Here, we propose a practical, feasible and reproducible evaluation guideline for computer-assisted reconstruction of the mandible in order to create uniformity between studies regarding postoperative accuracy evaluation. This protocol continues and specifies an earlier publication of this evaluation guideline.
Valid comparisons of postoperative accuracy results in computer-assisted reconstruction of the mandible are difficult due to heterogeneity in imaging modalities, mandibular defect classification, and evaluation methodologies between studies. This guideline uses a step-by-step approach guiding the process of imaging, classification of mandibular defects and volume assessment of three-dimensional (3D) models, after which a legitimized quantitative accuracy evaluation method can be performed between the postoperative clinical situation and the preoperative virtual plan. The condyles and the vertical and horizontal corners of the mandible are used as bony landmarks to define virtual lines in the computer-assisted surgery (CAS) software. Between these lines the axial, coronal, and both sagittal mandibular angles are calculated on both pre- and postoperative 3D models of the (neo)mandible and subsequently the deviations are calculated. By superimposing the postoperative 3D model to the preoperative virtually planned 3D model, which is fixed to the XYZ axis, the deviation between pre- and postoperative virtually planned dental implant positions can be calculated. This protocol continues and specifies an earlier publication of this evaluation guideline.
Computer-assisted surgery (CAS) in reconstructive surgery involves four consecutive phases: a virtual planning phase, a three-dimensional (3D) modeling phase, a surgical phase, and a postoperative evaluation phase1. The planning phase starts with the obtainment of a craniofacial computed tomography (CT) scan, and a donor site CT or CT angiography (CTA) scan. Diverse tissue types correspond to an amount of X-ray attenuation, which leads to scan voxels with a specific gray value ranged according to Hounsfield units (HU) (human bone [+1000 HU], water [0 HU], and air [-1000 HU]). These images are stored in Digital Imaging and Communications in Medicine (DICOM) file format. By selecting the regions of interest (ROIs) in segmentation software, 3D models can be generated2. The most popular and feasible segmentation technique is thresholding: voxels above a selected HU threshold value are enclosed in the ROI. These voxels are subsequently converted into 3D models in the Standard Tessellation Language (STL) file format3, and uploaded into CAS software to plan the osteotomies and to design 3D devices4. During the modeling phase, the designed devices are 3D printed and sterilized, followed by the surgical phase. The final evaluation phase consists of a postoperative CT scan of the patient's skull, followed by an accuracy analysis comparing the postoperative result with the preoperative virtual plan.
Our recently published systematic review regarding accuracy of computer-assisted mandibular reconstructions showed heterogeneity in image acquisition, classification of mandibular defects, and evaluation methodologies. This heterogeneity limits valid comparisons of postoperative hard tissue accuracy results between studies5. Standardization of CAS phases in the process of mandibular reconstruction is important due to the new European Union medical device regulation (MDR), which demands Conformité Européenne (CE) certification for all different CAS processes, and which will be operational from spring 20206. Here, we present a practical, feasible and reproducible evaluation guideline for computer-assisted reconstructions of the mandible in order to create uniformity between studies regarding postoperative accuracy evaluation. This protocol continues and specifies an earlier publication of this evaluation guideline7, which is currently being tested in a large multicenter cohort study in which all different types of mandibular reconstructions will be analyzed for their accuracy aiming to discover tolerable outcome ranges regarding functionality.
The Medical Ethics Review Committee of VU University Medical Center (registered with the US Office for Human Research Protections [OHRP] as IRB00002991) confirmed that the Medical Research Involving Human Subjects Act (WMO) does not apply to this study. The FWA number assigned to VU University Medical Center is FWA00017598.
NOTE: Validate all steps in this protocol independently by two different observers.
1. Skull and donor site imaging
2. Classification of the mandibular defect
3. Segmentation of the DICOM images of the postoperative CT scan
4. XYZ axis orientation
NOTE: The preoperative STL model includes the cranium, (neo)mandible, and the virtually planned dental implants (if planned). Note that the evaluation works easier with separated STL files of the cranium and skull, but still in fixed position to each other. When the preoperative STL model of the cranium and the mandible are merged, use the 3D medical software (following the steps described above) to split the mandible from the cranium.
5. Volume assessment of the pre- and postoperative STL models
NOTE: Examine pre- and postoperative STL models on volume similarity to rule out volume inaccuracies between the two models as much as possible, since they can influence accuracy measurements.
6. Superimposition of the condylar processes
7. Calculation of the coronal, axial, and sagittal mandibular angles
NOTE: The identification of bony landmarks is performed separately on the "Mandible Pre-op" and "Mandible Post-op" STL models. Deselect the Mandible Post-op while identifying bony landmarks in the "Mandible Pre-op", and vice versa.
8. Calculation of the XYZ deviations and distance XYZ of the virtually planned dental implants
NOTE: Use the correct dental implant diameter and height (including cover screw) during the preoperative planning for correct comparison.
A James Brown class III mandibular defect was reconstructed at our department with the fibula free flap as a donor site. Direct guided dental implant placement was performed with the use of a fibula cutting guide which also included dental implant guides. The reconstruction was evaluated with the presented guideline. The coronal, axial and sagittal mandibular angle deviations (°) and six dental implant XYZ distances (mm) were calculated and reported (Figure 4 and Figure 5).
Figure 1: Uniform orientation of the preoperative STL model of the skull on the XYZ axis with the Frankfurt plane projected to the Z axis (red line), the midsagittal plane projected to the Y axis (green line), and the nasion projected to the X axis (blue line). Please click here to view a larger version of this figure.
Figure 2: A part of the right side of the mandible (without involvement of osteosynthesis material which causes scattering) of the preoperative virtually planned STL model is superimposed on the postoperative STL model. Subsequently the CAS software is used to calculate the arithmetic mean. The 0.02 mm deviation between both volumes in this example falls within the norm (<0.5 mm) to proceed to the next step of the evaluation guideline. Please click here to view a larger version of this figure.
Figure 3: Superimposition of the postoperative STL model (grey) on the preoperative STL model, revised to the virtual plan (blue). Only both condylar processes are selected for the iterative closest-point algorithm (red). Please click here to view a larger version of this figure.
Figure 4: Reconstruction of a Brown class III defect using the fibula free flap as a donor site. In this example, six virtually planned dental implants are primary placed during the reconstruction using a 3D guide. The coronal, axial and sagittal angles are calculated on both the preoperative virtually planned 3D model and the postoperative 3D model. The deviations between the angles in degrees (°) are shown. CS, condyle superior; CP, condyle posterior; VC, vertical corner; HC, horizontal corner; ML, midsagittal line; FFF, fibula free flap. Please click here to view a larger version of this figure.
Figure 5: Dental implant deviations on the X, Y and Z axis and the distance XYZ (dXYZ) of the six guided placed dental implants. Please click here to view a larger version of this figure.
This postoperative evaluation guideline aims to facilitate increased uniformity of accuracy analysis of computer-assisted mandibular reconstructions. The focus is on four components determining the success of mandibular reconstruction: (1) the position of both condyles, (2) the angles of the osteotomy planes, (3) the size, position and fixation of the bone graft segments, and (4) the position of the guided dental implants (if immediate performed and included in the virtual planning).
In the first step of our proposed protocol, we recommend MDCT scanning for both pre- and postoperative imaging, because the quality of CT images affects the volume accuracy of segmented STL models. The largest volume deviations are found in STL models segmented out of cone beam computed tomography (CBCT) scanner DICOM data11. These volume deviations influence accuracy and fitting of 3D printed templates and guides, and thus also influences postoperative accuracy measurements between pre- and postoperative STL models. Therefore, we recommend the use of MDCT scanners in both pre- and postoperative imaging for mandibular reconstruction using CAS. Slice thickness is the most influencing factor in STL volume accuracy and should be set <1.25 mm. A higher slice thickness yields to loss of detail in the STL models and affects accuracy measurements12,13. A recently published systematic review on accuracy in mandibular reconstruction using CAS showed poor description in the materials and methods section of CT scanner parameters used by authors5. In our opinion, CAS studies should always specify the type and parameters of pre- and postoperative imaging modalities in the materials and methods section. In order to avoid long term changes in the volume, shape, and position of the segments of the bone graft, the postoperative MDCT scan should be performed within six weeks after reconstruction14. In case of adjuvant radiation therapy, use the first postoperative MDCT scan prior to the therapy to avoid radiation related pathology in the mandibular bone15.
Classification of mandibular defects is needed to compare reconstructions with similar complexity. In 2016, Brown et al.8 proposed a mandibular defect classification describing four classes, with a relationship between the class number and the complexity of the reconstruction. The alignment of pre- and postoperative STL models in the CAS software to evaluate the accuracy of the reconstruction introduces some difficulties. The superimposition software tool moves a selected part of an STL model (the source) to best match a fixed part of an STL model (the reference) using an iterative closest-point algorithm. However, superimposition of the entire (neo)mandible is inaccurate due to scattering of the reconstruction plate(s), which will lead to shifts of the entire reconstruction, not representing the postoperative clinical position of the mandible16. The same problem is introduced while superimposing isolated parts of the reconstruction17. Superimposition of the mandible including the maxilla and cranium is inaccurate because mouth opening will always be different during the pre- and postoperative scanning. Therefore, to evaluate the postoperative position of the (neo)mandible we decided to create mandibular angles (pioneered by De Maesschalck et al.18) on both pre-and postoperative STL models separately to bypass the superimposition problems. However, to evaluate the dental implant positions we necessarily needed to align both models, using the superimposition software tool. To align pre- and postoperative STL models with the closest approach to the clinical postoperative intermaxillary relation, we believe that superimposition of only both condylar processes is the most feasible, standardized and reproducible method. Although the postoperative position of both condyles can be affected by inaccurate neomandible reconstruction, the intermaxillary relation will accommodate to the midline and thus averages the position of both condyles around the midsagittal plane19. In our protocol, only the preoperative STL model is quickly fixed to the XYZ axis using a plane-line-point tool in the CAS software, representing a benchmark from which the postoperative deviations of the dental implants can be determined. The fixed skull position on the XYZ axis can lead to small cephalometric differences between cases. However, this has no influence on the dental implant measurements, because it has no consequences for the distance XYZ in mm between dental implant positions when the postoperative 3D model is superimposed onto the fixed preoperative 3D model with only both condyles selected for the iterative closest point algorithm.
As described above, De Maesschalck et al.18 pioneered an evaluation method for hard tissue accuracy of mandibular reconstruction using CAS, bypassing the need for osteotomy plane determination and bypassing the use of a superimposition tool. The most serious disadvantage of this method is that it failed to specify the method used to determine the midsagittal plane, which needs to be standardized and reproducible. Also, no virtually planned dental implants are included and a differentiation between complexity of mandibular reconstructions is lacking. We included the evaluation of postoperative positions of virtually planned dental implants in our protocol because the number of authors applying guided dental implants in the future is likely to increase. In 2016, Schepers et al.20 proposed an excellent postoperative evaluation method for virtually planned dental implants in mandibular reconstruction using CAS by measuring the center point deviation (mm) and angular deviation (°) per dental implant. The main limitation of this method is the quantity of measurements per implant which decreases the feasibility and results in loss of overview of accuracy of the entire reconstruction. We propose a more simplified method by determining one recapitulatory number per dental implant by measuring the distance XYZ (dXYZ in mm). With regard to dental rehabilitation, the position of the neck of the dental implant is decisive for future prosthetics. Therefore, our evaluation protocol recommends creating virtual points on the neck of the dental implants in the pre- and postoperative STL models. To keep the evaluation of the dental implants feasible we decided to skip angular deviation measurements, because angular deviations up to 15° can be corrected with angled implant abutments.
Our proposed guideline is applicable for all types of donor sites and allows for different bone graft fixation possibilities. Also, CT scattering of metal fixation parts in the postoperative imaging will not influence measurements of the guideline5. In this evaluation guideline, we used Mimics inPrint 3.0 and GOM Inspect Professional 2019. However, the protocol describes software tools which are available in all CAS software packages. This guideline aims to contribute to a much more standardized and uniform approach to objectify relationships between accuracy and all different approaches during the CAS phases. There is abundant room for further progress in determining acceptable mandibular angle deviations per Brown class, their relationship with the postoperative positions of virtually planned dental implants, and acceptable dental implant deviations (dXYZ) for future prosthetics. Currently, our department is conducting a multicenter study to validate this guideline in a large cohort, which also takes all the above-mentioned variables into account.
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Name | Company | Catalog Number | Comments |
GOM Inspect Professional 2019 | GOM | Evaluation software | |
Mimics inPrint 3.0 | Materialise | Image-based 3D medical software |
This article has been published
Video Coming Soon
ABOUT JoVE