Sign In

Principles Of Column Chromatography

The chromatography technique was first invented in 1901 by Michael S. Tswett, a Russian botanist, to separate plant pigments using organic solvents. Further, in 1941, Archer John Porter Martin and R. L. M. Synge modified the technique by packing silica gel into a column. A mixture of amino acids was then separated on the packed column using chloroform and water mixture as the mobile phase. This was the first report on column chromatography. At present, column chromatography is a widely used technique for separating various types of compounds from a sample mixture.

Factors Influencing Efficient Separation of Proteins

Various parameters like column material, packing, and operational conditions such as flow rates and temperature determine the efficiency of separation by column chromatography.

The choice of column material or matrix determines the extent of interaction with the sample. The matrix material must be tightly and uniformly packed in the column. Air bubbles, debris, large particles, and precipitates interfere with the uniform flow of the solvent through the column, affecting its separation efficiency. The column should also be free from particulate matter.

The sample injected into the column should be clear and free from aggregates that might clog the column, hindering solvent flow. The flow rate of solvent also affects the separation. Very high or very low flow rates of solvents result in inefficient separation of compounds and impure preparations. Very high rates may also disturb the column packing, affecting process efficiency. In addition, the composition of elution buffer is also an important factor. It should be non-corrosive and compatible with the sample and also the column material to prevent in situ precipitation or dissolution.

Another operational parameter, temperature, also plays an important role in the process. It decides the stability of the sample, column material, and solvent buffer. Also, a constant temperature throughout the column efficiently resolves compounds. After completing the separation process, the columns must be washed thoroughly by repeatedly passing a suitable solvent to avoid sample contamination in further runs. Occasionally, the solvent is passed in a column in the reverse direction to remove any clogged material.

Limitations

Though a very widely used technique, the method still has some limitations. It is a very time-consuming method as the flow rates need to be slower for better resolution of the compounds. Also, large quantities of highly pure solvents required in the mobile phase make the process expensive. This also raises the scaling-up cost when higher yields of pure compounds are needed.

Tags
Column ChromatographyMichael S TswettSilica GelAmino Acids SeparationMobile PhaseOperational ConditionsFlow RatesTemperatureColumn MaterialPacking EfficiencyElution BufferSolvent FlowSample PreparationSolvent ContaminationChromatographic Limitations

From Chapter undefined:

article

Now Playing

Principles Of Column Chromatography

Related Videos

4.8K Views

article

Overview Of Cell Separation And Isolation

Related Videos

4.2K Views

article

Cell Culture

Related Videos

12.5K Views

article

Cell Lines

Related Videos

5.3K Views

article

Hybridoma Technology

Related Videos

11.5K Views

article

Tissue Homogenization and Cell Lysis

Related Videos

5.6K Views

article

Subcellular Fractionation

Related Videos

5.2K Views

article

Flow Cytometry

Related Videos

8.5K Views

article

Types Of Column Chromatography

Related Videos

8.5K Views

article

Immunoprecipitation

Related Videos

4.1K Views

article

Tagging and Fusion Proteins

Related Videos

5.5K Views

article

SDS-PAGE

Related Videos

21.0K Views

article

Western Blotting

Related Videos

10.4K Views

article

Two-dimensional Gel Electrophoresis

Related Videos

4.7K Views

article

Enzyme-Linked Immunosorbent Assay

Related Videos

10.0K Views

See More

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2024 MyJoVE Corporation. All rights reserved