JoVE Logo

Sign In

A subscription to JoVE is required to view this content. Sign in or start your free trial.

In This Article

  • Summary
  • Abstract
  • Introduction
  • Protocol
  • Results
  • Discussion
  • Disclosures
  • Acknowledgements
  • Materials
  • References
  • Reprints and Permissions

Summary

We present a detailed calibration protocol for a commercial orthogonal superposition rheology technique using Newtonian fluids including end-effect correction factor determination methods and recommendations for best practices to reduce experimental error.

Abstract

Orthogonal superposition (OSP) rheology is an advanced rheological technique that involves superimposing a small-amplitude oscillatory shear deformation orthogonal to a primary shear flow. This technique allows the measurement of structural dynamics of complex fluids under non-linear flow conditions, which is important for the understanding and prediction of the performance of a wide range of complex fluids. The OSP rheological technique has a long history of development since the 1960s, mainly through the custom-built devices that highlighted the power of this technique. The OSP technique is now commercially available to the rheology community. Given the complicated design of the OSP geometry and the non-ideal flow field, users should understand the magnitude and sources of measurement error. This study presents calibration procedures using Newtonian fluids that includes recommendations for best practices to reduce measurement errors. Specifically, detailed information on the end-effect factor determination method, sample filling procedure, and identification of the appropriate measurement range (e.g., shear rate, frequency, etc.) are provided.

Introduction

Understanding the rheological properties of complex fluids is essential to many industries for the development and manufacture of reliable and reproducible products1. These “complex fluids” include suspensions, polymeric liquids, and foams that widely exist in our everyday life, for example, in personal care products, foods, cosmetics, and household products. The rheological or flow properties (e.g., viscosity) are key quantities of interest in establishing performance metrics for end use and processability, but flow properties are interconnected with the microstructures that exist within complex fluids. One prominent characteristic of complex fluids that distinguishes them from simple liquids is that they possess diverse microstructures spanning multiple length scales2. Those microstructures can be easily affected by different flow conditions, which, in turn, result in changes in their macroscopic properties. Unlocking this structure-property loop via non-linear viscoelastic behavior of complex fluids in response to flow and deformation remains a challenging task for experimental rheologists.

Orthogonal superposition (OSP) rheology3 is a robust technique to address this measurement challenge. In this technique, a small amplitude oscillatory shear flow is superimposed orthogonally to a unidirectional primary steady-shear flow, which enables the simultaneous measurement of a viscoelastic relaxation spectrum under the imposed primary shear flow. To be more specific, the small oscillatory shear perturbation can be analyzed using theories in linear viscoelasticity4, while the non-linear flow condition is achieved by the primary steady-shear flow. As the two flow fields are orthogonal and thus not coupled, the perturbation spectra can be directly related to the variation of the microstructure under the primary non-linear flow5. This advanced measurement technique offers an opportunity to elucidate structure-property-processing relationships in complex fluids to optimize their formulation, processing, and application.

The implementation of modern OSP rheology was not the result of a sudden epiphany; rather, it is based on many decades of development of custom devices. The first custom made OSP apparatus is dated back to 1966 by Simmons6, and many efforts were made thereafter7,8,9,10. Those early custom-built devices suffer from many drawbacks such as alignment issues, the pumping flow effect (due to the axial movement of the bob to provide orthogonal oscillation), and limits to instrument sensitivity. In 1997, Vermant et al.3 modified the force rebalance transducer (FRT) on a commercial separate motor-transducer rheometer, which enabled OSP measurements for fluids with a wider viscosity range than previous devices. This modification enables the normal force rebalance transducer to function as a stress-controlled rheometer, imposing an axial oscillation in addition to a measurement of the axial force. Recently, the geometries required for OSP measurements, after the methodology by Vermant, have been released for a commercial separate motor-transducer rheometer.

Since the advent of commercial OSP rheology, there is a growing interest in applying this technique for the investigation of various complex fluids. Examples include colloidal suspensions11,12, colloidal gels13,14 and glasses15,16,17. While the availability of the commercial instrument promotes OSP research, the complicated OSP geometry requires a deeper understanding of the measurement than other routine rheological techniques. The OSP flow cell is based on a double-wall concentric cylinder (or Couette) geometry. It features an open top and open bottom design to enable fluid to flow back and forth between the annular gaps and the reservoir. Despite the optimization made to the geometry design by the manufacturer, when undergoing OSP operation the fluid experiences an inhomogeneous flow field, geometric end effects, and residual pumping flow, all of which can introduce substantial experimental error. Our previous work18 reported important end-effect correction procedures using Newtonian fluids for this technique. To obtain correct viscosity results, appropriate end-effect factors in both primary and orthogonal directions should be applied. In this protocol, we aim to present a detailed calibration methodology for the OSP rheological technique and provide recommendations for best practices to reduce measurement errors. The procedures delineated in this paper on OSP geometry setup, sample loading, and OSP test settings should be easily adoptable and translated for non-Newtonian fluids measurements. We advise that users utilize the calibration procedures described here to determine the end-effect correction factors for their applications prior to OSP measurements on any fluid classification (Newtonian or Non-Newtonian). We note that the calibration procedures for end factors have not been reported previously. The protocol provided in the present paper also describes step-by-step guide and tips on how to perform accurate rheological measurements in general and the technical resource on the understanding of “raw” data versus “measured” data, which may be overlooked by rheometer users.

Protocol

1. Rheometer setup

NOTE: The protocol in this section describes basic steps to run a rheology experiment (for either a separate motor-transducer rheometer or a combined motor-transducer rheometer), including preparation of the setup, installation of appropriate geometry, loading the test material, setting up the experiment procedure, specifying the geometry, and starting the test. Specific instructions and notes are provided for OSP operation. To minimize thermal gradients in the transducer, it is recommended to power the rheometer for at least 30 min prior to the operation. The rheometer software used in this protocol for instrument control and data collection are noted in the Table of Materials. See Table 1 for rheometer specifications.

  1. Before setting up the rheometer, enable the Orthogonal Superposition feature in the rheometer software. Install a lower platinum resistance-thermometer (PRT) on the test station for temperature measurement and an environmental control device.
    NOTE: Lift the stage to maximum height for the installation process (Figure 1a). Install proper PRT before mounting the environmental control device. Be careful to not hit the PRT with the environmental control device during installation. Use the supplied spanner wrench to secure the environmental control device on the test station.
  2. Install the double wall concentric cylinder geometry.
    1. Assemble the inner and outer cylinders (Figure 1b) properly to complete the double-wall cup configuration.
      NOTE: Before assembling the cup, check the O-ring condition (for cracks, swelling or other damage) on the inner cylinder and replace if needed.
    2. Insert the cup into the environmental control device and align the geometry properly.
    3. Press the lower geometry (cup) downward to compress the spring-loaded PRT while tightening the thumbscrew using a torque screwdriver (0.56 N m fixed).
      NOTE: To check if the lower geometry is correctly installed, disable the motor power, and use a finger to spin the geometry. If the lower geometry spins freely in the environment control device, then it is installed properly and continue the next step. If it does not spin freely, remove the components from the test station in reverse order of the previous steps and then re-install the lower geometry. Verify that the temperature signal is received from the lower PRT. The rheometer should automatically recognize the temperature sensor by default; if not, select the lower PRT as the Temperature Control Sensor source in the temperature control options from the rheometer software.
    4. Install the upper geometry (bob) on the transducer shaft. Tare the normal force and torque by clicking the Tare Transducer button in the transducer control panel from the rheometer software or using Tare Torque and Tare Normal on the Instrument tab from the instrument touch screen. A picture of the complete rheometer setup is shown in Figure 1c.
    5. Zero the gap between the upper and lower geometries by clicking the Zero Fixture button in the gap control panel from either the rheometer software or from the instrument touch screen. Perform geometry mass calibration if needed.
      NOTE: Check the geometry documentation supplied by the manufacturer to see if the upper tool mass value is available. If not, perform geometry mass calibration at the end of this step. Follow the on-screen instructions to perform the upper tool mass calibration. Upon completion, confirm that the correct new fixture mass is accepted.

2. Loading the test material

  1. Lift the stage to provide enough workspace to load the test material into the cup.
  2. Use a pipette or a spatula to load the test material into the cup. Carefully handle the test material to minimize entrainment of air into the fluid.
    NOTE: For loading a low viscosity test material (e.g., less than 5 Pa s), use an adjustable volume pipette (Figure 2a). The minimum volume to fill the geometry can be found in the Geometry information under the Experiment panel in the rheometer software. Approximate volumes needed for the currently available OSP geometries, viz., 0.5 mm and 1.0 mm annular gap width, are 32 mL and 36 mL, respectively. For loading a higher viscosity test material (e.g., higher than 5 Pa s), use a spatula or a positive-displacement pipette (Figure 2b). Since precise volume control for a highly viscous liquid is difficult, fine adjustment based on the fluid volume is not recommended for loading a high viscosity liquid. In any case, it is expected to slightly underfill rather than overfill in this step. Follow the next step to ensure precise loading of material.
  3. Lower the bob into the cup to the geometry gap set point and lifted out to determine the fluid level in the loaded geometry. The goal is to achieve a fluid contact line that is slightly (approximately 2 mm) above the lower edge of the bob upper opening.
    NOTE: This process may require lengthy wait times to reach the desired fluid level because of the small annular gap width of the geometry and the relatively large volume of sample needed. The wait times mainly depend on the viscosity of the test material. For example, a highly viscous liquid takes longer to flow into the gaps between the cylinders and completely wet the bob surfaces.
  4. Lower the upper geometry carefully into the fluid to reach the geometry gap set point of 8 mm. This process is illustrated as Step 1 in Figure 2c. Wait a few minutes while the bob is being kept at position (iii) where the gap is set to 8 mm.
    NOTE: When the bob end surface contacts the fluid, reduce the downward velocity of the bob. For a high viscosity liquid or yield-stress fluid, closely monitor the normal force readings to prevent the transducer from overloading during this process.
  5. Lift the bob vertically using the slow slew speed of the instrument to a position where the wetted fluid contact line can be visually inspected (Figure 3). The contact line indicates the fluid level in the geometry at the gap set point. If the line on the bob is below the upper end of the bob (lower rim of the upper opening on the bob), it indicates that the fluid height is lower than the inner cylinder height and additional test material should be added to the geometry.
  6. Carefully lift the bob to the previous loading position to allow for enough workspace (Step 2 in Figure 2c) and load an additional amount of test material into the cup as needed. Slowly move the bob up or down to avoid cavitation. Add the test material carefully to prevent introducing additional air bubbles.
  7. Lower the upper geometry into the fluid and set to the final geometry gap again. Repeat Steps 1 and 2 (Figure 2c) until the wetted contact line on the bob is approximately 2 mm above the lower rim of the upper bob opening as shown in Figure 3a. Also check that the lower rim of the upper opening on the bob is properly wetted (Figure 3b). Move the bob to the geometry gap set point and allow the test material to relax.
    NOTE: The wait time depends on the viscosity of the standard material. For example, for a 1 Pa s liquid, a wait time of 15 min is sufficient; whereas for a 100 Pa s liquid, a much longer wait time (4 h) is needed. This process is illustrated as Step 3 in Figure 2c. The complete sample loading procedure is illustrated in Figure 2. High viscosity fluids require extended time and are difficult to load. To reduce the wait time, increasing the temperature by a few degrees may be helpful to facilitate the viscous calibration liquid to flow.

3. Running viscosity calibration measurements

NOTE: The calibration protocols provided in this paper are specific to the end-effect factors applied for the OSP technique. This does not include routine calibrations or verification checks, including torque and normal force calibrations, phase angle check, PDMS check, etc. that are recommended by individual rheometer manufacturers. Those procedures should be performed prior to the calibration protocols in the present paper. The readers should refer to the User Manual of the rheometer manufacturer for the procedures of performing routine calibrations or checks. The silicone viscosity standards used in this protocol are noted in the Table of Materials.

  1. Specify the geometry
    NOTE: Before setting up the experiment, make sure the correct geometry is selected in the rheometer software. For first-time use, create a new orthogonal double wall concentric cylinder geometry in the rheometer software following the steps below.
    1. Add a new orthogonal double wall concentric cylinder geometry.
    2. Enter the dimensions for the geometry as shown in Table 2.
      NOTE: The numbers and their corresponding symbols are inscribed on the bob and cup. The operating gap is 8 mm for the experimental geometry used here but should be specified by the manufacturer. Therefore, the inner cylinder height is equal to (immersed height + 8 mm).
  2. Specify the geometry constants. Fill in the fields of geometry inertia and geometry mass with proper values. Enter 1.00 for both the end-effect factor and orthogonal end-effect factor.
    NOTE: The geometry inertia for the 0.5 mm and 1.0 mm gap OSP geometries specified by the manufacturer are 15.5 µN m s2 and 10.3 µN m s2, respectively. Make sure that the correct value for the upper geometry mass is entered. This value can be found in the geometry documentation provided by the manufacturer. Alternatively, perform geometry mass calibration under the geometry calibration tab (Protocol step 1.2.5) and confirm that the correct new fixture mass is applied. The default end-effect factor (CL) is 1.065 and the orthogonal end-effect factor (CLo) is 1.04. Change both fields to 1.00. The stress constants are automatically calculated from the dimensions and end-effect factors. The strain constants are determined by the geometry dimensions only (expressions are provided in previous work18). The definitions of the dimensions are described in Table 2 and indicated in Figure 4. The expressions for the (primary) stress constant, Kτ, and orthogonal (linear) stress constant, Kτο, are:
    figure-protocol-11267
    figure-protocol-11344

4. Steady shear rate sweep tests

NOTE: Viscosity calibration measurements are performed independently in either the primary direction or the orthogonal direction to calibrate CL or CLo. For the primary direction, steady shear viscosity is measured by performing shear rate sweep tests. For the orthogonal direction, dynamic complex viscosity is measured by performing orthogonal frequency sweep tests.

  1. Condition the sample at 25 °C for 15 min to allow the test material to reach thermal equilibrium.
    NOTE: The calibration measurements are performed at the temperature at which the certified viscosity of the standard liquid is reported, i.e., 25 °C. The readers may use a different test temperature suitable for their Newtonian standard liquids. An equilibration time or soak time, i.e., 15 min, is recommended to ensure that the environmental control device, geometries, and sample to reach thermal equilibrium.
  2. Select the Flow Sweep Test under the Experiment Procedure in the rheometer software. Set the test temperature to 25 °C under Environment Control.
  3. Specify the shear rate range from 0.01 s−1 to 100.0 s−1 with data recording at 10 points per decade logarithmically. Enable automatic steady-state determination.
    NOTE: The shear rate range used here is based on the instrument torque sensitivity limits (Table 1) and the measuring liquid. For example, for a higher viscosity liquid (e.g., 300 Pa s), a lower shear rate range of 10−4 s−1 to 1 s−1 may be used, and vice versa.
  4. Start the experiment from the rheometer software.

5. Orthogonal frequency sweep tests

  1. Set the normal force transducer to FRT mode from the transducer control panel in the rheometer software.
    NOTE: The default transducer setting for the normal force transducer is spring mode for this separate motor-transducer rheometer. In the OSP operation, the normal force transducer operates as a stress-controlled or a combined motor-transducer rheometer to apply axial deformation while measuring the axial force simultaneously. The normal force transducer must be set in FRT mode to perform OSP tests.
  2. Condition the sample at 25 °C for 15 min to ensure thermal equilibration.
  3. Select the Orthogonal Frequency Sweep test under the Experiment Procedure in the rheometer software. Set the test temperature to 25 °C.
  4. Specify the desired normal strain and enter 0.0 s−1 for the shear rate in the rotational direction.
    NOTE: The maximum normal strain (axial strain amplitude) is dependent on the gap width of the OSP geometry and is limited by the maximum orthogonal oscillation displacement of the rheometer, i.e., 50 µm (Table 1).
  5. Specify the angular frequency range from 0.1 to 40 rad/s at 10 points per decade logarithmically.
    NOTE: The angular frequency range used here is a recommended range for OSP operation based on the instrument axial frequency sensitivity limits (Table 1) and the consideration of gap loading conditions18. See the Discussion section for more details.
  6. Start the experiment from the rheometer software.

6. Performing analysis

  1. Determination of the primary end-effect factor
    1. Export the steady shear rate sweep results (from Protocol step 4.4.) into an open file format such as .csv or .txt.
    2. Calculate the average value of the reported viscosities over the appropriate shear rate range in a spreadsheet software.
      NOTE: Only the viscosity data with corresponding torque values above the manufacture specified limits are used to calculate the average viscosity. The averaged viscosity value is defined as the uncorrected primary viscosity.
    3. Find the primary end-effect factor using the averaged viscosity value.
      NOTE: This section is provided here to show the derivation of the relationship between the primary end effect factor and the direct viscosity output from the rheometer software. An example of the calculation of the end factor from the experimental data is demonstrated in the Representative Results section. The primary steady shear viscosity is the ratio of shear stress τ to the shear rate figure-protocol-16126, which is calculated from the raw signals of torque M and rotational velocity Ω via the geometry constants (Kτ and Kγ). The expression is given by:
      figure-protocol-16419
      where Kτ is the primary stress constant (Equation 1) and Kγ is the primary strain constant which is solely dependent on the geometric dimensions. Therefore, substituting Equation 1 into Equation 3, the calculated primary viscosity, or the output viscosity values from the rheometer software, is shown to be inversely proportional to the primary end-effect factor CL (note that all other variables in Equation 3 are either geometric constants or raw measurement signals):
      figure-protocol-17046
      Note that Equation 3 is a general expression to any rotational rheometry where the measured viscosity is calculated from the raw data, i.e., torque and velocity, via the stress and strain constants that depend on different geometry used, e.g., cone-plate, parallel plate, concentric cylinder, etc.

7. Determination of the orthogonal end-effect factor

  1. Export the orthogonal frequency sweep results (from Protocol step 5.6.) into an open file format such as .csv or .txt.
  2. Calculate the average value of the reported OSP complex viscosity over the appropriate angular frequency range in a spreadsheet software.
    NOTE: Only the viscosity data with corresponding oscillation force values above the manufacturer specified limits are used to calculate the average viscosity. The averaged viscosity value is defined as the uncorrected orthogonal complex viscosity.
  3. Find the orthogonal end-effect factor using the averaged complex viscosity value.
    NOTE: This section is provided here to show the derivation of the relationship between the orthogonal end effect factor and the orthogonal complex viscosity output from the rheometer software. An example of the calculation of the orthogonal end factor from the experimental data is demonstrated in the Representative Results section. The orthogonal complex viscosity equals the orthogonal complex shear modulus figure-protocol-18585 divided by the orthogonal oscillatory frequency ωfigure-protocol-18709, which can be expressed as the equation below through the oscillation force Ffigure-protocol-18879, oscillation displacement θfigure-protocol-19004, frequency ωfigure-protocol-19114 (all three of which are raw signals), and the geometry constants (Kτο and Kγο):
    figure-protocol-19350
    where Kτο is the orthogonal stress constant (Equation 2) and Kγο is the orthogonal strain constant, which is solely related to the geometric dimensions. Therefore, substituting Equation 2 into Equation 5, the calculated orthogonal complex viscosity, or the output OSP complex viscosity values from the rheometer software, is shown to be directly proportional to the orthogonal end-effect factor CLo (note that all other variables in Equation 5 are either geometric constants or raw measurement signals):
    figure-protocol-20023
    Note that Equation 5 is a general expression for any linear motion measurements where the measured complex viscosity is calculated from the raw data, i.e., force, displacement, and frequency, via the stress and strain constants that depend on the geometry used, e.g., cone-plate, parallel plate, concentric cylinder, etc.

8. Viscosity verification check by OSP measurements

NOTE: This step is to verify if the corrections are valid using the calibrated end-effect factors obtained from the calibration experiments.

  1. Enter the calibrated values for the end-effect factor and orthogonal end-effect factor under the geometry constants, initially these values were set equal to 1.00. The stress constants are automatically updated, and the values are as shown in Table 3.
  2. Set up a same experimental procedure following the steps in the orthogonal frequency sweep tests. Enter 1.0 s−1 for the shear rate.
  3. Start the experiment.

Results

Representative results from the viscosity calibration measurements on a 12.2 Pa s silicone viscosity standard are represented in Figure 5 and Figure 6. Note that the primary end-effect factor and the orthogonal end-effect factor are both set to 1.00 for the calibration runs. Figure 5 shows the steady shear viscosity and the torque as a function of shear rate on a double y-axis plot. The silicone liquid is a Newtonian fluid; as expec...

Discussion

In this protocol, we present a detailed experimental procedure for performing viscosity calibration measurements using Newtonian fluids for a commercial orthogonal superposition rheology technique with a double-wall concentric cylinder geometry. The calibration factors, i.e., the primary end-effect factor CL and the orthogonal end-effect factor CLo, are determined independently by conducting steady shear rate sweep and orthogonal frequency sweep tests. After obtai...

Disclosures

The full description of the procedures used in this paper requires the identification of certain commercial products and their suppliers. The inclusion of such information should in no way be construed as indicating that such products or suppliers are endorsed by NIST or are recommended by NIST or that they are necessarily the best materials, instruments, software or suppliers for the purposes described.

Acknowledgements

Ran Tao would like to thank funding from the National Institute of Standards and Technology, U.S. Department of Commerce under grant 70NANB15H112. Funding for Aaron M. Forster was provided through congressional appropriations to the National Institute of Standards and Technology.

Materials

NameCompanyCatalog NumberComments
Advanced Peltier SystemTA Instruments402500.901Enviromental control device
ARES-G2 RheometerTA Instruments401000.501Rheometer
Brookfield Silicone Fluid, 12500cPAMTEK Brookfield12500 cpsViscosity standard liquid
OSP Slotted Bob, 33 mmTA Instruments402796.902Bob, upper geometry
OSP Slotted Double Gap Cup, 34 mmTA Instruments402782.901Double wall cup, lower geometry
Pipette (1 – 10 mL)Eppendorf3120000089To load test materials
Pipette (100 – 1,000 µL)Eppendorf3123000063To load test materials
Pipette Tips (0.5 – 10 mL)Eppendorf022492098To load test materials
Pipette Tips (50 – 1,000 µL)Eppendorf022491555To load test materials
SpatulaVWR82027-532To load test materials
TRIOSTA Instrumentsv4.3.1.39215Rheometer software

References

  1. Macosko, C. W. . Rheology: principles, measurements, and applications. , (1994).
  2. Larson, R. G. . The Structure and Rheology of Complex Fluids. , (1999).
  3. Vermant, J., Moldenaers, P., Mewis, J., Ellis, M., Garritano, R. Orthogonal superposition measurements using a rheometer equipped with a force rebalanced transducer. Review of Scientific Instruments. 68 (11), 4090-4096 (1997).
  4. Ferry, J. D. . Viscoelastic Properties of Polymers. , (1980).
  5. Yamamoto, M. Rate-dependent relaxation spectra and their determination. Transactions of the Society of Rheology. 15 (2), 331-344 (1971).
  6. Simmons, J. M. A servo-controlled rheometer for measurement of the dynamic modulus of viscoelastic liquids. Journal of Scientific Instruments. 43 (12), 887-892 (1966).
  7. Tanner, R. I., Williams, G. On the orthogonal superposition of simple shearing and small-strain oscillatory motions. Rheologica Acta. 10 (4), 528-538 (1971).
  8. Schoukens, G., Mewis, J. Nonlinear rheological behaviour and shear-dependent structure in colloidal dispersions. Journal of Rheology. 22 (4), 381-394 (1978).
  9. Zeegers, J., et al. A sensitive dynamic viscometer for measuring the complex shear modulus in a steady shear flow using the method of orthogonal superposition. Rheologica Acta. 34 (6), 606-621 (1995).
  10. Mewis, J., Schoukens, G. Mechanical spectroscopy of colloidal dispersions. Faraday Discussions of the Chemical Society. 65, 58-64 (1978).
  11. Lin, N. Y. C., Ness, C., Cates, M. E., Sun, J., Cohen, I. Tunable shear thickening in suspensions. Proceedings of the National Academy of Sciences. 113 (39), 10774-10778 (2016).
  12. Gracia-Fernández, C., et al. Simultaneous application of electro and orthogonal superposition rheology on a starch/silicone oil suspension. Journal of Rheology. 60 (1), 121-127 (2015).
  13. Sung, S. H., Kim, S., Hendricks, J., Clasen, C., Ahn, K. H. Orthogonal superposition rheometry of colloidal gels: time-shear rate superposition. Soft Matter. 14 (42), 8651-8659 (2018).
  14. Colombo, G., et al. Superposition rheology and anisotropy in rheological properties of sheared colloidal gels. Journal of Rheology. 61 (5), 1035-1048 (2017).
  15. Jacob, A. R., Poulos, A. S., Kim, S., Vermant, J., Petekidis, G. Convective Cage Release in Model Colloidal Glasses. Physical Review Letters. 115 (21), 218301 (2015).
  16. Jacob, A. R., Poulos, A. S., Semenov, A. N., Vermant, J., Petekidis, G. Flow dynamics of concentrated starlike micelles: A superposition rheometry investigation into relaxation mechanisms. Journal of Rheology. 63 (4), 641-653 (2019).
  17. Moghimi, E., Vermant, J., Petekidis, G. Orthogonal superposition rheometry of model colloidal glasses with short-ranged attractions. Journal of Rheology. 63 (4), 533-546 (2019).
  18. Tao, R., Forster, A. M. End effect correction for orthogonal small strain oscillatory shear in a rotational shear rheometer. Rheologica Acta. 59 (2), 95-108 (2020).
  19. Schrag, J. L. Deviation of velocity gradient profiles from the "gap loading" and "surface loading" limits in dynamic simple shear experiments. Transactions of the Society of Rheology. 21 (3), 399-413 (1977).
  20. Ewoldt, R. H., Johnston, M. T., Caretta, L. M., Spagnolie, S. . Complex Fluids in Biological Systems. Biological and Medical Physics, Biomedical Engineering. , 207-241 (2015).

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Explore More Articles

Orthogonal Superposition RheologyCalibration ProceduresComplex FluidsRadiological PropertiesMeasurement ErrorStandard Operating ProcedureEnd Effect FactorsRheometer SoftwareGeometry InstallationEnvironmental Control DeviceDouble Wall Cup ConfigurationPlatinum Resistance ThermometerTorque ScrewdriverTransducer Control PanelTest Material LoadingGeometry Gap Set Point

This article has been published

Video Coming Soon

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved