로그인

JoVE 비디오를 활용하시려면 도서관을 통한 기관 구독이 필요합니다. 전체 비디오를 보시려면 로그인하거나 무료 트라이얼을 시작하세요.

기사 소개

  • 요약
  • 초록
  • 서문
  • 프로토콜
  • 결과
  • 토론
  • 공개
  • 감사의 말
  • 자료
  • 참고문헌
  • 재인쇄 및 허가

요약

Here, we present a protocol to allow providers to perform focused cardiac ultrasound (FoCUS) in the clinical environment. We describe methods of transducer manipulation, review common pitfalls of transducer movements, and suggest tips to optimize phased array transducer use.

초록

Focused cardiac ultrasound (FoCUS) is a limited, clinician-performed application of echocardiography to add real-time information to patient care. These bedside exams are problem oriented, rapidly and repeatedly performed, and largely qualitative in nature. Competency in FoCUS includes mastery of the stereotactic and psychomotor skills required for transducer manipulation and image acquisition. Competency also requires the ability to optimize the setup, troubleshoot image acquisition, and understand the sonographic limitations because of complex clinical environments and patient pathology. This article presents concepts for successful, high-quality two-dimensional (B-mode) image acquisition in FoCUS.

Concepts of high-quality image acquisition can be applied to all established sonographic windows of the FoCUS exam: the parasternal long-axis (PLAX), parasternal short-axis (PSAX), apical four chamber (A4C), subcostal fourchamber (SC4C), and the inferior vena cava (IVC). The apical five-chamber (A5C) and subcostal short-axis (SCSA) views are mentioned, but are not discussed in-depth. A pragmatic figure illustrating the movements of the phased array transducer is also provided to serve as a cognitive aid during FoCUS image acquisition.

서문

Focused cardiac ultrasound (FoCUS) is a limited, clinician-performed application of echocardiography that provides immediate anatomic, physiologic, and functional information to patient care. These exams, consisting of five classic views, are problem oriented, performed in real time at the bedside, and do not replace comprehensive echocardiography exams1,2. Given the focused nature of these exams, they are often repeatedly performed when clinical status changes or serial monitoring is required. It is important to have standardized training and obtain adequate images of all five views, when possible, as some views may provide limited information depending on the individual patient and pathology.

The use of FoCUS is rapidly expanding. Many clinical landscapes, such as perioperative anesthesiology, critical care and emergency medicine1,2,3, now routinely employ FoCUS. Inpatient medical wards and outpatient clinical care settings are also adopting this tool to enhance clinical practice4,5,6. As a result, several societal bodies, such as the American Society of Echocardiography, the Society of Critical Care Medicine, and the American College of Emergency Physicians, have published guidelines and recommendations for FoCUS competency and scope of practice7,8,9. While these guidelines and recommendations are not codified, much of the content is consistent and influences FoCUS training curricula10.

Beyond didactics and image interpretation, competency in FoCUS includes mastery of stereotactic and psychomotor skill sets. Stereotactic skill refers to the accurate positioning of ultrasound transducers on the body, based on three-dimensional anatomic features. Psychomotor skill describes the relationship between cognitive function and physical movement that influences coordination, dexterity, and manipulation. Expanding knowledge and awareness about these skills supports FoCUS trainee development.

This article presents concepts for high-quality image acquisition in FoCUS, with both pragmatic considerations and attention to stereotactic and psychomotor skill sets. Specifically, it discusses optimal patient positioning, transducer manipulation, and suggests tips to optimize phased array transducer use. Finally, it examines image optimization for 2-dimensional (B-mode or 2-D mode) and motion modes (M-mode).

프로토콜

This material is the authors' original work, which has not been previously published elsewhere. The protocol described is for clinical use and not research purposes. De-identified images were obtained from a volunteer model in a non-clinical environment. The authors did not seek a formal "Not Regulated" determination from the IRB in accordance with institutional policy, as the activity falls outside of the Common Rule and FDA definitions of human subject research.

1. The transducer

  1. Use the phased array transducer. This is a 4-12 MHz transducer that penetrates deep into the thoracic space, due to its low frequency compared to other ultrasound transducers.
    1. Select the phased array transducer using the Exam button on the machine and selecting the cardiac exam or equivalent available exam.
  2. Practice with transducers from various vendors to gain experience and refine one's cognitive skill set for transducer manipulation.
  3. Practice transducer manipulation with both hands to develop ambidexterity, which may be required when working in limited clinical environments.
  4. Anchor the dominant hand while holding the ultrasound transducer on the patient, with the fat pad of the medial aspect of the hand.
    ​NOTE: This provides additional stability and reduces error caused by large movements. Holding the transducer by the base, without anchoring, results in unintended movements, preventing the ability to maintain the image axis and orientation.
  5. Use two hands for transducer manipulation to improve the ultra-fine adjustments often required for optimal views. Place the dominant hand on the base of the transducer with the non-dominant hand on the tail of the transducer, providing additional stability and guided movement.

2. Patient positioning

  1. Obtain the PLAX, PSAX, A4C, SC4C, and IVC views in the supine position.
  2. Instruct the patient to extend their left arm above their head and lie on their left side, and obtain images in this position if they are unable to be obtained in the supine position.
    NOTE: This results in expansion of the intercostal spaces for larger imaging windows.
    1. Place a wedge or blanket roll behind the patient's right upper torso if the patient cannot easily rotate to 45°, or reposition their limbs. Postoperative and intensive care unit patients often need support to maintain appropriate positioning for FoCUS examination.
  3. Drape the breast as desired by the patient, and make sure that the patient is aware of where the transducer will be placed before beginning the exam.
  4. Manipulate the breast tissue to allow for optimal image acquisition.
    1. If able, instruct the patient to assist moving the breast tissue with their right hand.
  5. Discuss the appropriateness of removal or relocation of monitors beforehand with bedside nurses and other relevant providers.
  6. Do not alter or remove tubes, lines, or drains for imaging purposes. Discuss the duration of the devices with the relevant providers and consider reimaging the patient following their removal.

3. Transducer manipulation

  1. Appreciate and understand the definitions of transducer movements to allow for optimal image acquisition, and provide consistent terminology for communication between providers, especially during teaching (Figure 1 and Table 1).

4. 2-D image optimization

  1. Adjust the depth (approximately 12-16 cm, depending on view) to see the structure of interest.
  2. Adjust the gain to optimize the brightness of the image.
  3. Adjust the focus to the depth of the structure of interest in order to improve resolution.

5. Motion mode (M-mode)

  1. Use M-mode to display a single scan line (wherever the cursor is placed) of the B-mode image (Y-axis) against time (X-axis).
    NOTE: This mode can help operators understand the dynamic relationship of different structures over time and is useful in many different assessments, including IVC size and variability and E-point septal separation (EPSS).
  2. Use the button with the letter "M" on it to turn on M-mode.
    ​NOTE: M-mode will be an on/off button unique to each machine.

6. Parasternal long axis (PLAX)

NOTE: The PLAX refers to obtaining an image that is along the long axis of the heart (Figure 2).

  1. Place the patient in the supine position. If there is difficulty obtaining the PLAX image, place the patient on their left side and extend their arm above their head, if possible.
  2. Place the transducer at an oblique angle between the third and fifth intercostal space of the left parasternal region, with the transducer marker pointing to the patient's right shoulder.
    1. Visualize the right ventricle, left ventricle, left atrium, mitral valve, left ventricular outflow tract, aortic valve, and descending thoracic aorta in this image.
    2. Visualize the mitral valve and aortic valve opening and closing together, to ensure that the image is not foreshortened. Foreshortening is where the ultrasound plane does not cut through the true apex of the structure, changing the perceived image.
  3. 2-D image optimization
    1. Start with an initial depth of approximately 15-20 cm. Adjust the depth so that the tip of the mitral valve is in the center of the image, and the descending thoracic aorta (deep to the left atrium) is visible.
    2. Adjust the gain to maximize the visibility of the myocardium and mitral valve.
    3. Move the focus to the region of interest, most focused at the depth of the mitral valve.
    4. Use M-mode for EPSS or fractional shortening.

