JoVE Logo

Sign In

A subscription to JoVE is required to view this content. Sign in or start your free trial.

In This Article

  • Summary
  • Abstract
  • Introduction
  • Protocol
  • Results
  • Discussion
  • Disclosures
  • Acknowledgements
  • Materials
  • References
  • Reprints and Permissions

Summary

This document describes a new dosimetry protocol for cell irradiations using low energy X-ray equipment. Measurements are performed in conditions simulating real cell irradiation conditions as much as possible.

Abstract

The importance of dosimetry protocols and standards for radiobiological studies is self-evident. Several protocols have been proposed for dose determination using low energy X-ray facilities, but depending on the irradiation configurations, samples, materials or beam quality, it is sometimes difficult to know which protocol is the most appropriate to employ. We, therefore, propose a dosimetry protocol for cell irradiations using low energy X-ray facility. The aim of this method is to perform the dose estimation at the level of the cell monolayer to make it as close as possible to real cell irradiation conditions. The different steps of the protocol are as follows: determination of the irradiation parameters (high voltage, intensity, cell container etc.), determination of the beam quality index (high voltage-half value layer couple), dose rate measurement with ionization chamber calibrated in air kerma conditions, quantification of the attenuation and scattering of the cell culture medium with EBT3 radiochromic films, and determination of the dose rate at the cellular level. This methodology must be performed for each new cell irradiation configuration as the modification of only one parameter can strongly impact the real dose deposition at the level of the cell monolayer, particularly involving low energy X-rays.

Introduction

The aim of radiobiology is to establish links between the delivered dose and the biological effects; dosimetry is a crucial aspect in the design of radiobiological experiments. For over 30 years, the importance of dosimetry standards and the harmonization of practices have been highlighted1,2,3,4,5. To establish a dose rate reference, several protocols exist6,7,8,9,10; however, as shown by Peixoto and Andreo11 , there can be differences of up to 7% depending on the dosimetric quantity used for the dose rate determination. Moreover, even if protocols exist, it is sometimes difficult to know which protocol is the most suitable for a particular application, if any, because the dose rate for the cells depends on parameters such as the cell container, quantity of cell culture media or beam quality, for example. The scattering and the backscattering for this type of irradiation is also a very important parameter to take into account. Indeed, for low and medium energy X-rays, in the AAPM TG-61 reference protocol10, the absorbed dose in water is measured at the surface of a water phantom. Taking into account the very specific cell irradiation conditions, the small volume of cell culture media surrounded by air is closer to kerma conditions than those defined for an absorbed dose with a large water equivalent phantom as in the TG-61 protocol. Therefore, we have chosen to use the kerma in water as a dosimetric quantity for reference rather than the absorbed dose in water. Thus, we are proposing a new approach to provide a better determination of the actual dose delivered to cells.

Moreover, another crucial aspect for radiobiological studies is the complete reporting of the methods and protocols used for irradiation in order to be able to reproduce, interpret and compare experimental results. In 2016, Pedersen et al.12 highlighted the inadequate reporting of dosimetry in preclinical radiobiological studies. A larger recent study from Draeger et al.13 highlighted that even though some dosimetry parameters such as the dose, energy, or source type are reported, a large part of the physics and dosimetry parameters that are essential to properly replicate the irradiation conditions are missing. This large scale review, of more than 1,000 publications covering the past 20 years, shows a significant lack of the reporting of the physics and dosimetry conditions in radiobiological studies. Thus, a complete description of the protocol and the method utilized in radiobiological studies is mandatory in order to have robust and reproducible experiments.

Taking into account these different aspects, for the radiobiological experiments carried out at IRSN (Institute of Radiation Protection and Nuclear Safety), a stringent protocol was implemented for cell irradiations in an orthovoltage facility. This dosimetry protocol was designed in order to simulate the real cell irradiation conditions as much as possible and thus, to determine the actual dose delivered to cells. To this end, all the irradiation parameters are listed, and the beam quality index was evaluated by measuring the half value layer (HVL) for which some adaptations have been made as the standard recommendations from the AAPM protocol10 cannot be followed. The absolute dose rate measurement was then performed with the ionization chamber inside the cell container used for cell irradiations, and the attenuation and the scattering of the cell culture media was also quantified with EBT3 radiochromic films. As the modification of only one single parameter of the protocol can significantly impact the dose estimation, a dedicated dosimetry is performed for each cell irradiation configuration. Moreover, the HVL value must be calculated for each voltage-filter combination. In this present work, a voltage of 220 kV, an intensity of 3 mA, and an inherent and an additional filtration of 0.8 mm and 0.15 mm of beryllium and copper, respectively, are used. The cell irradiation configuration chosen is on a T25 flask, where cells were irradiated with 5 mL of cell culture media.

Access restricted. Please log in or start a trial to view this content.

Protocol

1. Irradiation platform and determination of irradiation parameters

  1. Use an irradiation platform delivering low to medium energy X-rays. Determine the parameters of the experiment to ensure the robustness and the reproducibility of the radiobiological experiment: High voltage, Intensity, Filtration (inherent and additional), Half Value Layer (HVL), Effective energy, Detector used for dosimetry measurements, Source Sample Distance (SSD), Irradiation field (shape, size, geometry), Dosimetry quantity, Dosimetry method, Dose rate, Cell container and Quantity of cell culture media. All parameters used in this protocol are given in Table 1.

2. Beam quality index: determination of the half value layer

NOTE: The HVL is defined as the thickness of an attenuator (usually copper or aluminum) to reduce the intensity of the beam by a factor of two compared with the original value.

  1. Set up the equipment (support, collimator, diaphragm, ionization) inside the irradiation enclosure by following the instructions in Figure 1. No attenuator material is used at this step.
  2. Ensure that all the distances reported in Figure 1 are correct. Measure these with a tape measure.
  3. Place that the ionization chamber in the horizontal position. For this work, we used a 31002 (equivalent to 31010) cylindrical ionization chamber calibrated in air kerma.
  4. Pre-irradiate the ionization chamber for 5 min and measure the background (this step can be performed without a collimator).
  5. Perform 10 measurements of 1 min each in charge collection mode corresponding to the Mraw value (in coulombs).
  6. Take the temperature and pressure with suitable calibrated equipment placed inside the irradiation enclosure in our case (if it is not possible, place it near to the experiment). Correct the Mraw reading on the electrometer by the temperature and pressure correction factor given as follows:
    figure-protocol-2133
    where: T (°C) and P (hPa) are the actual temperature and pressure, respectively. Tref and Pref are the reference temperature and pressure when the ionization was calibrated by the standards laboratory. The pressure and temperature must be measured with calibrated instruments. The obtained value in charge mode is the average reference value M (in coulombs).
    NOTE: This step is not strictly necessary for HVL measurement, but it is recommended.
  7. Place an attenuator of certain thickness above the diaphragm. The HVL set is composed of foils with different thicknesses (0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5 and 10 mm of copper) with a dimension allowing to cover the entire beam (80 x 80 mm here).
  8. Take a measurement of 1 min (Mraw corrected by the KT,P as described before).
    1. If the dose rate is divided by a factor of 2 with respect to the starting value, the HVL value is found. Take 5 measurements of 1 min to estimate the average dose rate.
    2. If the dose rate is not divided by a factor of 2 with respect to the starting value, increase or decrease the attenuator thickness and take another measurement. Adjust the thickness of the attenuator as necessary.
  9. Once the thickness of the attenuator that decreases the intensity of the beam by a factor two is found, take 5 measurements of 1 min to confirm the HVL.
    NOTE: In most cases, the exact thickness of the attenuator cannot be found from the foils available. In this case, proceed by bisection and interpolate the HVL.

3. Evaluation of the irradiation field (no dose estimation)

  1. Place an EBT3 film on the support used for irradiation.
  2. Irradiate this film to obtain a well-marked irradiation field (at least 2 Gy).
  3. Scan the EBT3 film using a dedicated scanner.
  4. Plot the dose profile using Image J using the Analyze and then Plot Profile option (Figure 2).
  5. Determine the size of irradiation field usage for irradiation (homogeneous area, excluding penumbra regions, see Figure 2).
  6. Make marks on the support used for irradiation to ensure that the cell container is in the right position.
    NOTE: In this step, the size of the irradiation field is determined, and the dose is not estimated. The complete procedure for film reading and analysis is given in section 5. Also, take margins in order to avoid errors due to the cell container positioning.

4. Dose rate measurement with ionization chamber

  1. Take the cell container and break a little part on the side or at the bottom (depending on the particular container and ionization chamber used) to be able to place the ionization chamber inside (Figure 3, upper section) or below (Figure 3, lower section) the container. The examples are given in Figure 3 with different ionization chambers (cylindrical or plane parallel) and cell containers. In this case, a T25 flask was used (Figure 3, red box).
    NOTE: a soldering iron or heated scalpel is a good alternative to make holes in plastic ware
  2. Place the container inside the enclosure on the support used for irradiation (carbon plate here).
  3. Place the ionization chamber in the container (Figure 3, red box), in the correct position and connect it to the electrometer.
  4. Ensure that all irradiation parameters listed in section 1 are correct (high voltage, intensity, additional filtrations, source sample distance, etc.).
  5. Pre-irradiate the ionization chamber for 5 min and perform the zeroing of the electrometer.
  6. Take 10 measurements of 1 min to determine the average dose rate in air kerma (Gy.min-1). Calculate the determination of the dose rate in Kair as follows:
    figure-protocol-6332
    where M is the reading of the dosemeter corrected by temperature, pressure, polarity effect, ion recombination, and electrometer calibration. NKair and Kq are the calibration and correction factors for radiation quality, whose values are specific to each ionization chamber.

5. Measurement of cell culture media attenuation and scattering

NOTE: Handle EBT3 films with gloves throughout the procedure.

  1. Preparation of the experiment
    1. Cut small pieces of EBT3 films at least 24 h before irradiation.
    2. Determine the size of the films as a function of the cell container used for radiobiology experiments (4 x 4 cm for a T25 flask, for example).
      Cut two sets of radiochromic films: One set for the calibration curves composed of three pieces of EBT3 radiochromic film by dose or time point (nine points in total for this work) ; and one set for the quantification of the cell culture media attenuation, also three pieces per point.
    3. Number all the films for identification (upper-right corner here) and scan them on the same position on the scanner.
    4. Keep the films away from light.
    5. Prepare the cell container used for the EBT3 film measurements and, if necessary, cut a part to put the film inside (an example with a T25 is given in Figure 4).
  2. Dose rate estimation
    1. Measure the dose rate for the configuration as described in the previous section.
    2. Keep this configuration in place for the irradiation of the EBT3 radiochromic films and use the same type of cell container.
  3. Construction of the calibration curve
    1. Take the pre-cut EBT3 films for the calibration curve.
    2. Do not irradiate three pieces (0 Gy).
    3. Place the first film inside the cell container, in the same configuration as for cell irradiation.
    4. Irradiate it to obtain the first dose points.
    5. Repeat this operation to obtain three pieces of EBT3 films irradiated with the same dose.
    6. Perform this for each dose point (nine dose points in this work (0, 0.25, 0.5, 0.75, 1, 1.5, 2, 2.5, and 3 Gy) as illustrated in Figure 5).
  4. Evaluation of the attenuation of cell culture media and scattering.
    1. Chose the same irradiation time for all irradiations (60 s, for example).
    2. Irradiate three pieces of EBT3 films in the container without water.
    3. Irradiate three pieces of EBT3 films in the container with water as follows.
      1. Place the film inside the container.
      2. Fill the container with the exact quantity of water to represent the cell culture media (5 mL here). Use small pieces of tape if the films do not remain submerged properly.
      3. Place the cell container inside the enclosure and ensure that the film is correctly immersed.
      4. When the irradiation is complete, take the EBT3 films, dry them with absorbent paper, and store them away from light.

