Zaloguj się

Aby wyświetlić tę treść, wymagana jest subskrypcja JoVE. Zaloguj się lub rozpocznij bezpłatny okres próbny.

W tym Artykule

  • Podsumowanie
  • Streszczenie
  • Wprowadzenie
  • Protokół
  • Wyniki
  • Dyskusje
  • Ujawnienia
  • Podziękowania
  • Materiały
  • Odniesienia
  • Przedruki i uprawnienia

Podsumowanie

Dual camera emission splitting systems for two-color fluorescence microscopy generate real-time image sequences with exceptional optical and temporal resolution, a requirement of certain live cell assays including parallel plate flow chamber adhesion assays. When software is employed to merge images from simultaneously acquired emission channels, pseudocolored image sequences are produced.

Streszczenie

Multi-color immunofluorescence microscopy to detect specific molecules in the cell membrane can be coupled with parallel plate flow chamber assays to investigate mechanisms governing cell adhesion under dynamic flow conditions. For instance, cancer cells labeled with multiple fluorophores can be perfused over a potentially reactive substrate to model mechanisms of cancer metastasis. However, multi-channel single camera systems and color cameras exhibit shortcomings in image acquisition for real-time live cell analysis. To overcome these limitations, we used a dual camera emission splitting system to simultaneously capture real-time image sequences of fluorescently labeled cells in the flow chamber. Dual camera emission splitting systems filter defined wavelength ranges into two monochrome CCD cameras, thereby simultaneously capturing two spatially identical but fluorophore-specific images. Subsequently, psuedocolored one-channel images are combined into a single real-time merged sequence that can reveal multiple target molecules on cells moving rapidly across a region of interest.

Wprowadzenie

Methods for analysis of molecules on the cell surface, such as immunostaining, employ probes that are chemically conjugated to fluorophores, allowing detection of target molecules. Live cell imaging and hydrodynamic flow-based cell adhesion assays are typically recorded with monochrome CCD cameras designed to capture physiological processes at the cellular and/or molecular level 1, 2. These cameras are highly sensitive, deliver fast frame rates (greater than 30 frames per second), and provide exceptional temporal resolution (due to fast frame rates and short exposure times). However, monochrome cameras can only capture a single emission channel (detecting a single fluorophore) to collect images. Single camera emission splitting systems can be incorporated to capture multiple emission channels but often reduce the field of view and require the same exposure time for imaging all channels. To capture the full color spectrum from cells labeled with multiple fluorophores, a color camera can be used as an alternative. However, color cameras are not generally capable of providing the temporal resolution desired for live cell imaging in certain applications. Another imaging device is needed for applications in which it is advantageous to image live cells at multiple wavelengths while retaining a high temporal resolution. A prime experimental application is the parallel plate flow chamber adhesion assay, in which cells are perfused at physiologically relevant conditions over a potentially reactive substrate 1, 3. Cells in flow that express specific cell surface molecules may adhere and roll on the substrate, such as a cell monolayer expressing adhesion molecules or surface-adsorbed extracellular matrix proteins 4, 5. Rolling cells may undergo rotational and translational movement in fractions of a second. Molecular features on rolling and adherent cells, such as clusters of cell surface molecules, also have the potential to undergo active reorganization on the cell surface. Thus, imaging systems must provide an exceptional temporal resolution (30 frames per second or greater and "near zero" exposure times) to generate an image sequence that illustrates the step-by-step progression of cell rolling 6, 7. Dual camera emission splitting systems are capable of meeting these demands for imaging cells labeled with multiple fluorophores.

Dual camera emission splitting systems split and filter fluorescence channels into two similar cameras to simultaneously capture two spatially identical but fluorophore-specific images while retaining the full field of view. This technology enables direct comparison of the image captured in real-time in each channel and allows the user to quickly switch between camera models with different imaging capabilities. This feature is useful for making adjustments to image capture settings in one camera that better allow the system to capture fluorophores with different intensities, lifetimes, and extinction coefficients 8. Coupled with imaging software, dual camera emission splitting systems allow the real-time recording of live cell imaging assays in multiple wavelengths and may enhance in vitro assays that use fluorescence to study cell behavior.

Protokół

1. Software Installation

  1. Purchase StreamPix 5 Multi-Camera software with SimulPix module or other imaging software capable of collecting and combining images captured by monochrome CCD cameras.
  2. Minimum requirements for StreamPix 5 include: A PC equipped with Core 2 Duo with 2.4 GHz or better with 2 GB RAM or higher. A supported IEEE digital or analog camera and compatible frame grabber. Windows XP/7/Vista 32 or 64 bit. A monitor supporting resolution 1,024 x 768 or higher. A graphic adapter with good 2D performance (OCU express 16x or better is recommended) and 7,200 RPM hard disk drive for recording with RAID-O configuration.

2. Installation of Dual Camera System Capable of Two Color Simultaneous Real-time Imaging

  1. Connect the DC2 Photometrics emission splitter, or similar dual camera emission splitting device, to the microscope by sliding the DC2 C-mount adapter into the video port of the microscope such that the housing of the dichroic on the device is accessible.
  2. Insert two monochrome CCD cameras into female C-mount 1 and female C-mount 2. Identical cameras are preferable. Similar cameras may be used, but compatibility is dependent on internal hardware, most importantly the CCD image sensor.
  3. Use imaging software to display images from both cameras. The following steps (2.3.1 - 2.3.4) apply to StreamPix 5 with SimulPix module and must be adapted for other multi-camera imaging software.
    1. Open StreamPix 5 and select "workspace" in the task ribbon.
    2. Open "workspace manager" located at far right of the screen and select "new workspace".
    3. After naming a camera, follow software prompts to select a grabber from a pre-populated list. Select the grabber matching the camera model and repeat for the second camera workspace. For the merged workspace, repeat once more but "all cameras are in use" error message will appear. This message is normal.
    4. Once the merged workspace is loaded, assign cameras from workspace 1 and workspace 2 to the merged workspace in the docking panel located on the right-hand side of the viewing screen.
  4. Align cameras in the dual camera emission splitting system.
    1. Align cameras in bright field or phase contrast with the calibration grid provided by the manufacturer using a PlanApo 40X objective (or greater).
    2. Send light to the microscope eyepieces, place the metallic-coated calibration grid that was provided by the manufacturer on the microscope stage, and square the grid on the microscope stage.
    3. Bring the grid into focus.
    4. Displace, but do not remove, the dichroic housing to align camera 1 (bypass camera). Display the grid without binning and without autoscaling. Focus the image using the focus adjustment dial on C-mount port 1.
    5. Push the dichroic housing into the dual channel acquisition device fully, so that the image is split by the dichroic to both cameras.
    6. Loosen the three 5/64" set screws and rotate the second camera on the female C-mount 2 until the orientation of the grid is identical in both images. Using the focus adjustment dial on C-mount port 2, bring the image from camera 2 into focus.
    7. Finalize the alignment using the combined image in the merged workspace of the imaging software. This allows one to see a real-time pseudocolor overlay of the two images of the calibration grid. Adjust the combined image positions using only the R/L knob on the right side of the DC2. Adjust this knob until right/left vertical boundary of the images is precisely aligned.
    8. Use the U/D knob to adjust the up/down alignment of the second image until the horizontal grid bars are properly aligned.
    9. If during the up/down alignment a slight camera rotation occurs, correct this issue by loosening the three 5/64" inch screws for camera 2 and rotate until the displayed image is properly aligned.

3. Preparation of Cancer Cells for Parallel Plate Flow Chamber Adhesion Assay

  1. Culture BT-20 breast cancer cells in minimum essential medium (MEM) supplemented to complete medium with 10% fetal bovine serum (FBS) and 1% penicillin-streptomycin at 37 °C and 5.0% CO2 as previously described 2.
  2. Harvest cancer cells with a dilute trypsin treatment and count cells with a hemocytometer.
  3. Wash and suspend cells at 107 cells/ml in 0.1% albumin from bovine serum (BSA) in DPBS with calcium and magnesium (DPBS+).
  4. Dilute primary antibodies, purified mouse anti-human CD24 and purified rat HECA-452 (anti-sialofucosylated antigens), to a predetermined optimal concentration in 0.1% BSA DPBS+. Incubate cells for 30 min 9, 10.
  5. Centrifuge cells at 1,200 rpm to obtain a cell pellet. Aspirate primary antibody, and wash cells twice with 0.1% BSA DBPS+.
  6. Dilute secondary antibodies, goat anti-mouse IgG AlexaFluor 568 (emission max, 603 nm), and goat anti-rat IgM AlexaFluor 488 (emission max, 519 nm), to a predetermined optimal concentration in 0.1% BSA DPBS+. Incubate cells for 30 min.
  7. Wash cells twice and suspend in 0.1% BSA DPBS+ at 106 cells/ml in 0.1% BSA DPBS+.

4. Preparation of Reactive Substrate for Parallel Plate Flow Chamber Adhesion Assay

  1. Connect two separate lengths of tubing to the inlet and outlet ports of the parallel plate flow chamber that fits inside a 35 mm tissue culture dish. One tube will connect to the reservoir of labeled cells suspended in 0.1% BSA DPBS+ and the other to the syringe pump, which will withdraw fluid at a specified flow rate.
  2. Connect a vacuum line to the flow chamber 1.
  3. Position the flow chamber over the 35 mm tissue culture dish containing a monolayer of Chinese hamster ovary cells expressing E-selectin (CHO-E). CHO-E cells were grown in MEM complete medium at 37 °C and 5.0% CO2 2.
  4. Adjust the flow chamber and/or flow chamber gasket to ensure an adequate vacuum seal 1.
  5. Withdraw fluid from reservoir, monitoring the flow chamber assembly for proper seal and no visual evidence of air bubble formation 1.
  6. Place flow chamber assembly on microscope (Leica DMI 6000B)1.

5. Two-color Image Acquisition

  1. Power a Leica EL-6000 compact light source, or similar device, to generate high intensity light and couple it into a light guide. Use an inverted microscope to pass this light through a combination filter cube, i.e. G/R (green/red) filter cube, to excite fluorophores of a desired color/intensity.
  2. Ensure dichroic housing is in correct operating position (depressed) and is not in bypass mode.
  3. Operate microscope in fluorescence mode and select proper fluorescence filter cube (green/red).
  4. Determine the exposure time and gain in camera 1 (red; 620±60 nm) and camera 2 (green; 535±40 nm).
    1. Depress the dichroic to obtain images in both the red and green emission channels. Use a flow chamber assay with cells labeled with the red fluorophore only, and obtain an image in the red channel by adjusting the exposure time of camera 1 in "live settings" of the imaging software.
    2. Match the exposure time of camera 2 to the exposure time of camera 1. If an image is observed in the green channel, bleed-through artifacts are present and exposure time in both cameras needs to be reduced.
    3. Keeping the exposure time equivalent in camera 1 and camera 2, reduce the exposure time until the bleed-through artifact is no longer visible in the green channel. Adjust the gain of camera 1 to improve image visibility in the red channel if necessary.
    4. Repeat steps 5.4.1 - 5.4.3 with cells labeled with the green fluorophore only, but define exposure time with camera 2 to image cells in the green channel without bleed-through artifacts in the red channel collected by camera 1.
    5. Retitrate antibodies if bleed-through artifacts cannot be eliminated 11, or if exposure times of camera 1 and camera 2 are substantially different such that merged images may be temporally misaligned.
  5. Use imaging software to view images simultaneously captured from two monochrome CCD cameras in the flow chamber experiment.
  6. Merge images collected from both cameras using the SimulPix module of Streampix 5 or an alternative software package.
  7. Use digital correction adjustments in SimulPix or alternative software to spatially align the merged images, if necessary. Images can be corrected with the SimulPix module for horizontal, vertical, and rotational adjustments.

Wyniki

A parallel plate flow chamber adhesion assay was used to demonstrate a dual camera emission splitting system that simultaneously captured real-time image sequences in two emission channels (Figure 1). The dual camera emission splitting system detected BT-20 cells that were fluorescently labeled with anti-human CD24 and HECA-452 (detecting sialofucosylated antigens) monoclonal antibodies and appropriate secondary antibodies (Figure 2). Some rolling cells displayed red signals (CD24) yet u...

Dyskusje

The dual camera emission splitting system has the spatial, temporal, and optical resolution needed to capture high quality images in applications where cell or molecular movement is rapid. In generating the representative results, parameters in the dual camera emission splitting system, including software settings and camera settings, were optimized to obtain a merged image sequence in which rolling and adherent cells were spatially and temporally aligned. This optimization step is critical, as misalignment can result in...

Ujawnienia

The authors declare that they have no competing financial interests.

Podziękowania

The authors wish to thank Dr. Douglas Goetz (Department of Chemical and Biomolecular Engineering, Ohio University) and Dr. Fabian Benencia (Department of Biomedical Sciences, Ohio University) for insightful discussions and manuscript review. We also thank Dr. Christopher Huppenbauer for helpful technical discussions (W. Nuhsbaum Inc.). This work was supported by grants from the National Science Foundation (CBET-1106118) and the National Institutes of Health (1R15CA161830-01).

Materiały

NameCompanyCatalog NumberComments
Reagent/Material
BT-20 cellsATCCHTB-19
CHO-E cellsGift from Dr. R. Sackstein (Brigham and Women's Hospital, Harvard Medical School)
MEM Thermo ScientificSH30024.01
FBSThermo ScientificSH30396.03
Penicillin-streptomycinThermo ScientificSV30010
0.25% Trypsin / 0.1% EDTAThermo ScientificSV30031.01
DPBSThermo Scientific SH30028.02
DPBS+Life Technologies14080-055
BSASigmaA9647
HECA-452 monoclonal antibodyBD Biosciences555946
Anti-human CD24 monoclonal antibodyBD Biosciences555426
Anti-rat IgM AlexaFluor 488InvitrogenA21212
Anti-mouse IgG AlexaFluor 568InvitrogenA11004
[header]
Equipment
EXi Blue Fluorescence Microscopy Digital CCD CameraQ ImagingEXI-BLU-R-F-M-14-C
Retiga EXi FAST 1394 Digital CCD CameraQ ImagingRET-EXi-F-M-12-C
DC2 Emission SplitterPhotometricsDC2
Inverted Fluorescence MicroscopeLeica DMI6000 B
Streampix 5 softwareNorpix

Odniesienia

  1. Wiese, G., Barthel, S. R., Dimitroff, C. J. Analysis of physiologic E-selectin-mediated leukocyte rolling on microvascular endothelium. J. Vis. Exp. (24), e1009 (2009).
  2. Shirure, V. S., Reynolds, N. M., Burdick, M. M. Mac-2 binding protein is a novel e-selectin ligand expressed by breast cancer cells. PLoS One. 7 (9), e44529 (2012).
  3. Burdick, M. M., McCaffery, J. M., Kim, Y. S., Bochner, B. S., Colon Konstantopoulos, K. carcinoma cell glycolipids, integrins, and other glycoproteins mediate adhesion to HUVECs under flow. Am. J. Physiol. Cell Physiol. 284 (4), C977-C987 (2003).
  4. Resto, V. A., Burdick, M. M., Dagia, N. M., McCammon, S. D., Fennewald, S. M., Sackstein, R. L-selectin-mediated lymphocyte-cancer cell interactions under low fluid shear conditions. J. Biol. Chem. 283 (23), 15816-15824 (2008).
  5. Shirure, V. S., Henson, K. A., Schnaar, R. L., Nimrichter, L., Burdick, M. M. Gangliosides expressed on breast cancer cells are E-selectin ligands. Biochem Biophys. Res. Commun. 406 (3), 423-429 (2011).
  6. Tees, D. F., Goetz, D. J. Leukocyte adhesion: an exquisite balance of hydrodynamic and molecular forces. News Physiol. Sci. 18, 186-190 (2003).
  7. Chang, K. C., Tees, D. F., Hammer, D. A. The state diagram for cell adhesion under flow: leukocyte rolling and firm adhesion. Proc. Natl. Acad. Sci. U.S.A. 97 (21), 11262-11267 (2000).
  8. Lichtman, J. W., Conchello, J. A. Fluorescence microscopy. Nat. Methods. 2 (12), 910-919 (2005).
  9. Kummitha, C. M., Shirure, V. S., Delgadillo, L. F., Deosarkar, S. P., Tees, D. F., Burdick, M. M., Goetz, D. J. HECA-452 is a non-function blocking antibody for isolated sialyl Lewis x adhesion to endothelial expressed E-selectin under flow conditions. J. Immunol. Methods. 384 (1-2), 43-50 (2012).
  10. Burdick, M. M., Henson, K. A., Delgadillo, L. F., Choi, Y. E., Goetz, D. J., Tees, D. F., Benencia, F. Expression of E-selectin ligands on circulating tumor cells: cross-regulation with cancer stem cell regulatory pathways?. Front. Oncol. 2, 103 (2012).
  11. Hoffman, G. E., Le, W. W., Sita, L. V. The Importance of Titrating Antibodies for Immunocytochemical Methods. Current Protocols in Neuroscience. , (2001).
  12. Weinberg, S., Lipke, E. A., Tung, L. In vitro electrophysiological mapping of stem cells. Methods Mol. Biol. 660, 215-237 (2010).
  13. Kong, K. V., Leong, W. K., Lim, L. H. Osmium carbonyl clusters containing labile ligands hyperstabilize microtubules. Chem. Res. Toxicol. 22 (6), 1116-1122 (2009).
  14. Vermot, J., Fraser, S. E., Liebling, M. Fast fluorescence microscopy for imaging the dynamics of embryonic development. HFSP J. 2 (3), 143-155 (2008).
  15. Bullen, A., Friedman, R. S., Krummel, M. F. Two-photon imaging of the immune system: a custom technology platform for high-speed, multicolor tissue imaging of immune responses. Curr. Top. Microbiol. Immunol. 334, 1-29 (2009).
  16. Worth, D. C., Parsons, M. Advances in imaging cell-matrix adhesions. J. Cell Sci. 123, 3629-3638 (2010).
  17. Xiao, Z., Visentin, G. P., Dayananda, K. M., Neelamegham, S. Immune complexes formed following the binding of anti-platelet factor 4 (CXCL4) antibodies to CXCL4 stimulate human neutrophil activation and cell adhesion. Blood. 112 (4), 1091-1100 (2008).

Przedruki i uprawnienia

Zapytaj o uprawnienia na użycie tekstu lub obrazów z tego artykułu JoVE

Zapytaj o uprawnienia

Przeglądaj więcej artyków

Dual Camera Emission Splitting SystemMulti color Immunofluorescence MicroscopyParallel Plate Flow ChamberReal time Live Cell AnalysisFluorescently Labeled CellsCell AdhesionCancer Metastasis

This article has been published

Video Coming Soon

JoVE Logo

Prywatność

Warunki Korzystania

Zasady

Badania

Edukacja

O JoVE

Copyright © 2025 MyJoVE Corporation. Wszelkie prawa zastrzeżone