JoVE Logo

Zaloguj się

Aby wyświetlić tę treść, wymagana jest subskrypcja JoVE. Zaloguj się lub rozpocznij bezpłatny okres próbny.

W tym Artykule

  • Podsumowanie
  • Streszczenie
  • Wprowadzenie
  • Protokół
  • Wyniki
  • Dyskusje
  • Ujawnienia
  • Podziękowania
  • Materiały
  • Odniesienia
  • Przedruki i uprawnienia

Podsumowanie

Stimulated Raman scattering (SRS) microscopy is a powerful, nondestructive, and label-free imaging technique. One emerging application is stimulated Raman histology, where two-color SRS imaging at the protein and lipid Raman transitions are used to generate pseudo-hematoxylin and eosin images. Here, we demonstrate a protocol for real-time, two-color SRS imaging for tissue diagnosis.

Streszczenie

Stimulated Raman scattering (SRS) microscopy has emerged as a powerful optical imaging technique for tissue diagnosis. In recent years, two-color SRS has been shown to be able to provide hematoxylin and eosin (H&E)-equivalent images that allow fast and reliable diagnosis of brain cancer. Such capability has enabled exciting intraoperative cancer diagnosis applications. Two-color SRS imaging of tissue can be done with either a picosecond or femtosecond laser source. Femtosecond lasers have the advantage of enabling flexible imaging modes, including fast hyperspectral imaging and real-time, two-color SRS imaging. A spectral-focusing approach with chirped laser pulses is typically used with femtosecond lasers to achieve high spectral resolution.

Two-color SRS acquisition can be realized with orthogonal modulation and lock-in detection. The complexity of pulse chirping, modulation, and characterization is a bottleneck for the widespread adoption of this method. This article provides a detailed protocol to demonstrate the implementation and optimization of spectral-focusing SRS and real-time, two-color imaging of mouse brain tissue in the epi-mode. This protocol can be used for a broad range of SRS imaging applications that leverage the high speed and spectroscopic imaging capability of SRS.

Wprowadzenie

Traditional tissue diagnostics rely on staining protocols followed by examination under an optical microscope. One common staining method used by pathologists is H&E staining: hematoxylin stains cell nuclei a purplish blue, and eosin stains the extracellular matrix and cytoplasm pink. This simple staining remains the gold standard in pathology for many tissue diagnoses tasks, particularly cancer diagnosis. However, H&E histopathology, particularly the frozen sectioning technique used in an intraoperative setting, still has limitations. The staining procedure is a laborious process involving tissue embedding, sectioning, fixation, and staining1. The typical turnaround time is 20 min or longer. Performing H&E during frozen sectioning can sometimes become more challenging when multiple sections are processed at once due to the need to evaluate cellular features or growth patterns in 3D for margin assessment. Moreover, intraoperative histological techniques require skilled technicians and clinicians. Limitation in the number of board-certified pathologists in many hospitals is a constraint for intraoperative consultation in many cases. Such limitations may be alleviated with the fast development interests in digital pathology and artificial intelligence-based diagnosis2. However, the H&E staining results are variable, depending on the experience of the technician, which presents additional challenges for computer-based diagnosis2.

These challenges can potentially be addressed with label-free optical imaging techniques. One such technique is SRS microscopy. SRS uses synchronized pulsed lasers—pump and Stokes—to excite molecular vibrations with high efficiency3. Recent reports have demonstrated that SRS imaging of proteins and lipids can generate H&E-equivalent images (also known as stimulated Raman histology or SRH) with intact fresh tissue, which bypasses the need for any tissue processing, significantly shortens the time needed for diagnosis, and has been adapted intraoperatively4. Moreover, SRS imaging can provide 3D images, which offers additional information for diagnosis when 2D images are insufficient5. SRH is unbiased and generates digital images that are readily available for computer-based diagnosis. It quickly emerges as a possible solution for intraoperative cancer diagnosis and tumor margin analysis, especially in brain cancer6,7,8. More recently, SRS imaging of chemical changes of tissue has also been suggested to provide useful diagnostic information that can further help clinicians stratify different cancer types or stages9.

Despite its tremendous potential in tissue diagnosis applications, SRS imaging is mostly limited to academic laboratories specialized in optics due to the complexity associated with the imaging platform, which includes ultrafast lasers, the laser scanning microscope, and sophisticated detection electronics. This protocol provides a detailed workflow to demonstrate the use of a common femtosecond laser source for real-time, two-color SRS imaging and the generation of pseudo-H&E images from mouse brain tissue. The protocol will cover the following procedures:

Alignment and chirp optimization
Most SRS imaging schemes use either picosecond or femtosecond lasers as the excitation source. With femtosecond lasers, the bandwidth of the laser is much larger than the Raman linewidth. To overcome this limitation, a spectral focusing approach is used to chirp the femtosecond lasers to a picosecond timescale to achieve narrow spectral resolution10. Optimal spectral resolution is only achieved when the temporal chirp (also known as the group delay dispersion or just dispersion) is properly matched for the pump and the Stokes lasers. The alignment procedure and the steps needed to optimize the dispersion of the laser beams using highly dispersive glass rods are demonstrated here.

Frequency calibration
An advantage of spectral focusing SRS is that the Raman excitation can be quickly tuned by changing the time delay between the pump and the Stokes lasers. Such tuning affords fast imaging and reliable spectral acquisition compared to tuning laser wavelengths. However, the linear relationship between excitation frequency and time delay requires external calibration. Organic solvents with known Raman peaks are used to calibrate the Raman frequency for spectral focusing SRS.

Real-time, two-color imaging
It is important to increase the imaging speed in tissue diagnosis applications to shorten the time needed for analyzing large tissue specimens. Simultaneous two-color SRS imaging of lipids and proteins obviates the need to tune the laser or time delay, which increases the imaging speed by more than two-fold. This is achieved by using a novel orthogonal modulation technique and dual-channel demodulation with a lock-in amplifier11. This paper describes the protocol for orthogonal modulation and dual-channel image acquisition.

Epi-mode SRS imaging
The majority of SRS imaging shown to date is performed in transmission mode. Epi-mode imaging detects backscattered photons from tissue12. For pathology applications, surgical specimens can be quite large. For transmission mode imaging, tissue sectioning is often necessary, which undesirably requires extra time. In contrast, epi-mode imaging can work with intact surgical specimens. Because the same objective is used to collect backscattered light, there is also no need for aligning a high numerical-aperture condenser required for transmission imaging. Epi-mode is also the only option when tissue sectioning is difficult, such as with bone. Previously we have demonstrated that for brain tissue, epi-mode imaging offers superior imaging quality for tissue thickness > 2 mm13. This protocol uses a polarizing beam splitter (PBS) to collect scattered photons depolarized by tissue. It is possible to collect more photons with an annular detector at the expense of the complexity of customized detector assembly12. The PBS approach is simpler to implement (similar to fluorescence), with the standard photodiode already being used for transmission mode detection.

Pseudo-H&E image generation
Once two-color SRS images are collected, they can be recolored to simulate H&E staining. This paper demonstrates the procedure for converting lipid and protein SRS images to pseudo-H&E SRS images for pathology applications. The experimental protocol details critical steps needed to generate high-quality SRS images. The procedure shown here is not only applicable to tissue diagnosis but also can be adapted for many other hyperspectral SRS imaging applications such as drug imaging and metabolic imaging14,15.

General system requirements
The laser system for this protocol must be able to output 2 synchronized femtosecond laser beams. Systems ideally feature an Optical Parametric Oscillator (OPO) for broad wavelength tuning of one of the laser beams. The setup in this protocol uses a commercial laser system Insight DS+ that outputs two lasers (one fixed beam at 1,040 nm and one OPO-based tunable beam, ranging from 680 to 1,300 nm) with a repetition rate of 80 MHz. Laser scanning microscopes, either from major microscope manufacturers or home-built, can be used for SRS imaging. The utilized microscope is an upright laser scanning microscope built on top of a commercial upright microscope frame. A pair of 5 mm galvo mirrors are used to scan the laser beam. For users choosing to adopt a homebuilt laser scanning microscope, refer to a previously published protocol for the construction of a laser scanning microscope16.

Protokół

All experimental animal procedures were conducted with 200 µm, fixed, sectioned mouse brains, in accordance with the protocol (# 4395-01) approved by the Institute of Animal Care and Use Committee (IACUC) of the University of Washington. Wild-type mice (C57BL/6J strain) are euthanized with CO2. Then, a craniotomy is performed to extract their brains for fixation in 4% paraformaldehyde in phosphate-buffered saline. The brains are embedded in a 3% agarose and 0.3% gelatin mixture and sectioned into 200 μm-thick slices by a vibratome.

1. Initial alignment

NOTE: Ensure that the beam size and divergence of both arms are matched for best sensitivity and resolution. Collimate the pump and the Stokes beams and adjust their sizes before they enter the laser scanning microscope. To do this, use a pair of achromatic lenses for each beam before combining them on the dichroic mirror. Always wear proper laser goggles for beam alignment.