6. Reading of EBT3 radiochromic films

  1. Read EBT3 films at least 24 h after irradiation.
  2. Scan the films on a dedicated scanner.
  3. Set the scanner parameters as: 48 bit red-green-blue tiff format, 150 dpi in transmission mode, and no image correction.
  4. Perform a warm up of the scanner as follows.
    1. Place a non-irradiated film on the scanner.
    2. Launch a preview of the scan.
    3. Launch a timer and wait for 30 s.
    4. Launch the scan.
    5. At the end of the scan, launch a timer, and wait for 90 s.
    6. At the same time, register the scan, open the image with ImageJ, trace a square ROI (always the same size and in the same position), and take a measurement of the average red pixel level of the area.
    7. At the end of the 90 s, repeat the procedure from step 2 (without touching the film inside the scanner).
    8. Repeat this at least 30 times to warm up and stabilize the scanner (no variations in the average red pixel level of the area selected on the non-irradiated films). If the scanner, i.e., the average red pixel value, is not stabilized, continue the procedure.
  5. Scanning of the EBT3 films
    1. Place the first film in the center of the scanner bed. Delimitate an area to always place the film in the same place and in the same orientation.
    2. Launch a preview of the scan.
    3. Launch a timer and wait for 30 s.
    4. Launch the scan.
    5. At the end of the scan, launch a timer, and wait for 90 s. During these 90 s change the EBT3 film.
      NOTE: An analysis of the EBT3 radiochromic films was performed using a self-programmed C++ program. Different methods can be used for the EBT3 film analysis, such as the red channel method or the three channels method14,15. In this case, we have used the red channel method with no background subtraction, and the images were converted to optical densities and then to the dose using our program. As this method is already well defined, our C++ program was not included here. Moreover, dedicated software16 can also be used for EBT3 film analysis.

7. Determination of the dose rate at the level of the cell monolayer

  1. Convert the average dose rate obtained with the ionization chamber corrected by the attenuation and scattering of the cell culture media (K) to the water kerma using the ratio of the mean mass energy absorption coefficient for water to air evaluated over the photon fluence spectrum (μen/ρ).
    figure-protocol-12358
    A dedicated software17 was used to calculate the photon energy spectrum in air with no phantom, and we used the NIST table18 to calculate the mean mass energy absorption coefficient.

Access restricted. Please log in or start a trial to view this content.

Results

In this work, we used a platform dedicated to small animal irradiation19; however, this platform can be used to irradiate other types of samples such as cells. The irradiation source is a Varian X-ray tube (NDI-225-22) having an inherent filtration of 0.8 mm of beryllium, a large focal sport size of 3 mm, a high voltage range of about 30 to 225 kV and a maximal intensity of 30 mA.

The parameters used for this study are reported in Table 1. We have chose...

Access restricted. Please log in or start a trial to view this content.

Discussion

This work presents the protocol used and implemented for cell irradiations using low energy X-ray facility. Nowadays, many radiobiology experiments are performed with this type of irradiator as they are easy to use, cost effective and with very few radioprotection constraints, compared to cobalt source for example. Although these setups have many advantages, as they use a low X-ray energy source, a modification of only one irradiation parameter can significantly impact the dosimetry. Several studies have already highligh...

Access restricted. Please log in or start a trial to view this content.

Disclosures

The authors have nothing to disclose.

Acknowledgements

None

Access restricted. Please log in or start a trial to view this content.

Materials

NameCompanyCatalog NumberComments
31010 ionization chamberPTWionization Radiation, Detectors including code of practice, catalog 2019/2020, page 14https://www.ptwdosimetry.com/fileadmin/user_upload/DETECTORS_Cat_en_16522900_12/blaetterkatalog/index.html?startpage=1#page_14
EBT3 radiochromic filmsMeditestquote requesthttps://www.meditest.fr/produit/ebt3-8x10/
electrometer UNIDOSEweblinePTWonline catalog, quote requesthttps://www.ptwdosimetry.com/en/products/unidos-webline/?type=3451&downloadfile=1593&
cHash=
6096ddc2949f8bafe5d556e931e6c865
HVL material (filter, diaphragm)PTWonline catalog, page 70, quote requestthickness foils: 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5 and 10 mm of copper, https://www.ptwdosimetry.com/fileadmin/user_upload/Online_Catalog/Radiation_Medicine_Cat_en_
58721100_11/blaetterkatalog/index.html#page_70
scanner for radiochromic filmsEpsonquote requestEpson V700, seiko Epson corporation, Suwa, Japan
temperature and pressure measurements, Lufft OPUS20lufftquote requesthttps://www.lufft.com/products/in-room-measurements-291/opus-20-thip-1983/

References

  1. Zoetelief, J., Broerse, J. J., Davies, R. W. Protocol for X-ray dosimetry EULEP. Report No. Report EUR 9507. Commission of the European Communities. , (1985).
  2. Zoetelief, J., et al. Protocol for X-ray dosimetry in radiobiology. International Journal of Radiation Biology. 77 (7), 817-835 (2001).
  3. Zoetelief, J., Jansen, J. T. Calculated energy response correction factors for LiF thermoluminescent dosemeters employed in the seventh EULEP dosimetry intercomparison. Physics in Medicine and Biology. 42 (8), 1491-1504 (1997).
  4. Coleman, C. N., et al. Education and training for radiation scientists: radiation research program and American Society of Therapeutic Radiology and Oncology Workshop, Bethesda, Maryland. Radiation Research. 160 (6), 729-737 (2003).
  5. Desrosiers, M., et al. The importance of dosimetry standardization in radiobiology. Journal of Research of National Institute of Standards and Technology. 118, 403-418 (2013).
  6. DIN. Klinische Dosimetrie: Teil 4. Anwendung von Röntgenstrahlen mit Röhrenspannungen von 10 bis 100 kV in der Strahlentherapie und in der Weichteildianostik. , Report No. DIN 6809 (1988).
  7. DIN. Klinische Dosimetrie: Teil 5. Anwendung von Röntgenstrahlen mit Röhrenspannungen von 100 bis 400 kV in der Strahlentherapie. , Report No. DIN 6809-5 (1996).
  8. NCS. Dosimetry of low and medium energy x-rays: A code of practice for use in radiotherapy and radiobiology. NCS. , Report No. 10 (1997).
  9. International Atomic Energy Agency. Absorbed Dose Determination in External Beam Radiotherapy. International Atomic Energy Agency. , (2000).
  10. Ma, C. M., et al. AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology. Medical Physics. 28 (6), 868-893 (2001).
  11. Peixoto, J. G., Andreo, P. Determination of absorbed dose to water in reference conditions for radiotherapy kilovoltage x-rays between 10 and 300 kV: a comparison of the data in the IAEA, IPEMB, DIN and NCS dosimetry protocols. Physics in Medicine and Biology. 45 (3), 563-575 (2000).
  12. Pedersen, K. H., Kunugi, K. A., Hammer, C. G., Culberson, W. S., DeWerd, L. A. Radiation biology irradiator dose verification survey. Radiation Research. 185 (2), 163-168 (2016).
  13. Draeger, E., et al. A dose of reality: how 20 years of incomplete physics and dosimetry reporting in radiobiology studies may have contributed to the reproducibility crisis. International Journal of Radiation Oncology, Biology, Physics. 106 (2), 243-252 (2020).
  14. Devic, S., et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Medical Physics. 32 (7), 2245-2253 (2005).
  15. Micke, A., Lewis, D. F., Yu, X. Multichannel film dosimetry with nonuniformity correction. Medical Physics. 38 (5), 2523-2534 (2011).
  16. Filmqa Software. GAF Chromic.com. , Available from: http://www.gafchromic.com/filmqa-software/filmqapro/index.asp (2020).
  17. Poludniowski, G., Landry, G., DeBlois, F., Evans, P. M., Verhaegen, F. SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes. Physics in Medicine and Biology. 54 (19), 433-438 (2009).
  18. Hubbell, J. H., Seltzer, S. M. X-Ray mass attenuation coefficients - Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients 1 keV to 20 MeV for elements Z = 1 to 92 and 48 additional substances of dosimetric interest (version 1.4). NIST Standard Reference Database. , 126(1995).
  19. Wong, J., et al. High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities. International Journal of Radiation Oncology, Biology, Physics. 71 (5), 1591-1599 (2008).
  20. Trompier, F., et al. Investigation of the influence of calibration practices on cytogenetic laboratory performance for dose estimation. International Journal of Radiation Biology. , 1-9 (2016).
  21. Dos Santos, M., et al. Importance of dosimetry protocol for cell irradiation on a low X-rays facility and consequences for the biological response. International Journal of Radiation Biology. , 1-29 (2018).
  22. Noblet, C., et al. Underestimation of dose delivery in preclinical irradiation due to scattering conditions. Physica Medica. 30 (1), 63-68 (2014).
  23. Paixao, L., et al. Monte Carlo derivation of filtered tungsten anode X-ray spectra for dose computation in digital mammography. Radiologia Brasileira. 48 (6), 363-367 (2015).

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Explore More Articles

DosimetryCell IrradiationOrthovoltage X rayRadiobiological StudiesDose MeasurementIrradiation ParametersIonization ChamberAverage Dose RateCalibration CurveSelf developing FilmDose ProfileImageJElectromagnetic MeasurementsX ray Facilities

This article has been published

Video Coming Soon

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved