Electrochemical Measurements of Supported Catalysts Using a Potentiostat/Galvanostat

Overview

Source: Laboratory of Dr. Yuriy Román — Massachusetts Institute of Technology

A potentiostat/galvanostat (often referred to as simply a potentiostat) is an instrument that measures current at an applied potential (potentiostatic operation) or measures potential at an applied current (galvanostatic operation) (Figure 1). It is the most commonly used instrument in the electrochemical characterization of anode and cathode materials for fuel cells, electrolyzers, batteries, and supercapacitors.

Conventionally, these anode and cathode materials are interfaced with a potentiostat via a three-electrode electrochemical cell. The electrode leads from the potentiostat are connected to the reference electrode, the counter electrode (often called the auxiliary electrode), and the working electrode (which contains the test material of interest). The electrochemical cell is then filled with a high ionic strength electrolyte solution, such as an acidic, alkaline, or salt solution. The media for this high ionic strength solution is typically aqueous; however, for applications necessitating higher operating cell potential windows, such as batteries and supercapacitors, non-aqueous media is often used. The cell media is degassed with an inert gas (to prevent unwanted side reactions) or with a test gas (if the test reaction involves a gas at one of the electrodes).

Alternatively, a salt bridge or membrane is employed to maintain ionic contact if the two half cells are to be measured in different electrolytes. In heterogeneous electrocatalysis, this type of "two compartment" cell is often used if the test molecule at the working electrode is also reactive at the counterelectrode. This happens frequently as the counterelectrode typically employed is platinum, which is a highly active catalyst for many reactions. Here, single compartment cells will be used, where all three electrodes are in the same media.

This video will explain the process of polishing a working electrode, preparing a catalyst ink, mounting the catalyst ink onto the working electrode, preparing the electrochemical cell, and then performing electrochemical measurements. The measurements that are performed include: cyclic voltammetry (CV), linear sweep voltammetry (LSV), chronopotentiometry (CP), and chronoamperometry (CA).

Figure 1
Figure 1. An example of a single compartment electrochemical cell. a.) Teflon cap, b.) glass cell, c.) Pt wire counter electrode, d.) working electrode, e.) Ag/AgCl reference electrode, f.) 0.5 M aqueous sulfuric acid electrolyte solution.

Procedure

1. Catalyst Ink and Working Electrode Preparation

Safety Precautions: Metals supported on carbon black must be handled in a fume hood or balance enclosure until it is in suspension form as these powders are inhalation hazards.

  1. Using an enclosed balance, weigh out 5–10 mg of metal/carbon black catalyst and add to a glass vial with a cap.
  2. Using a micropipette, dilute the catalyst with water such that the final concentration is 7.5 mg of catalyst per m

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Results

This procedure will result in figures containing plots of measured current vs. potential for each of the four techniques. By convention for CV and LSV, the plots will also be outputted as measured current vs. potential despite the reality that these are transient techniques that measure current vs. the time derivative of potential.

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Application and Summary

CV, LSV, CP, and CA are indispensable techniques for determining the efficacy of new electrode materials for fuel cells, electrolyzers, batteries, and supercapacitors as well as for developing fields such as the selective partial oxidation or reduction of commodity chemicals. These methods allow for determining overpotentials of reactions on different electrode materials as compared to their thermodynamic equilibrium potentials. These methods also allow the volumetric or gravimetric capacitance of supercapacitors to be d

Log in or to access full content. Learn more about your institution’s access to JoVE content here

Tags
Electrochemical MeasurementsSupported CatalystsPotentiostatGalvanostatElectrochemical CharacterizationElectrical ChangesRedox ReactionsOxidation reductionElectrode SurfaceTransfer Of ElectronsApplied PotentialThree electrode CellWorking ElectrodeCounter ElectrodeReference Electrode

跳至...

0:00

Overview

1:09

Principles of Voltammetry

4:36

Working Electrode Preparation

5:51

Electrochemical Cell Preparation

6:53

Electrochemical Analysis

8:38

Applications

10:17

Summary

此集合中的视频:

article

Now Playing

Electrochemical Measurements of Supported Catalysts Using a Potentiostat/Galvanostat

Analytical Chemistry

50.9K Views

article

Sample Preparation for Analytical Characterization

Analytical Chemistry

82.4K Views

article

Internal Standards

Analytical Chemistry

201.7K Views

article

Method of Standard Addition

Analytical Chemistry

316.6K Views

article

Calibration Curves

Analytical Chemistry

779.7K Views

article

Ultraviolet-Visible (UV-Vis) Spectroscopy

Analytical Chemistry

608.7K Views

article

Raman Spectroscopy for Chemical Analysis

Analytical Chemistry

50.1K Views

article

X-ray Fluorescence (XRF)

Analytical Chemistry

25.1K Views

article

Gas Chromatography (GC) with Flame-Ionization Detection

Analytical Chemistry

276.2K Views

article

High-Performance Liquid Chromatography (HPLC)

Analytical Chemistry

377.0K Views

article

Ion-Exchange Chromatography

Analytical Chemistry

257.6K Views

article

Capillary Electrophoresis (CE)

Analytical Chemistry

91.8K Views

article

Introduction to Mass Spectrometry

Analytical Chemistry

109.9K Views

article

Scanning Electron Microscopy (SEM)

Analytical Chemistry

85.8K Views

article

Cyclic Voltammetry (CV)

Analytical Chemistry

122.2K Views

JoVE Logo

政策

使用条款

隐私

科研

教育

关于 JoVE

版权所属 © 2025 MyJoVE 公司版权所有,本公司不涉及任何医疗业务和医疗服务。