Here, we present a protocol for evaluating the surface characteristics of endodontic retreatment files after repeated use in retreatment procedures, utilizing scanning electron microscopy to identify and analyze potential surface defects.
This study aimed to evaluate surface defects of Remover rotary Nickel-Titanium (NiTi) files after single and multiple uses in conventional endodontic retreatment procedures using scanning electron microscopy (SEM). Eighty acrylic blocks, simulating root canals with a 1.5 mm internal diameter, a 5 mm radius of curvature, and a 55° curvature, were utilized. After chemomechanical preparation and obturation, 24 new Remover files (N30, 7%, L23) were randomly assigned to three groups: single use, triple use, and six uses. The files were operated at 600 rpm with a torque of 2.5 Ncm, cleaned, and sterilized after each use.
SEM analysis at magnifications of 100x, 250x, and 500x revealed surface defects, including tip deformation, microcracks, fracture, unwinding, surface pitting, and blade disruption. Deformation was observed in 75% of the files after a single use and in 100% of the files after three and six uses. Microcracks were absent after single use but appeared in 25% and 87.5% of files after three and six uses, respectively, showing a statistically significant increase (p < 0.001). Surface pitting also significantly increased among groups (p = 0.004).
No fractures were observed in any group. The most common defects were tip deformation (91.7%) and surface pitting (70.8%). The findings suggest that repeated use of NiTi files significantly increases surface defects, elevating the risk of fatigue fractures. Thus, the results recommend limiting the reuse of Remover files to a maximum of 3x. Further research is needed to correlate defect types with anatomical factors and to assess file effectiveness in retreatment scenarios.
Endodontic retreatment is a procedure performed when a previously treated tooth fails to heal or develops new pathologies, such as persistent infection, reinfection, or missed anatomy. The procedure involves the removal of the existing root canal filling material, thorough cleaning and disinfection of the canal system and subsequent refilling1,2.
Nickel-titanium (NiTi) instruments are of great importance in improving and facilitating endodontic procedures due to their flexibility and high cutting efficiency3,4. The superelas....
1. Sample procurement
2. Cleaning and shaping procedure
Deformation was observed in 75% of files after single use and in 100% of files after three and six uses, but the differences between the groups were not statistically significant (Table 1). The evaluation of deformation types among groups is shown in Table 2. When the types of deformation were evaluated separately, no microcracks were observed after a single use, while microcracks were observed in 25% of the files after three uses and in 87.5% of the files after six uses; this difference.......
This study evaluated the presence and types of microscopic defects on the external surfaces of Remover files after single, triple, and six-times use in acrylic blocks simulating curved canals. Ideally, human teeth are recommended for use in studies evaluating the fracture resistance of files to better simulate clinical use28. In their study, Peters and Barbakow29 found an increase in fracture initiation and propagation rates in instruments used in blocks compared to extract.......
We would like to express our sincere gratitude to Bogazici University for providing the laboratory facilities and technical support necessary for this research. We also thank Dr. Demet Sezgin Mansuroglu, Dr. Eda Karadogan, and Dr. Mustafa Enes Ozden for their valuable assistance in data collection and analysis. The research was funded by the authors. No external financial support was obtained.
....Name | Company | Catalog Number | Comments |
Acrylic block | ArdaDent Medical,Ankara,Turkey | for obturation | |
DiaRoot Biosealer | DiaDent, South Korea | BS23101161 | for obturation |
DualMove Endomotor | MicroMega, Coltene, France | 52002023 | for preparation |
EndoArt Smart Gold | EndoArt, Inci Dental, Turkey | SGK10114 | for initial preparation |
Gutta Percha | EndoArt, Inci Dental, Turkey | GD23080701 | for obturation |
Quattro ESEM | Thermo Fisher Scientific, USA | SEM analysis | |
Paper Points | Dentsply Maillefer, Ballaigues, Switzerland | 1I0305 | for dry to root canal |
Remover File | MicroMega, Besançon, France | 891144/873757/ | for retreatment procedure |
Sodium Hypochlorite | Saba Chemical&Medical, Turkey | 3010225 | for irrigation |
SPSS v29 | IBM SPSS Corp, Armonk, New York, USA | Statistical analysis |