Atomic fluorescence spectroscopy (AFS) is an analytical technique that involves the electronic transitions of atoms in a flame, furnace, or plasma being excited by electromagnetic (EM) radiation. When these atoms absorb energy, they become excited and subsequently release energy as they return to their original state. This emitted light, or "fluorescence," is observed at a right angle to the incident beam. Both absorption and emission processes transpire at distinct wavelengths, which are characteristic of the specific atomic species present. AFS is particularly useful for determining mercury (Hg) and other elements that form volatile hydrides, such as arsenic (As) and selenium (Se).
The instrumentation required for atomic fluorescence measurements includes a high-intensity light source, an atomizer, a wavelength selector, and a detector. While a continuum source would be desirable, it is rarely used due to its low power output. Instead, pulsed hollow-cathode lamps, electrodeless-discharge lamps, xenon or mercury arc lamps, and lasers serve as potential light sources.
The fluorescence signal intensity is proportional to the target element's concentration and irradiation intensity, making high-intensity sources and minimal interfering radiation essential. Various chemicals, such as releasing and protective agents, can be introduced into the matrix to minimize chemical and spectral interferences that arise during atomization.
From Chapter 14:
Now Playing
Atomic Spectroscopy
164 Views
Atomic Spectroscopy
478 Views
Atomic Spectroscopy
186 Views
Atomic Spectroscopy
351 Views
Atomic Spectroscopy
326 Views
Atomic Spectroscopy
212 Views
Atomic Spectroscopy
232 Views
Atomic Spectroscopy
337 Views
Atomic Spectroscopy
193 Views
Atomic Spectroscopy
388 Views
Atomic Spectroscopy
192 Views
Atomic Spectroscopy
88 Views
Atomic Spectroscopy
328 Views
Atomic Spectroscopy
123 Views
Atomic Spectroscopy
97 Views
See More
Copyright © 2025 MyJoVE Corporation. All rights reserved