Oturum Aç

Scanning Electron Microscopy (SEM)

Genel Bakış

Source: Laboratory of Dr. Andrew J. Steckl — University of Cincinnati

A scanning electron microscope, or SEM, is a powerful microscope that uses electrons to form an image. It allows for imaging of conductive samples at magnifications that cannot be achieved using traditional microscopes. Modern light microscopes can achieve a magnification of ~1,000X, while typical SEM can reach magnifications of more than 30,000X. Because the SEM doesn’t use light to create images, the resulting pictures it forms are in black and white. 

Conductive samples are loaded onto the SEM’s sample stage. Once the sample chamber reaches vacuum, the user will proceed to align the electron gun in the system to the proper location. The electron gun shoots out a beam of high-energy electrons, which travel through a combination of lenses and apertures and eventually hit the sample. As the electron gun continues to shoot electrons at a precise position on the sample, secondary electrons will bounce off of the sample. These secondary electrons are identified by the detector. The signal found from the secondary electrons is amplified and sent to the monitor, creating a 3D image. This video will demonstrate SEM sample preparation, operation, and imaging capabilities.

Prosedür

1. Preparation of the Sample

  1. Place sample onto sample stub. If necessary, carbon tape may be used to adhesively bond the sample to the stub.
  2. Place the sample into a gold sputtering system. Using a mini-gold sputter, sputter gold for 30 s at ~ 70 mTorr pressure. A different gold layer thickness may be necessary depending on the geometry of the sample. More rough or porous surface requires a longer sputtering time.
  3. Remove stub from gold sputtering system.

.css-f1q1l5{display:-webkit-box;display:-webkit-flex;display:-ms-flexbox;display:flex;-webkit-align-items:flex-end;-webkit-box-align:flex-end;-ms-flex-align:flex-end;align-items:flex-end;background-image:linear-gradient(180deg, rgba(255, 255, 255, 0) 0%, rgba(255, 255, 255, 0.8) 40%, rgba(255, 255, 255, 1) 100%);width:100%;height:100%;position:absolute;bottom:0px;left:0px;font-size:var(--chakra-fontSizes-lg);color:#676B82;}

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Sonuçlar

The SEM, seen in Figure 2a, has been used for making measurements and acquiring sample photos. The sample consisted of sodium chloride (NaCl) salt. It was placed onto the stub as seen in Figure 2b, then a few nanometers of gold was sputtered onto it to make it conductive. The conductive sample was then placed into the SEM sample area as seen in Figure 2c.

SEM images were obtained at 50X, 200X, 500X, 1,000X, and 5,000X magnification levels as s

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Başvuru ve Özet

The SEM is a very powerful tool that is common in most research institutions because of its ability to image any object that is conductive, or has been treated with a conductive coating. The SEM has been used to image objects such as semiconductor devices,2 biological membranes,3 and insects,4 among others. We have also used the SEM to analyze nanofibers and paper-based materials, biomaterials, micropatterned structures. Of course, there are materials, such as liquids, that can’t be

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Referanslar
  1. Goldstein, J., Newbury, D., Joy, D., Lyman, C., Echlin, P., Lifshin, E., Sawyer, L., Michael, J. Scanning Electron Microscopy and X-ray Microanalysis. 3rd Ed. Springer, New York, NY. (2003).
  2. Purandare, S., Gomez, E.F., Steckl, A.J. High brightness phosphorescent organic light emitting diodes on transparent and flexible cellulose films. Nanotechnology. 25, 094012 (2014).
  3. Masuda, Y., Yamanaka, N., Ishikawa, A., Kataoka, M., Aral, T., Wakamatsu, K., Kuwahara, N., Nagahama, K., Ichikawa, K., Shimizu, A. Glomerular basement membrane injuries in IgA nephropathy evaluated by double immunostaining for a5(IV) and a2(IV) chains of type IV collagen and low-vacuum scanning electron microscopy. Clinical and Experimental Nephrology. 1-9. (2014).
  4. Kang, J.H., Lee, Y.J., Oh, B.K., Lee, S.K. Hyun, B.R. Lee, B.W, Choi, Y.G., Nam, K.S., Lim, J.D. Microstructure of the water spider (Argyroneta aquatic) using the scanning electron microscope Journal of Asia-Pacific Biodiversity. 7 484-488 (2014).
Etiketler
Scanning Electron Microscopy SEMSEM TechniqueElectron BeamSurface Structure AnalysisChemical Composition AnalysisLight Microscopy LimitationsDiffractionLateral ResolutionMagnificationSub nanometer ResolutionDepth Of FieldSEM In Chemistry And Material AnalysisNanoscale AnalysisSEM Sample PreparationSEM OperationConductive Samples

Atla...

0:00

Overview

1:49

Principles of Scanning Electron Microscopy

4:10

Sample Preparation and Loading

5:54

SEM Operation

8:22

Image Analysis

9:17

Applications

11:13

Summary

Bu koleksiyondaki videolar:

article

Now Playing

Scanning Electron Microscopy (SEM)

Analytical Chemistry

85.8K Görüntüleme Sayısı

article

Sample Preparation for Analytical Characterization

Analytical Chemistry

82.5K Görüntüleme Sayısı

article

Internal Standards

Analytical Chemistry

201.8K Görüntüleme Sayısı

article

Method of Standard Addition

Analytical Chemistry

316.7K Görüntüleme Sayısı

article

Calibration Curves

Analytical Chemistry

780.4K Görüntüleme Sayısı

article

Ultraviolet-Visible (UV-Vis) Spectroscopy

Analytical Chemistry

609.7K Görüntüleme Sayısı

article

Raman Spectroscopy for Chemical Analysis

Analytical Chemistry

50.1K Görüntüleme Sayısı

article

X-ray Fluorescence (XRF)

Analytical Chemistry

25.1K Görüntüleme Sayısı

article

Gas Chromatography (GC) with Flame-Ionization Detection

Analytical Chemistry

277.1K Görüntüleme Sayısı

article

High-Performance Liquid Chromatography (HPLC)

Analytical Chemistry

377.4K Görüntüleme Sayısı

article

Ion-Exchange Chromatography

Analytical Chemistry

258.0K Görüntüleme Sayısı

article

Capillary Electrophoresis (CE)

Analytical Chemistry

91.9K Görüntüleme Sayısı

article

Introduction to Mass Spectrometry

Analytical Chemistry

110.0K Görüntüleme Sayısı

article

Electrochemical Measurements of Supported Catalysts Using a Potentiostat/Galvanostat

Analytical Chemistry

50.9K Görüntüleme Sayısı

article

Cyclic Voltammetry (CV)

Analytical Chemistry

122.3K Görüntüleme Sayısı

JoVE Logo

Gizlilik

Kullanım Şartları

İlkeler

Araştırma

Eğitim

JoVE Hakkında

Telif Hakkı © 2020 MyJove Corporation. Tüm hakları saklıdır