Here, we present a comprehensive capillary zone electrophoresis protocol for the assessment of intrinsic physicochemical heterogeneity of monoclonal antibodies as a quality attribute.
Biotherapeutic proteins, such as monoclonal antibodies (mAbs), are feasible alternatives for the treatment of chronic-degenerative diseases. The biological activity of these proteins depends on their physicochemical properties. The use of high-performance techniques like chromatography and capillary electrophoresis has been described for the analysis of physicochemical heterogeneity of mAbs. Nowadays, capillary zone electrophoresis (CZE) technique constitutes one of the most resolutive and sensitive assays for the analysis of biomolecules. Besides, the electro-driven separation in CZE is governed by extensive properties of matter and offers the advantage of analyzing proteins close to their native state. However, the successful implementation of this technique for routine analysis depends on the skills of the analyst at the critical steps during sample and system preparation. The purpose of this tutorial is to detail the steps to succeed in the CZE analysis of mAbs. Further, this protocol can be used for the development and improvement of skills of the personnel involved in protein analytical chemistry laboratories.
Os anticorpos monoclonais (mAbs) são proteínas bioterapêuticos com crescente interesse devido à sua capacidade de agir contra várias doenças crónicas e degenerativas 1. Como outras biomoléculas, os mAbs são propensos a sofrer várias alterações físico-químicos em todas as fases do seu ciclo de vida (ou seja, a partir de biossíntese para o produto final). Tais modificações incluem, mas não estão limitados a: desamidação, glicosilação, oxidação, ciclização, isomerização, a agregação e a clivagem proteolítica 2. Assim, técnicas analíticas capazes de resolver isoformas intrínsecas são necessários para monitorar mAbs heterogeneidade e estabilidade a fim de estabelecer especificações de qualidade.
A electroforese capilar (CE) é uma tecnologia de separação de alta performance realizada outinside de um tubo de sílica fundida estreita (gama uM) cheio com um electrólito de fundo (BGE). Após a aplicação de um campo eléctrico (até 30.000 V), molécula carregadas migrar para o eletrodo com carga oposta (ie, separação orientada por electro). O uso de altas tensões no CE permite análises rápidas e uma maior eficiência, que são superiores a electroforese em gel clássica. Eletroforese capilar de zona (CZE) é uma técnica baseada CE rotineiramente utilizado na indústria biofarmacêutica para a avaliação da qualidade do produto 3-9. Ao contrário de outros modos de Ce (por exemplo, electroforese capilar em gel, isoeléctrica capilar de focagem) ou métodos baseados em cromatografia, CZE pode ser realizado sem a utilização de agentes desnaturantes ou interfaces de fase sólida, permitindo a análise da heterogeneidade inerente de mAbs próximo ao seu estado nativo 10 . CZE separação de isoformas de mAb ocorre dentro de um capilar de sílica fundida coberta com um polímero hidrofílico (capilar neutro) e baseia-se na sua mobilidade electroforética diferente, que é regida por carga, massa, tamanho e forma (ou volume hidrodinâmico) 11. porções de mAb são detectados quandoeles são mobilizadas e passe através da janela de detecção, que é detectada por um detector de ultravioleta (UV) a absorvância a 214 nm de 4.
O sucesso da implementação desta técnica analítica dependerá a devida atenção aos detalhes antes e durante o experimento. Agindo de outra forma vai aumentar o custo eo tempo para realizar a análise, levando a falha constante e frustração.
Aqui, apresentamos um guia passo-a-passo para realizar uma análise bem-sucedida de mAb heterogeneidade por CZE através da explicação detalhada sobre a preparação de soluções e amostras, a preparação do sistema CE, os métodos instrumento criado, a aquisição de dados e o processamento. Para o propósito deste tutorial, um factor alfa de necrose anti-tumoral totalmente humano recombinante (anti-TNF) é o mAb usado como modelo de proteína; no entanto, este protocolo pode ser facilmente customizado para a análise de outras proteínas considerando breves modificações. UMAdditionally, várias recomendações para mitigar possíveis problemas são propostos. O leitor é encorajado a seguir rigorosamente o protocolo proposto, como a probabilidade de sucesso irá aumentar.
1. Preparação de Soluções
Preparação 2. Amostra
3. Preparação do sistema CE.
Figura 1: Posição e nível de enchimento de frascos universais nas bandejas de entrada e saída de buffer. Encha níveis: 1/1 = 1,400 mL, 3/4 = 1300 mL e 1000 mL = 1/2. Abreviaturas: W = água, HCl = ácido clorídrico 0,1 M. Por favor clique aqui para ver uma versão maior desta figura.
4. Métodos Instrumento Set-up
programa de tempo método de condicionamento | |||||||
Tempo (min) | Evento | Valor | Duração | frasco de entrada | frasco de saída | Resumo | Comentários |
- | Lavar - Pressão | 2068 mbar | 1,0 min | BI: C1 | BO: C1 | para a frente | HCl 0,1 M |
- | Lavar - Pressão | 2068 mbar | 2,0 min | BI: B1 | BO: C1 | para a frente | Água |
- | Lavar - Pressão | 2068 mbar | 10,0 min | BI: E1 | BO: C1 | para a frente | BGE |
0.00 | Separado - Tensão | 15,0 KV | 10,0 min | BI: D1 | BO: D1 | 0,17 rampa Min, polaridade normal, | Separado |
10.00 | Lavar - Pressão | 2758 mbar | 10,0 min | BI: E1 | BO: C1 | para a frente | BGE |
20.01 | Esperar | - | 0,0 min | BI: A1 | BO: A1 | - | dicas de enxaguamento |
Correndo método de programa de tempo | |||||||
Tempo (min) | Evento | Valor | Duração | frasco de entrada | frasco de saída | Resumo | Comentários |
- | Lavar - Pressão | 2068 mbar | 1,0 min | BI: C1 | BO: C1 | Adiante, In / Out inc frasco 6 | HCl 0,1 M |
- | Lavar - Pressão | 2068 mbar | 2,0 min | BI: B1 | BO: C1 | Adiante, In / Out inc frasco 6 | água |
- | Lavar - Pressão | 2068 mbar | 4,0 min | BI: E1 | BO: C1 | Adiante, In / Out inc frasco 6 | BGE |
- | Injectar - Pressão | 34 mbar | 20,0 seg | SI: A1 | BO: B1 | Substituir, Avançado | injecção da amostra |
- | Esperar | - | 0,4 min | BI: A1 | BO: A1 | In / Out inc frasco 6 | dicas de lavagem |
0 | Separado - Tensão | 15,0 KV | 30,0 min | BI: D1 | BO: D1 | 0,17 Min rampa, polaridade normal, In / Out inc frasco 6 | Separação amostra |
0,5 | autozero | - | - | - | - | - | - |
30.01 | pare de dados | - | - | - | - | - | - |
30.02 | Lavar - Pressão | 2068 mbar | 2,0 min | BI: B1 | BO: C1 | Adiante, In / Out inc frasco 6 | água |
32.03 | Esperar | - | 0,0 min | BI: A1 | BO: A1 | In / Out inc frasco 6 | dicas de enxaguamento |
32.04 | Fim | - | - | - | - | - | Fim |
programa de tempo método de desligamento | |||||||
Tempo (min) | Evento | Valor | Duração | frasco de entrada | frasco de saída | Resumo | Comentários |
- | Lavar - Pressão | 2068 mbar | 1,0 min | BI: E6 | BO: E6 | para a frente | HCl 0,1 M |
- | Lavar - Pressão | 2068 mbar | 6,0 min | BI: F6 | BO: F6 | para a frente | água |
- | Lamp - Off | - | - | - | - | - | - |
- | Esperar | - | 0,0 min | BI: A1 | BO: A1 | - | dicas de enxaguamento |
Tabela 1: Condicionado, correndo e programas de tempo de desligamento.
5. Aquisição de Dados e Processamento
A Figura 2 mostra o perfil de corrente eléctrica típica de um EACA 200 mM, 30 mM de acetato de lítio, pH 4,8 com BGE amostra anti-TNFa mAb diluído com tampão Tris (50 mM, pH 8,0). Como pode ser observado, a corrente é estável durante toda a análise e pode oscilar entre os valores de 30 a 35 uA. A Figura 3 mostra o electroferograma CZE de uma amostra em branco onde o pico detectado corresponde ao padrão interno de histamina. Espera-se para a hist...
Neste tutorial, destacamos a importância de práticas adequadas na condução CZE análises de mAbs, a fim de aumentar a probabilidade de sucesso. No entanto, quando CZE é usado de forma rotineira, questões surgem inevitavelmente 12.
Para melhores resultados, é importante seguir as notas que foram incluídos ao longo do protocolo, como eles vão ajudar o analista a superar e resolver problemas difíceis passos. Uma consideração importante para obter uma melhor resolução a ...
The authors Carlos E. Espinosa-de la Garza, Rodolfo D. Salazar-Flores, Néstor O. Pérez, Luis F. Flores-Ortiz and Emilio Medina-Rivero are employees of Probiomed S.A. de C.V., which is developing, manufacturing and marketing biosimilar products. All authors are involved in the development of biosimilar products for Probiomed.
The authors thank Wiley for the granted permission to use the concepts of the following publication for this tutorial. Carlos E. Espinosa-de la Garza, Francisco C. Perdomo-Abúndez, Jesús Padilla-Calderón, Jaime M. Uribe-Wiechers, Néstor O. Pérez, Luis F. Flores-Ortiz, Emilio Medina-Rivero: Analysis of recombinant monoclonal antibodies by capillary zone electrophoresis. Electrophoresis. 2013. 34. 1133-1140. Copyright Wiley-VCH Verlag GmbH & Co. KGaA. This work was supported by CONACyT, Mexico, grant 230551.
Name | Company | Catalog Number | Comments |
Glacial acetic acid | Tecsiquim | AT0035-7 | |
ACS grade hydrochloric acid | J.T. Baker | 9535-05 | |
Histamine dihydrochloride | Fluka | 53300 | |
(Hydroxypropyl) methyl cellulose | Fluka | 09963 | |
Lithium acetate | Sigma-Aldrich | 517992 | |
6-Aminocaproic acid | Sigma-Aldrich | A2504 | |
eCAP Tris Buffer, 50.0 mM, pH 8 | Beckman Coulter | 477427 | |
PA 800 Plus Pharmaceutical Analysis System | Beckman Coulter | A66528 | |
eCAP Neutral capillary | Beckman Coulter | 477441 | |
Vial, Micro, 200 µl | Beckman Coulter | 144709 | |
Universal Vial Caps | Beckman Coulter | A62250 | |
Universal Vials | Beckman Coulter | A62251 | |
Cable, Optics, UV/Vis | Beckman Coulter | 144093 | |
UV/Vis Detector Module | Beckman Coulter | 144733 | |
Cartridge Assembly Kit, Blank | Beckman Coulter | 144738 |