È necessario avere un abbonamento a JoVE per visualizzare questo. Accedi o inizia la tua prova gratuita.
This video demonstrates a technique for stimulating line-patterned neuronal cultures with a uniform, unidirectional electric field.
1. Imaging of spontaneous or evoked activity in neuronal cultures with fluorescent dyes.
2. Electric Stimulation of Cultures
NOTE: The basic setup for electric stimulation is shown in Figure 2. A cover slip on which the neuronal culture has been grown for about 14 days is placed in a Petri dish under a fluorescence microscope. Electrical activity of the neurons is imaged using calcium sensitive dyes while a voltage is applied via two pairs of bath electrodes that are positioned outside the culture. The electrodes are driven by a signal generator whose output is amplified by a dual channel amplifier. Voltage control for stimulation is preferred over the more standard current control, because the electric field vectors are determined directly, thus enabling straightforward vector addition and combination. This does require a careful check of the uniformity of the electric field, which can be performed over the whole sample for the case of voltage control. When using voltage control care should be taken to avoid any ground loops and the homogeneity of the electric field should be verified.
Figure 1: Example Traces of Calcium Transients Imaged During Synchronous Network Bursts. A. An image of neurons that were dyed previous to the experiment with a calcium dye. B. Traces of intensity vs. time of the ROIs in A with the color of the trace representing the color of the border of the ROI in A. A large increase in intensity synchronized within the three ROIs represents a...
Name | Company | Catalog Number | Comments |
Calcium chloride , 1M | Fluka | 21098 | Extracellular recording solution. |
D-(+)-Glucose, 1M | Sigma-Aldrich | 65146 | Plating medium, Extracellular recording solution. |
Hepes, 1M | Sigma-Aldrich | H0887 | Extracellular recording solution. |
KCl, 3M | Merck | 1049361000 | Extracellular recording solution. |
Magnesium chloride , 1M | Sigma-Aldrich | M1028 | Extracellular recording solution. |
NaCl, 4M | Bio-Lab | 19030591 | Extracellular recording solution . |
Sucrose, 1M | Sigma-Aldrich | S1888 | Extracellular recording solution. |
Electrodes wires | A-M Systems, Carlsborg WA | 767000 | Electric stimulation of neuronal cultures. |
Signal generator | BKPrecision | 4079 | Shaping of the electric signal. |
Amplifier | Homemade | Voltage amplification of the signal from the signal generator to the electrodes. | |
Power supply | Matrix | MPS-3005 LK-3 | Power supply to the sputtering machine. |
Platinum wires 0.005'' thick; A-M Systems, | Carlsborg WA | 767000 | Electric stimulation of neuronal cultures. |