7. Parasternal short axis (PSAX; Figure 3)

  1. Place the patient in the same positioning for the PSAX as used for the PLAX.
  2. Place the transducer approximately 90° relative to the transducer in the PLAX.
    1. Obtain an optimal PLAX and rotate the transducer slowly in a clockwise fashion without lifting the transducer from the patient's chest, until the transducer is obliquely angled across the third to fifth intercostal space of the parasternal region, with the transducer marker pointing to the patient's left shoulder.
      NOTE: Over-rotation beyond 90° can lead to interventricular septal flattening, and falsely appear as right ventricular volume or pressure overload.
    2. Tilt the transducer until the mid-papillary muscles are visualized for the FoCUS assessment.
      NOTE: The papillary muscles should move in synchrony with the left ventricle wall. If the papillary muscles appear to be bouncing or fluttering independent of the left ventricular wall, it may signify that the image is capturing the mitral valve leaflet as a result of being off axis.
    3. Tilt the transducer toward the base of the heart, to visualize the bi-leaflet mitral valve initially followed by the tri-leaflet aortic valve.
  3. 2-D image optimization
    1. Start with a deeper image (approximately 16 cm) to identify any pleural effusion.
    2. Adjust the depth to include the full depth of the left ventricle and a few centimeters beyond, to ensure that a pericardial effusion would be fully visualized.
    3. Adjust the gain to maximize visualization of the septum and papillary muscles.
    4. Adjust the focus to the papillary muscles.

8. Apical four chamber view (A4C; Figure 4)

NOTE: Images in patients with chronic obstructive pulmonary disease (COPD), and otherwise inflamed thoracic cavities, are obtained more medially, and images in patients with left ventricular hypertrophy (LVH) or heart failure with reduced ejection fraction (HFrEF) have their view more lateral.

  1. Position the patient with their left arm extended above their head and lying on their left side. If there is significant artifact present, have the patient exhale and hold their breath to minimize pulmonary artifact.
  2. Position the transducer in the fourth through sixth intercostal space along the left anterior axillary line (inferolateral to the left pectoral muscle), with the transducer marker pointed toward the left axilla. Move the transducer lateral, medial, or caudal, as needed to obtain an optimal A4C view.
    1. Lift the breast tissue and push superior along the inframammary fold as needed.
    2. If the left ventricular apex is not fully visualized, move the transducer lateral while orienting the transducer toward the right shoulder.
    3. Position the marker on the phased array transducer between the two and three o'clock positions. In the normal heart, the apex of the left ventricle is at the top and center of the sector, the right ventricle is triangular and smaller, and the myocardium should be uniform from the apex to the atrioventricular valves. If this is not the case, the image may be foreshortened, and should be optimized and acquired from a lower intercostal space.
    4. Tilt the transducer cephalad approximately 60°, so that the A4C view can capture an A4C cardiac image that includes both atria, ventricles, the interventricular septum, and the lateral portions of the tricuspid and mitral annuli. The aortic valve and left ventricular outflow tract should not be present in an A4C view, and are only present in an apical five chamber view.
    5. Visualize the mitral valve, the tricuspid valves, and the interventricular septum on the A4C image. If both valves and the interventricular septum are not visualized, the image should be further optimized.
    6. Slide the transducer up or down a rib space and tilt the base of the transducer down (cranially) to improve the image of the valves. If the base of the transducer is tilted too far down (cranially), an apical five chamber view, including the aortic valve, will appear, and the transducer should be tilted back up (caudally) to optimize the A4C view. If the base of the transducer is tilted too far up (caudally), the coronary sinus will appear, and the transducer should be tilted back down (cranially).
    7. Rotate the base of the transducer toward the patient's midline to optimize the interventricular septum position, which should be present vertical in the center of the image. Minimal rotation should be required. If over-rotated, a two chamber view will be observed.
  3. 2-D image optimization
    1. Increase the depth to include both atria at the image's deepest point, in addition to accommodating the left and right ventricular free walls (approximately an initial depth of 20 cm).
    2. Adjust the gain to maximize visibility, often resulting in increased echogenicity, of the myocardium, the mitral valvular annulus, and the tricuspid valvular annulus.
    3. Adjust the focus to the depth of the valvular annuli (the tricuspid annulus is most commonly used). This depth will also be appropriate when transitioning to an apical five view, if desired.

9. Subcostal four chamber view (SC4C; Figure 5)

  1. Position the patient supine for the subcostal four chamber view. Flex the patient's knees, and support them to maintain the flexed position, to reduce the abdominal muscle tone for easier transducer compression.
  2. Place the transducer nearly flat on the patient's subxiphoid abdomen, with the operator hand on top of the transducer providing cephalad pressure. Find the liver, and then drop the transducer to a shallower angle (often less than 30°) against the subxiphoid portion of the patient's abdomen in a cephalad fashion, with the transducer marker to the patient's left side, approximately pointing to three o'clock. Include the liver in the image so that it can be utilized as an acoustic window to obtain this image.
    1. Hold the lateral aspects of the transducer with the fingers, not underneath the transducer, to flatten the transducer appropriately. Use the index finger to provide downward pressure.
  3. 2-D image optimization
    1. Start with an initial depth of 18-24 cm. Adjust the depth to include the liver as a sonographic window for the ultrasound. The optimal depth varies from patient to patient based on patient body habitus and liver size.
    2. Decrease the gain, as the liver often provides a good medium to transmit sound waves.
    3. Increase the focal points to focus on the cardiac structures of interest below the liver.
    4. Instruct non-intubated patients to perform an inspiratory hold, as this often increases the quality of the image.

10. Inferior vena cava (IVC; Figure 6)

  1. Position the patient supine for the IVC view.
  2. Begin with the subcostal view, then rotate the transducer counterclockwise until confluence of the right atrium and IVC is appreciated, and the transducer is placed longitudinally in the upper abdomen, to the right of the patient's midline. Optimize the image by tilting the transducer until the IVC is fully visualized, and then by rocking the transducer cephalad until the confluence of the IVC and the right atrium is fully visualized. Adjust the transducer (most frequently with minimal rotation) until the IVC is visualized across the entire screen.
    1. Visualize the right atrium, IVC, liver, and hepatic vein in an optimal view.
    2. Do not confuse the IVC with the aorta. The aorta is left lateral of the IVC and will not touch the liver. The IVC will always touch the liver, and is often surrounded by liver on both sides.
      1. Appreciate that hepatic veins can often be seen coming into the IVC, providing another way to prove that the structure being imaged is the IVC.
      2. Use pulse wave velocity to visualize the venous (vs. arterial) waveform, and confirm the vessel being imaged is the IVC and not the aorta.
  3. Measure IVC diameters at midline and mid-expiration. Do not measure the diameter of the IVC to the side which could underestimate the diameter of the IVC.
  4. 2-D image optimization
    1. Minimize the depth to only include the IVC. Do not include the spine in the image.
    2. Adjust the gain to be the same as for the subcostal view.
    3. Adjust the focus to the depth of the IVC.
    4. Place the cursor 1-3 cm from the confluence of the IVC and the right atrium, and apply M-mode to evaluate respiratory variation of the IVC.

결과

Representative images obtained from the focused cardiac ultrasound protocol presented above are presented in Figure 2, Figure 3, Figure 4, Figure 5, and Figure 6, demonstrating the feasibility of the technique described. These images were captured with the phased array 5-1 MHz transducer. The parasternal long axis (PLAX) image obtained from protocol se...

토론

The aim of this publication is to provide practical recommendations and best practices to achieve optimal FoCUS images in challenging clinical environments. Formal ultrasound seminars, clinical experience, and observations of learners during hands-on teaching have given insight into pitfalls and less optimal tendencies. As a result, many factors that influence the stereotactic and psychomotor skills have become apparent. Although this material is described in relation to FoCUS exams, many of the principles can be applied...

공개

The authors whose names are listed immediately below certify that they have NO affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers' bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge, or beliefs) in the subject matter or materials discussed in this manuscript.

감사의 말

We would like to thank the University of Michigan Department of Anesthesia, Max Harry Weil Institute for Critical Care Research and Innovation, and Katelyn Murphy for their administrative and graphic design support.

자료

NameCompanyCatalog NumberComments
Aquasonic ultrasound gelParker30592052https://dr.graphiccontrols.com/en/catalog/ultrasound-gel/parker-laboratories-01-50-aquasonic-100-gel-5l-1332e66e/
Philips Sparq ultrasound machinePhillipshttps://www.usa.philips.com/healthcare/product/HC795090CC/sparq-ultrasound-system#documents

참고문헌

  1. Birch, M. S., Marin, J. R., Liu, R. B., Hall, J., Hall, M. K. Trends in diagnostic point-of-care ultrasonography reimbursement for medicare beneficiaries among the US emergency medicine workforce, 2012 to 2016. Annals of Emergency Medicine. 76 (5), 609-614 (2020).
  2. Moore, C. L., Copel, J. A. Point-of-care ultrasonography. The New England Journal of Medicine. 364 (8), 749-757 (2011).
  3. Su, E., Dalesio, N., Pustavoitau, A. Point-of-care ultrasound in pediatric anesthesiology and critical care medicine. Canadian Journal of Anaesthesiology. 65 (4), 485-498 (2008).
  4. Niblock, F., Byun, H., Jabbarpour, Y. Point-of-care ultrasound use by primary care physicians. The Journal of the American Board of Family Medicine. 34 (4), 859-860 (2021).
  5. Coritsidis, G. N., et al. Point-of-care ultrasound for assessing arteriovenous fistula maturity in outpatient hemodialysis. The Journal of Vascular Access. 21 (6), 923-930 (2020).
  6. Gundersen, G. H., et al. Adding point of care ultrasound to assess volume status in heart failure patients in a nurse-led outpatient clinic. A randomised study. Heart. 102 (1), 29-34 (2016).
  7. Kirkpatrick, J. N., et al. Recommendations for echocardiography laboratories participating in cardiac point of care cardiac ultrasound (POCUS) and critical care echocardiography training: report from the American Society of Echocardiography. Journal of the American Society of Echocardiography. 33 (4), 409-422 (2020).
  8. Annals of Emergency Medicine. Ultrasound Guidelines: Emergency, point-of-care and clinical ultrasound guidelines in medicine. Annals of Emergency Medicine. 69 (5), 27-54 (2017).
  9. Levitov, A., et al. Guidelines for the appropriate use of bedside general and cardiac ultrasonography in the evaluation of critically ill patients-part II: cardiac ultrasonography. Critical Care Medicine. 44 (6), 1206-1227 (2016).
  10. Neskovic, A. N., et al. Focus cardiac ultrasound core curriculum and core syllabus of the European Association of Cardiovascular Imaging. European Heart Journal. Cardiovascular Imaging. 19 (5), 475-481 (2018).
  11. Mitchell, C., et al. Guidelines for Performing a Comprehensive Transthoracic Echocardiographic Examination in Adults: Recommendations from the American Society of Echocardiography. Journal of the American Society of Echocardiography. 32 (1), 1-64 (2019).
  12. Dinh, V. A., et al. Measuring cardiac index with a focused cardiac ultrasound examination in the ED. The American Journal of Emergency Medicine. 30 (9), 1845-1851 (2012).
  13. Expert Round Table on Echocardiography in ICU. International consensus statement on training standards for advanced critical care echocardiography. Intensive Care Medicine. 40 (5), 654-666 (2014).
  14. Expert Round Table on Echocardiography in ICU. International expert statement on training standards for critical care ultrasonography. Intensive Care Medicine. 37 (7), 1077-1083 (2011).
  15. Spencer, K. T., et al. Focused cardiac ultrasound: recommendations from the American Society of Echocardiography. Journal of the American Society of Echocardiography. 26 (6), 567-581 (2013).
  16. Nelson, B. P., Sanghvi, A. Point-of-care cardiac ultrasound: feasibility of performance by noncardiologists. Global Heart. 8 (4), 293-297 (2013).

재인쇄 및 허가

JoVE'article의 텍스트 или 그림을 다시 사용하시려면 허가 살펴보기

허가 살펴보기

더 많은 기사 탐색

Cardiac UltrasoundImage AcquisitionPatient PositioningTransducer ManipulationImage OptimizationHemodynamic AssessmentClinical ExaminationParasternal Long AxisMitral ValveAortic ValveLeft VentriclePleural EffusionPericardial EffusionTransducer AdjustmentClinical Setting

This article has been published

Video Coming Soon

JoVE Logo

개인 정보 보호

이용 약관

정책

연구

교육

JoVE 소개

Copyright © 2025 MyJoVE Corporation. 판권 소유