Diffusion tensor imaging (DTI) was performed to try to depict the major parts of the visual pathway. The goal was to use an FDA approved standard commercial workstation which could be used for everyday routine in order to try to reduce postoperative damage of the visual pathway in the patients.
DTI is a technique that identifies white matter tracts (WMT) non-invasively in healthy and non-healthy patients using diffusion measurements. Similar to visual pathways (VP), WMT are not visible with classical MRI or intra-operatively with microscope. DTI will help neurosurgeons to prevent destruction of the VP while removing lesions adjacent to this WMT. We have performed DTI on fifty patients before and after surgery between March 2012 to January 2014. To navigate we used a 3DT1-weighted sequence. Additionally, we performed a T2-weighted and DTI-sequences. The parameters used were, FOV: 200 x 200 mm, slice thickness: 2 mm, and acquisition matrix: 96 x 96 yielding nearly isotropic voxels of 2 x 2 x 2 mm. Axial MRI was carried out using a 32 gradient direction and one b0-image. We used Echo-Planar-Imaging (EPI) and ASSET parallel imaging with an acceleration factor of 2 and b-value of 800 s/mm². The scanning time was less than 9 min.
The DTI-data obtained were processed using a FDA approved surgical navigation system program which uses a straightforward fiber-tracking approach known as fiber assignment by continuous tracking (FACT). This is based on the propagation of lines between regions of interest (ROI) which is defined by a physician. A maximum angle of 50, FA start value of 0.10 and ADC stop value of 0.20 mm²/s were the parameters used for tractography.
There are some limitations to this technique. The limited acquisition time frame enforces trade-offs in the image quality. Another important point not to be neglected is the brain shift during surgery. As for the latter intra-operative MRI might be helpful. Furthermore the risk of false positive or false negative tracts needs to be taken into account which might compromise the final results.
Diffusion tensor imaging (DTI) is used to portray WMT non-invasively in the human brain1. It has been used in the past decade to reduce the risk of harming eloquent areas of the brain during surgery1.
DTI was performed in fifty patients between March 2012 and January 2014 to portray the visual pathway. DTI might improve preservation of eloquent areas of the brain during surgery by providing important information about the anatomical location of white matter tracts. It has been incorporated into strategic planning for resection of complex brain lesions1. However, the portrayal of the visual pathway remains a challenge because there is no standard for the parameters of DTI, the placement of the seed volumes and interpretation of results12.
Different algorithms have been implemented so far 19-21. Some approaches concentrated on deterministic methods 19, 22-25. Others were using probabilistic methods,26,27,29. More recently, techniques using Q-ball tensor fields, diffusion spectral imaging and High Angular Resolution Diffusion Imaging (HARDI) are being used to depict white matter tracts among others the visual pathway 1,13-15,18 . Nevertheless, the necessary time for HARDI is significantly longer with 45 min, the software is not commercially available and emphasizes scientific applications18. The teaching period for HARDI seems to be longer than for DTI18.
The presented protocol is easy feasible and can be used for everyday routine in neurosurgical operations in order to avoid morbidity and improve the postoperative outcome. The additional time for this protocol is less than 9 min which is significantly faster than other protocols1,9,12,16. Acknowledging the fact that many sophisticated algorithms have been developed recently the paper restricts itself to the use of a commercially available and FDA approved software. However it is mandatory to take into account the limitations of this technique which are mentioned above.
NOTE: This protocol follows the guidelines of the Centre Hospitalier de Luxembourg in Luxembourg.
1. Preparation of Diffusion Tensor Imaging for the Visual Pathway for Neurosurgery and Follow Up
2. Using the Planning Station
3. Fibertracking
NOTE: Anatomic knowledge of the visual pathway is very important for the successful result.
This protocol enables the physician to portray adequately the major parts of the VP. It can be used with a little amount of time in order to prevent damages in patients with cerebral lesions next to eloquent areas. Postoperative controls show also good results. VP is portrayed in Figure 7 after the patient was operated from a glioblastoma. Figure 2 shows the VP after recurrence of a glioblastoma. The authors recognize the fact of the difficulties presented by this protocol to depict the ...
DTI is a technique enabling the neurosurgeon to visualize white matter tracts in vivo8. The visual pathway is one of these tracts. Although this method provides physicians with new possibilities regarding the treatment of patients with lesions concerning eloquent regions of the brain we have to say that some limitations of this technique do still exist. The first and most obvious challenge is brain shift, which remains an issue under investigation4. After opening the dura mater and after ma...
The authors declare that they have no competing financial interests.
We would like to thank the whole Service of Neuroradiology. We would like to thank Lis Prussen for her work in the library.
Name | Company | Catalog Number | Comments |
Name of Material/ Equipment | Company | Catalog Number | Comments/Description |
3-Tesla-MRI | General Electric | Signa LX version 9.1 | |
Surgical Navigation System Srogram | Medtronic | 9734478 | |
Surgical Navigation System Srogram | Medtronic | 4500810331 20016318 |