High-performance liquid chromatography (HPLC) is an important analytical method commonly used to separate and quantify components of liquid samples. In this technique, a solution (first phase) is pumped through a column that contains a packing of small porous particles with a second phase bound to the surface. The different solubilities of the sample components in the two phases cause the components to move through the column with different average velocities, thus creating a separation of these components. The pumped solution is called the mobile phase, while the phase in the column is called the stationary phase.
There are several modes of liquid chromatography, depending upon the type of stationary and/or mobile phase employed. This experiment uses reversed-phase chromatography, where the stationary phase is non-polar, and the mobile phase is polar. The stationary phase to be employed is C18 hydrocarbon groups bonded to 3-µm silica particles, while the mobile phase is an aqueous buffer with a polar organic modifier (acetonitrile) added to vary its eluting strength. In this form, the silica can be used for samples that are water-soluble, providing a broad range of applications. In this experiment, the mixtures of three components frequently found in diet soft drinks (namely caffeine, benzoate, and aspartame) are separated. Seven prepared solutions containing known amounts of the three species are used, and their chromatograms are then recorded.
During an HPLC experiment, a high-pressure pump takes the mobile phase from a reservoir through an injector. It then travels through a reverse-phase C18-packed column for component separation. Finally, the mobile phase moves into a detector cell, where the absorbance is measured at 220 nm, and ends in a waste bottle. The amount of time it takes for a component to travel from the injector port to the detector is called the retention time.
A liquid chromatograph is used in this experiment, where separation is performed on a reverse-phase column. The column dimensions are 3 mm (i.d.) x 100 mm, and the silica packing (3-µm particle size) is functionalized with C18 octadecylsilane (ODS). A Rheodyne 6-port rotary injection valve is used to initially store the sample in a small loop and introduces the sample to the mobile phase upon rotation of the valve.
Detection is by absorption spectroscopy at a wavelength of 220 nm. This experiment can be run at 254 nm, if a detector is not variable. Data from the detector has an analog voltage output, which is measured using a digital multimeter (DMM), and read by a computer loaded with a data acquisition program. The resulting chromatogram has a peak for every component in the sample. For this experiment, all three components elute within 5 min.
This experiment uses a single mobile phase and pump, which is called an isocratic mobile phase. For samples that are difficult to separate, a gradient mobile phase can be used. This is when the initial mobile phase is primarily an aqueous one, and over time, a second organic mobile phase is gradually added to the overall mobile phase. This method raises the polarity of this phase over time, which lowers the retention times of the components and works similarly to a temperature gradient on a gas chromatograph. There are some instances where the column is heated (usually to 40 °C), which takes away any retention time errors associated with a change of ambient temperature.
In reverse-phase HPLC, the column stationary phase packing is usually either a C4, C8, or C18 packing. The C4 columns are primarily for proteins with large molecular weights, whereas the C18 columns are for peptides and basic samples with lower molecular weights.
Detection by absorption spectroscopy is overwhelmingly the detection method of choice, as the absorption spectra of the components are all readily available. Some systems use electrochemical measurements, such as conductivity or amperometry, as their detection method.
For this experiment, the mobile phase is primarily 20% acetonitrile and 80% purified deionized (DI) water. A small amount of acetic acid is added to lower the pH of the mobile phase, which keeps the silanol in the stationary packing phase in an undissociated state. This reduces the adsorption peak from tailing, giving narrower peaks. Then, the pH is adjusted with 40% sodium hydroxide to raise the pH and help decrease the retention times of the components.
Each group uses a set of the 7 vials containing different concentrations of the standard solutions (Table 1). The first 3 are used to identify each peak, and the last 4 are for creating a calibration chart for each component. Standards 1–3 are also used for the calibration chart.
Number | Caffeine (mL) | Benzoate (mL) | Aspartame (mL) |
1 | 4 | 0 | 0 |
2 | 0 | 4 | 0 |
3 | 0 | 0 | 4 |
4 | 1 | 1 | 1 |
5 | 2 | 2 | 2 |
6 | 3 | 3 | 3 |
7 | 5 | 5 | 5 |
Table 1. Volumes of stock standards used to prepare the 7 provided working standards (total volume of each standard is 50 mL).
1. Making the Mobile Phase
2. Creating the Component Solutions
The three components that need to be made are caffeine (0.8 mg/mL), potassium benzoate (1.4 mg/mL), and aspartame (L-aspartyl-L-phenylalanine methyl ester) (6.0 mg/mL). These concentrations, once diluted in the same fashion, put the standards at the levels found in the soda samples.
3. Making the 7 Standard Solutions
The three components all have differing distribution coefficients, which affects how each interacts with both of the phases. The larger the distribution coefficient, the more time the component spends in the stationary phase, resulting in longer retention times in reaching the detector.
4. Checking the Initial Settings of the HPLC System
5. Manually Injecting the Sample and Data Collection
Figure 1. The chromatogram of the 3 components. From left to right, they are caffeine, aspartame, and benzoate.
6. The Samples of Diet Sodas
Diet Coke, Diet Pepsi, and Coke Zero are the "unknowns." They have been left out in open containers overnight to get rid of the carbonation, as bubbles are not good for the HPLC system. This sufficiently gets rid of any gases in the samples.
7. Calculations
Figure 2. A basic example of a curve's peak height and width, which are to be multiplied (peak height times width at ½ height).
The HPLC chromatograms are able to quantify each of the 3 components for all the samples based upon the calibration curves of the standards (Figure 3).
From this set of experiments, it was determined that a 12-oz can of these diet sodas contained the following amounts of each component:
Diet Coke: 50.5 mg caffeine; 217.6 mg aspartame; 83.6 mg benzoate.
Coke Zero: 43.1 mg caffeine; 124.9 mg aspartame; 85.3 mg benzoate.
Diet Pepsi: 34.1 mg caffeine; 184.7 mg aspartame; 79.5 mg benzoate.
Not surprisingly, all 3 had roughly the same amount of benzoate, as it is just a preservative. The Coke products had a bit more caffeine, and the Coke Zero had much less aspartame than the other two sodas, as it also includes citric acid for some flavoring.
The following numbers are the actual amounts of caffeine and aspartame in a 12-oz can of the 3 diet sodas (The caffeine content was obtained from the Coca-Cola and Pepsi websites. The aspartame content was obtained from both LiveStrong.com and DiabetesSelfManagement.com.):
Diet Coke: 46 mg caffeine; 187.5 mg aspartame
Coke Zero: 34 mg caffeine; 87.0 mg aspartame
Diet Pepsi: 35 mg caffeine; 177.0 mg aspartame
Sample Calculations (Table 2):
Concentration of caffeine in STD#1: The component solution for caffeine had 0.400 g of caffeine diluted to 500 mL = 0.500 L → 0.800 g/L = 0.800 mg/mL.
STD#1 had 1 mL of this solution diluted to 50.0 mL
0.800 mg/mL * (1.0 mL / 50.0 mL) = 0.016 mg/mL = 16.0 mg/L.
STD#2 had 2 mL of this solution diluted to 50.0 mL
0.800 mg/mL * (2.0 mL / 50.0 mL) = 0.032 mg/mL = 32.0 mg/L.
The results from the three calibration graphs (Figure 4) yielded the following equations:
Caffeine Peak Area = 0.1583*[Caffeine mg/L] - 0.574
Aspartame Peak Area = 0.02696*[Aspartame mg/L] - 0.405
Benzoate Peak Area = 0.1363*[Benzoate mg/L] - 1.192
Diet Coke: Caffeine Peak Area = 10.68 = 0.1583*[Caffeine mg/L] - 0.574
[Caffeine mg/L] = (10.68 + 0.574)/ (0.1583) = 71.1 mg/L in the injected sample.
Since the sample was diluted by a factor of 2, the Diet Coke had 141.2 mg/L caffeine.
The amount per 12-oz can = (141.2 mg/L)(0.3549 mL/12-oz can) = 50.5 mg caffeine/can.
Figure 3. The HPLC chromatograms of the 5 standards and the 3 samples.
Figure 4. The calibration curves for each of the 3 components.
Table 2. The data tables for the HPLC trials used for generating the calibration curves.
HPLC is a widely-used technique in the separation and detection for many applications. It is ideal for non-volatile compounds, as gas chromatography (GC) requires that the samples are in their gas phase. Non-volatile compounds include sugars, vitamins, drugs, and metabolites. Also, it is non-destructive, which allows each component to be collected for further analysis (such as mass spectrometry). The mobile phases are practically unlimited, which allows changes to the polarity of pH to achieve better resolution. The use of gradient mobile phases allows for these changes during the actual trials.
There has been concern over the possible health issues that may be associated with the artificial sweetener aspartame. Current product labeling does not show the amount of these components inside of the diet beverages. This method allows for quantifying these amounts, along with the caffeine and benzoate.
Other applications include determining the amounts of pesticides in water; determining the amount of acetaminophen or ibuprofen in pain reliever tablets; determining whether there are performance-enhancing drugs present in the bloodstream of athletes; or simply determining the presence of drugs in a crime lab. While the concentrations of these samples, and often the identity of the components, can be readily determined, the one limitation is that several samples could have close to identical retention times, resulting in co-eluting.