A subscription to JoVE is required to view this content. Sign in or start your free trial.

In This Article

  • Summary
  • Abstract
  • Introduction
  • Protocol
  • Results
  • Discussion
  • Disclosures
  • Acknowledgements
  • Materials
  • References
  • Reprints and Permissions

Summary

Here, we describe a correlative workflow for the excision, pressurization, fixation, and imaging of the murine pulmonary valve to determine the gross conformation and local extracellular matrix structures.

Abstract

The underlying causes of heart valve related-disease (HVD) are elusive. Murine animal models provide an excellent tool for studying HVD, however, the surgical and instrumental expertise required to accurately quantify the structure and organization across multiple length scales have stunted its advancement. This work provides a detailed description of the murine dissection, en bloc staining, sample processing, and correlative imaging procedures for depicting the heart valve at different length scales. Hydrostatic transvalvular pressure was used to control the temporal heterogeneity by chemically fixing the heart valve conformation. Micro-computed tomography (µCT) was used to confirm the geometry of the heart valve and provide a reference for the downstream sample processing needed for the serial block face scanning electron microscopy (SBF-SEM). High-resolution serial SEM images of the extracellular matrix (ECM) were taken and reconstructed to provide a local 3D representation of its organization. µCT and SBF-SEM imaging methods were then correlated to overcome the spatial variation across the pulmonary valve. Though the work presented is exclusively on the pulmonary valve, this methodology could be adopted for describing the hierarchical organization in biological systems and is pivotal for the structural characterization across multiple length scales.

Introduction

The pulmonary valve (PV) serves to ensure unidirectional blood flow between the right ventricle and the pulmonary artery. Pulmonary valve malformations are associated with several forms of congenital heart disease. The current treatment for congenital heart valve disease (HVD) is valvular repair or valve replacement, which can necessitate multiple invasive surgeries throughout a patient's lifetime1. It has been widely accepted that the function of the heart valve is derived from its structure, often referred to as the structure-function correlate. More specifically, the geometric and biomechanical properties of the heart dictate its function. The mechanical properties, in turn, are determined by the composition and organization of the ECM. By developing a method for determining the biomechanical properties of murine heart valves, transgenic animal models can be used to interrogate the role of the ECM on heart valve function and dysfunction2,3,4,5.

The murine animal model has long been regarded as the standard for molecular studies because transgenic models are more readily available in mice compared to other species. Murine transgenic models provide a versatile platform for researching heart valve-related diseases6. However, the surgical expertise and instrumentation requirements to characterize both the geometry and ECM organization have been a major hurdle in progressing HVD research. Hstological data in literature provides a picture into murine heart valve extracellular matrix content, but only in the form of 2D images, and are unable to describe its 3D architecture7,8. Additionally, the heart valve is both spatially and temporally heterogeneous, making it difficult to draw conclusions across experiments regarding ECM organization if the sampling and conformation are not fixed. Conventional 3D characterization methods, such as MRI or 3D echocardiography, do not provide the resolution necessary to resolve ECM components9,10.

This work details a fully correlative workflow where the temporal heterogeneity due to the cardiac cycle was addressed by fixing the conformation of the murine PV with hydrostatic transvalvular pressure. The spatial heterogeneity was controlled precisely by sampling regions of interest and registering data sets from different imaging modalities, specifically µCT and serial block face scanning electron microscopy, across different length scales. This method of scouting with µCT for guiding downstream sampling has been proposed previously, but because the pulmonary valve exhibits temporal variation, an additional level of control was needed on the surgical level11.

In vivo studies describing murine heart valve biomechanics are sparse and, instead, rely on computational models when describing the deformation behavior. It is of critical importance that local extracellular data on the nanometer length scale be related to the geometry and location of the heart valve. This, in turn, provides quantifiable, spatially mapped distributions of mechanically contributing ECM proteins, which can be used to reinforce existing biomechanical heart valve models12,13,14.

Protocol

The use of animals in this study was in accordance with Nationwide Children's Hospital institutional animal care and use committee under protocol AR13-00030.

1. Pulmonary valve excision

  1. Autoclave the necessary tools needed for the mouse dissection. This includes fine scissors, micro forceps, micro vascular clamps, clamp applying forceps, microneedle holders, spring scissors, and retractors.
  2. Acclimate all mice for a minimum of 2 weeks before the operation. Remove C57BL/6 mice, approximately 1 year of age, from their cages and weigh, then euthanize with a ketamine/xylazine cocktail (3:1 ketamine:xylazine, 0.01 mL per g) overdose.
  3. Place the mouse in a dorsal recumbence position on a tray and secure its limbs with tape. Once secured, perform the thoracotomy.
    1. Expose the heart by removing any excess adipose tissue and fascia.
    2. Remove the right atrium and perfuse through the left ventricle with room temperature saline solution (0.9% NaCl). The perfusion should take approximately 20 mL over 30 s. This results in the exsanguination of the mouse.
  4. Remove the entire heart by severing the superior vena cava, inferior vena cava, pulmonary artery, and aorta. Special care should be taken for the pulmonary artery. Cut approximately 2 mm above the ventriculo-arterial junction, as this will serve as the conduit for pressurization.
  5. Remove the left and right ventricles to expose chambers to atmospheric pressure. Ensure that the structure of the pulmonary trunk should be unaffected by the removal of the ventricles.

2. Pressure fixation of pulmonary valve

  1. Anastomose pressurization tubing with pulmonary artery, leaving approximately 1 mm distance between the sino-tubular junction and the end of the tubing to accommodate for large movements of the leaflets and pulmonary trunk.
  2. Elevate the reservoir to an analogous physiological pressure and fill it with the saline solution. Test the flow-through system to ensure there are no blockages or air bubbles.
  3. Attach a stopcock to the anastomosed pulmonary valve and ensure adequate flow through the tubing (i.e., no air bubbles) by switching the outflow tract. Once the flow is adequate, switch the outflow to the anastomosed pulmonary valve and ensure pressurization of the pulmonary trunk. This is identified by pulmonary trunk distention.
  4. Once the pressurization of the pulmonary trunk is confirmed, gradually incorporate primary fixative solution (1.25% glutaraldehyde, 1.0% paraformaldehyde in 0.15 M cacodylate) until the saline solution is purged. This is done by removing a portion, approximately 25% of the reservoir capacity, of the saline solution and replacing it with the primary fixative.
    CAUTION: The fixatives used (paraformaldehyde and glutaraldehyde) are highly toxic and appropriate personal protective equipment (PPE) should be worn to ensure safety.
  5. Place a fixative-soaked gauze over the tissue sample to prevent drying.
  6. Perfuse the fixative for 3 h, refilling the reservoir as needed to maintain a constant pressure. Throughout fixation, it is not uncommon for the pulmonary valve to shrink because of the chemical fixation. If this is the case, constantly replenish the reservoir with primary fixative to maintain physiological pressure.
  7. Store the heart valve in the fixative solution at 4 °C until use. Samples were stored for up to 1 week without any discernable difference.

3. En bloc sample staining and embedment15,16

CAUTION: The staining reagents used in this section (potassium ferrocyanide, osmium tetroxide, thiocarbohydrazide, lead aspartate, and uranyl acetate) are highly toxic and should be handled with extreme care. Use of a fume hood and proper PPE is advised.

  1. Staining
    1. Wash the fixed heart valve sample for 5 min with cold 0.15 M cacodylate buffer. Repeat the wash two more times.
    2. Completely submerge the heart valve in a solution of 1.5% potassium ferrocyanide, 0.15 M cacodylate, 2 mM calcium chloride, and 2% osmium tetroxide, on ice for 1 h.
    3. While the sample is incubating, prepare thiocarbohydrazide (TCH) solution by dissolving 0.1 g of TCH in 10 mL of ddH2O. Place the solution in a 60 °C oven for 1 h. Gently agitate periodically to ensure TCH is dissolved completely. Filter the solution through a 0.22 µm syringe filter immediately before use.
    4. Wash the samples with room temperature ddH2O by placing them in a tube of ddH2O for 5 min and slightly agitate it by shaking the container. Repeat this process three times.
    5. Place it in filtered TCH solution for 20 min at room temperature. Perform the washing step three times (5 min each) with room temperature ddH2O as described in step 3.1.4.
    6. Once done, place the sample in 2% osmium tetroxide for 30 min at room temperature. Then wash it again for a total of three times for 5 min each with room temperature ddH2O.
    7. Incubate the sample in 1% uranyl acetate overnight at 4 °C.
    8. During this time, make a solution of 0.066 g of lead nitrate in 10 mL of aspartic acid stock solution. Adjust the pH to 5.5 with 1 N KOH. Place the solution in a 60 °C oven to dissolve lead nitrate.
    9. After the overnight incubation, perform the washing step as described in step 3.1.4. Repeat three times. Then, incubate the heart valve tissue in lead aspartate solution from step 3.1.8 in a 60 °C oven for 30 min.
  2. Dehydration
    1. Wash the tissues for 5 min at room temperature with ddH2O. Repeat three times.
    2. Make fresh solutions of 20%, 50%, 70%, 90%, and 100% ethanol in ddH2O. Keep on ice until use.
    3. To dehydrate the heart valve tissue. Perform subsequent treatments of 20%, 50%, 70%, and 90% ethanol on ice for 5 min each. Then perform two subsequent treatments of 100% ethanol on ice for 5 min.
    4. Move the tissue to ice-cold acetone for 10 min. Then place in fresh acetone at room temperature for 10 min.
  3. Embedding
    1. Make the resin mixture (see Table of Materials) according to the manufacturer's specifications: 11.4 g of component A, 10 g of component B, 0.3 g of component C, and 0.05-0.1 g of component D. Initially mix components A and B by heating each of the components to 60 °C before adding components C and D. The mixture will turn into an amber color upon adding components C and D. It will be uniform when properly mixed.
    2. Make mixtures with volume ratios of 25:75, 50:50, and 75:25 resin:acetone. Mix thoroughly.
    3. Place tissues in subsequent treatments of the 25:75, 50:50, and 75:25 resin:acetone mixture for 2 h each at room temperature.
    4. Place tissues in 100% resin overnight at room temperature.
    5. The following day, place tissues in fresh 100% resin for 2 h at room temperature.
    6. Transfer the heart valve tissues to an embedding capsule, substitute with fresh 100% resin, and place in a 60 °C oven for 48 h to cure.
  4. Perform correlative imaging as shown below.

4. Micro-computed tomography imaging

  1. Mount the resin sample block onto a µCT sample holder with an adhesive (glue or double-sided tape works well).
  2. Place the holder into µCT chamber and fix it onto the stage by screwing the collet such that there is no motion of the holder. Sample movement during the scan will decrease the image quality.
  3. Open the µCT user interface by double-clicking on the icon. Add a project by selecting the + sign next to Projects and fill in the necessary fields.
  4. Once this is created, find the created project, and select it by clicking on it. This will open up a second column Acquisitions. Select the + next to Acquisitions and fill out the necessary fields.
  5. Close the chamber doors and arm the system by hitting the arming button on the front panel or the µCT. Warmup x-rays from the user interface by selecting the button indicating Warmup.
    NOTE: The system will automatically turn off the x-rays following a successful warm-up procedure. Turn on x-rays by selecting the button.
  6. Adjust the stage rotation such that the center of the rotation of the sample does not deviate from the center of the monitor (i.e., the sample is within the field of view for the entire scan). For the system used for this experiment, this involves adjusting the region of interest (ROI) stage under the drop-down tab as detailed below.
    1. Set the ROI stage adjustment by setting the rotation angle to 0° and marking the edge of the sample. The sample is then rotated to 180° and the edge of the sample is marked again.
    2. Adjust the ROI stage x-axis position such that the edge of the sample is between these two extremes. Repeat this process for rotation angles of 90° and 270° for the y-position.
    3. In the case where the rotation of the sample causes the edges to move out of the field of view, decrease the magnification of the µCT needs until the edges of the sample are visible at the above set angles and a coarse ROI stage adjustment can be made.
    4. Once adjusted at the lower magnification, move the sample or the detector to increase the magnification and the ROI positions can be fine adjusted.
      NOTE: This process may need to be repeated to ensure alignment. Proper alignment results in a sample that shows little to no horizontal movement through all sample rotation angles.
  7. Adjust the µCT to desired scanning parameters using the manufacturer's software. These parameters include tube potential, tube current, detector distance, sample distance, exposure time, trajectory, and the number of projections (see Table S1 for the µCT acquisition parameters used in this study).
    NOTE: Calibration parameters, such as clear field and dark field scans, should be kept at the manufacturer's specifications unless otherwise stated. For example, if a sample is too large and the system cannot move the sample out of the field view, then the sample needs to be removed to perform clear field calibration scans.
  8. Once the parameters are set, an approximation of the duration of the scan can be seen by hitting the Time Estimate button. Select the Start button to begin the scan.
  9. Following the scan completion, ensure that the x-rays are turned off, unscrew the collet, and carefully remove the sample from the µCT chamber.

5. Sample processing and image correlation

  1. Reconstruct µCT projections using a filtered back projection algorithm with the software provided by the manufacturer (see Table of Materials).
    1. Select the Recon tab at the top of the screen. Select the Project and Acquisition tab.
    2. Select the recon template associate with filtered back projection. Hit the Start Recon button.
  2. Identify and segment the pulmonary valve using image processing software. This requires prior knowledge of the anatomy of the pulmonary valve17. At elevated arterial pressures, the leaflets coapt and block the lumen of the pulmonary valve.
  3. Identify the slicing orientation with respect to the scanning direction and sample. Reorient the sample such that the slicing direction is aligned with the desired axis.
  4. Identify regions of interest for high-resolution imaging. In this experiment, the belly (middle) of the leaflet and the arterio-valvular junction was chosen.
  5. Remove the excess resin and sample by either razor blade or sanding for large pieces of material, or by microtome for finer pieces.
  6. Once at a location of interest, compare the physical specimen with virtual µCT slices to confirm location. This was done by comparing anatomical features at the cross section.

6. Serial block face scanning electron microscopy18

  1. Reduce the cross-section of the specimen to accommodate SBF-SEM, approximately 2.0 x 1.5 x 1.8 mm.
  2. Mount the trimmed sample onto the SBF-SEM aluminum stub using epoxy.
  3. Coat the specimen block with 35 nm gold. Spin the sample on the platform to apply an even layer of coating.
  4. Vent the SBF-SEM chamber and adjust the height of the knife blade to microscope eucentric height.
  5. Insert the sample and level the sample stub.
  6. Image under low vacuum conditions to prevent charging and with backscatter detector (see Table S2 for SBF-SEM acquisition parameters).
  7. Image and stitch multiple regions of interest at different magnifications using automated software (see Table of Materials).

Results

Anastomosis of the pulmonary artery to the pressurization tubing is shown in Figure 1A. Following the application of hydrostatic pressure, the pulmonary trunk distends radially (Figure 1B) indicating that the pulmonary valve leaflets are in a closed configuration. Pulmonary valve conformation was confirmed by µCT. In this case, the leaflets were coapt (closed) and the annulus was circular (Figure 2A). Figur...

Discussion

Removal of the ventricles serves two purposes. First, exposing the ventricle side to the atmospheric pressure, thereby only needing to apply a transvalvular pressure from the arterial side of the pulmonary valve to close, and second, providing a stable base to prevent twisting of the pulmonary trunk. During pressurization, the pulmonary trunk distends radially and inferiorly, making it prone to twisting, causing the collapse of the pulmonary trunk. Preloading the pulmonary valve with a saline solution offers an additiona...

Disclosures

The authors have nothing to disclose.

Acknowledgements

This work is supported, in part, by R01HL139796 and R01HL128847 grants to CKB and RO1DE028297 and CBET1608058 for DWM.

Materials

NameCompanyCatalog NumberComments
25% glutaraldehyde (aq)EMS16210Primary fixative component
0.9% sodium chloride injectionHospira Inc.NDC 0409-4888-10
1 mL syringeBD309659
10 mL syringeBD309604
200 proof ethanolEMS15055
22G needleBD305156
3 mL syringeBD309657
3-way stopcockSmiths Medical ASD, Inc.MX5311L
4% osmium tetroxideEMS19150Staining component
4% paraformaldehyde (aq)EMS157-4-100Primary fixative component
Absorbable hemostatEthicon1961
AcetoneEMS10012
Black polyamide monofilament suture, 10-0AROSurgical instruments CorporationTI38402
Black polyamide monofilament suture, 6-0AROSurgical instruments CorporationSN-1956
C57BL/6 miceJackson Laboratories664Approximately 1 yo
Calcium chlorideSigma-Aldrich10043-52-4
Clamp applying forcepFST00072-14
Cotton tip applicatorsFisher Scientific23-400-118
DPBSGibco14190-144
Dumont #5 forcepFST11251-20
Dumont #5/45 forcepsFST11251-35
Dumont #7 fine forcepFST11274-20
Durcupan ACM resinEMS14040For embedding
Fine scissorFST14028-10
Heliscan microCTThermo Fisher ScientificMicro-CT
Ketamine hydrochloride injectionHospira Inc.NDC 0409-2053
L-aspartic acidSigma-Aldrich56-84-8Staining component
Lead nitrateEMS17900Staining component
low-vacuum backscatter detectorThermo Fisher ScientificVSDBSSEM backscatter detector
Micro-adson forcepFST11018-12
Millex-GP filter, 0.22 um, PES 33mm, non-sterileEMD MilliporeSLGP033NS
Non-woven songesMcKesson Corp.94442000
Potassium hexacyanoferrate(II) trihydrateSigma-Aldrich14459-95-1Staining component
Potassium hydroxideSigma-Aldrich1310-58-3
Pressure monitor lineSmiths Medical ASD, Inc.MX562
Saline solution (sterile 0.9% sodium chloride)Hospira Inc.NDC 0409-0138-22
Size 3 BEEM capsuleEMS69910-01Embedding container
Sodium cacodylate trihydrateSigma-Aldrich6131-99-3Buffer
Solibri retractorsFST17000-04
Sputter, carbon and e-beam coaterLeicaEM ACE600Gold coater
Surgical microscopeLeicaM80
Thiocarbohydrazide (TCH)EMS21900Staining component
Tish needle holder/forcepMicrinsMI1540
TrimmerWahl9854-500
Uranyl acetateEMS22400Staining component
Volumescope scanning electron microscopeThermo Fisher ScientificVOLUMESCOPESEMSerial Block Face Scanning Electron Microscope
Xylazine sterile solutionAkorn Inc.NADA# 139-236

References

  1. Azari, S., et al. A systematic review of the cost-effectiveness of heart valve replacement with a mechanical versus biological prosthesis in patients with heart valvular disease. Heart Failure Reviews. 25 (3), 495-503 (2020).
  2. Ng, C. M., et al. TGF-β-dependent pathogenesis of mitral valve prolapse in a mouse model of Marfan syndrome. Journal of Clinical Investigation. 114 (11), 1586-1592 (2004).
  3. Cheek, J. D., Wirrig, E. E., Alfieri, C. M., James, J. F., Yutzey, K. E. Differential activation of valvulogenic, chondrogenic, and osteogenic pathways in mouse models of myxomatous and calcific aortic valve disease. Journal of Molecular and Cellular Cardiology. 52 (3), 689-700 (2012).
  4. Jiménez-Altayó, F., et al. Stenosis coexists with compromised α1-adrenergic contractions in the ascending aorta of a mouse model of Williams-Beuren syndrome. Scientific Reports. 10 (1), 889 (2020).
  5. Thacoor, A. Mitral valve prolapse and Marfan syndrome. Congenital Heart Disease. 12 (4), 430-434 (2017).
  6. McAnulty, P., Dayan, A., Ganderup, N. -. C., Hastings, K., Dawson, H. A Comparative Assessment of the Pig, Mouse and Human Genomes. The Minipig in Biomedical Research. , (2011).
  7. Hinton, R. B., Yutzey, K. E. Heart valve structure and function in development and disease. Annual Review of Physiology. 73, 29-46 (2011).
  8. Hinton, R. B., et al. Extracellular matrix remodeling and organization in developing and diseased aortic valves. Circulation Research. 98 (11), 1431-1438 (2006).
  9. Sacks, M. S., Merryman, W. D., Schmidt, D. E., David Merryman, D. W., Schmidt, D. E. On the biomechanics of heart valve function. Journal of Biomechanics. 42 (12), 1804-1824 (2009).
  10. Sacks, M. S., Yoganathan, A. P. Heart valve function: a biomechanical perspective. Philosophical Transactions of the Royal Society B-Biological Sciences. 362 (1484), 1369-1391 (2007).
  11. Morales, A. G., et al. Micro-CT scouting for transmission electron microscopy of human tissue specimens. Journal of Microscopy. 263 (1), 113-117 (2016).
  12. Sacks, M. S., Smith, D. B., Hiester, E. D. The aortic valve microstructure: Effects of transvalvular pressure. Journal of Biomedical Materials Research. 41 (1), 131-141 (1998).
  13. Ayoub, S., et al. Heart valve biomechanics and underlying mechanobiology. Comprehensive Physiology. 6 (4), 1743-1780 (2016).
  14. Stella, J. A., Liao, J., Sacks, M. S. Time-dependent biaxial mechanical behavior of the aortic heart valve leaflet. Journal of Biomechanics. 40 (14), 3169-3177 (2007).
  15. Korn, E. D., Weisman, R. A. I. loss of lipids during preparation of amoebae for electron microscopy. Biochimica et Biophysica Acta (BBA)/Lipids and Lipid Metabolism. 116 (2), 309-316 (1966).
  16. Tapia, J. C., et al. High-contrast en bloc staining of neuronal tissue for field emission scanning electron microscopy. Nature Protocols. 7 (2), 193-206 (2012).
  17. Hinton, R. B., et al. Mouse heart valve structure and function: Echocardiographic and morphometric analyses from the fetus through the aged adult. American Journal of Physiology - Heart and Circulatory Physiology. 294 (6), 2480-2488 (2008).
  18. Denk, W., Horstmann, H. Serial block-face scanning electron microscopy to reconstruct three-dimensional tissue nanostructure. Plos Biology. 2 (11), 1900-1909 (2004).
  19. Lincoln, J., Florer, J. B., Deutsch, G. H., Wenstrup, R. J., Yutzey, K. E. ColVa1 and ColXIa1 are required for myocardial morphogenesis and heart valve development. Developmental Dynamics. 235 (12), 3295-3305 (2006).
  20. Hamatani, Y., et al. Pathological investigation of congenital bicuspid aortic valve stenosis, compared with atherosclerotic tricuspid aortic valve stenosis and congenital bicuspid aortic valve regurgitation. PLoS One. 11 (8), (2016).

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Explore More Articles

Murine Pulmonary ValveCorrelative ImagingSurgical ProtocolC57BL 6 MouseThoracotomyAnatomical DissectionMicrosurgeryPressurization TubingSaline SolutionFixative SolutionTissue FixationStructural AnalysisFlow through SystemPulmonary Trunk Distension

This article has been published

Video Coming Soon

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved