We describe a single-molecule approach to antigen-antibody affinity measurements using mass photometry (MP). The MP-based protocol is fast, accurate, uses a very small amount of material, and does not require protein modification.
Measurements of the specificity and affinity of antigen-antibody interactions are critically important for medical and research applications. In this protocol, we describe the implementation of a new single-molecule technique, mass photometry (MP), for this purpose. MP is a label- and immobilization-free technique that detects and quantifies molecular masses and populations of antibodies and antigen-antibody complexes on a single-molecule level. MP analyzes the antigen-antibody sample within minutes, allowing for the precise determination of the binding affinity and simultaneously providing information on the stoichiometry and the oligomeric state of the proteins. This is a simple and straightforward technique that requires only picomole quantities of protein and no expensive consumables. The same procedure can be used to study protein-protein binding for proteins with a molecular mass larger than 50 kDa. For multivalent protein interactions, the affinities of multiple binding sites can be obtained in a single measurement. However, the single-molecule mode of measurement and the lack of labeling imposes some experimental limitations. This method gives the best results when applied to measurements of sub-micromolar interaction affinities, antigens with a molecular mass of 20 kDa or larger, and relatively pure protein samples. We also describe the procedure for performing the required fitting and calculation steps using basic data analysis software.
Antibodies have become ubiquitous tools of molecular biology and are used extensively in both medical and research applications. In medicine, they are critically important in diagnostics, but their therapeutic applications are also expanding and new antibody-based therapies are constantly being developed1,2,3,4. The scientific applications of antibodies include many indispensable laboratory techniques such as immunofluorescence5, immunoprecipitation6, flow cytometry7, EL....
1. Prepare the flow chambers
We have previously examined the interaction of human α-thrombin (HT) and mouse monoclonal anti-human thrombin antibody (AHT) using the MP based assay11. Since the molecular mass of the HT (37 kDa) is below the 40 kDa detection limit, the maximum sample concentration can exceed the 50 nM MP concentration limitation without negatively affecting the resolution of mass distributions. The experiment was planned as a titration series with the AHT antibody at a fixed 25 nM concentration, a.......
The Mass Photometry based protocol outlined here provides a fast and accurate method of measuring antigen-antibody binding affinities. MP analysis uses a very small amount of material, and additional information—including stoichiometry, oligomerization, and purity—can be assessed from the same data (Figure 5). Without modifications, this method is applicable to the measurements of dissociation constants in the approximately 5 nM to 500 nM range, and for ligand molecules with mole.......
The authors have nothing to disclose.
We thank Keir Neuman for his critical reading of the manuscript. This work was supported by the intramural program of the NHLBI, NIH.
....Name | Company | Catalog Number | Comments |
AcquireMP | Refeyn | MP data collection software | |
Anti-human thrombin | Haematologic Technologies | AHT-5020 | RRID: AB_2864302 |
Cotton-tipped applicators | Thorlabs | CTA10 | cotton optical swabs for lens cleaning |
Coverslips 24x24 mm | Globe Scientific | 1405-10 | |
Coverslips 24x50 mm | Fisher Scientific | 12-544-EP | |
DiscoverMP | Refeyn | MP data processing software | |
Forceps | Electron Microscopy Sciences | 78080-CF | soft-tipped forceps for coverslips handling |
Human α-thrombin | Haematologic Technologies | HCT-0020 | |
Immersion oil | Thorlabs | MOIL-30 | |
Isopropanol | Alfa Aesar | 36644 | |
Microsoft Excel | Microsoft | spreadsheet | |
OneMP | Refeyn | Mass Photometry instrument | |
Origin | OriginLab | scientific graphing software | |
PBS | Corning | 46-013-CM | 10x stock |
Syringe filter | Millipore | SLGSR33SS | buffer and sample filtering |