An unknown compound can be established by identifying the molecular ion peak in the mass spectrum. The molecular ion peak is often weak or absent due to the predominance of fragmentation in high-energy electron beams. In such cases, a low-energy electron beam can be used to scan the spectrum to enhance the intensity of the molecular ion peak. Additionally, chemical ionization, field ionization, and desorption ionization spectra are used to obtain a relatively intense molecular ion peak.

To determine the empirical formula from the molecular ion mass, a high-resolution mass spectrometer like a double-focusing mass spectrometer, a time-of-flight mass spectrometer, an orbitrap mass spectrometer, or a Fourier transform ion cyclotron resonance mass spectrometer is required.

Comparing the intensities of isotope peaks (M+1 and M+2), which arise from the natural abundances of higher-mass isotopes, with the molecular ion peak (M+) gives a fair idea of the empirical formula of an unknown compound.

The number of carbon atoms present in the unknown compound can be deduced by comparing the M+1 to M+ peak intensity ratio to the expected M+1 contribution per carbon atom.

The nitrogen rule, stating that a molecule having an even molecular weight contains zero or even number of nitrogen atoms, predicts the number of nitrogen atoms present in the unknown compound.

Further, the degree of unsaturation predicted from the formula

Figure1

indicates the presence of multiple bonds or rings in the unknown compound. Lastly, the study of fragmentation patterns for compounds based on a series of guidelines helps identify signature fragments at characteristic m/z values, suggesting the presence of certain structural features.

Tags

Mass SpectrumMolecular Ion PeakFragmentationLow energy Electron BeamChemical IonizationHigh resolution Mass SpectrometerEmpirical FormulaIsotope PeaksCarbon AtomsNitrogen RuleDegree Of UnsaturationFragmentation PatternsM z Values

Aus Kapitel 6:

article

Now Playing

6.6 : Mass Spectrum: Interpretation

Principles of Mass Spectrometry

1.0K Ansichten

article

6.1 : Mass Spectrometry: Overview

Principles of Mass Spectrometry

3.4K Ansichten

article

6.2 : Mass Spectrometry: Isotope Effect

Principles of Mass Spectrometry

1.7K Ansichten

article

6.3 : Mass Spectrometry: Molecular Fragmentation Overview

Principles of Mass Spectrometry

2.7K Ansichten

article

6.4 : Mass Spectrometers

Principles of Mass Spectrometry

4.6K Ansichten

article

6.5 : Mass Spectrum

Principles of Mass Spectrometry

1.5K Ansichten

article

6.7 : Mass Analyzers: Overview

Principles of Mass Spectrometry

515 Ansichten

article

6.8 : Mass Analyzers: Common Types

Principles of Mass Spectrometry

510 Ansichten

article

6.9 : High-Resolution Mass Spectrometry (HRMS)

Principles of Mass Spectrometry

1.1K Ansichten

article

6.10 : Mass Spectrometry: Complex Analysis

Principles of Mass Spectrometry

637 Ansichten

article

6.11 : Tandem Mass Spectrometry

Principles of Mass Spectrometry

730 Ansichten

article

6.12 : Gas Chromatography–Mass Spectrometry (GC–MS)

Principles of Mass Spectrometry

3.6K Ansichten

article

6.13 : Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview

Principles of Mass Spectrometry

536 Ansichten

article

6.14 : Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Principles of Mass Spectrometry

302 Ansichten

JoVE Logo

Datenschutz

Nutzungsbedingungen

Richtlinien

Forschung

Lehre

ÜBER JoVE

Copyright © 2025 MyJoVE Corporation. Alle Rechte vorbehalten