Quantitative Strain Mapping of an Abdominal Aortic Aneurysm

Biomedical Engineering

4.6K Visualizzazioni

article

Photoacoustic Tomography to Image Blood and Lipids in the Infrarenal Aorta

Biomedical Engineering

5.6K Visualizzazioni

article

Cardiac Magnetic Resonance Imaging

Biomedical Engineering

14.4K Visualizzazioni

article

Computational Fluid Dynamics Simulations of Blood Flow in a Cerebral Aneurysm

Biomedical Engineering

11.5K Visualizzazioni

article

Near-infrared Fluorescence Imaging of Abdominal Aortic Aneurysms

Biomedical Engineering

8.2K Visualizzazioni

article

Noninvasive Blood Pressure Measurement Techniques

Biomedical Engineering

11.6K Visualizzazioni

article

Acquisition and Analysis of an ECG (electrocardiography) Signal

Biomedical Engineering

100.5K Visualizzazioni

article

Tensile Strength of Resorbable Biomaterials

Biomedical Engineering

7.4K Visualizzazioni

article

Micro-CT Imaging of a Mouse Spinal Cord

Biomedical Engineering

7.9K Visualizzazioni

article

Visualization of Knee Joint Degeneration after Non-invasive ACL Injury in Rats

Biomedical Engineering

8.1K Visualizzazioni

article

Combined SPECT and CT Imaging to Visualize Cardiac Functionality

Biomedical Engineering

10.9K Visualizzazioni

JoVE Logo

Riservatezza

Condizioni di utilizzo

Politiche

Ricerca

Didattica

CHI SIAMO

Copyright © 2025 MyJoVE Corporation. Tutti i diritti riservati