An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.

The atomizer used in AAS can be either a flame atomizer or an electrothermal atomizer. The gas-phase atoms generated by the atomizer often get thermally excited and emit radiation, causing an interference with the radiation source signal. A circular metal disk, called a chopper, is used to eliminate such interference between the radiation source and the flame atomizer. The metal chopper isolates the radiation source by creating an alternating signal, unlike the continuous signal from flame.

In a single-beam spectrophotometer, the radiation passes directly through the atomized sample, while in a double-beam instrument, the source beam is split, with one beam passing through the atomized sample and the other being the reference beam. The two beams are then recombined using a half-silvered mirror. The combined beam then goes through the monochromator, which allows only a narrow wavelength range to reach the detector. The detector is typically a photomultiplier tube that processes and converts the radiation into amplified electrical signals.

From Chapter 14:

article

Now Playing

14.4 : Atomic Absorption Spectroscopy: Instrumentation

Atomic Spectroscopy

108 Views

article

14.1 : Atomic Spectroscopy: Absorption, Emission, and Fluorescence

Atomic Spectroscopy

203 Views

article

14.2 : Atomic Spectroscopy: Effects of Temperature

Atomic Spectroscopy

79 Views

article

14.3 : Atomic Absorption Spectroscopy: Overview

Atomic Spectroscopy

148 Views

article

14.5 : Atomic Absorption Spectroscopy: Radiation and Light Sources

Atomic Spectroscopy

99 Views

article

14.6 : Atomic Absorption Spectroscopy: Atomization Methods

Atomic Spectroscopy

76 Views

article

14.7 : Atomic Absorption Spectroscopy: Interference

Atomic Spectroscopy

129 Views

article

14.8 : Atomic Absorption Spectroscopy: Lab

Atomic Spectroscopy

59 Views

article

14.9 : Atomic Emission Spectroscopy: Overview

Atomic Spectroscopy

164 Views

article

14.10 : Atomic Emission Spectroscopy: Instrumentation

Atomic Spectroscopy

63 Views

article

14.11 : Atomic Emission Spectroscopy: Interference

Atomic Spectroscopy

26 Views

article

14.12 : Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle

Atomic Spectroscopy

204 Views

article

14.13 : Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation

Atomic Spectroscopy

60 Views

article

14.14 : Atomic Emission Spectroscopy: Lab

Atomic Spectroscopy

43 Views

article

14.15 : Atomic Fluorescence Spectroscopy

Atomic Spectroscopy

37 Views

See More

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved