Scanning Electron Microscopy (SEM)

Overview

Source: Laboratory of Dr. Andrew J. Steckl — University of Cincinnati

A scanning electron microscope, or SEM, is a powerful microscope that uses electrons to form an image. It allows for imaging of conductive samples at magnifications that cannot be achieved using traditional microscopes. Modern light microscopes can achieve a magnification of ~1,000X, while typical SEM can reach magnifications of more than 30,000X. Because the SEM doesn’t use light to create images, the resulting pictures it forms are in black and white. 

Conductive samples are loaded onto the SEM’s sample stage. Once the sample chamber reaches vacuum, the user will proceed to align the electron gun in the system to the proper location. The electron gun shoots out a beam of high-energy electrons, which travel through a combination of lenses and apertures and eventually hit the sample. As the electron gun continues to shoot electrons at a precise position on the sample, secondary electrons will bounce off of the sample. These secondary electrons are identified by the detector. The signal found from the secondary electrons is amplified and sent to the monitor, creating a 3D image. This video will demonstrate SEM sample preparation, operation, and imaging capabilities.

Procedure

1. Preparation of the Sample

  1. Place sample onto sample stub. If necessary, carbon tape may be used to adhesively bond the sample to the stub.
  2. Place the sample into a gold sputtering system. Using a mini-gold sputter, sputter gold for 30 s at ~ 70 mTorr pressure. A different gold layer thickness may be necessary depending on the geometry of the sample. More rough or porous surface requires a longer sputtering time.
  3. Remove stub from gold sputtering system.

.css-f1q1l5{display:-webkit-box;display:-webkit-flex;display:-ms-flexbox;display:flex;-webkit-align-items:flex-end;-webkit-box-align:flex-end;-ms-flex-align:flex-end;align-items:flex-end;background-image:linear-gradient(180deg, rgba(255, 255, 255, 0) 0%, rgba(255, 255, 255, 0.8) 40%, rgba(255, 255, 255, 1) 100%);width:100%;height:100%;position:absolute;bottom:0px;left:0px;font-size:var(--chakra-fontSizes-lg);color:#676B82;}

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Results

The SEM, seen in Figure 2a, has been used for making measurements and acquiring sample photos. The sample consisted of sodium chloride (NaCl) salt. It was placed onto the stub as seen in Figure 2b, then a few nanometers of gold was sputtered onto it to make it conductive. The conductive sample was then placed into the SEM sample area as seen in Figure 2c.

SEM images were obtained at 50X, 200X, 500X, 1,000X, and 5,000X magnification levels as s

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Application and Summary

The SEM is a very powerful tool that is common in most research institutions because of its ability to image any object that is conductive, or has been treated with a conductive coating. The SEM has been used to image objects such as semiconductor devices,2 biological membranes,3 and insects,4 among others. We have also used the SEM to analyze nanofibers and paper-based materials, biomaterials, micropatterned structures. Of course, there are materials, such as liquids, that can’t be

Log in or to access full content. Learn more about your institution’s access to JoVE content here

References
  1. Goldstein, J., Newbury, D., Joy, D., Lyman, C., Echlin, P., Lifshin, E., Sawyer, L., Michael, J. Scanning Electron Microscopy and X-ray Microanalysis. 3rd Ed. Springer, New York, NY. (2003).
  2. Purandare, S., Gomez, E.F., Steckl, A.J. High brightness phosphorescent organic light emitting diodes on transparent and flexible cellulose films. Nanotechnology. 25, 094012 (2014).
  3. Masuda, Y., Yamanaka, N., Ishikawa, A., Kataoka, M., Aral, T., Wakamatsu, K., Kuwahara, N., Nagahama, K., Ichikawa, K., Shimizu, A. Glomerular basement membrane injuries in IgA nephropathy evaluated by double immunostaining for a5(IV) and a2(IV) chains of type IV collagen and low-vacuum scanning electron microscopy. Clinical and Experimental Nephrology. 1-9. (2014).
  4. Kang, J.H., Lee, Y.J., Oh, B.K., Lee, S.K. Hyun, B.R. Lee, B.W, Choi, Y.G., Nam, K.S., Lim, J.D. Microstructure of the water spider (Argyroneta aquatic) using the scanning electron microscope Journal of Asia-Pacific Biodiversity. 7 484-488 (2014).
Tags
Scanning Electron Microscopy SEMSEM TechniqueElectron BeamSurface Structure AnalysisChemical Composition AnalysisLight Microscopy LimitationsDiffractionLateral ResolutionMagnificationSub nanometer ResolutionDepth Of FieldSEM In Chemistry And Material AnalysisNanoscale AnalysisSEM Sample PreparationSEM OperationConductive Samples

跳至...

0:00

Overview

1:49

Principles of Scanning Electron Microscopy

4:10

Sample Preparation and Loading

5:54

SEM Operation

8:22

Image Analysis

9:17

Applications

11:13

Summary

此集合中的视频:

article

Now Playing

Scanning Electron Microscopy (SEM)

Analytical Chemistry

86.6K Views

article

分析制备的样品制备

Analytical Chemistry

83.7K Views

article

内部标准

Analytical Chemistry

203.4K Views

article

标准加入的方法

Analytical Chemistry

318.8K Views

article

校准曲线

Analytical Chemistry

789.1K Views

article

(紫外-可见) 的紫外-可见光谱法

Analytical Chemistry

616.5K Views

article

拉曼光谱化学分析

Analytical Chemistry

50.8K Views

article

X 射线荧光光谱 (XRF)

Analytical Chemistry

25.3K Views

article

气相色谱 (GC) 与火焰电离检测

Analytical Chemistry

279.9K Views

article

高性能液相色谱法 (HPLC)

Analytical Chemistry

381.8K Views

article

离子交换色谱法

Analytical Chemistry

263.0K Views

article

毛细管电泳 (CE)

Analytical Chemistry

93.0K Views

article

质谱分析法导论

Analytical Chemistry

111.4K Views

article

负载型催化剂,用恒电位仪/结合电化学测量

Analytical Chemistry

51.2K Views

article

循环伏安法 (CV)

Analytical Chemistry

123.4K Views

JoVE Logo

政策

使用条款

隐私

科研

教育

关于 JoVE

版权所属 © 2025 MyJoVE 公司版权所有,本公司不涉及任何医疗业务和医疗服务。