Summary

A procedure is described for manipulating the activity of cerebral cortical pyramidal neurons optogenetically while the electroencephalogram, electromyogram, and cerebral lactate concentration are monitored. Experimental recordings are performed on cable-tethered mice while they undergo spontaneous sleep/wake cycles. Optogenetic equipment is assembled in our laboratory; recording equipment is commercially available.

Abstract

Although the brain represents less than 5% of the body by mass, it utilizes approximately one quarter of the glucose used by the body at rest1. The function of non rapid eye movement sleep (NREMS), the largest portion of sleep by time, is uncertain. However, one salient feature of NREMS is a significant reduction in the rate of cerebral glucose utilization relative to wakefulness2-4. This and other findings have led to the widely held belief that sleep serves a function related to cerebral metabolism. Yet, the mechanisms underlying the reduction in cerebral glucose metabolism during NREMS remain to be elucidated.

One phenomenon associated with NREMS that might impact cerebral metabolic rate is the occurrence of slow waves, oscillations at frequencies less than 4 Hz, in the electroencephalogram5,6. These slow waves detected at the level of the skull or cerebral cortical surface reflect the oscillations of underlying neurons between a depolarized/up state and a hyperpolarized/down state7. During the down state, cells do not undergo action potentials for intervals of up to several hundred milliseconds. Restoration of ionic concentration gradients subsequent to action potentials represents a significant metabolic load on the cell8; absence of action potentials during down states associated with NREMS may contribute to reduced metabolism relative to wake.

Two technical challenges had to be addressed in order for this hypothetical relationship to be tested. First, it was necessary to measure cerebral glycolytic metabolism with a temporal resolution reflective of the dynamics of the cerebral EEG (that is, over seconds rather than minutes). To do so, we measured the concentration of lactate, the product of aerobic glycolysis, and therefore a readout of the rate of glucose metabolism in the brains of mice. Lactate was measured using a lactate oxidase based real time sensor embedded in the frontal cortex. The sensing mechanism consists of a platinum-iridium electrode surrounded by a layer of lactate oxidase molecules. Metabolism of lactate by lactate oxidase produces hydrogen peroxide, which produces a current in the platinum-iridium electrode. So a ramping up of cerebral glycolysis provides an increase in the concentration of substrate for lactate oxidase, which then is reflected in increased current at the sensing electrode. It was additionally necessary to measure these variables while manipulating the excitability of the cerebral cortex, in order to isolate this variable from other facets of NREMS.

We devised an experimental system for simultaneous measurement of neuronal activity via the elecetroencephalogram, measurement of glycolytic flux via a lactate biosensor, and manipulation of cerebral cortical neuronal activity via optogenetic activation of pyramidal neurons. We have utilized this system to document the relationship between sleep-related electroencephalographic waveforms and the moment-to-moment dynamics of lactate concentration in the cerebral cortex. The protocol may be useful for any individual interested in studying, in freely behaving rodents, the relationship between neuronal activity measured at the electroencephalographic level and cellular energetics within the brain.

Protocol

1. Surgical Preparation of Animals

1. Experimental Subjects

Use mice of the B6.Cg-Tg(Thy1-COP4/eYFP)18Gfng/J transgenic line9; JAX strain #7612) or other mice expressing the blue light-sensitive cation channel, Channelrhodopsin-2, in cerebral cortical neurons. Application of blue light to the cerebral cortex of the B6.Cg-Tg(Thy1-COP4/eYFP)18Gfng/J transgenic line causes the pyramidal neurons expressing Channelrhodopsin-2 to depolarize and undergo action potentials9,10. As a consequence of pyramidal cell activation, local interneurons are activated, and the change in potential is propagated to neurons beyond the site of stimulation10. Perform surgery under sterile conditions in adherence to applicable regulatory guidelines: autoclave surgical tools and softpacks (surgical field, gauze pads); hot bead sterilize all heat-tolerant implanted metal devices (cannulas, screws and wires) for 30-sec; Disinfect heat-intolerant implantable devices (fiber optic cable, plastic connector) with a 10-sec dip in ethanol; blow ethanol-dipped items dry by application of canned air.

2. Anesthesia, Stereotaxic Placement and Clearance of Skull

Use the IMPAC6 Integrated Multi Patient (Vet Equip Inc) system to anesthetize the mouse. Use 5% Isoflurane/95% Oxygen for induction and 3% Isoflurane/97% oxygen for maintenance during surgery. Place the animal on a circulating-water blanket set to 37 C. It is not necessary to monitor body temperature if the blanket is kept at this temperature.

3. Preparing the Skull for Implantation

Prepare the incision site by shaving all fur from the top of the skull. The shaved area should extend laterally from ear to ear and anterior-posterior from the eyes to the posterior end of the skull. Disinfect the exposed skin with three consecutive swabs of betadine followed by ethanol. Make a medial incision on the top of the skull, from between the eyes to the back of the skull. Clean the skull with hydrogen peroxide and sterile saline (0.9% NaCl). Control bleeding with a cauterizing tool. Locate and mark bregma and lambda for determining stereotaxic coordinates. Stereotaxic coordinates for electrode/optogenetic stimulus configurations that we have used are presented in Table 1 and shown schematically in Figure 1A.

4. Preparation of the Implantation Sites on Skull

Use a high speed dental drill with a 0.5 mm ball burr bit to establish each implantation site in the skull. Chamfer the external margin of each hole using a 0.7 mm ball burr bit to facilitate screw insertion.

5. Insertion and Securing of Cannulas, EMG Leads and EEG Leads

Insert EEG screws (0.8 mm in length, with 2 cm of uncoated 30-gauge tinned-copper bus wire pre-soldered to screw head) into the holes and drive them in with a slotted hand screwdriver approximately 4-5 revolutions to obtain the desired depth. Hold guide cannulas in place with a stereotaxic cannula holder and then affix them to the skull and anchors with acrylic cement. Each guide cannula should contain a dummy cannula/stylet (Plastics One, part # C312 DC\1) from the time of surgery to the time of experimentation (10-14 days) to maintain patency.

6. Closing Surgical Site

Once EEGs and guide cannulas are positioned in the skull, bond them together with dental acrylic cement (Lang Dental - Ortho-Jet cement). After cement sets, position the plastic connector (Pinnacle Technology part # 8402) above the dried cement mound. Solder the ends of the wires emanating from EEG leads to the contacts on the plastic connector. Encase the wire leads in cement. Then draw the EMG wires through the nuchal muscles by sliding them into the barrel of a 21ga needle pierced through muscle. Tie a double surgeon's knot of 5-0 nylon suture around these wires just distal to where they exit the muscle. Suture back together the skin that was retracted to access the muscle tissue with single interrupted surgeon knots, using a reverse cutting P-3 needle and 5-0 nylon suture.

7. Post-operative Analgesia and Recovery

Administer buprenorphine as an analgesic (0.1 mg/kg, subcutaneous) and flunixin meglumine (0.1 mg/kg, subcutaneous) as an anti-inflammatory immediately after surgery, and daily on days 1-2 post-surgery. Allow animals to recover in standard colony housing conditions for at least seven days before experimentation. With no additional care beyond standard colony housing, cannulas can be expected to remain patent for at least two weeks with the indwelling stylets secured in place. EEG and EMG sensors can also be expected to retain functionality for this duration.

2. Generation of Timed Light Pulses

1. Programming of MC Stimulus Unit

Connect the MC-Stimulus unit (Multi-Channel Systems) via a USB connection to a desktop computer. Use the spreadsheet-like MC-Stimulus user interface to program the MC Stimulus unit. Enter the desired TTL (transistor-transistor logic) voltage for the on period, the durations of stimulus on and off periods, the number of cycles per second desired, and the total number of repeated cycles for the entire stimulus session. For further details see MC-Stimulus technical manual:

http://www.multichannelsystems.com/sites/multichannelsystems.com/files/documents/manuals/STG4002_Manual.pdf

3. Manufacture/Assembly of Fiber Optic Probes for Light Delivery to the Cerebral Cortex

  1. Supplies and equipment needed are found at: http://www.thorlabs.com/Thorcat/1100/1166-D02.pdf
  2. Cutting and polishing of fiber optic cables. http://www.thorlabs.com/Thorcat/1100/1166-D02.pdf

4. Assembly of Rodent Optogenetic Stimulation System

1. Interface of MC-Stimulus Unit and Laser

Once the MC-Stimulus unit is programmed, run it as a stand-alone signal generator of a 5-Volt binary ON/OFF signal. Connect the MC-Stimulus unit to the TTL enabled power unit of the laser with BNC connectors.

2. Producing Laser Output via Fiber Optic Cable

Use the digital display and variable power output dial of the laser power unit to manually adjust the power output of the laser. Connect the laser power unit to the laser via a ribbon cable. Connect the female FC (ferrule connector) fiber adapter of the laser to the male FC connector on the fiber optic patch cable (a 3 m length, 200 μm diameter glass conduction fiber encased in light reflecting acrylate polymer cladding).

3. Conveying the Laser to the Animal: Rotary Commutator

Connect the originating fiber optic cable (3 m in length) and the delivery fiber optic cable (45 cm in length) to opposing ends of the optic rotary joint encasing two collimating lenses (Doric lenses). The fiber optic rotary joint serves as a commutator: as the animal moves about its cage, the rotary joint rotates to prevent breakage of the fiber optic cables. Use plastic cable ties to affix the commutator to a metal stand placed above the cylindrical cage in which the animal is housed.

4. Attaching Delivery Fiber Optic Cable to Head

Restrain the mouse by using one hand to pin the mouse under a cupped palm. Orient the head between the experimenter's middle and index finger. Remove the dummy cannula from the guide cannula placed for the fiber optic cannula. Clear the guide cannula of debris using a sterile 25ga needle. Insert the fiber optic cable by hand and fasten it to the fiber optic guide cannula with a threaded screw cap salvaged from a dummy cannula. Control the depth of insertion of the fiber optic cable in the brain by a suture knot tied on the fiber optic cable at a fixed distance from the flat-cleaved end. The entire experimental apparatus is shown in Figure 1B.

5. Measurement of EEG-defined Sleep and Lactate Concentration during Optogenetic Stimulation

1. Precalibration of the Lactate Sensor

Pre-calibration of the lactate sensor is performed with an in vitro kit (Pinnacle Technology part # 7000-K1-T). The sensor is equilibrated in phosphate buffered saline (pH 7.4), then exposed to three concentrations of L-lactate in stepwise fashion per manufacturer protocols.

2. Insertion of the Lactate Sensor and EEG Cable

Insert the pre-calibrated sensor into the skull-mounted lactate guide cannula in a manner identical to the fiber optic cable insertion procedure (section 4.4). Connect the lactate sensor to the preamplifier of the Pinnacle 8,400 biosensor with bipolar plug connectors. Attach this preamplifier to the 8-pin connector on the surgically-implanted headmount.

3. Establishing Optogenetic Stimulus Intensity

Before collecting data, use the laser intensity control knob to adjust the intensity of the optogenetic stimulus. The amplitude of the EEG response will vary across animals, due to factors that have not been studied systematically, and must be verified by visual inspection at the onset of stimulation. In practice, an intensity of 80-120 μWatts will typically produce an increase in EEG amplitude of approximately 50% at the frequency of stimulation. Post-hoc fast Fourier transform analysis (see section 5.5) is necessary to quantify the absolute magnitude of the response.

4. Data Collection

Collect data using the Pinnacle 8400 system: http://pinnaclet.com/pal-8400.html.

5. Data Processing with Neuroscore

Classify sleep states by visual inspection of the EEG and EMG data with the Neuroscore interface (DataSciences, International: http://www.datasci.com/products/software/NeuroScore_CNS_Software.asp) or the Sirenia interface (Pinnacle Technology: http://www.pinnaclet.com/sirenia.html). Process data in ten-second epochs as wake, non-rapid eye movement sleep, or rapid eye movement sleep based on EEG and EMG. Save data as Microsoft Excel spreadsheets for additional analysis and statistical testing.

Results

As shown in Figure 2, a mouse equipped for optogenetic stimulation and lactate/EEG/EMG data collection underwent spontaneous sleep/wake state transitions while EEG, EMG and cerebral lactate concentration were monitored continuously. Current at the lactate sensor increased during periods of low amplitude EEG and decreased during periods of high amplitude EEG. As shown in Figure 3, both channels of the EEG are responsive to optogenetic stimuli delivered in the frontal cort...

Discussion

The methods presented here allow one to measure the relationship between sleep and changes in the brain concentration of the glycolytic intermediate lactate on a time scale not previously possible. Animals undergo spontaneous transitions between wake, NREMS and REMS. Furthermore, we are able to apply optogenetic stimuli while animals undergo these transitions. Data collected to date demonstrate that both spontaneous and induced waves impact on the readout of a lactate oxidase-based biosensor.

Disclosures

No conflicts of interest declared.

Acknowledgements

Research funded by Department of Defense (Defense Advanced Research Projects Agency, Young Faculty Award, Grant Number N66001-09-1-2117) and NINDS (R15NS070734).

Materials

NameCompanyCatalog NumberComments
BASi Mouse Guide CannulaPinnacle Technology/BASi Inc7032
Lactate BiosensorPinnacle Technology7004
Head MountPinnacle Technology8402
Sleep/Biosensor Recording systemPinnacle Technology8400-K1-SL2 EEG channels, 1 EMG channel, & 1 biosensor
Tethered Mouse in-vitro Calibration kitPinnacle Technology7000-K1-T
Fiber Optic Guide CannulaPlastics OneC312G21 Gauge Guide Cannula
Dummy CannulaPlastics OneC312DC21 Gauge Dummy
Diamond Fiber ScribeThorlabsS90W
Fiber Connector Crimp ToolThorlabsCT042
Furcation TubingThorlabsFT03003.0 mm
ThorlabsT10S13Max Dia. 0.012
Furcation Tube StripperThorlabsFTS3
Bare Hard Cladding Multimode FiberThorlabsBFL37-200200 μm Core, 0.37 NA
Wire Snips/Kevlar ShearsThorlabsT865
Fiber Optic EpoxyThorlabsF112
Fiber Stripper ToolThorlabs
Glass Polishing PlateThorlabsCTG913
Rubber Polishing PadThorlabsNRS913
Eye LoupeThorlabsJEL10
Kim WipesThorlabsKW32
Compressed AirThorlabsCA3
Polishing PuckThorlabsD50-xx
Fiber Inspection scopeThorlabsCL-200
Polishing FilmsThorlabsLFG5P, LFG3P, LFG1P, LFG03P
FC/PC connector endThorlabs30126G2-240240 μm Bore, SS Ferrule
MC Stimulus UnitMulti-Channel SystemsSTG-4002
MC Stimulus SoftwareMulti-Channel SystemsMC-Stimulus V 2.1.5
Blue LaserCrystaLaserCL473-050-0
Laser Power supplyCrystaLaserCL2005
Fiber Optic Rotary JointDoric LensesFRJ-v4
Table 2. Supplies and equipment.

References

  1. Magistretti, P., Zigmond, M. J., Bloom, F. E., Landis, S. C., Roberts, J. L., Squire, L. R. Brain Energy Metabolism. Fundamental Neuroscience. , 389-413 (1999).
  2. Maquet, P., et al. Cerebral glucose utilization during sleep-wake cycle in man determined by positron emission tomography and [18F]2-fluoro-2-deoxy-D-glucose method. Brain Res. 513 (1), 136-143 (1990).
  3. Buchsbaum, M. S., et al. Regional cerebral glucose metabolic rate in human sleep assessed by positron emission tomography. Life Sci. 45 (15), 1349-1356 (1989).
  4. Kennedy, C. Local cerebral glucose utilization in non-rapid eye movement sleep. Nature. 297 (5864), 325-327 (1982).
  5. Pappenheimer, J. R., Koski, G., Fencl, V., Karnovsky, M. L., Krueger, J. Extraction of sleep-promoting factor S from cerebrospinal fluid and from brains of sleep-deprived animals. J. Neurophysiol. 38 (6), 1299-1311 (1975).
  6. Borbely, A. A., Achermann, P., Kryger, M. H., Roth, T., Dement, W. C. Sleep homeostasis and models of sleep regulation. Principles and Practice of Sleep Medicine. , 377-390 (2004).
  7. Destexhe, A., Contreras, D., Steriade, M. Spatiotemporal analysis of local field potentials and unit discharges in cat cerebral cortex during natural wake and sleep states. J. Neurosci. 19 (11), 4595-4608 (1999).
  8. Astrup, J., Sorensen, P. M., Sorensen, H. R. Oxygen and glucose consumption related to Na+-K+ transport in canine brain. Stroke. 12 (6), 726-730 (1981).
  9. Arenkiel, B. R., et al. In vivo light-induced activation of neural circuitry in transgenic mice expressing channelrhodopsin-2. Neuron. 54 (2), 205-218 (2007).
  10. Mateo, C. In Vivo Optogenetic Stimulation of Neocortical Excitatory Neurons Drives Brain-State-Dependent. Curr. Biol. , (2011).
  11. Wisor, J. P., Clegern, W. C. Quantification of short-term slow wave sleep homeostasis and its disruption by minocycline in the laboratory mouse. Neurosci. Lett. 490 (3), 165-169 (2011).
  12. El Yacoubi, M., et al. Behavioral, neurochemical, and electrophysiological characterization of a genetic mouse model of depression. Proc. Natl. Acad. Sci. U.S.A. 100 (10), 6227-6232 (2003).
  13. Tsunematsu, T., et al. Acute optogenetic silencing of orexin/hypocretin neurons induces slow-wave sleep in mice. J. Neurosci. 31 (29), 10529-10539 (2011).
  14. Le, S., Gruner, J. A., Mathiasen, J. R., Marino, M. J., Schaffhauser, H. Correlation between ex vivo receptor occupancy and wake-promoting activity of selective H3 receptor antagonists. J. Pharmacol. Exp. Ther. 325 (3), 902-909 (2008).
  15. Burmeister, J. J., Palmer, M., Gerhardt, G. A. L-lactate measures in brain tissue with ceramic-based multisite microelectrodes. Biosens. Bioelectron. 20 (9), 1772-1779 (2005).
  16. Cardin, J. A. Targeted optogenetic stimulation and recording of neurons in vivo using cell-type-specific expression of Channelrhodopsin-2. Nat. Protoc. 5 (2), 247-254 (2010).
  17. Destexhe, A., Contreras, D., Steriade, M. Cortically-induced coherence of a thalamic-generated oscillation. Neuroscience. 92 (2), 427-443 (1999).
  18. Liu, Z. W., Faraguna, U., Cirelli, C., Tononi, G., Gao, X. B. Direct evidence for wake-related increases and sleep-related decreases in synaptic strength in rodent cortex. J. Neurosci. 30 (25), 8671-8675 (2010).
  19. Iwai, Y., Honda, S., Ozeki, H., Hashimoto, M., Hirase, H. A simple head-mountable LED device for chronic stimulation of optogenetic molecules in freely moving mice. Neurosci. Res. 70 (1), 124-127 (2011).

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Explore More Articles

Simultaneous ElectroencephalographyReal time MeasurementLactate ConcentrationOptogenetic ManipulationNeuronal ActivityRodent Cerebral CortexGlucose UtilizationNon Rapid Eye Movement SleepCerebral MetabolismSlow WavesElectroencephalogramDepolarized up StateHyperpolarized down StateAction Potentials

This article has been published

Video Coming Soon

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved