Summary

The accuracy and sensitivity of protein determination by the rapid and convenient Bradford assay is compromised by intrinsic nonlinearity. We show a simple linearization procedure that greatly increases the accuracy, improves the sensitivity of the assay about 10-fold, and significantly reduces interference by detergents.

Abstract

Determination of microgram quantities of protein in the Bradford Coomassie brilliant blue assay is accomplished by measurement of absorbance at 590 nm. This most common assay enables rapid and simple protein quantification in cell lysates, cellular fractions, or recombinant protein samples, for the purpose of normalization of biochemical measurements. However, an intrinsic nonlinearity compromises the sensitivity and accuracy of this method. It is shown that under standard assay conditions, the ratio of the absorbance measurements at 590 nm and 450 nm is strictly linear with protein concentration. This simple procedure increases the accuracy and improves the sensitivity of the assay about 10-fold, permitting quantification down to 50 ng of bovine serum albumin. Furthermore, the interference commonly introduced by detergents that are used to create the cell lysates is greatly reduced by the new protocol. A linear equation developed on the basis of mass action and Beer's law perfectly fits the experimental data.

Protocol

Linearization of the Bradford Protein Calibration Graph:

The Coomassie brilliant blue protein assay, commonly known as the Bradford assay 1, is widely used because of its rapid and convenient protocol as well as its relative sensitivity. Unfortunately, there is a large degree of curvature over a broad range of protein concentrations (Fig. 1). Therefore, only a narrow range of relatively high protein concentrations, 2-10 μg/ml BSA, is used for the calibration graph, which then better fits linear regression (Fig. 1, green). However, the nonlinearity requires protein concentration of the unknown samples to fall within the limited range of the calibration graph in order to avoid a large error, and it also reduces the accuracy within the limited range. The nonlinearity presents a serious problem in particular when microgram amounts of protein are not available, and it often requires multiple dilutions of the unknown samples.

As noted in the original Bradford paper, "the source of the nonlinearity is in the reagent itself since there is an overlap in the spectrum of the two different color forms of the dye." 1. In fact, three forms of the Coomassie brilliant blue dye are in acid-base equilibrium at the usual acidic pH of the assay 2. The red, blue, and green forms have absorbance maxima at 470, 590, and 650 nm, respectively (Fig. 2). The blue is the form that binds the protein, forming a complex that intensely absorbs light at 594 nm 3, 4 (Fig. 2). Bradford also noted that "the background value for the reagent is continually decreasing as more dye is bound to protein" 1 (Fig. 3). Therefore, we attempted to calculate the reduction of the 590 nm background as increasing protein quantities are added, by measuring the change of absorbance at 450 nm, where the protein-dye complex does not absorb. We found that the decreasing background partially, but not fully, accounts for the nonlinearity (Fig. 4).

We then hypothesized that the decrease in free dye concentration produces another distortion of the linear response, because as protein-dye binding is in equilibrium 5, complex formation depends not only on the concentration of the free protein, but also on that of the free dye (Fig. 5). Taking into account both issues related to the variable concentration of the free dye, we developed a mathematical equation that describes a linear relationship between protein concentration and the ratio of absorbance measurements, 590 nm over 450 nm (Fig. 6). A detailed description of the theoretical and experimental study can be found in our 1996 publication in Analytical Biochemistry 6. The mathematical equation was experimentally tested and found to yield a linear calibration curve over the entire protein concentrations range (Fig. 7). Furthermore, the equation was validated also by an independent determination of the correct pH-dependent value of the Y-axis intercept 6.

Detailed Protocol for the Improved Bradford Protein Assay, using a Microplate Absorbance Reader:

  1. Prepare a 0.1 mg/ml stock solution of the standard, bovine serum albumin. Any other standard may be chosen, but note that the same standard must be used in all experiments.
  2. Dilute the unknown samples in deionized water. Aim to 5-50 μg/ml. Yet, higher or lower protein concentrations are acceptable, since there is no apparent limit for the linear range of the assay. However, the measurement must be within the linear range of the absorbance reader.
  3. Dilute the Bradford reagent (Bio-Rad) 2.5-fold in deionized water.
  4. Add 0 and 10-50 μl of BSA stock solution to triplicate wells (creating a 0-5 μg BSA calibration curve). Complement with deionized water to reach 100 μl/well.
  5. In different wells, add 100 μl of unknown sample in triplicates. Several concentrations of the unknown sample may be used to increase accuracy.
  6. Add 100 μl of the diluted Bradford reagent to all wells. Total volume is 200 μl/well.
  7. Wait at least 5 min, but not more than 60 min for color development.
  8. The Blank must be 200 μl of deionized water (and not the zero protein dye well).
  9. Measure the absorbance at 590 nm and at 450 nm.
  10. Prepare a calibration graph by dividing the net absorbance values at 590 nm and at 450 nm. Note that the zero protein (dye only) value should be included as a data point (Fig. 8).
  11. Calculate the concentration of the unknown sample based on the linear equation of the calibration curve (Fig. 9).

