Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions. While fluorometers are suitable for quantitative work and can measure fluorescence intensity, they cannot record excitation or emission spectra.

In contrast, spectrofluorometers use a monochromator to select excitation and emission wavelengths. These instruments employ a high-pressure xenon arc lamp as the excitation source, which produces a continuum emission spectrum. Spectrofluorometers can record excitation or emission spectra, allowing for a more detailed fluorescence analysis. They are also suitable for quantitative measurements.

In the case of molecular phosphorescence, specialized instrumentation, known as phosphorimetery, is required to discriminate between phosphorescence and fluorescence. Phosphorescence has a longer lifetime than fluorescence, so a delay is incorporated between excitation and measuring phosphorescent emission. This can be achieved using two choppers that rotate out of phase. The choppers block the fluorescent emission when the excitation source is focused on the sample and block the excitation source when measuring phosphorescent emission. To prevent deactivation of the excited state by external conversion, a form of radiationless relaxation in which energy is transferred to the solvent or sample matrix instead of being emitted as light. Samples for phosphorescence measurements are typically dissolved in a suitable organic solvent and frozen at liquid nitrogen temperatures. This forms an optically clear solid matrix that minimizes external conversion. Alternatively, the sample can be immobilized on a solid substrate, enabling phosphorescence measurement at room temperature.

From Chapter 12:

article

Now Playing

12.14 : Fluorescence and Phosphorescence: Instrumentation

Introduction to Molecular Spectroscopy

128 Views

article

12.1 : Dual Nature of Electromagnetic (EM) Radiation

Introduction to Molecular Spectroscopy

1.6K Views

article

12.2 : Interaction of EM Radiation with Matter: Spectroscopy

Introduction to Molecular Spectroscopy

908 Views

article

12.3 : Molecular Spectroscopy: Absorption and Emission

Introduction to Molecular Spectroscopy

637 Views

article

12.4 : Spectrophotometry: Introduction

Introduction to Molecular Spectroscopy

2.3K Views

article

12.5 : Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview

Introduction to Molecular Spectroscopy

2.0K Views

article

12.6 : UV–Vis Spectroscopy: Molecular Electronic Transitions

Introduction to Molecular Spectroscopy

928 Views

article

12.7 : UV–Vis Spectrometers

Introduction to Molecular Spectroscopy

825 Views

article

12.8 : UV–Vis Spectrum

Introduction to Molecular Spectroscopy

764 Views

article

12.9 : UV–Vis Spectroscopy: Beer–Lambert Law

Introduction to Molecular Spectroscopy

1.1K Views

article

12.10 : Photoluminescence: Fluorescence and Phosphorescence

Introduction to Molecular Spectroscopy

136 Views

article

12.11 : Variables Affecting Phosphorescence and Fluorescence

Introduction to Molecular Spectroscopy

58 Views

article

12.12 : Deactivation Processes: Jablonski Diagram

Introduction to Molecular Spectroscopy

123 Views

article

12.13 : Photoluminescence: Applications

Introduction to Molecular Spectroscopy

82 Views

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved