Source: Laboratory of Dr. B. Jill Venton - University of Virginia
The goal of many chemical analyses is a quantitative analysis, where the amount of a substance in a sample is determined. In order to accurately calculate the concentration of an unknown from a sample, careful sample preparation is key. Every time a sample is handled or transferred, some of the sample can be lost. There are strategies however, for minimizing sample loss. There are also strategies for coping with sample loss and still making accurate measurements of concentration.
To minimize sample loss, the ideal is to minimize the number of sample handling and transfer steps. For example, massing a solid sample directly into a flask that a solution will be made in reduces a transfer step. If it's necessary to transfer from one flask to another and a dilution is being made, then triple rinsing the glassware helps ensure all the sample is transferred. Other strategies are more specific to the sample. For example, samples that adsorb to glass, such as proteins, might better be handled in polypropylene disposable tubes. The tubes are not hydrophilic, so if a small amount of sample is to be pipetted in water, it is best to have already added the water to the tube, so the sample can be pipetted directly into the solvent. It may be better to concentrate, rather than completely dry a sample, due to losses from insolubilities after rehydration.
Another source of sample loss is through incomplete sample manipulations. For example, if a derivatization procedure is used and the derivatization is incomplete, then the full amount of sample is not observed. Errors such as this are systematic errors, and can be solved by correcting the problem, such as changing the derivatization procedure. Another cause of systematic error in measurements is matrix effects. These can interfere with measurement of certain substances and performing calibrations in the same matrix as the sample can reduce this effect.
Quantitative analysis is typically carried out using either external or internal standards. For external standards, a calibration curve is made by measuring different known concentrations of the analyte of interest. Then, the sample is run separately from the standard. For internal standards, the standard is in the same sample as the analyte of interest, allowing the measurement to be taken simultaneously. Typically, a different species is added called the internal standard and the ratio of the response for that internal standard and the analyte is calculated. The idea is that the ratio of the response, called the response factor, is proportional to their concentration. While the method must be able to distinguish between the analyte of interest and the internal standard, any sample losses that occur after the internal standard is added should be similar for both substances and thus the ratio of the response stays the same. A special case of using internal standards is the method of standard additions, where increasing amounts of the analyte is added to the solution and the original amount of analyte is back-calculated. Internal standards can be used in chromatography, electrochemistry, and spectroscopy.
An internal standard is a substance added in a constant amount to a sample, blank, and standard in an analysis. An internal standard can compensate for both systematic and random errors. For example, if there are instrument fluctuations that cause random errors in the measurement, these fluctuations are expected to be the same for both the internal standard and the analyte and thus the ratio of signals does not change. For systematic errors, such as matrix effects of the solvent, as long as the effect is equal for both the standard and the analyte, the ratio is again unaffected.
The disadvantage of internal standards is that it is hard to find a suitable internal standard. The internal standard must have a signal that is similar to the analyte, but different enough that it can be distinguished by the instrument. Also, the internal standard should not be present in the sample matrix, so that the added standard is the only source of the standard. Occasionally, a major constituent of a sample that is constant in concentration can be used as the internal standard instead of an added standard, but the concentration must be well known for that constituent. The internal standard should also not suppress or enhance the signal of the analyte.
In chromatography, one of the largest sources of error is often the injection. If manual injections are used, there can be errors in loading the syringe properly. Volumes are typically small (~1 µL) so there are uncertainties in reproducibly injecting this small a volume, often a couple of percent relative standard deviation (RSD). In gas chromatography (GC), the sample is injected into a heated port and evaporation from the needle tip can result in variations in volume injected. Auto-samplers help with both the error in loading the syringe and the error in injecting quickly to avoid evaporation in GC, but the error can still be 1–2% RSD. With chromatography, peak area is generally used instead of peak height, as the peaks get wider and shorter with time but the peak area is constant. Thus, for internal standards, ratios of peak areas are used in chromatography instead of peak heights.
For a calibration with an internal standard, a response factor is calculated. The response factor (R) is the ratio of the peaks compared to the ratio of the concentrations where X is the analyte and IS is the internal standard.
For chromatography area (A) is used. The response factor can be calculated from a calibration plot of AX/AIS vs CX/CIS, where the response factor is the slope and the y-intercept is assumed to be 0. Once the response factor is known for standards, then the response of the unknown can be calculated from the measured area ratio from the experiment.
1. Proper Sample Handling: Making a Solution
2. Preparation of an Internal Standard Calibration Curve
3. Preparation of a Real Sample with Internal Standard for Gas Chromatography
4. Run the Sample and Calculate the Concentration
5. Results: GC Analysis of Caffeine with Internal Standard
Figure 1. Calibration plot using an internal standard. A plot of the area ratios vs concentration ratios for 3 standard samples of caffeine (1, 0.5, and 0.2 mg/mL) with 0.33 mg/mL adenine internal standard added to each. The slope of the line is 1.8, which is the response factor.
Figure 2. Chromatogram of coffee with adenine internal standard. A plot of the response of the FID detector to samples. The three main peaks are adenine (IS), caffeine, and palmitic acid.
Internal standards are used in many fields, including spectroscopy and chromatography. In spectroscopy, internal standards can help correct for random errors due to changes in light source intensity. If a lamp or other light source has variable power, it will affect the absorption and consequently, emission of a sample. However, the ratio of an internal standard to analyte will stay constant, even if the light source does not. One example of this is using lithium (Li) as an internal standard for the analysis of sodium in a blood sample by flame spectroscopy. Li is chemically similar to sodium but is not natively found in blood.
For chromatography, internal standards are often used in both gas chromatography and liquid chromatography. For applications with mass spectrometry as the detector, the internal standard can be an isotopically-labeled analyte, so that the molecular weight (MW) will be different than the analyte of interest. Internal standards are commonly used in pharmaceutical or environmental analyses.
Перейти к...
Видео из этой коллекции:
Now Playing
Analytical Chemistry
204.6K Просмотры
Analytical Chemistry
84.6K Просмотры
Analytical Chemistry
319.8K Просмотры
Analytical Chemistry
796.0K Просмотры
Analytical Chemistry
622.9K Просмотры
Analytical Chemistry
51.2K Просмотры
Analytical Chemistry
25.4K Просмотры
Analytical Chemistry
281.7K Просмотры
Analytical Chemistry
384.0K Просмотры
Analytical Chemistry
264.3K Просмотры
Analytical Chemistry
93.7K Просмотры
Analytical Chemistry
112.2K Просмотры
Analytical Chemistry
87.1K Просмотры
Analytical Chemistry
51.3K Просмотры
Analytical Chemistry
124.9K Просмотры
Авторские права © 2025 MyJoVE Corporation. Все права защищены