Our goals were to design, manufacture and test ferromagnetic stents for endothelial cell capture. Ten stents were tested for fracture and 10 more stents were tested for retained magnetism. Finally, 10 stents were tested in-vitro and 8 more stents were implanted in 4 pigs to show cell capture and retention.
In the protocol, we present a method to manufacture a small caliber stent-graft by sandwiching a balloon expandable stent between two electrospun nanofibrous polyurethane layers.