Representative Results:

Unlike the absorbance at a single wavelength of 590 nm, the ratio of absorbance values, 590 nm over 450 nm, is linear with protein concentration (Fig. 10).

The protein concentration of the unknown sample may be simply calculated using the linear equation of the calibration curve (Fig. 10, equation).

However, increased accuracy is obtained by measuring several unknown sample dilutions. To this end, prepare two graphs. The first is a calibration graph for the standard with μg protein on the X axis (Fig. 11, right). The second graph is for the unknown sample, with μl of the original undiluted sample on the X axis (Fig. 11, left). The dye only value should be included in both graphs (Fig. 11). The protein concentration of the unknown sample is derived by dividing the slopes of the unknown sample and the standard (Fig. 11, equation).

BSA standard curve graph; absorbance vs concentration; protein quantification; spectrophotometry.
Figure 1. Conventional Bradford calibration graph. The linear range is represented by green symbols.

Protein-Dye interaction, absorbance vs. wavelength graph, Coomassie Brilliant Blue analysis.
Figure 2. Spectra of the protein-dye complex and of the dye alone.

Protein aggregation diagram; depicts equilibrium process with molecular interactions and complex formation.
Figure 3. Protein-dye equilibrium.

Absorbance vs. BSA concentration graph; comparison of blank reduction methods; spectroscopic analysis.
Figure 4. Reduction of the correct calculated background partially decreases nonlinearity.

Protein binding equilibrium diagram, K_eq equation, illustrating protein-ligand interaction.
Figure 5. Protein-dye equilibrium.

Equilibrium equation for protein-ligand binding analysis, with color-coded variables and constants.
Figure 6. Mathematical basis for the linearization of the Bradford protein assay.

Absorbance ratio vs. BSA graph; linear fit equation and R² value; spectrophotometry analysis.
Figure 7. Linearization of the Bradford calibration graph.

Protein quantification; BSA absorbance data table; comparison at 590 nm and 450 nm using spectrometry.
Figure 8. Calculation of the absorbance ratio values.

Absorbance ratio graph, linear equation, BSA quantification, spectrophotometer data analysis.
Figure 9. A linear Bradford calibration graph.

BSA protein absorbance analysis; graphs of absorbance at 590 nm and absorbance ratio 590/450 nm.
Figure 10. Linearization of the Bradford calibration graph.

Protein concentration graphs; left: lysate linear fit, right: BSA standard; absorbance ratio data.
Figure 11. Unknown sample- concentration calculation.

Access restricted. Please log in or start a trial to view this content.

Discussion

The Bradford protein assay is popular due to its ease of performance and relative sensitivity. The linearization over the entire protein concentrations range obtained by the protocol presented here further simplifies the assay, as the unknown samples do not need to fall within the range of the calibration graph.

Importantly, the improved protocol further provides the following advantages over the original Bradford protocol:

  1. Increased accuracy.
  2. The sensitivity is incre...

Access restricted. Please log in or start a trial to view this content.

Acknowledgements

This paper is dedicated to the memory of the late Dr. Zvi Selinger, who hosted the original research described here.

Access restricted. Please log in or start a trial to view this content.

Materials

NameCompanyCatalog NumberComments
Bradford reagentBio-Rad500-0006
Bovine Serum AlbuminAmersco0332
Multiplate absorbance readerBioTekSynergy2

References

  1. Bradford, M. M. A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Anal Biochem. 72, 248-254 (1976).
  2. Chial, H. J., Thompson, H. B., Splittgerber, A. G. A spectral study of the charge forms of Coomassie blue G. Anal Biochem. 209, 258-266 (1993).
  3. Chial, H. J., Splittgerber, A. G. A comparison of the binding of Coomassie brilliant blue to proteins at low and neutral pH. Anal Biochem. 213, 362-369 (1993).
  4. Compton, S. J., Jones, C. G. Mechanism of dye response and interference in the Bradford protein assay. Anal Biochem. 151, 369-374 (1985).
  5. Congdon, R. W., Muth, G. W., Splittgerber, A. G. The binding interaction of Coomassie blue with proteins. Anal Biochem. 213, 407-413 (1993).
  6. Zor, T., Selinger, Z. Linearization of the Bradford protein assay increases its sensitivity: theoretical and experimental studies. Anal Biochem. 236, 302-308 (1996).

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Explore More Articles

Biology590 NmCoomassie Brilliant Blue AssayProtein QuantificationCell LysatesCellular FractionsRecombinant Protein SamplesNormalization Of Biochemical MeasurementsNonlinearitySensitivityAccuracyStandard Assay ConditionsAbsorbance Ratio450 NmLinear EquationMass ActionBeer s Law

This article has been published

Video Coming Soon

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved