6.11 : Tandem Mass Spectrometry

Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and signal-to-noise ratio for the analyte. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.

Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced dissociation (CID). Other techniques involve interaction with the surface through surface-induced dissociation (SID); interaction with low-energy electrons, called electron-capture dissociation (ECD); or through an intense laser beam, called photo-induced dissociation (PID).

Tandem mass spectrometry can be used for various experiments by varying the mode of operation of the two mass analyzers. When the first mass analyzer selects a particular precursor ion and the second mass analyzer scans all the fragmented product ions, it is called a product-ion scan, and the spectra obtained by this experiment are called product-ion spectra. Alternately, when the first mass analyzer scans all the precursor ions and the second mass analyzer selects a particular product ion, it is called 'precursor-ion mode', and the spectra are termed precursor-ion spectra. Another experiment can be performed by scanning both the analyzers simultaneously with an offset of a mass equal to a neutral fragment lost by the ion to be monitored. Such a mode is called 'neutral-loss scanning mode', and the corresponding spectra are called neutral-loss spectra.

The instrumentation of tandem mass spectrometry can be varied. Tandem-in-space spectrometers use two independent mass analyzers in two different regions in space, for example, a triple quadrupole mass spectrometer and TOF–TOF spectrometer. On the other hand, a tandem-in-time spectrometer, such as a quadrupole ion-trap instrument, uses the same spatial region to fragment the ions and dissociate and analyze a selected ion of interest, but at different times by expelling the unwanted ions. Tandem-in-time spectrometers can be used to perform MSn experiments.

Tags
Tandem Mass SpectrometryMass AnalyzersSelectivitySignal to noise RatioSecondary FragmentationCollision induced Dissociation CIDSurface induced Dissociation SIDElectron capture Dissociation ECDPhoto induced Dissociation PIDProduct ion ScanPrecursor ion ModeNeutral loss Scanning ModeTandem in space SpectrometersTandem in time SpectrometersMSn Experiments

From Chapter 6:

article

Now Playing

6.11 : Tandem Mass Spectrometry

Principles of Mass Spectrometry

390 Views

article

6.1 : Mass Spectrometry: Overview

Principles of Mass Spectrometry

2.4K Views

article

6.2 : Mass Spectrometry: Isotope Effect

Principles of Mass Spectrometry

1.1K Views

article

6.3 : Mass Spectrometry: Molecular Fragmentation Overview

Principles of Mass Spectrometry

2.0K Views

article

6.4 : Mass Spectrometers

Principles of Mass Spectrometry

3.2K Views

article

6.5 : Mass Spectrum

Principles of Mass Spectrometry

921 Views

article

6.6 : Mass Spectrum: Interpretation

Principles of Mass Spectrometry

540 Views

article

6.7 : Mass Analyzers: Overview

Principles of Mass Spectrometry

248 Views

article

6.8 : Mass Analyzers: Common Types

Principles of Mass Spectrometry

268 Views

article

6.9 : High-Resolution Mass Spectrometry (HRMS)

Principles of Mass Spectrometry

783 Views

article

6.10 : Mass Spectrometry: Complex Analysis

Principles of Mass Spectrometry

407 Views

article

6.12 : Gas Chromatography–Mass Spectrometry (GC–MS)

Principles of Mass Spectrometry

2.6K Views

article

6.13 : Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview

Principles of Mass Spectrometry

379 Views

article

6.14 : Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Principles of Mass Spectrometry

229 Views

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved