Riepilogo

Here, we demonstrate a robust and standardized protocol for fabricating high-quality factor (Q-factor) Whispering Gallery Mode (WGM) Microbubble resonators (MBRs) with a precision glass processing machine (PGP).

Abstract

We demonstrate a robust and standardized method for the fabrication of high-quality factor (Q-factor) Whispering Gallery Mode (WGM) Microbubble resonators (MBRs) with a precision glass processing machine (PGP). Microbubble resonators are a unique class of WGM devices with integrated fluidic channels, making them ideal for diverse sensing applications. Herein, we show a standardized protocol to fabricate high-Q microbubble resonators through the optimization of key performance metrics, such as Q-factor and wall thickness. We also show methods to improve the sensitivity of the platform to refractive index changes and other sensing targets through Hydrofluoric acid (HF) wet etching. Lastly, a brief analysis of the resistance of microbubbles to fluid flow is discussed, showing that smaller-diameter microbubbles exhibit greater resistance to flow for analyte delivery - a factor that should be considered for analyte delivery. The implementation of this refined fabrication protocol not only increases the success rate of device production but also reduces fabrication time. Moreover, the protocol can be expanded to other techniques used to produce MBRs, such as CO2 laser-based methods.

Introduzione

Whispering Gallery Mode (WGM) microresonators are a class of optical sensors that have demonstrated enormous potential not only for the detection of single molecules and nano-particles1,2,3,4,5,6 but also for sensing a wide range of physical phenomena such as magnetic7 and electric fields8, temperature9, and ultrasonic waves10,11. Under optical resonance conditions, light is trapped within the device, leading to a significant power amplification12,13. Any localized change to the resonator (such as the binding of a biomolecule or changes in the refractive index of the surrounding media) induces changes in the local optical environment, therefore shifting the resonant frequency or wavelength. By monitoring the shifts in resonance wavelength or frequency, one can detect and characterize analytes in real time.

WGM microresonators can be designed in a variety of geometries. Common geometries include but are not limited to, microtoroids14, microrings15, and microbubble16 resonators (MBR). Here, we focus on MBRs due to their great potential in optofluidic sensing applications. A key advantage of MBRs is their fluidic integration17,18,19,20, which is enabled by the fabrication of the device from a microcapillary. In this design, the inline capillary facilitates the easy delivery of small volumes (i.e., microliters) of analytes in solution to the sensing area without the need for external fluidic channels, as shown in Figure 1. With their unique fluidic handling capabilities, MBRs are well suited for a wide range of sensing applications that are not easily achievable with other WGM platforms. For example, MBRs have been filled with magnetic fluids, thereby imbuing sensitivity to external magnetic fields21. Additionally, MBRs have also been used to control the specific orientation of gold nanorods in solution through optical torques22.

The fabrication of MBRs can be summarized as follows: Aerostatic pressure is applied inside the capillary while a small area of the capillary is locally heated. The combination of localized heating and internal pressure inflates the heated section into a spherical geometry capable of supporting high-Q WGMs, as illustrated in Figure 2. Various methods can be employed to achieve localized heating of the capillary, such as using a CO2 laser23, a fiber optic splicer24, a hydrogen flame source25, and a precision glass processing machine (PGP). The methods presented here can be expanded to other heating sources, including a CO2 laser. The PGP is similar to an optical fiber splicer but offers enhanced control over heating time, power setting, and the positioning of fibers or capillaries26. PGPs often include built-in microscopes adjacent to the heating elements, enabling real-time monitoring of the fabrication process. Typically, light from a tunable diode laser is coupled into the MBR via a tapered optical fiber that is in contact with the equator of the MBR. The fiber is tapered (to ~1 μm) to enable efficient coupling of light into and out of the MBR. The resulting transmission spectra from the MBR are then captured by a photodetector through the optical fiber and visualized on an oscilloscope.

Sensing with WGM MBRs relies on the interaction of the WGM field with the target analyte. The strength of this interaction is directly proportional to the fraction of the WGM field that penetrates the MBR's hollow cavity where liquid or gas phase samples can flow through27. As shown in Figure 3, COMSOL simulations illustrate how the penetration of the WGM field into the inner cavity varies with the MBR's wall thickness. Maximum field penetration of the WGM field occurs as the wall thickness is reduced to less than 1 μm, with these simulations conducted using light in the 780 nm band. Achieving such reduced wall thickness through the standard heat-and-inflate fabrication protocol alone is challenging. To further thin the walls of the MBR and to make the device more sensitive, we incorporate additional wet etching steps using Hydrofluoric (HF) acid.

Using a PGP, we will focus on the fabrication of MBRs in line with a silica capillary. A detailed description of the fabrication process and methods to enhance sensitivity to refractive index changes through wet etching will also be presented.

Protocollo

1. Microbubble fabrication

  1. Start with a polymer-coated silica glass capillary (250 μm inner diameter and 360 μm outer diameter) that is ~75 cm in length. The length of the capillary can vary by user needs; ensure that the pressure described below is reached at longer capillary lengths.
  2. Burn off ~2.5 cm of the polymer coating at one end of the capillary with a butane torch and clean the end with a delicate task wipe and isopropyl alcohol (IPA).
  3. Place the clean end of the capillary in the PGP and click the splice only button on the software to heat for 5 s at 180 W to seal the end of the capillary. Set the heating time and duration with a right-click on the splice only button.
  4. To ensure a proper seal, examine the sealed capillary end, Figure 1B.
  5. At ~25 cm from the sealed end (or 1/3 the capillary length from the sealed end), burn off a ~2.5 cm long strip of polymer coating from the capillary and clean the resulting area with IPA.
    NOTE: The position described above is not critical as pressure is evenly distributed along the capillary length. If the user's experimental setup warrants a different placement of the MBR, this should not affect the rest of the procedure.
  6. Repeat this step until no polymer is visible under the imaging system, which displays capillary image with sub-micron resolution.
  7. Place this newly cleaned section of the capillary in the glass processor instrument over the heating element and under the microscope described earlier.
  8. Using a gas-tight syringe and a syringe pump, inject air into the capillary such that the internal pressure reaches 10 bar (i.e., decreasing the volume in a syringe from 5 mL to 0.5 mL, with corresponding pressures calculated with Boyle's Law). Fit the capillary to a syringe with a luer lock to a 360 μm adapter. Alternatively, apply pressure via a pressure regulator and compressed air/inert gas.
  9. Heat the capillary with the PGP using a filament power of ~100-110 W for 1.5-2 s.
  10. Conduct additional heating steps with 90-100 W and 0.1-1 s firings of the filament. This method will slowly and controllably increase the MBR's diameter to the desired range. However, this process may result in asymmetric MBRs. If an asymmetric MBR is observed (the sphere is not symmetric about the capillary axis), rotate it about the capillary axis between successive heating stepsto promote symmetry.
  11. Inspect the MBR under a microscope for quality control, looking for things such as dust, cracks, or deformations. Additionally, use the microscope here to estimate the wall thickness of the MBR.

2. Wet etching with hydrofluoric acid

CAUTION: Hydrofluoric acid is very dangerous, toxic, and corrosive. Calcium gluconate should be kept nearby as this chemical can neutralize hydrofluoric acid. Wear appropriate personal protective equipment and follow all safety precautions in the Material Safety Data Sheet (MSDS).

  1. Remove the sealed end of the capillary by cutting 2 cm below the sealed end.
  2. Place one capillary end inside a blunt tip to the luer lock adapter. Make sure the blunt tip's inner diameter is as close as possible to the outer diameter of the capillary while making sure the capillary can still fit inside.
    NOTE: The adapter material must not be reactive with HF acid.
  3. Apply UV curable epoxy to the end of the blunt tip with the capillary inside to join the two, then cure with UV light (50-100 W UV light source; higher wattage will speed up the curing).
    CAUTION: Appropriate eye protection must be worn when operating a UV light source28.
  4. Once the epoxy is cured, attach the luer lock tip to a 25 mL syringe that is inert to HF.
  5. Place the syringe in a syringe pump with a withdrawal rate of 50 μL/min.
  6. With one end of the capillary connected to the syringe, place the other end of the capillary into deionized (DI) water.
  7. Set the syringe pump to withdraw water through the capillary into the syringe to check that proper flow has been established (i.e., the system is free from air bubbles with liquid consistently flowing through the capillary).
  8. Transfer the capillary end from the DI water to a container of HF acid once proper flow is established.
  9. Calculate an approximate etching time based on the MBR wall thickness before etching, considering a measured etching rate of 8.18 μm/h with 12% (w/v) hydrofluoric acid. Wall thickness can be measured under an imaging system that displays MBRs with sub-micron resolution.
    NOTE: The etching rate for acids of other concentrations can be empirically determined. This can be done by measuring the capillary wall thickness before and after a set etching time.
  10. Run the syringe pump at 50 μL/min for the calculated etching time.
  11. Rinse the capillary with DI water for 10-15 min to remove all acid from the capillary after the etching time has elapsed. Calculate the rinsing time to rinse the entire capillary volume 3-5 times.
  12. Measure the MBR wall thickness after etching. Use this information to update the etching rate for the acid.
  13. Repeat this process can be repeated until the desired wall thickness is achieved.

Risultati

A representative MBR fabricated with the PGP machine is shown in Figure 1C. Given our starting capillary outer diameter (OD) of 360 μm, we expand the capillary ~2x in the fabrication process. Expanding the capillary to ~700 μm results in wall thicknesses between 5 μm and 15 μm. It has been shown that the optimal wall thickness for biosensing with MBRs is on the order of the wavelength of light used to excite the WGM27. MBRs can theoretically achiev...

Discussione

Here, we described the protocol to fabricate high-quality whispering gallery mode (WGM) microbubble resonators (MBRs) using a precision glass processor. We present critical steps in the fabrication protocol, including the heat and expand steps. Here, a combination of overheating, heating too long, or injecting too much internal air pressure can lead to unsuccessful fabrication. To address these issues, adjustments such as lowering the heating power or heating duration in the software user interface of the PGP machine can...

Divulgazioni

The authors have nothing to disclose.

Riconoscimenti

This project was supported in part by R41AI152745. AJQ was funded by the T32 Cancer Biology Award (NIH CA009547) and K08EB033409.

Materiali

NameCompanyCatalog NumberComments
Blunt tip to luer lock adapterEllsworth Adhesives8001286
Gas-tight syringeHamilton81520
Luer Lock to 360 µm adapterIDEXp-662
Silica CapillaryBGB AnalytikTSP250350
Syringe Pump Universalna
UV GlueAmazonB09H7BJKT1
Vytran Glass Processor Thorlabs/VytranGPX3000PGP instrument with software

Riferimenti

  1. Dantham, V. R., Holler, S., Barbre, C., Keng, D., Kolchenko, V., Arnold, S. Label-free detection of single protein using a nanoplasmonic-photonic hybrid microcavity. Nano Lett. 13 (7), 3347-3351 (2013).
  2. Chen, W., Özdemir, &. #. 3. 5. 0. ;. K., Zhao, G., Wiersig, J., Yang, L. Exceptional Points enhance sensing in an optical microcavity. Nature. 548 (7666), 192-195 (2017).
  3. Zhu, J., et al. On-chip single nanoparticle detection and sizing by mode splitting in an ultrahigh-Q microresonator. Nat Photonics. 4 (1), 46-49 (2010).
  4. Dantham, V. R., Holler, S., Kolchenko, V., Wan, Z., Arnold, S. Taking whispering gallery-mode single virus detection and sizing to the limit. Appl Phys Lett. 101, 043704 (2012).
  5. Vollmer, F., Arnold, S. Whispering-gallery-mode biosensing: Label-free detection down to single molecules. Nat Methods. 5 (7), 591-596 (2008).
  6. Wang, Y., Zeng, S., Humbert, G., Ho, H. P. Microfluidic whispering gallery mode optical sensors for biological applications. Laser Photonics Rev. 14 (12), 2000135 (2020).
  7. Li, B. B., Bulla, D., Prakash, V., Forstner, S., Dehghan-Manshadi, A., Rubinsztein-Dunlop, H., Foster, S., Bowen, W. P. Invited Article: Scalable high-sensitivity optomechanical magnetometers on a chip. APL Photonics. 3, 120806 (2018).
  8. Ioppolo, T., Ayaz, U., Ötügen, M. V. Tuning of whispering gallery modes of spherical resonators using an external electric field. Opt Express. 17 (19), 16465 (2009).
  9. Xu, L., et al. High-Q silk fibroin whispering gallery microresonator. Opt Express. 24 (18), 20825 (2016).
  10. Kim, K. H., et al. Air-coupled ultrasound detection using capillary-based optical ring resonators. Sci Rep. 7, 109 (2017).
  11. Pan, J., et al. Microbubble resonators combined with a digital optical frequency comb for high-precision air-coupled ultrasound detectors. Photonics Res. 8 (3), 303 (2020).
  12. Foreman, M. R., Swaim, J. D., Vollmer, F. Whispering gallery mode sensors. Opt Express. 7 (2), 168-240 (2015).
  13. Sumetsky, M., Dulashko, Y., Windeler, R. S. Optical microbubble resonator. Opt Lett. 35 (7), 898 (2010).
  14. Jiang, X. F., Xiao, Y. F., Yang, Q. F., Shao, L., Clements, W. R., Gong, Q. Free-space coupled, ultralow-threshold Raman lasing from a silica microcavity. Appl Phys Lett. 103, 101102 (2013).
  15. Xu, Q., Schmidt, B., Pradhan, S., Lipson, M. Micrometre-scale silicon electro-optic modulator. Nature. 435 (7040), 325-327 (2005).
  16. Henze, R., Seifert, T., Ward, J., Benson, O. Tuning whispering gallery modes using internal aerostatic pressure. Opt Lett. 36 (23), 4536 (2011).
  17. Wade, J. H., Jones, J. D., Lenov, I. L., Riordan, C. M., Sligar, S. G., Bailey, R. C. Microfluidic platform for efficient nanodisc assembly, membrane protein incorporation, and purification. Lab Chip. 17 (17), 2951-2959 (2017).
  18. Zhu, H., White, I. M., Suter, J. D., Zourob, M., Fan, X. Opto-fluidic micro-ring resonator for sensitive label-free viral detection. Analyst. 133 (3), 356-360 (2008).
  19. White, I. M., Gohring, J., Sun, Y., Yang, G., Lacey, S., Fan, X. Versatile waveguide-coupled optofluidic devices based on liquid core optical ring resonators. Appl Phys Lett. 91, 241104 (2007).
  20. Wang, Z., et al. Monitoring and identifying pendant droplets in microbottle resonators. Photonics Res. 10 (3), 662 (2022).
  21. Liu, W., et al. Magnetic sensor based on WGM hollow microbubble resonator filled with magnetic fluid. Opt Commun. 497, 127148 (2021).
  22. Hogan, L. T., Horak, E. H., Ward, J. M., Knapper, K. A., Nic Chormaic, S., Goldsmith, R. H. Toward real-time monitoring and control of single nanoparticle properties with a microbubble resonator spectrometer. ACS Nano. 13 (11), 12743-12757 (2019).
  23. Yu, J., et al. A tellurite glass optical microbubble resonator. Opt Express. 28 (22), 32858-32868 (2020).
  24. Guo, W., Liu, J., Liu, J., Wang, G., Wang, G., Huang, M. A single-ended ultra-thin sphericalmicrobubble based on the improved critical-state pressure-assisted arc discharge method. Coatings. 9 (2), 144 (2019).
  25. Jiang, J., et al. Wall-thickness-controlled microbubble fabrication for WGM-based application. Appl Opt. 59 (16), 5052 (2020).
  26. Wang, B., Mies, E. Review of fabrication techniques for fused fiber components for fiber lasers. Fiber Lasers VI Technol Syst Appl. 7195, 71950A (2009).
  27. Yang, Y., Ward, J., Chormaic, S. N. Quasi-droplet microbubbles for high resolution sensing applications. Opt Express. 22 (6), 6881 (2014).
  28. . University of Rochester Ultraviolet Light Safety Guidelines Available from: https://www.safety.rochester.edu/ih/guidelines/pdf/Ultraviolet_Light_Guidelines.pdf (2021)
  29. Sun, Y., Shopova, S. I., Frye-mason, G., Fan, X. Rapid chemical-vapor sensing using optofluidic ring resonators. Opt. Lett. 33 (8), 788-790 (2008).
  30. Giannetti, A., et al. Optical micro-bubble resonators as promising biosensors. Proc SPIE 9506, Optical Sensors. 2015, 950617 (2015).
  31. Guo, Z., Lu, Q., Zhu, C., Wang, B., Zhou, Y., Wu, X. Ultra-sensitive biomolecular detection by external referencing optofluidic microbubble resonators. Opt Express. 27 (9), 12424 (2019).
  32. Berneschi, S., et al. High Q silica microbubble resonators fabricated by arc discharge. Opt Lett. 36 (17), 3521 (2011).
  33. Yang, Y., Saurabh, S., Ward, J. M., Nic Chormaic, S. High-Q, ultrathin-walled microbubble resonator for aerostatic pressure sensing. Opt Express. 24 (1), 294-299 (2016).
  34. Jiang, X., Qavi, A. J., Huang, S. H., Yang, L. Whispering-gallery sensors. Matter. 3 (2), 371-392 (2020).
  35. Yu, D., Humar, M., Meserve, K., Bailey, R. C., Chormaic, S. N., Vollmer, F. Whispering-gallery-mode sensors for biological and physical sensing. Nat Rev Methods Prim. 1, 83 (2021).

Ristampe e Autorizzazioni

Richiedi autorizzazione per utilizzare il testo o le figure di questo articolo JoVE

Richiedi Autorizzazione

Esplora altri articoli

Engineering

This article has been published

Video Coming Soon

JoVE Logo

Riservatezza

Condizioni di utilizzo

Politiche

Ricerca

Didattica

CHI SIAMO

Copyright © 2025 MyJoVE Corporation. Tutti i diritti riservati