Atomic fluorescence spectroscopy (AFS) is an analytical technique that involves the electronic transitions of atoms in a flame, furnace, or plasma being excited by electromagnetic (EM) radiation. When these atoms absorb energy, they become excited and subsequently release energy as they return to their original state. This emitted light, or "fluorescence," is observed at a right angle to the incident beam. Both absorption and emission processes transpire at distinct wavelengths, which are characteristic of the specific atomic species present. AFS is particularly useful for determining mercury (Hg) and other elements that form volatile hydrides, such as arsenic (As) and selenium (Se).
The instrumentation required for atomic fluorescence measurements includes a high-intensity light source, an atomizer, a wavelength selector, and a detector. While a continuum source would be desirable, it is rarely used due to its low power output. Instead, pulsed hollow-cathode lamps, electrodeless-discharge lamps, xenon or mercury arc lamps, and lasers serve as potential light sources.
The fluorescence signal intensity is proportional to the target element's concentration and irradiation intensity, making high-intensity sources and minimal interfering radiation essential. Various chemicals, such as releasing and protective agents, can be introduced into the matrix to minimize chemical and spectral interferences that arise during atomization.
From Chapter 14:
Now Playing
Atomic Spectroscopy
227 Views
Atomic Spectroscopy
731 Views
Atomic Spectroscopy
267 Views
Atomic Spectroscopy
861 Views
Atomic Spectroscopy
500 Views
Atomic Spectroscopy
309 Views
Atomic Spectroscopy
341 Views
Atomic Spectroscopy
589 Views
Atomic Spectroscopy
294 Views
Atomic Spectroscopy
940 Views
Atomic Spectroscopy
310 Views
Atomic Spectroscopy
144 Views
Atomic Spectroscopy
471 Views
Atomic Spectroscopy
171 Views
Atomic Spectroscopy
136 Views
See More
Copyright © 2025 MyJoVE Corporation. All rights reserved