Podsumowanie

This manuscript describes a protocol for the in vivo imaging of the mouse retina with high-resolution spectral domain optical coherence tomography (SD-OCT). It focuses on retinal ganglion cells (RGC) in the peripapillary region, with several scanning and quantifying approaches described.

Streszczenie

Structural changes in the retina are common manifestations of ophthalmic diseases. Optical coherence tomography (OCT) enables their identification in vivo—rapidly, repetitively, and at a high resolution. This protocol describes OCT imaging in the mouse retina as a powerful tool to study optic neuropathies (OPN). The OCT system is an interferometry-based, non-invasive alternative to common post mortem histological assays. It provides a fast and accurate assessment of retinal thickness, allowing the possibility to track changes, such as retinal thinning or thickening. We present the imaging process and analysis with the example of the Opa1delTTAG mouse line. Three types of scans are proposed, with two quantification methods: standard and homemade calipers. The latter is best for use on the peripapillary retina during radial scans; being more precise, is preferable for analyzing thinner structures. All approaches described here are designed for retinal ganglion cells (RGC) but are easily adaptable to other cell populations. In conclusion, OCT is efficient in mouse model phenotyping and has the potential to be used for the reliable evaluation of therapeutic interventions.

Wprowadzenie

OCT is a diagnostic tool that facilitates the examination of retinal structures1, including the optic nerve head (ONH). Over the years it has become a dependable indicator of disease progression in humans2,3, as well as in rodents4,5. It uses interferometry to create cross-sectional images of retinal layers at a 2-µm axial resolution. The innermost layer is the retinal nerve fiber layer (RNFL), containing RGC axons, which is followed by the ganglion cell layer (GCL), containing mostly RGC bodies. Next is the inner plexiform layer (IPL), where RGC dendrites meet bipolar, horizontal, and amacrine cell axons. These, together with horizontal cells, form the inner nuclear layer (INL), and their protrusions connect with photoreceptor axons in the outer plexiform layer (OPL). This is followed by the outer nuclear layer (ONL), with photoreceptor cell bodies, and is separated from the photoreceptor layer by the outer limiting membrane (OLM), also called the inner segment/outer segment (IS/OS) layer. Finally, the last observable layers in the mouse retina are the retinal pigment epithelium (RPE) and the choroid (C). The RNFL alone is normally too thin to be measured in mice; thus, analyzing the RNFL/GCL instead is preferable4,5. Another possibility is the GC complex layer, which contains the latter in addition to the IPL, making it thicker and thus even easier to measure on OCT scans4. Consequently, OCT can provide insight into the pathological status of the retina, such as in OPNs.

Alternatively, the thickness of the mouse retina is often analyzed with post mortem histology. However, this technique faces limitations relating to tissue collection, fixation, cutting, staining, mounting, etc. Hence, some defects, such as subtle thickness changes, cannot be detected. Finally, because the same mouse cannot be tested at several time points, the number of animals per study greatly increases, unlike for OCT. All in all, the non-invasiveness, high-resolution, possibility for repetition, time monitoring in time, and ease of use of the OCT technology make it the method of choice in retinal disease studies.

Mouse models are used to identify gene defects and to elucidate molecular mechanisms underlying retinopathies6. OPN is a form of retinopathy with substantial damage to the optic nerve (ON), which is made up of approximately 1.2 million RGC axons. OPN can be focused on the ON or can be secondary to other disorders, inborn or not7, leading to visual field loss and later, blindness. Characteristic traits of OPN are RGC loss and ON damage, which can be observed in human OCT as RNFL and GCL thinning2,3. Meanwhile, the pathophysiology of OPN is still poorly understood, and hence the need to test mouse retinas remains.

This manuscript describes the imaging and quantification of retinal layer thickness, using the example of the Opa1delTTAG mouse line8,9, a model of dominant optic atrophy (DOA)10. To assess RGC pathophysiology, radial, rectangular, and annular scans were quantified. This was done either with standard calipers provided by the OCT software or with a homemade macro developed for an open-source image processing program. The standard calipers are difficult to manipulate and often thicker than the RNFL/GCL, while the homemade calipers are easy to use, reproducible, and more precise. The macro performs a measurement for an automatically detected layer, in 5 points and at fixed positions, on both sides of the ONH in the peripapillary region. The goal of the presented protocol is to describe OCT scan acquisition to specify retinal positioning, with a focus on RGCs.

Protokół

The experimental protocol was approved by the Institut national de la santé et de la recherche médicale (Inserm; Montpellier, France), is consistent with the European directives, and complies with the ARVO Statement for the Use of Animals in Ophthalmic Research. It was carried out under the agreement of the Languedoc Roussillon Comity of Ethics in Animal Experimentation (CEEALR; nuCEEA-LR-12123).

1. Equipment Setup and Pre-imaging Preparation

NOTE: Here, OCT was performed on mouse retinas using the spectral domain (SD) ophthalmic imaging system (Figure 1A). The SD-OCT apparatus consists of a base and an animal imaging mount (AIM) with a rodent alignment stage (RAS) (Figure 1B). The base includes the computer, the OCT engine, the SD-OCT probe, and the mouse-specific lens. The probe is mounted onto the AIM, which includes the Z-translator. The RAS is used for mouse positioning thanks to the table with the X- and Y-translator, the cassette that can be rotated and swiveled, and the removable bite bar with the nose band. The software provided by the manufacturer allows for the acquisition and analysis of OCT files, although the latter may also be done with an open-source image processing program.

  1. Firmly place the probe in the AIM.
  2. Connect the mouse-specific lens to the probe.
  3. Adjust the reference arm settings for the specific lens (here, power at 086 and position at 964).
  4. Making sure that the mouse-specific lens is sufficiently distant from the cassette, attach the bite bar with the nose band onto the cassette.
    NOTE: The distance between the lens and the cassette may be adjusted with the Z-translator screw, located at the back of the AIM. The nose band is a commercially available elastic band, and its tension should be adjusted depending upon the size of the mouse.
  5. Turn on the power supply (lower right corner of the cart) and then the computer.
  6. To launch the imaging program, double-click on the appropriate shortcut on the screen.
    NOTE: This depends on the type of lens; here, Mouse Retina.
  7. Create a new Clinical Study and add the desired Protocols. Otherwise, use the existing Clinical Study and/or Protocols.
    1. Add a new study by choosing the "Study Name"; "ID"; "Treatment Arm Specifications" (here, "Uncategorized"); and predefined "Scan Protocols", if they exist.
  8. Choose an "Examiner" in the "Clinical Study" section.
    NOTE: For a new "Examiner", go to "Setup Examiners & Physicians" to define it.
  9. Add a new Patient in the Patient/Exam section by click on Add Patient; entering the "ID", "name", "surname", "sex", and "date of birth" (optional).
    NOTE: Do this just before testing each mouse, if more convenient. Make sure that the ID is no longer than 10 characters. For the mouse, the refractive error for both eyes is 0 and the axial length is 23.0.
  10. Click on "Add Exam" and select the desired "Protocol" by adding "Preset Scans" from the list, starting with the eye that will be measured first (here, the right eye), or by customizing the scans.
  11. To customize a scan, either use an existing scan as a template and edit it, or create it from scratch through the "Add Custom Scan" option. After adding all the scans to the list, define a new protocol using the "Enter "New Protocol Name" option (OS: oculus sinister, left eye; OD: oculus dexter, right eye).
    NOTE: Here, the protocol involves three scans: (i) a radial scan, with a 1.4 mm diameter, 0 mm horizontal and vertical offsets, 1,000 lines of A-scans/B-scan, 100 B-scans/volume, 1 frame/B-scan, 80 lines of inactive A-scans/B-scan, and 1 volume; (ii) a rectangular scan, with a 1.4-mm length and width, 0° angle, 0-mm horizontal and vertical offsets, 1,000 lines of A-scans/B-scan, 100 B-scans, 1 frame/B-scan, 80 lines of inactive A-scans/B-scan, and 1 volume; and (iii) an annular scan, with a 0.4-mm minimum and 0.6-mm maximum diameter, 0-mm horizontal and vertical offsets, 1,000 lines of A-scans/B-scan, 3 B-scans/volume, 48 frames/B-scan, 80 lines of inactive A-scans/B-scan, and 1 volume.

2. Mouse Preparation

  1. Eye dilation and immobilization
    1. At least 15 min before data acquisition, instill 10% phenylephrine eye drops into the mouse eyes, remove the excess, and instill 0.5% tropicamide eye drops.
      NOTE: The first dilatation can be done at once for all the mice that will be tested in the session. Make sure that the liquids are at room temperature.
    2. Immediately after inducing general anesthesia (step 2.2), administer 0.4% oxybuprocaine hydrochloride eye drops, keeping them in place for 3 s to anesthetize and immobilize the eyes. Afterwards, wipe off the drops and repeat the instillation of 10% phenylephrine and 0.5% tropicamide to ascertain that the eyes are well dilated.
      NOTE: Make sure that the liquids are at room temperature and that the mouse does not swallow the oxybuprocaine.
  2. General anesthesia
    1. Prepare an anesthetic solution of ketamine (20 mg/mL) and xylazine (1.17 mg/mL) in saline. Store at 4 °C for a maximum of 2 weeks.
    2. Approximately 5 min before testing, intraperitoneally inject 5 - 10 µL of anesthetic solution per 1 g of body mass, depending upon the age and size of the mouse.
      NOTE: Young and/or thin mice need less anesthetic, take less time to anesthetize, but also awaken faster. Here, 8 µL/g (160 mg/kg of ketamine, 9.33 mg/kg of xylazine).
    3. Optionally, lubricate the eyes with viscous eye drops or an ophthalmic gel to avoid corneal dryness.

3. Mouse Positioning

  1. Lubricate the eyes with viscous glycol-based eye drops to provide corneal hydration.
    NOTE: If the eyes seem dry during the exam, re-apply the eye drops.
  2. Wrap the mouse in a sheet of surgical gauze to keep it warm.
  3. Using a sponge or cotton wick, apply a thin layer of an ophthalmic gel with 0.3% hypromellose onto each eye, to minimize light refraction, avoid opacities, and secure corneal hydration. While doing so, move the eyelashes and whiskers aside.
  4. Position the mouse in the cassette, with the head straight and pointing forward.
  5. Gently open the bite bar clamp and place the bar in the mouth; use the nose band to secure the position.
    NOTE: Ensure that the bite bar is in the middle of the cassette.
  6. Place a cotton roll under the right (left) side if the right (left) eye is being tested.
  7. While keeping the clip-on aiming tip on the mouse-specific lens, bring it towards the eye by rotating the Z-translator screw counterclockwise.
  8. By rotating and swiveling the cassette and by turning the bite bar and the X-translator screws, pre-set the position of the mouse; the goal is to have the right eye look directly into the lens, and thus to align the optical axes of the eyes and the lens.
    NOTE: If the mouse is too high or too low, the Y-translator screw should be used first.
  9. Choose the first scan in the "Exam", click on "Start Aiming", and make any further manual adjustments for retinal imaging.
    1. Using the Z-translator screw, move the retina vertically on the left panel (Figure 2, Horizontal B-scan Alignment) and horizontally on the right panel (Figure 2, Vertical B-scan Alignment).
    2. Rotate the cassette to bring the ONH in the middle of the right panel by moving the ONH up or down. Use the bite bar screw to straighten the retina on the right panel. Swivel the cassette to position the ONH in the middle of the left panel.
    3. Use the X-translator screw to level the retina on the left panel. Keeping in mind the major function of each modulator, further adjust the position of the retina to centralize the ONH.
      NOTE: Use the Y-translator screw to move the ONH up and down on the right panel, if necessary. The position may be refined at any time between scans.

4. SD-OCT Imaging of the ONH and Retina

  1. Once content with the adjustments, click on "Start Snapshot" to begin SD-OCT scanning.
    1. If a 3D imaging is not required, uncheck the OCU option.
  2. Save the scan and the report.
  3. Proceed with the next scans.
  4. To image the second eye, after retracting the lens, turn the cassette accordingly and repeat steps 3.6 - 4.3.

5. Acquisition Completion

  1. When the acquisition is complete, remove the mouse from the cassette, apply ophthalmic gel with 0.3% hypromellose to each eye, and place the mouse on a heating plate to wake up.
  2. After the last acquisition, close the software and switch off the computer and the OCT machine (power supply button).
  3. Clean the cassette with disinfectant.

6. Analysis

  1. For the retinal layer thickness measurement, use the automated segmentation software provided by the manufacturer.
    1. Click on the "Patient", choose the desired "Exam" from then list, and click on the "Review Exam" option.
    2. Load the desired OCT data onto the OCT software by right-clicking on the folder icon in the desired scan.
    3. Right click on the B-scan; configure the calipers by enabling up to 10 measuring calipers; and enter their names, angles, and colors.
    4. Choosing the desired caliper by a right-clicking on the B-scan; place it accordingly onto the retina for measurement.
      NOTE: For the scans centered on the ONH, set 5 calipers on each side of it, equidistant from each other. For the peripapillary analysis, make sure that the caliper is not placed too far from the ONH. For radial scanning, analyze 10 pictures per scan that were chosen by the OCT software. Their numbers may be found in the "Reports" folder.
    5. Save the results for analysis in a spreadsheet software program by right-clicking on the B-scan and clicking "Save results".
      NOTE: The results can be found in the same folder as the scanning data.
  2. Alternatively, use the homemade caliper macro, "MRI Retina Tool", developed for an open-source image processing program (see the Table of Materials).
    1. Make sure that the "MRI Retina Tool" and the "Modify Polygon Section Tool" macros are active. Load the image. Click on the m button to start the measurement.
      NOTE: The macro automatically creates one 0.2 mm-long measuring cassette on both sides of the ONH. Each cassette contains 5 measuring points that function as calipers. Their lateral position is unchangeable and is customized for the peripapilla in the radial scans. The horizontal position of the cassettes is predefined to measure the RNFL/GCL thickness, but it is easily modifiable to measure the GC complex layer instead. If the scan quality is poor, the horizontal position needs to be adjusted, or the picture needs to be excluded from the analysis.
    2. For horizontal adjustment, click on the e button. In the newly opened "ROI Manager" window, choose the first cassette.
    3. Click on the blue polygon button and adjust the position of the first cassette by clicking on the borders of the measured layer in the picture.
    4. Repeat for the second cassette by first choosing it in the "ROI Manager" window and then clicking on the appropriate image. Click on the r button to re-measure. View the results in the "Measurements" window.
      NOTE: The results may be copied to a spreadsheet software program at any time. They contain the following values for the right cassette (r), left cassette (l), and total: Intden - integrated density within the cassette; Area: area of the measurement, in mm2; Len: length of the caliper within the cassette, in mm; Mean: mean intensity of the signal within the cassette; and Std: standard error of the mean intensity of the signal.
    5. Proceed with the next image.
    6. Further analyze the data in a spreadsheet software program.
      NOTE: To find the mean thickness of the layer, take the ten Len values.

Wyniki

The SD-OCT technology enables retinal imagining and thickness analysis that is comparable to histology, but is faster and more detailed (Figure 3). As presented with wildtype C57Bl/6 mice, even though the quality of an SD-OCT scan (Figure 3A, right) is not as good as that of an image of a retinal cross-section (Figure 3A, left), it visualizes more layers (e.g., OLM). Moreover, it takes only ...

Dyskusje

The OCT system, a non-invasive in vivo imaging method, provides high-resolution retinal cross-section-like scans. Thus, its main advantage is its potential for detailed analysis, with the wonderful opportunity to transpose protocols routinely applied to humans to mouse models.

In the example of Opa1delTTAG mutant mice, SD-OCT results showed an increase of RNFL and GC complex layer thickness, which allowed for further exploration of DOA pathophysiology

Ujawnienia

The authors have nothing to disclose.

Podziękowania

This work was supported by Inserm, Université Montpellier, Retina France, Union National des Aveugles et Déficients Visuels (UNADEV), Association Syndrome de Wolfram, Fondation pour la Recherche Médicale, Fondation de France, and the Laboratory of Excellence EpiGenMed program.

Materiały

NameCompanyCatalog NumberComments
Mice
Opa1delTTAG mouseInstitute for Neurosciences in Montpellier, INSERM UMR 1051, France-Opa1 knock-in mice carrying  OPA1 c.2708_2711delTTAG mutation on C57Bl6/J background
NameCompanyCatalog NumberComments
Equipment
EnVisu R2200 SD-OCT Imaging SystemBioptigen, Leica Microsystems, Germany-Spectral-Domain Optic Coherence Tomography system
EnVisu R2200 SD-OCT Imaging System SoftwareBioptigen, Leica Microsystems, Germany-Software for OCT acquisition and analysis
ImageJ 1.48vWayne Rasband, National Institutes of Health, USA-Software for analysis, requires downloading and installing two hommade macros: http://dev.mri.cnrs.fr/projects/imagej-macros/wiki/Retina_Tool
Self-regulating heating plateBioseb, FranceBIO-062Protection against hypothermia
NameCompanyCatalog NumberComments
Supplies
Nose Band--Elastic band
Gauze pads 3"x3"Curad, USACUR20434ERBProtection against hypothermia
Dual Ended Cotton tip applicatorEssence of Beauty, CVS Health Corporation, USA-Gel application
Cotton TwistsCentraVet, FranceT.7979C.CSMouse positioning
NameCompanyCatalog NumberComments
Reagents and Drugs
Néosynéphrine Faure 10%Laboratoires Europhtha, Monaco-Eye dilatation
Mydriaticum 0.5%Laboratoires Théa, France3397908Eye dilatation
Cebesine 0.4%Laboratoire Chauvin, Bausch&Lomb, France3192342Local anesthesia
Imalgene 1000Merial, France/CentraVet, FranceIMA004General anesthesia
RompunBayer Healthcare, Germany/CentraVet, FranceROM001General anesthesia, analgesia, muscle relaxation
NaCl 0,9%Laboratoire Osalia, France 103697114Physiological serum
Systene UltraAlcon, Novartis, USA-Hydration of eyes
GenTeal'Alcon, Novartis, USA-Ophtalmic gel to minimize light refraction and opacities
Aniospray Surf 29Laboratoires Anios, France59844Desinfectant

Odniesienia

  1. Drexler, W., Fujimoto, J. G. State-of-the-art retinal optical coherence tomography. Prog Retin Eye Res. 27 (1), 45-88 (2008).
  2. Grenier, J., et al. WFS1 in Optic Neuropathies: Mutation Findings in Nonsyndromic Optic Atrophy and Assessment of Clinical Severity. Ophthalmology. 123 (9), 1989-1998 (2016).
  3. Zmyslowska, A., et al. Retinal thinning as a marker of disease progression in patients with Wolfram syndrome. Diabetes Care. 38 (3), e36-e37 (2015).
  4. Fischer, M. D., et al. Noninvasive, in vivo assessment of mouse retinal structure using optical coherence tomography. PLoS One. 4 (10), e7507 (2009).
  5. Grieve, K., Thouvenin, O., Sengupta, A., Borderie, V. M., Paques, M. Appearance of the Retina With Full-Field Optical Coherence Tomography. Invest Ophthalmol Vis Sci. 57 (9), OCT96-OCT104 (2016).
  6. Chang, B., et al. Retinal degeneration mutants in the mouse. Vision Res. 42 (4), 517-525 (2002).
  7. Mustafa, S., Pandit, L. Approach to diagnosis and management of optic neuropathy. Neurol India. 62 (6), 599-605 (2014).
  8. Sarzi, E., et al. The human OPA1delTTAG mutation induces premature age-related systemic neurodegeneration in mouse. Brain. 135 (Pt 12), 3599-3613 (2012).
  9. Sarzi, E., et al. Increased steroidogenesis promotes early-onset and severe vision loss in females with OPA1 dominant optic atrophy. Hum Mol Genet. 25 (12), 2539-2551 (2016).
  10. Delettre-Cribaillet, C., Hamel, C. P., Lenaers, G., Pagon, R. A., et al. Optic Atrophy Type 1. Gene Reviews. 7 (2007), (2007).
  11. Lenaers, G., et al. Dominant optic atrophy. Orphanet J Rare Dis. 7 (46), (2012).
  12. Delettre, C., et al. Nuclear gene OPA1, encoding a mitochondrial dynamin-related protein, is mutated in dominant optic atrophy. Nat Genet. 26 (2), 207-210 (2000).
  13. Liu, Y., et al. Monitoring retinal morphologic and functional changes in mice following optic nerve crush. Invest Ophthalmol Vis Sci. 55 (6), 3766-3774 (2014).

Przedruki i uprawnienia

Zapytaj o uprawnienia na użycie tekstu lub obrazów z tego artykułu JoVE

Zapytaj o uprawnienia

Przeglądaj więcej artyków

Optical Coherence TomographyOCTRetinal Ganglion CellsNon invasiveOptic NeuropathyRetinal ThicknessMouseEyePhenylephrineTropicamideAnesthesiaOxybuprocaine HydrochlorideGlycolHypromelloseImaging

This article has been published

Video Coming Soon

JoVE Logo

Prywatność

Warunki Korzystania

Zasady

Badania

Edukacja

O JoVE

Copyright © 2025 MyJoVE Corporation. Wszelkie prawa zastrzeżone