Capillary electrophoresis (CE) is a separation technique that separates molecules in an electric field according to size and charge. CE is performed in a small glass tube called a capillary that is filled with an electrolyte solution. Analytes are separated due to differences in electrophoretic mobility, which varies with charge, solvent viscosity, and size. Traditional electrophoresis in gels is limited in the amount of voltage that can be applied because Joule heating effects will ruin the gel and the separation. Capillaries have a large surface area-to-volume ratio and thus dissipate heat better. Therefore, the voltages applied for a capillary electrophoresis experiment are quite large, often 10,000–20,000 V.

Capillary electrophoresis is useful for high-performance separations. Compared to liquid chromatography, CE separations are often faster and more efficient. However, capillary electrophoresis works best to separate charged molecules, which is not a limitation of liquid chromatography. CE has a greater peak capacity than high-performance liquid chromatography (HPLC), meaning the separations are more efficient and more peaks can be detected. The instrumentation can be very simple. However, HPLC is more versatile and many stationary and mobile phases have been developed for different types of molecules.

Процедура

1. CE Instrumentation Setup

  1. Turn on the CE instrument and computer. Using the computer software, turn on the light source for UV analysis to allow it to warm up. Some software has an indicator when the lamp is ready for use (lamp icon turns color).
  2. Make a methods file. Set the important parameters for running the CE. In this analysis the temperatures of the cartridge and sample storage are 35 °C. The wavelength for UV detection is 214 nm.
  3. Write a time program. The program general

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Результаты

Electropherograms collected for diet Pepsi and Pepsi samples are shown in Figures 1 and 2, respectively. The three peaks for caffeine, aspartame, and benzoic acid are observed in diet Pepsi and have similar migration times as the standards. For regular Pepsi, the caffeine peak is present but not the aspartame and benzoic acid peaks. The CE analysis is fast as the migration times are only 3–4 min.

The calibration curve for caffeine is shown in Fig

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Заявка и Краткое содержание

Capillary electrophoresis is used for many specialty separations. For example, it is used in the pharmaceutical industry for quality testing, to make sure there are no side products or interferents. CE is particularly useful for separating drugs with a basic amino group, as the walls of the capillary can be made neutral with an acidic pH and thus the drug will not stick to the capillary.

A mode of CE was also used to sequence the human genome and separate DNA. This mode of CE is capillary gel

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Теги
Capillary ElectrophoresisCEChemical AnalysisSeparate MoleculesElectric FieldSizeChargeSub millimeter Diameter TubeCapillaryFlowing Electrolyte SolutionVelocityChargeSizeSolvent s ViscosityResolutionHigh performance Liquid ChromatographyEfficiencySensitivity

Перейти к...

0:00

Overview

1:06

Principles of Capillary Electrophoresis

3:19

CE Instrumentation Setup

4:26

Preparation of the Standards and Soda Samples

5:23

Running the Samples

6:50

Applications

8:26

Summary

Видео из этой коллекции:

article

Now Playing

Capillary Electrophoresis (CE)

Analytical Chemistry

92.7K Просмотры

article

Sample Preparation for Analytical Characterization

Analytical Chemistry

83.1K Просмотры

article

Internal Standards

Analytical Chemistry

202.8K Просмотры

article

Method of Standard Addition

Analytical Chemistry

318.2K Просмотры

article

Calibration Curves

Analytical Chemistry

787.3K Просмотры

article

Ultraviolet-Visible (UV-Vis) Spectroscopy

Analytical Chemistry

614.7K Просмотры

article

Raman Spectroscopy for Chemical Analysis

Analytical Chemistry

50.5K Просмотры

article

X-ray Fluorescence (XRF)

Analytical Chemistry

25.2K Просмотры

article

Gas Chromatography (GC) with Flame-Ionization Detection

Analytical Chemistry

279.1K Просмотры

article

High-Performance Liquid Chromatography (HPLC)

Analytical Chemistry

380.4K Просмотры

article

Ion-Exchange Chromatography

Analytical Chemistry

262.4K Просмотры

article

Introduction to Mass Spectrometry

Analytical Chemistry

111.0K Просмотры

article

Scanning Electron Microscopy (SEM)

Analytical Chemistry

86.3K Просмотры

article

Electrochemical Measurements of Supported Catalysts Using a Potentiostat/Galvanostat

Analytical Chemistry

51.1K Просмотры

article

Cyclic Voltammetry (CV)

Analytical Chemistry

122.9K Просмотры

JoVE Logo

Исследования

Образование

О JoVE

Авторские права © 2025 MyJoVE Corporation. Все права защищены