5.15 : Methods of Sterilization I: Physical Methods
As used in a healthcare facility, sterilization destroys all microorganisms through physical or chemical methods. The physical method includes steam, dry heat, boiling water, and radiation.
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary temperature and pressure. Steam sterilization is influenced by four factors: steam, pressure, temperature, and time. The ideal steam for sterilization is dry saturated steam and entrained water with a dryness fraction of ≥97%. Pressure is used to generate the high temperatures needed to destroy microorganisms quickly. The presence of microbicidal activity requires a particular temperature to be reached. Standard steam sterilization temperatures range from 121°C (250 °F) to 132°C (270 °F). A minimum exposure time of 4 minutes at 132°C (270 °F) in a pre-vacuum sterilizer or 30 minutes at 121°C (250 °F) in a gravity displacement sterilizer is required to sterilize wrapped medical supplies. Moist heat destroys microorganisms through the irreversible coagulation and denaturation of enzymes and structural proteins. Items that can be steam-sterilized include all critical and semi-critical, heat- and moisture-resistant items.
Dry heat sterilization uses moisture-free heat to remove microorganisms and is only suitable for moisture-sensitive materials. It is non-toxic and has no adverse environmental effects; it penetrates materials and is non-corrosive to metal objects and sharps.
Boiling water is typically used in the home because it is easy and affordable and controls or kills most microorganisms. However, it does not eliminate spores and some viruses.
Sterilization by radiation exposes items or areas to different types of radiation, including non-ionizing radiation and ionizing radiation. Non-ionizing radiation uses ultraviolet radiation applied to the surface of aseptic work areas. Ionizing radiation is used for several medical products, including pharmaceuticals, and utilizes cobalt 60 gamma rays or electron accelerators to sterilize.
From Chapter 5:
Now Playing
5.15 : Methods of Sterilization I: Physical Methods
Infection Prevention and Control
20.1K Views
5.1 : Infection
Infection Prevention and Control
8.1K Views
5.2 : Stages of Infection
Infection Prevention and Control
56.8K Views
5.3 : Defense Mechanism Against Infection
Infection Prevention and Control
6.6K Views
5.4 : Inflammatory Response I: Vascular and Cellular
Infection Prevention and Control
11.2K Views
5.5 : Inflammatory Response II: Inflammatory Exudate and Tissue Repair
Infection Prevention and Control
5.0K Views
5.6 : Factors Affecting the Risk of Infection
Infection Prevention and Control
11.9K Views
5.7 : Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
Infection Prevention and Control
4.2K Views
5.8 : Healthcare Associated Infections II: Preventive Measures
Infection Prevention and Control
2.7K Views
5.9 : Asepsis
Infection Prevention and Control
2.2K Views
5.10 : Hand hygiene
Infection Prevention and Control
3.3K Views
5.11 : Handwashing I: Introduction and Types of Equipment
Infection Prevention and Control
4.1K Views
5.12 : Handwashing II: Pre-procedure and Initial Procedure Steps
Infection Prevention and Control
942 Views
5.13 : Handwashing III: During the Procedure and Post-Procedure Steps
Infection Prevention and Control
1.5K Views
5.14 : Cleaning, Sterilization, and Disinfection
Infection Prevention and Control
7.0K Views
See More