A subscription to JoVE is required to view this content. Sign in or start your free trial.
Method Article
A novel microfluidic system has been developed using the phenomenon of passive pumping and a user controlled fluid delivery system. This microfluidic system has the potential to be used in a wide variety of biological applications given its low cost, ease of use, volumetric precision, high speed, repeatability and automation.
In this report we demonstrate a fluid delivery method that uses small droplet surface tension to pump a desired volume through a microfluidic channel in order to achieve a number of different fluid phenomena. For example, the user may wish to flow a single fluid as fast as possible, or deliver multiple fluids in rapid succession to create specific fluidic patterns. In order to do this, the user must first have an application built around a microfluidic device. The microflluidic device does not need to be bonded, but should be made from a hydrophilic material. Therfore, the method can be utilized with almost any microfluidic device, with performance largely dictated by the geometrical constraints of the microfluidic channel. To help navigate the geometrical constraints of this method, an introduction to the relevant numerical analysis is presented first.
Example: System activated for a minute (60 seconds). The frequency was 15 Hz (15 pulses in one second). The per-pulse open time was 20 milliseconds (ms).
(20ms)(15Hz)(60s)=18000ms.
This means that out of the 60,000 ms in one minute, the valve was actually open for 18,000ms.
Let’s assume the volume of fluid delivered weighed 5 grams. Then,
5 grams / 18000 ms = 2.78e-4 grams/ms.
In the case of water, with its density being one gram per milliliter (mL),
2.78e-4 grams/ms = 2.78e-4 mL/ms.
After calibration, the volume of a drop is dependent on the open time. For example, with an open time of 20ms, and all the parameters remaining the same as in the previous example,
(2.78e-4 mL/ms)(20 ms) = 5.56e-3 mL = 5.56 μL.
To find the open time y needed to make a drop of x microliter (μL) volume,
(x μL) / [(2.78e-4 mL/ms)(1000 μL/mL)] = y ms
8) Aim one or more nozzles to the inlet of the PDMS device (figure 3). Having calibrated the system, calculate the volume coming out of each valve, based on microfluidic device dimensions. For high speed passive pumping (to obtain maximum flow rate), calculate the inlet drop volume necessary to create an inlet drop which possesses a 90deg contact angle with inlet surface [2]. For packet creation, calculate valve frequency and open times and the valve timings necessary to activate two valves in sequence. As seen in Figure 3, the two nozzles can be pointed at the inlet. This can extended to multiple nozzles, all aimed at the channel inlet.
Representative Results:
When properly calibrated, with valve open times correctly calculated and the nozzles properly aimed at the inlet, the user should be able to see flow passively pumped (figure 4). A burst of liquid should come out of the valve and reach the inlet. As liquid reaches the inlet, there is an instant collapse of the inlet drop into the channel, towards the outlet. Liquid within the channel moves only during the collapse of an inlet drop. Complete fluid movement within the channel stops at the end of the drop collapse, providing for instantaneous fluid stop and well defined fluidic boundaries (in the case that the user is flowing multiple liquids). The duration of drop collapse depends on the inlet port radius and the volume of the inlet drop [1]. In our experimental setup and design, inlet drop collapse occurs in a matter of a few milliseconds.
Figure 1. PDMS microfluidic device with one inlet, left, and one outlet, right. Please click here to see a larger version of figure 1.
Figure 2. Reservoir system and valve setup. Please click here to see a larger version of figure 2.
Figure 3. Two valves, both aimed at a single inlet of a microfluidic device. Please click here to see a larger version of figure 3.
Figure 4. Time-step sequence (33 millisecond) of inlet drop collapse following fluid ejection from a valve. Please click here to see a larger version of figure 4.
Funding was provided by the Wisconsin Institute of Discovery.
Name | Company | Catalog Number | Comments |
Sylgard 184 Silicone elastometer base | Dow Corning | MSDS No.: 01064291 | |
Sylgard 184 Silicone elastometer curing agent | Dow Corning | MSDS No.: 01064291 | |
VHS Microdispensing Starting kit | The Lee Company | IKTX0322000A | |
Miniature Holders | Bioscience Tools | MH-2 | |
LabVIEW | National Instruments | Control System | |
1.14mm I.D. tubing | Scientific Commodities Inc. | BB31695-PE/7 | |
1.57mm I.D. tubing | Scientific Commodities Inc. | BB31695-PE/10 | |
20 mL BD™ Luer-Lok Tip Syringe, non-sterile | BD Biosciences | 301032 | |
Request permission to reuse the text or figures of this JoVE article
Request PermissionThis article has been published
Video Coming Soon
Copyright © 2025 MyJoVE Corporation. All rights reserved