  1. Beam collimation
    1. Install a pair of achromatic lenses for the pump beam. As a starting point, use a 100 mm lens and 200 mm lens to magnify the laser beam size by 2-fold. Ensure that the distance of the two lenses is roughly 300 mm. Align the pump beam through the center of both lenses.
    2. Place a mirror after the second lens to send the beam toward a wall (>1 m away). Take care when sending the beam across long distances. Trace the beam from the mirror to the wall with an IR card and check if the beam changes in size. Collimate the beam if the beam changes in size as a function of distance.
      1. If the beam is converging (decreasing size with propagation), move the two lenses closer.
      2. If the beam is diverging (increasing size with propagation), move the two lenses further apart. Adjust the distance until the beam is collimated.
    3. Repeat steps 1.1.1 and 1.1.2 for the Stokes beam to collimate the beam.
  2. Beam size adjustment
    1. If a beam profiler is available, measure the collimated beam size for each beam. Alternatively, estimate the beam size using the IR card and a ruler to obtain a beam diameter of 4-5 mm.
    2. If the beam size is too small or too large, change the lens pair used in step 1.1. Adjust the lens pair until both beams have a diameter of 4-5 mm.
      NOTE: The magnification of the beam is the ratio between the focal length of the second lens to the first lens (f2/f1).
  3. Spatial overlap
    NOTE: SRS imaging requires both laser beams to be combined in space and time to excite molecular vibrations. A schematic of the spectral-focusing SRS imaging is shown in Figure 1.
    1. Combine the two laser beams by installing a dichroic mirror with several steering mirrors for adjustment. Optimize spatial overlap of the pump and Stokes by monitoring beams after the dichroic mirror at two different positions far apart (~1 m). Iteratively adjust the steering mirror before the dichroic and the dichroic mirror to align the Stokes beam with the pump beam.
      NOTE: If the two beams are spatially overlapping at both positions, they are sufficiently overlapped.
    2. Ensure that the combined beams are sent to the center of the scan mirrors of the laser scanning microscope by adjusting a pair of steering mirrors when the scan mirrors are in the parked position. Ensure both beams travel through the center of the microscope objective and the condenser.
    3. After the condenser, use another pair of lenses with focal lengths of 100 mm and 30 mm, respectively, to relay the transmitted beam onto the photodiode. Ensure both beams are contained within the photodiode and install two low-pass filters to block out the modulated Stokes beam.

2. SRS signal detection

  1. Electrooptical modulation (EOM)
    NOTE: EOM of 20 MHz is used to modulate the Stokes amplitude. As discussed later, the EOM is derived from the 80 MHz laser pulse train, which is required for orthogonal modulation. Other modulation frequencies can be used if only single-color or hyperspectral SRS is performed. In that case, synchronization of modulation frequency to laser frequency is unnecessary. A frequency generator with an RF power amplifier can be used to drive the EOM. As a result, steps 2.1.1-2.1.4 can be skipped.
    1. Place a beam sampler in the Stokes beam path to pick up 10% of the beam and send it to a fast photodiode to detect the 80 MHz pulse train.
      NOTE: The photodiode signal is sent to a frequency divider to generate a 20 MHz TTL output. This output is further sent into a fanout buffer to replicate the output into four identical 20 MHz outputs. One of the outputs is used to trigger the oscilloscope.
    2. Take one of the outputs of the fanout buffer and filter it with a bandpass filter to get a 20 MHz sinusoidal wave. Use an RF attenuator to adjust the output peak-to-peak voltage to ~500 mV.
    3. Send the resulting output to a phase shifter, which allows fine adjustment of the RF phase with a voltage source. Send this output to an RF power amplifier and connect the output of the amplifier to the EOM.
    4. Unblock the Stokes beam and optimize the modulation depth of EOM1 by placing a photodiode in the beam path. Adjust the EOM voltage and quarter-wave plate until the modulation depth (valley to peak ratio) appears satisfactory.
      NOTE: At 20 MHz modulation (1/4 of the laser repetition rate), two pulses are expected every 50 ns.
  2. Temporal overlap
    NOTE: Temporal overlap of the pump and Stokes is achieved by delaying one of the two laser pulse trains with a retro-reflector mounted onto a delay stage (Figure 1 shows the Stokes being delayed). Coarse overlap is monitored with the oscilloscope, and fine overlap is monitored by the SRS signal. Fine temporal overlap can also be achieved with an autocorrelator if available.
    1. Place a photodiode after the dichroic mirror to detect the laser beam. Block the Stokes beam first. Zoom in on one of the pump pulse peaks on the oscilloscope. Place a vertical cursor to mark the temporal position of this peak with the oscilloscope.
    2. Block the pump beam and unblock the Stokes beam. Translate the delay stage to temporally match the peak position on the oscilloscope to the marked position in the previous step. See Figure 2 for a display of the temporal overlap of two beams.
      1. (OPTIONAL) If the translation of the delay stage is insufficient to temporally match the two beams, then move the delay stage to the middle of its movement range.
      2. Calculate the delay distance required to match the two beams by taking the temporal difference between the two beams and multiplying the difference by the speed of light to find the amount of distance needed to match the two beams temporally.
      3. Elongate the beam path of the faster beam or shorten the beam path of the slower beam to roughly match the temporal delay accordingly.
    3. Prepare a microscope slide sample with DMSO and double-sided tape as a spacer to hold the sample between the slide and a coverslip.
    4. Place the sample on the microscope with the coverslip side facing the microscope objective. Change the microscope to brightfield illumination and observe the sample from the eyepiece. Find the focus of the sample by first finding the focus at both the top and bottom layer of air bubbles at the glass-tape interface, and then move the focus to be in between the two layers of tape.
      NOTE: Ensure the laser beams are blocked before looking into the eyepiece.
    5. Set the tunable beam output to 798 nm. Based on the optical throughput of the condenser, adjust the optical power to be ~40 mW each for the pump and Stokes beams at the objective focus.
    6. Open ScanImage in MATLAB (or other scanning software that controls the microscope) and click on the button labeled FOCUS to start scanning.
      NOTE: The laser beams will be raster-scanned through the sample to generate an image. The low-frequency signal output from the photodiode (<100 kHz) is directly sent into channel 1 of the data acquisition card (referred to as the DC channel). The high-frequency output (>100 kHz) from the photodiode is sent into the lock-in amplifier, and the X-output of the lock-in amplifier signal is sent into channel 2 of the data acquisition card (referred to as the AC channel).
    7. Adjust the steering mirror before the galvo scanner to center the DC signal on the channel 1 display. Move the motorized delay stage and closely observe the lock-in output shown on the channel 2 (i.e., AC channel) display.
      NOTE: When the pump and Stokes coincide in time, a signal will show up on the AC channel. It is helpful to adjust the color scale of the AC channel to display the small intensity change.
    8. Maximize the AC signal intensity by finely adjusting the time delay. Adjust the dichroic mirror to center the SRS signal on the AC channel (while keeping the DC channel centered). Adjust the phase of the lock-in amplifier to maximize the signal. See Figure 3 for a satisfactory signal.

3. Spectral resolution optimization

NOTE: The pump and Stokes beams reaching the sample should have the same amount of group delay dispersion (GDD) to maximize spectral resolution. The dispersion depends heavily on the experimental setup. The experimental setup described here utilizes femtosecond pulses at 1,040 nm and 800 nm as Stokes and pump, respectively. Dense flint glass rods (H-ZF52A) are used as the pulse-stretching medium.

  1. Insert 48 cm of a highly dispersive glass rod (H-ZF52A or equivalent dense flint glass) into the 800 nm beam path. Estimate the GDD using Eq (1):
    figure-protocol-10330 (1)
    NOTE: GVD of various glass materials at different wavelengths can be found from the refractive index database resource. For example, H-ZF52A has a GVD of 220.40 fs2/mm at 800 nm. The total GDD is 105792 fs2.
  2. Calculate how many cm of the dispersive glass rod is required to add to the 1,040 nm beam path to match the GDD of the pump. Insert the appropriate length of dispersive glass rods to the 1,040 nm beam path to roughly match the GDD of the 800 nm beam. Note that the addition of glass rods will change the temporal overlap of the two beams, and adjustment of delay may be necessary.
  3. Calibration of spectral resolution
    1. Make a microscope slide sample with DMSO. Place the slide onto the microscope and check the power of the beams coming out of the microscope condenser. Adjust the power accordingly to have ~40 mW each at sample focus.
    2. Open ScanImage from MATLAB. Find the maximal SRS signal by scanning through the delay stage, which corresponds to the 2,913 cm-1 Raman peak of DMSO. Estimate the stage position based on the previous stage position with the increased optical path length due to the insertion of rods. Realign the beam spatial overlap because of the small deviation of the beam when glass rods are added.
    3. Save a hyperspectral SRS scan by sequentially taking a series of SRS images while moving the motorized stage.
      NOTE: The delay scan range covers two Raman peaks, corresponding to the 2,913 cm-1 and 2,994 cm-1 Raman peaks of DMSO, respectively. These two transitions are observed when utilizing an 800 nm pump and 1,040 nm Stokes laser.
    4. Plot out the SRS spectra of the DMSO solution using either ImageJ or MATLAB. Fit the large DMSO 2,913 cm-1 peak to a Gaussian or Lorentzian function in MATLAB to calculate the Full Width at Half Maximum (FWHM) of the peak.
      NOTE: Representative results are shown in Figure 4. If only one broad peak is present, that means either the spectral resolution is too poor to distinguish the two peaks and more glass rods are required, or the scanned range was too small to detect the second peak. Typically, an acceptable spectral resolution DMSO is ~20-25 cm-1 when glass rods length of >60 cm are used. A lower resolution is often used for tissue imaging to trade for higher signals with shorter pulses17.
  4. (OPTIONAL) Use an autocorrelator or a FROG (Frequency-Resolved Optical Gating) to determine the pulse duration of each arm to calculate exactly the amount of GDD and the length of rods needed to match the GDD between the pump and the Stokes.
  5. Repeat steps 3.3.2-3.3.4 for different rods lengths on the Stokes beam to find the optimal spectral resolution, which means the best GDD match has been found experimentally. Use multiple sets of glass rods differing in length to achieve optimal spectral resolution.

4. Signal to noise (SNR) characterization

  1. Ensure that step 4.2 is performed after complete spatial and temporal alignment.
  2. Acquire an SRS image corresponding to the 2,913 cm-1 Raman peak of DMSO. Open the image in ImageJ and select a small area in the center of the frame. Use the measure function to calculate the mean and standard deviation of values in the selected area.
  3. Divide the mean value of the selected area by the standard deviation to find the SNR value, as in Eq (2).
    figure-protocol-14086 (2)
    NOTE: A good SNR for the system (with a lock-in time constant of 4 μs) using DMSO at 40 mw/40 mw at focus for both arms is >800. Lower concentrations of DMSO or lower power can be used for a more accurate estimation of the SNR if the data acquisition card has a limited bit depth.
  4. If the SNR is too low, realign the laser pulses to optimize spatial overlap, temporal overlap, beam size/collimation matching, and/or dispersion matching. For an objective with an aberration correction collar, optimize the signal by adjusting the correction collar.

5. Frequency axis calibration

NOTE: This step is performed to relate the delay stage position to the scanned Raman transition. Careful selection of solvents is required to generate an appropriate "Raman Ruler." DMSO is an effective solvent for CH bonds as it has two sharp Raman peaks at 2,913 cm-1 and 2,994 cm-1.

  1. Save a hyperspectral scan with the delay-stage range covering the 2,913 cm-1 and 2,994 cm-1 Raman peaks of DMSO. Save the stage positions corresponding to the hyperspectral dataset.
    NOTE: The global maximum peak of the spectrum corresponds to the DMSO 2,913 cm-1 Raman shift and the second maximum peak corresponds to the DMSO 2,994 cm-1 Raman shift.
  2. Perform linear regression for the stage positions and Raman shifts at 2,913 cm-1 and 2,994 cm-1. Using the linear regression equation relating the stage position to the Raman shift, convert the delay positions to the corresponding Raman frequencies.

6. Orthogonal modulation and two-color imaging

NOTE: The orthogonal modulation step is only necessary when real-time two-color imaging is needed. A schematic of this scheme is shown in Figure 5. The orthogonal modulation uses a pair of EOMs driven at a quarter of the laser frequency (20 MHz for 80 MHz laser) with a 90° phase shift between the two. This orthogonal modulation step can be skipped for single-color SRS imaging or hyperspectral SRS imaging.

  1. EOM1 modulation
    1. Install a PBS (PBS2), a quarter-wave plate (QWP2), and a second EOM (EOM2) into the Stokes beam path after the first EOM. Unplug the signal input to EOM2. Plug in the signal input to EOM1 and turn it on.
    2. Modulate the Stokes beam (fixed at 1,040 nm) at 20 MHz (f0/4) by sending the beam through the first EOM. Adjust the tilt and position of EOM1 to ensure that the beam is hitting straight and centered through the EOM crystal.
    3. Monitor the modulation depth by observing both polarizations coming out of PBS1 with two photodiodes and displaying the modulation on an oscilloscope.
    4. Adjust the QWP1, EOM1 voltage, and phase of the 20 MHz input (using a phase shifter) to optimize the modulation depth of the transmitted beam to be close to 100%. See Figure 2B for an illustration of good modulation depth.
  2. EOM2 modulation
    1. Unplug EOM1 and plug in EOM2.
    2. Send in the high voltage output of the second amplifier at 20 MHz to EOM2. Adjust the tilt and position of EOM2 to ensure that the beam is hitting straight and centered through the EOM crystal.
    3. Once again, monitor the modulation depth by looking at both polarizations coming out of PBS2 with an oscilloscope. Adjust the QWP2, EOM2 voltage, and the phase shifter as needed to achieve close to 100% modulation for both polarizations.
    4. Ensure that the pulse train modulation has a 90° phase shift from the first modulation.
      NOTE: If the two pump pulse trains are not 90° orthogonal, crosstalk between the two channels will be a problem.
    5. Test the orthogonality of the modulation by turning on and plugging in both EOM1 and EOM2. Monitor both polarizations being split by PBS2 with an oscilloscope. Reoptimize EOM1 and EOM2 individually if the pulse train after the second PBS does not resemble Figure 2C.
    6. Install a 20 mm birefringent quartz crystal (BRC) and HWP downstream of EOM2. Plug in both EOMs at once and monitor the pulse train such that it resembles Figure 2D.
      NOTE: For the chirp used in this experiment, 20 mm BRC induces a time delay that corresponds to an 80 cm-1 Raman shift. A different BRC length may be needed if a different chirp is used.
  3. Calibration
    1. Calibrate the system using DMSO by detecting signals from the lock-in amplifier X and Y channel output (sent to channels 2 and 3 of the data acquisition card).
    2. Check whether the signals generated by the 2,913 cm-1 peak from the faster polarization and that from the slower polarization are close to 90° out-of-phase on the lock-in amplifier. If this is not the case, adjust the EOM alignment until the two signals are close to 90° out-of-phase.
    3. Once calibration is complete, find the delay position that probes the protein transition at 2,930 cm-1 for one of the orthogonal beams. Ensure that the other polarization probes the lipid transition of 2,850 cm-1.

7. Epi-mode SRS imaging

NOTE: In the transmission mode imaging scheme, the objective focuses the laser into the sample, and then a condenser lens directs the transmitted beam to a photodiode for lock-in detection. In the epi-mode imaging scheme, light that is backscattered and depolarized by the sample is recollected by the focusing objective and isolated using a polarizing beam splitter. The isolated and backscattered photons are sent to a photodiode through a pair of relay lenses for lock-in detection. Figure 6 depicts the epi-mode imaging scheme.

  1. Install an HWP before the beam enters the microscope to change the polarization of the beam going into the microscope. Place a PBS above the objective to allow the depolarized back-reflected beam to reach the detector.
  2. Use a pair of lenses consisting of a 75 mm achromat lens and a 30 mm aspheric lens to relay the backscattered photons from the back aperture of the objective to the photodetector. Mount the detector to collect the backscattered light directed by the PBS. Install a filter to block out the modulated beam from entering the detector.
  3. Place the tissue sample under the objective. As the condenser is unnecessary for epi-mode imaging, remove it if more space is required.
  4. Imaging
    1. Block off the beam with a shutter; shine a white light source onto the sample from the side; and use brightfield to find objective focus.
    2. Unblock both beams and use the precalibrated delay positions to acquire lipid and protein SRS images from tissue from the two outputs of the lock-in amplifier.
    3. Adjust the lock-in gain and pixel bin factor to acquire good-quality images.

8. False-color staining

  1. Open the image stack with ImageJ.
  2. Pull out the two images that correspond to the lipid (2,850 cm-1) and protein (2,930 cm-1) species by right-clicking on the image and clicking on Duplicate.
  3. Rename the lipid image to lipids and the protein image to proteins.
  4. Go to Process | Image Calculator and perform proteins subtract lipids.
  5. Combine the images by going to Image | Color | Merge Channels, setting lipids to green and proteins to blue. Open the image channels tool (Image | Color | Channels Tool) and adjust the brightness and contrast (Image | Adjust | Brightness/Contrast).
  6. Adjust the brightness and contrast for each channel using the channels tool. For the lipid channel, adjust the contrast until the cellular features appear dark. For the protein channel, adjust the contrast until the cellular features appear blue. Convert the merged channel green/blue image to an RGB image by going to Image | Type | RGB Color. Export this image by File | Save As | Tiff.
    NOTE: For false H&E staining, the color scheme shows pink cytoplasm, while the nuclei are dark blue-purple.
  7. Download the HE.m MATLAB script from the false H&E staining script resource in the Table of Materials.
  8. Run the HE.m script in MATLAB. Select the exported RGB image from the previous step to generate an artificially H&E stained image.
  9. (OPTIONAL) Normalize the image intensity for large field-of-view imaging because the image appears darker in the periphery than in the center.
    1. To perform field normalization of the images, average as many images as possible. Then, remove the intensity features with ImageJ (Process | Filters | Gaussian Blur | Radius=50).
    2. Measure the maximum intensity of the blurred image (Ctrl+M). Divide the blurred image by the maximum intensity (Process | Math | Divide). Divide the raw SRS image by the blurred image (Process | Image | Calculator).

Wyniki

Optimizing spectral resolution:
Dispersion through a material is affected by the dispersive medium (length and material) and wavelength. Changing the dispersion rod length affects the spectral resolution and the signal size. It is a give-and-take relationship that can be weighed differently depending on the application. The rods stretch out the beam pulse from being wide in frequency and narrow in time to being narrow in frequency and broad in time. Figure 7 shows the ...

Dyskusje

The two-color SRS imaging scheme presented in this protocol hinges on the proper implementation of one-color SRS imaging. In one-color SRS imaging, the critical steps are spatial alignment, temporal alignment, modulation depth, and phase shift. Spatially combining the two beams is accomplished by a dichroic mirror. Several steering mirrors are used for fine adjustment when sending the beams to the dichroic mirror. Once the beams are combined with the dichroic mirror, spatial alignment can be confirmed by picking off the ...

Ujawnienia

The authors declare that there are no conflicts of interest.

Podziękowania

This study was supported by NIH R35 GM133435 to D.F.

Materiały

NameCompanyCatalog NumberComments
100 mm Achromatic LensTHORLABSAC254-100-BBroadband, 650 - 1,050 nm, achromatic lens focal length, 100 mm
20 MHz bandpass filterMinicircuitsBBP-21.4+Lumped LC Band Pass Filter, 19.2 - 23.6 MHz, 50 Ω
200 mm Achromatic LensTHORLABSAC254-200-BBroadband, 650 - 1,050 nm, achromatic lens focal length, 200 mm
Achromatic Half WaveplateUnion OpticWPA2210-650-1100-M25.4Broadband half waveplate
Achromatic Quarter WaveplateUnion OpticWPA4210-650-1100-M25.4Broadband quarter waveplate
Beam SamplerTHORLABSBSN1110:90 Plate Beamsplitter
Dichroic MirrorTHORLABSDMSP1000Other dichroics with a center wavelength around 1,000 nm can be used.
DMSO (Dimethyl sulfoxide)Sigma Aldrich472301Solvent for calibration of Raman shift. Other solvents with known Raman peaks can be used.
Electrooptic Amplitude ModulatorTHORLABSEO-AM-NR-C1Two EOMs are needed for orthogonal modulation and dual-channel imaging. Resonant version is recommended so lower driving voltage can be used.
False H&E Staining ScriptMatlabhttps://github.com/TheFuGroup/HE_Staining
Fanout BufferPRL-414BPulse Research Lab1:4 TTL/CMOS Fanout Buffer and Line Driver, for generating the EOM driving frequency and the reference to the lock-in
Fast PhotodiodeTHORLABSDET10A2Si Detector, 1 ns Rise Time
Frequency DividerPRL-220APulse Research LabTTL Freq. Divider (f/2, f/4, f/8, f/16), for generating 20MHz from the laser output.
Highly Dispersive Glass RodsUnion OpticCYLROD01High dispersion H-ZF52A Rod lens 120 mm, SF11 Rod lens 100 mm
Insight DS+NewportLaser system capable of outputting two synchronzied pulsed lasers (one fixed beam at 1, 040 nm and one tunable beam, ranging from 680-1,300 nm) with a repetition rate of 80 MHz. 
Lock-in AmplifierLiquid InstrumentsMoku LabLock-in amplifier to extract SRS signal from the photodiode. A Zurich Instrument HF2LI or similar instrument can be used as well.
MirrorsTHORLABSBB05-E03-10Broadband Dielectric Mirror, 750 - 1,100 nm. Silver mirrors can also be used.
Motorized Delay StageZaberX-DMQ12P-DE52Delay stage for fine control of the temporal overlap of the pump and the Stokes lasers. Any other motorized stage should work.
Oil Immersion CondensorNikonCSC10031.4 NA. Other condensers with NA>1.2 can be used.
OscilloscopeTektronixTDS7054Any other oscilloscope with 400 MHz bandwdith or higher should work.
Phase ShifterSigaTekSF50A2For shifting the phase of the modulation frequency
PhotodiodeHamamatsu CorpS3994-01Silicon PIN diode with large area (10 x 10 cm2). Other diodes with large area and low capacitance can be used.
Polarizing Beam SplitterUnion OpticPBS9025-620-1000Broadband polarizing beamsplitter
Refactive Index Databaserefractiveindex.info
Retro-reflectorEdmund Optics34-408BBAR Right Angle Prism. Other prisms or retroreflector can be used.
RF Power AmplifierMinicircuitsZHL-1-2W+Gain Block, 5 - 500 MHz, 50 Ω
Scan MirrorsCambridge Technologies6215HWe used a 5mm mirror set with silver coating
ScanImageVidrioScanImage BasicLaser scanning microscope control software
Shortpass FilterTHORLABSFESH100025.0 mm Premium Shortpass Filter, Cut-Off Wavelength: 1,000 nm. For efficient suppression of the Stokes, two filters may be necessary.
Upright MicroscopeNikonEclipse FN1Any other microscope frame can be used. If a laser scanning microscope is available, it can be used directly. Otherwise, a galvo scanner and scan lens needed to be added to the microscope.
Water Immersion ObjectiveOlympusXLPLN25XWMP2The multiphoton 25X Objective has a NA of 1.05. Other similar objectives can be used.

Odniesienia

  1. Fischer, A. H., Jacobson, K. A., Rose, J., Zeller, R. Hematoxylin and eosin staining of tissue and cell sections. Cold Spring Harbor Protocols. 2008, (2008).
  2. Cui, M., Zhang, D. Y. Artificial intelligence and computational pathology. Laboratory Investigation. 101 (4), 412-422 (2021).
  3. Freudiger, C. W., et al. Label-free biomedical imaging with high sensitivity by stimulated Raman scattering microscopy. Science. 322 (5909), 1857-1861 (2008).
  4. Orringer, D. A., et al. Rapid intraoperative histology of unprocessed surgical specimens via fibre-laser-based stimulated Raman scattering microscopy. Nature Biomedical Engineering. 1, 0027 (2017).
  5. Ji, M., et al. label-free detection of brain tumors with stimulated Raman scattering microscopy. Science Translational Medicine. 5 (201), (2013).
  6. Hollon, T. C., et al. Near real-time intraoperative brain tumor diagnosis using stimulated Raman histology and deep neural networks. Nature Medicine. 26 (1), 52-58 (2020).
  7. Shin, K. S., et al. Intraoperative assessment of skull base tumors using stimulated Raman scattering microscopy. Scientific Reports. 9 (1), 20392 (2019).
  8. Lu, F. -. K., et al. Label-free neurosurgical pathology with stimulated Raman imaging. Cancer Research. 76 (12), 3451-3462 (2016).
  9. Shin, K. S., et al. Quantitative chemical imaging of breast calcifications in association with neoplastic processes. Theranostics. 10 (13), 5865-5878 (2020).
  10. Fu, D., Holtom, G., Freudiger, C., Zhang, X., Xie, X. S. Hyperspectral imaging with stimulated Raman scattering by chirped femtosecond lasers. The Journal of Physical Chemistry B. 117 (16), 4634-4640 (2013).
  11. Figueroa, B., Hu, R., Rayner, S. G., Zheng, Y., Fu, D. Real-time microscale temperature imaging by stimulated Raman scattering. The Journal of Physical Chemistry Letters. 11 (17), 7083-7089 (2020).
  12. Saar, B. G., et al. Video-rate molecular imaging in vivo with stimulated Raman scattering. Science. 330 (6009), 1368-1370 (2010).
  13. Hill, A. H., Hill, A. H., Manifold, B., Manifold, B., Fu, D. Tissue imaging depth limit of stimulated Raman scattering microscopy. Biomedical Optics Express. 11 (2), 762-774 (2020).
  14. Fu, D., et al. Imaging the intracellular distribution of tyrosine kinase inhibitors in living cells with quantitative hyperspectral stimulated Raman scattering. Nature Chemistry. 6 (7), 614-622 (2014).
  15. Zhang, L., et al. Spectral tracing of deuterium for imaging glucose metabolism. Nature Biomedical Engineering. 3 (5), 402-413 (2019).
  16. Tsai, P. S., et al., Frostig, R. D., et al. Principles, design, and construction of a two-photon laser-scanning microscope for in vitro and in vivo brain imaging. In Vivo Optical Imaging of Brain. , 113-171 (2002).
  17. Francis, A., Berry, K., Chen, Y., Figueroa, B., Fu, D. Label-free pathology by spectrally sliced femtosecond stimulated Raman scattering (SRS) microscopy. PLoS One. 12 (5), 0178750 (2017).
  18. Figueroa, B., et al. Broadband hyperspectral stimulated Raman scattering microscopy with a parabolic fiber amplifier source. Biomedical Optics Express. 9 (12), 6116-6131 (2018).
  19. Kong, L., et al. Multicolor stimulated Raman scattering microscopy with a rapidly tunable optical parametric oscillator. Optics Letters. 38, 145-147 (2013).
  20. He, R., et al. Dual-phase stimulated Raman scattering microscopy for real-time two-color imaging. Optica. 4 (1), 44-47 (2017).
  21. Pence, I. J., Kuzma, B. A., Brinkmann, M., Hellwig, T., Evans, C. L. Multi-window sparse spectral sampling stimulated Raman scattering microscopy. Biomedical Optics Express. 12 (10), 6095-6114 (2021).
  22. Wei, M., et al. Volumetric chemical imaging by clearing-enhanced stimulated Raman scattering microscopy. Proceedings of the National Academy of Sciences of the United States of America. 116 (14), 6608-6617 (2019).
  23. Wright, A. J., et al. Adaptive optics for enhanced signal in CARS microscopy. Optics Express. 15 (26), 18209-18219 (2007).

Przedruki i uprawnienia

Zapytaj o uprawnienia na użycie tekstu lub obrazów z tego artykułu JoVE

Zapytaj o uprawnienia

Przeglądaj więcej artyków

Stimulated Raman ScatteringSRS MicroscopyReal time ImagingTissue DiagnosisSpectral FocusingHistology ApplicationsAchromatic LensesLaser AlignmentDichroic MirrorSpatial OverlapPhotodiode DetectionLow Pass FiltersBeam SamplerRF Phase Adjustment

This article has been published

Video Coming Soon

JoVE Logo

Prywatność

Warunki Korzystania

Zasady

Badania

Edukacja

O JoVE

Copyright © 2025 MyJoVE Corporation. Wszelkie prawa zastrzeżone